VOLUME 44
Advances in CHROMATOGRAPHY
VOLUME 44
Advances in CHROMATOGRAPHY EDITORS:
ELI GRUSHK A Hebrew University ...
30 downloads
819 Views
5MB Size
Report
This content was uploaded by our users and we assume good faith they have the permission to share this book. If you own the copyright to this book and it is wrongfully on our website, we offer a simple DMCA procedure to remove your content from our site. Start by pressing the button below!
Report copyright / DMCA form
VOLUME 44
Advances in CHROMATOGRAPHY
VOLUME 44
Advances in CHROMATOGRAPHY EDITORS:
ELI GRUSHK A Hebrew University of Jerusalem Jerusalem, Israel
NELU GRINBERG Boehringer-Ingelheim Pharmaceutical, Inc. Riverside, Connecticut, U.S.A.
Boca Raton London New York
A CRC title, part of the Taylor & Francis imprint, a member of the Taylor & Francis Group, the academic division of T&F Informa plc.
Published in 2005 by CRC Press Taylor & Francis Group 6000 Broken Sound Parkway NW, Suite 300 Boca Raton, FL 33487-2742 © 2005 by Taylor & Francis Group, LLC CRC Press is an imprint of Taylor & Francis Group No claim to original U.S. Government works Printed in the United States of America on acid-free paper 10 9 8 7 6 5 4 3 2 1 International Standard Book Number-10: 0-0000-0000-0 (Hardcover) International Standard Book Number-13: 978-0-0000-0000-0 (Hardcover) Library of Congress Card Number XX-XXXXX This book contains information obtained from authentic and highly regarded sources. Reprinted material is quoted with permission, and sources are indicated. A wide variety of references are listed. Reasonable efforts have been made to publish reliable data and information, but the author and the publisher cannot assume responsibility for the validity of all materials or for the consequences of their use. No part of this book may be reprinted, reproduced, transmitted, or utilized in any form by any electronic, mechanical, or other means, now known or hereafter invented, including photocopying, microfilming, and recording, or in any information storage or retrieval system, without written permission from the publishers. For permission to photocopy or use material electronically from this work, please access www.copyright.com (http://www.copyright.com/) or contact the Copyright Clearance Center, Inc. (CCC) 222 Rosewood Drive, Danvers, MA 01923, 978-750-8400. CCC is a not-for-profit organization that provides licenses and registration for a variety of users. For organizations that have been granted a photocopy license by the CCC, a separate system of payment has been arranged. Trademark Notice: Product or corporate names may be trademarks or registered trademarks, and are used only for identification and explanation without intent to infringe.
Library of Congress Cataloging-in-Publication Data Authorxxxxxxxxxxxxxxxxxxxxxxxxx titlexxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx p. cm. xxxxxxxxxxxxx Includes bibliographical references and index. ISBN 0-0000-0000-0 xxxxxxxxxx 1xxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxxx xxxxxxxxxxx xxxxxxxxxxx xxx xxx xxx xxxx xxxx xxxxxxxxxxxxx 2004 xxxxxx'xx--dc22
XX-XXXXX
Visit the Taylor & Francis Web site at http://www.taylorandfrancis.com Taylor & Francis Group is the Academic Division of T&F Informa plc.
and the CRC Press Web site at http://www.crcpress.com
Foreword With Volume 44 of the Advances in Chromatography series I say goodbye, reluctantly, to Phyllis R. Brown, who served as one of the editors of the series. Advances in Chromatography was started by J. Calvin Giddings and Roy A. Keller in 1965. Very quickly, the series became an important tool for quick dissemination of new developments in the field. In Volume 12 of the series (1975) Jack Cazes and I joined Giddings and Keller as editors. In 1976 (Volume 14) Phyllis replaced Roy Keller. Jack Cazes left the series in the mid 1980s and Cal Giddings stepped down from his position as executive editor in 1993. Over the last 10 years or so Phyllis and I were the sole editors of the Advances in Chromatography series. Several years ago Phyllis retired from her position as a chemistry professor at the University of Rhode Island, and now she has decided that the time has come to hand over her editorial responsibilities to the younger generation. Nelu Grinberg was selected to replace Phyllis, and with this volume he makes his debut as an editor. Phyllis served as an editor for almost 30 years. In many respects she was the primary moving force of the series. The fact that Advances in Chromatography is still going strong after 40 years is due to her foresight, determination, diligence, and perseverance. Phyllis has a very good eye for recognizing the latest developments in chromatography and related techniques. Her purpose was always to “bring to our readers the latest developments and advances at the forefront of the field.” Using her many acquaintances and contacts, we succeeded in staying on top of the field and providing our readers with up-to-date reviews in all areas of chromatography and other elution techniques. Based on her own research, Phyllis recognized very early the usefulness of HPLC in life sciences. She saw to it that the Advances in Chromatography series educates its readers about the power of the technique as applied to the biological field. I want to thank Phyllis very much for her friendship as well as for her help, encouragement, and patience all through our years together as editors. I would like to welcome Nelu Grinberg as my new co-editor. I have known Nelu for many years and I am well familiar with the caliber of his work and with his commitment, as a scientist, to excellence. Nelu and I will strive to maintain the excellence of the series that was so important to Phyllis. Judging from the cooperation between us in producing the present volume of the Advances in Chromatography, the future of the series remains bright. Eli Grushka Jerusalem
Contributors Andrew J. Alpert PolyLC, Inc. Columbia, Maryland Gabriela Cimpan Sirius Analytical Instruments, Ltd. Forest Row, U.K. John Comer Sirius Analytical Instruments, Ltd. Forest Row, U.K. Mauricio Dantus Merck and Company Rahway, New Jersey Andrew G. Ewing Departments of Chemistry and Neural and Behavioral Sciences The Pennsylvania State University University Park, Pennsylvania Simion Gocan Analytical Chemistry Department Babes-Bolyai University Cluj-Napoca, Romania Michael J. Gray Nanoscale Organisation and Dynamics Group University of Western Sydney Sydney, Australia
Tyge Greibrokk Department of Chemistry University of Oslo Oslo, Norway Y.V. Kazakevich Seton Hall University South Orange, New Jersey R. LoBrutto Novartis Pharmaceutical Corporation East Hanover, New Jersey Elsa Lundanes Department of Chemistry University of Oslo Oslo, Norway Tracy L. Paxon Department of Chemistry The Pennsylvania State University University Park, Pennsylvania R. Andrew Shalliker Nanoscale Organisation and Dynamics Group University of Western Sydney Sydney, Australia Rodger W. Stringham Chiral Technologies, Inc. West Chester, Pennsylvania
Table of Contents Chapter 1 Separations in Multiple-Channel Microchips ...........................................................1 Tracy L. Paxon and Andrew G. Ewing Chapter 2 Temperature Effects in Liquid Chromatography ....................................................45 Elsa Lundanes and Tyge Greibrokk Chapter 3 Lipophilicity Measurements by Liquid Chromatography.......................................79 Simion Gocan, Gabriela Cimpan, and John Comer Chapter 4 Concepts and Practice of Multidimensional High-Performance Liquid Chromatography ....................................................................................................177 R. Andrew Shalliker and Michael J. Gray Chapter 5 High-Performance Liquid Chromatography in the Pharmaceutical Industry: Application, Validation, and Regulatory Issues Under the PAT Framework .......237 Mauricio Dantus Chapter 6 The Use of Polysaccharide Phases in the Separation of Enantiomers.................257 Rodger W. Stringham Chapter 7 Chaotropic Effects in RP-HPLC ...........................................................................291 R. LoBrutto and Y.V. Kazakevich Chapter 8 Chromatography of Difficult and Water-Insoluble Proteins with Organic Solvents....................................................................................................317 Andrew J. Alpert Index......................................................................................................................331
1
Separations in MultipleChannel Microchips Tracy L. Paxon The Pennsylvania State University, University Park
Andrew G. Ewing The Pennsylvania State University, University Park
CONTENTS 1.1 1.2
1.3
1.4 1.5
1.6
1.7
Introduction ......................................................................................................2 Microfabrication of Multiple-Channel Microchips .........................................3 1.2.1 Glass Substrates ...................................................................................3 1.2.2 Polymer Substrates...............................................................................3 Multiple-Channel Microchip Designs .............................................................4 1.3.1 Rectilinear Microchip Designs ............................................................4 1.3.2 Radial Microchip Designs ...................................................................6 1.3.3 Bioassay Microchip Designs ...............................................................8 1.3.4 Continuous Sample Introduction Microchip Designs .........................8 Channel Filling and Sample Loading..............................................................8 Injection Methods ............................................................................................9 1.5.1 T-Type Injections................................................................................10 1.5.2 Optically Gated Injection...................................................................13 1.5.3 Capillary Sample Introduction...........................................................15 Detection Methods .........................................................................................17 1.6.1 Single-Point Confocal Detection .......................................................18 1.6.2 Translation-Stage Confocal Detection...............................................18 1.6.3 Scanning Methods of Detection ........................................................19 1.6.3.1 Confocal Fluorescence Detection with a Galvanometric Scanner.......................................................19 1.6.3.2 Confocal Fluorescence Detection with a Rotary Scanner....................................................................19 1.6.3.3 Fluorescence Detection with an Acousto-Optical Deflection Scanner..............................................................22 1.6.4 Fluorescence Detection with a Charge-Coupled Device (CCD) ......22 Applications of Multiple-Channel Microchips..............................................23 1.7.1 DNA Separations ...............................................................................24
1
2
Advances in Chromatography
1.7.1.1
Separations Involving Restriction Enzyme Digests — DNA Fragment Analysis ....................................................24 1.7.1.2 Genotyping..........................................................................25 1.7.1.3 Sequencing in Multiple-Channel Microchips ....................27 1.7.2 Affinity Electrophoresis for Immunoassays in Microchips ..............29 1.7.3 Dynamic Analysis of Reactions with Multiple-Channel Electrophoresis ...................................................................................30 1.7.3.1 Application to DNA............................................................30 1.7.3.2 Application to RNA Reactions...........................................32 1.8 Conclusions and Future Perspectives ............................................................32 Acknowledgments....................................................................................................35 References................................................................................................................35
1.1 INTRODUCTION Capillary electrophoresis (CE) has been reviewed extensively for a variety of applications [1]. The technique provides fast separation times and small sample requirements; however, CE is a serial technique that significantly limits throughput. To address this, capillary array electrophoresis, in which separations are performed in an array of parallel silica capillaries, was introduced by Mathies [2] and subsequently has been used to perform high-speed, high-throughput deoxyribonucleic acid (DNA) sequencing [3, 4] and DNA fragment sizing [5]. This method has proven to be an invaluable technique for the completion of the human genome project. However, sample introduction into a large number of capillaries and the physical manipulation of these capillaries can be difficult and cumbersome. The miniaturization of capillary electrophoresis onto a microchip was thus realized in the early 1990s [6] and miniaturization has been used to increase performance by reducing analysis times and reagent volumes [6−9]. Since then, extensive advances have been made in the area of microchip technology, allowing for the analysis of a variety of substances including amino acids [10], small drug molecules [11], peptides [12], oligonucleotides [13], proteins [14, 15], and DNA fragments [16]. Initial microchip separation devices consisted of a single separation channel and a T-type injector. Developments have since been made to multiplex many separation channels on a single microchip, thereby providing a means to achieve high throughput separations [16−25]. However, multiple channels require multiple reservoirs, thus increasing the complexity of multi-channel microchip designs. Other factors that must be considered when dealing with multiplexed microchips include fabrication, bonding, filling, sample introduction, and complex electronics required to perform injections and separations, as well as simultaneous detection for multiple separations. The objective of this chapter is to present recent advances in multichannel microchip design, sample introduction, and detection of various analytes for high throughput analyses.
Separations in Multiple-Channel Microchips
3
1.2 MICROFABRICATION OF MULTIPLE-CHANNEL MICROCHIPS Microchip fabrication is based on techniques developed by the semiconductor industry and has traditionally been performed on glass substrates but polymeric microchips have also been produced. The multi-channel microchip fabrication process is similar to the process for single channel microchips; however, the masks become increasingly more complex with an increase in channel number. Additionally, fabrication of large numbers of channels on a single microchip requires the substrates to be essentially free of defects. Thus, the challenge is to fabricate defect-free multiplechannel microchips in a reproducible, cost-effective manner. Developments to this effect have been reviewed by Gao et al. [26], and therefore will be only briefly discussed here.
1.2.1 GLASS SUBSTRATES Channels and other microchip features are typically defined on glass plates using photolithography and wet chemical etching techniques [7, 9, 27−29]. Microchips have been fabricated on soda lime glass slides and, more commonly for multiplechannel microchips, on Borofloat glass wafers. The traditional photolithographic process normally begins with the deposition of various metals followed by layers of photoresist onto a clean glass substrate; however, amorphous silicon has recently been used as a sacrificial mask prior to photoresist deposition [16, 30]. In the modified protocol, glass substrates are pre-etched in hydrofluoric acid followed by deposition of a thin layer of amorphous silicon using a plasma-enhanced chemical vapor deposition system [30]. Spin-coating a layer of photoresist onto the plate and subsequent illumination of the substrate through a mask containing the multiplechannel pattern follows. The photoresist is developed and the channel pattern is transferred to the silicon film by reactive etching with SF6 plasma. The channels are then etched into the glass by a concentrated hydrofluoric acid solution. The photoresist layer is stripped off with high-pressure O2 plasma. Holes are drilled into either the etched plate or a second glass plate that has not been subjected to this procedure, to provide access holes to each channel. The plates are finally aligned and thermally bonded, thereby enclosing the channels [31].
1.2.2 POLYMER SUBSTRATES Polymeric materials have been employed as alternative substrates as they lend themselves to mass fabrication of affordable microfluidic devices [32−37]. Currently, the most widely used replication process to fabricate channel structures for separations is hot embossing [32, 33, 38, 39]. After fabrication of the master, the master and planar polymer substrate are heated, placed under vacuum and brought together with a controlled force. This force presses the channel structure into the polymer, creating the desired microstructures. Alternatively, injection molding is used in polymer microchip fabrication. In this process, the raw polymer material in granular form is melted and injected under high pressure into an evacuated cavity containing
4
Advances in Chromatography
the master [32, 40–42]. The molten polymer cools in a few seconds, allowing for rapid production of identical structures. Casting is another polymer microchip fabrication method in which a mixture of an elastomer precursor and its curing agent are poured over the master template [32, 34, 36, 37]. After a few hours, the solidified polymer is removed from the master, exposing the desired channel structure. Regardless of the fabrication technique used, the open channels must be bonded to a planar surface to form closed channels that can be used for separations. There are a variety of methods for bonding including lamination, gluing, heat, and pressure application, laser welding, and ultrasonic welding. For more information on these polymer methods, refer to Becker and Gartner [32].
1.3 MULTIPLE-CHANNEL MICROCHIP DESIGNS Significant advances have been made over the last 10 years in the area of multiplechannel microchips. Channel layouts have progressed from single channels, with as many as four reservoirs for sample introduction, to as many as 384 channels with 481 reservoirs for separation and sample introduction, respectively, as demonstrated by Mathies and coworkers [31]. Since the fabrication of multiple features on a glass microchip does not significantly increase the cost of production, as compared to fabrication of a single feature, parallelization is a cost effective way to increase throughput. All of the components of a microchip are fabricated on a single substrate. However, as the number of separation channels and reservoirs increases, so does the complexity of sample loading, electrical connections to apply injection and separation potentials, as well as parallel channel detection. Thus, microchip design is continually being improved to achieve higher throughput and address the challenges of chip complexity. Several chip designs have emerged from such development and include rectilinear, radial, and bioassay microchips.
1.3.1 RECTILINEAR MICROCHIP DESIGNS Parallel analysis on a microchip was demonstrated in 1994 by Wooley and Mathies when they performed ultra-high-speed DNA fragment separations on a parallelchannel microchip [43]. In this microchip design, 15 channels are etched into soda lime glass. The channel dimensions vary in size to include 3.5 cm long by 30, 50, and 70 μm wide channels, with 30, 70, and 120 μm cross-channels. Their work has characterized two different injection methods — stack injection and plug injection — and addresses both the effect of channel geometry and the effect of electric field on separation. Procedures to fill these channels with hydroxyethyl cellulose- (HEC) sieving matrices has been developed, and separations of FX174 Hae III fragments with excellent resolution achieved in less than 2 minutes. Additionally, ultra-highspeed DNA sequencing has been accomplished on similar microchips using fourcolor fluorescence detection [44]. In an attempt to simplify detection, Woolley and coworkers [24] have designed and fabricated a capillary array electrophoresis microchip (Figure 1.1a) that has the capacity to analyze 12 different samples in parallel in under 160 seconds. This microchip design consists of 12 independent pairs of injection (A, 8 mm long) and
Separations in Multiple-Channel Microchips
5
FIGURE 1.1 (a) Mask design for the 12-channel capillary array electrophoresis chip for highspeed DNA genotyping. (From Woolley, 1997.) (b) Mask design of the 16-channel capillary electrophoresis chip for parallel DNA sequencing. Sixteen identical 250 μm twin-T injectors are used for all channels. (Inset) Detail of the compensation of different channel lengths. (From Liu, 2000.) (c) Mask pattern for the 96-sample capillary array electrophoresis microplate for genetic analysis. (From Simpson, 1998.)
separation (B, 8 μm deep, 60 μm wide, and 60 mm long) channels, as well as two optical alignment channels (C), which are patterned and etched on a glass microscope slide. With just 12 channels, 41 access holes are required to provide sample, buffer, and waste reservoirs on this microchip. These studies provide a separation method in which high-speed, parallel separations of DNA restriction fragments and PCR products can be analyzed. Furthermore, this microchip design demonstrates that microfabricated multi-channel microchips provide a powerful method for rapid highthroughput DNA diagnostics in which ten base pair (bp) resolution has been achieved. Studies to perform DNA sequencing on a 16-channel microchip (Figure 1.1b) has been completed by Liu et al. [45] This microchip is fabricated on a 10-cm diameter wafer with two eight-channel groups, each with a common anode reservoir. Each channel has its own sample and waste reservoir in the twin-T injector configuration as well as its own cathode reservoir. The Mathies group has expanded upon previous work by creating a capillary array electrophoresis microchip in which two different samples are serially separated in 48 parallel channels for a total of 96 sample analyses [30]. The microchip design
6
Advances in Chromatography
employed in these experiments consists of 48 individual separation channels etched in a 150 μm periodic array in the detection region (Figure 1.1c) [30]. The channels branch out to form an 8 × 12 array of reservoirs. These arrays of reservoirs are spaced 9 mm apart such that samples and buffer can be loaded via an eight-channel pipettor. The separation channels are 10 cm in length, spanning the area between the injection region and the anode. Two injection reservoirs (B and C) are coupled to each separation channel, thus allowing for individual serial injections into each channel. In order to simplify the microchip design, the researchers group the channels such that only one waste reservoir (D) is required for every four channels. By the same rationale, the cathode reservoirs (E) are connected to groups of six or twelve channels. The anode (F) is placed off-center and is common to all 48 channels. Overall, 127 holes are drilled into the etched plate to provide access to the channels, followed by thermal bonding to a flat glass wafer. The investigators report high quality bonding over the entire surface of the microchip. Using this device, 96 samples have been separated in two runs of 48 samples (corresponding to the samples in reservoirs B and C) in less than 8 minutes. Nineteen different samples have been dispensed among the 96 sample reservoirs, thereby providing five-fold redundancy in sample analysis. This work further multiplexes separation channels on a microchip by combining reservoirs to reduce the number of access holes thereby creating a more simplified device; however, this design can result in sample cross-contamination. Serial injections are an effective method of increasing the sample throughput with a limited number of separation channels and will be discussed in detail later in this review.
1.3.2 RADIAL MICROCHIP DESIGNS The design, fabrication, and operation of a radial capillary array electrophoresis microchip and scanner for high-throughput DNA analysis was presented in 1999 and has been applied in various DNA applications including genotyping, sequencing and single nucleotide polymorphism typing [16, 30, 46–49]. A 96-channel radial microplate (Figure 1.2a) has more recently been used for the detection of 2,4,6trinitrotoluene (TNT) [50]. These applications are discussed later in this review. This microchip consists of 96 individual channels in a radial design with a common anode reservoir in the center of a circular 10 cm-diameter wafer and an array of channels extending toward injector units at the perimeter of the wafer [16]. The distance from the injector to the detection point is 33 mm. Channels are grouped into four 24channel quadrants, each terminating at the center of the microchip. Access holes for each of the reservoirs have been drilled prior to thermal bonding. This radial design is preferred over former rectilinear designs [24, 30] because it is easier to fabricate and all of the channels are straight, thereby eliminating resolution problems associated with curved channels [51]. Additionally, the radial design allows for the use of a radial scanner for detection purposes. This will be discussed further in the section on detectors [16]. Additional studies by the Mathies group have led to the development of a 96folded-channel radial microchip (Figure 1.2b) in order to improve separation efficiency [49]. Initial studies considered the parameters that cause turn-induced band
Separations in Multiple-Channel Microchips
7
FIGURE 1.2 (a) Mask pattern for the 96-channel radial capillary array electrophoresis microchip utilized for various DNA applications including genotyping and single nucleotide polymorphism typing, as well as the detection of 2,4,6-trinitrotoluene. (From Medintz, 2000.) (b) Mask pattern used to form a 96-channel radial capillary array electrophoresis microplate with folded channels for increased (16 cm) separation length. Enlarged view of the folded-channel layout and the novel hyper-turn geometry. (From Medintz, 2001.) (c) Layout of the 384-lane capillary array electrophoresis microchip on a 200 mm diameter wafer for ultra-high throughput genetic analysis. (From Emrich, 2002.)
broadening and investigated ways to modify channel geometry such that the broadening could be reduced [52]. These devices are a modified version of the aforementioned 96-channel microchips and have been fabricated using the same method. By using optimized folded channels, separation quality is maintained while the separation efficiency is significantly enhanced due to the increased channel length. The investigators suggest these devices will be useful in the development of optimized DNA-sequencing multi-channel microchips with enhanced performance, throughput, and device packing [47, 49, 52, 53]. In fact, the 96-folded-channel radial microchip has been used for DNA genotyping and sequencing [47, 49]. The pioneering work with 96-channel radial microchips has enabled the design, fabrication, and implementation of a 384-channel radial microchip (Figure 1.2c)
8
Advances in Chromatography
[31] for massively parallel genetic analysis, providing a new tool for accelerating gene mapping, pharmacogenomic screening, forensics, and proteomics. This microchip design consists of 4 quadrants of 24 quartets distributed radially around a common anode reservoir. Each channel has its own unique sample reservoir and each quartet shares a common cathode. This microchip contains no waste reservoirs; therefore, injections are made directly onto the separation lanes. The elimination of waste reservoirs allows for more tightly packed channels but also results in less well-defined sample plugs, as described below. The researchers report that using a multiple-step electrokinetic injection method along with short injection times helps to minimize the size-biased injection effect of the less well-defined sample plug [31].
1.3.3 BIOASSAY MICROCHIP DESIGNS To date, DNA analysis has been the major focus of the multiplexed microchip designs presented but work to integrate immunoassays onto microchips is underway in several laboratories. Initial studies have focused on single channel devices [54–61], although there has been a recent desire to incorporate multiple bioassay channels onto a single device [50, 62–66]. To this end, Cheng et al. [62] have designed a sixchannel microfluidic immunoassay device. This device consists of six independent mixing, reaction, and separation channels as well as two channels for alignment purposes (Figure 1.3) [62]. The separation channels are 9.3 cm in length and bent around two or three corners in order to fit on the substrate. The corners are rounded to a curvature of 300 μm such that sample plug distortion is minimized [51]. Using this device, simultaneous direct immunoassays for ovalbumin and for anti-estradiol have been performed. Mixing, reaction, and separation have been performed within 60 seconds in all cases and in less than 30 seconds when optimized conditions are employed. Another high-throughput homogeneous immunoassay for the sensitive detection of TNT has been developed by Bromberg and Mathies using the 96-channel radial capillary array electrophoresis microdevices previously described [50].
1.3.4 CONTINUOUS SAMPLE INTRODUCTION MICROCHIP DESIGNS Several microchip designs have been developed to facilitate sample introduction and will be discussed in detail in the injection methods section. Briefly, the Ewing group has utilized straight channels for multiplexing separations using optically gated sample introduction as well as straight, open channels with a specialized capillary sample introduction scheme [66–69]. These unique methods of sample introduction have allowed dynamic reactions to be studied with minimal user intervention.
1.4 CHANNEL FILLING AND SAMPLE LOADING In single channel microchip electrophoresis, channel filling is straightforward. The separation media is placed in one reservoir and is either vacuum- or pressure-pumped through the lane. Multiplexing this process is not a trivial task but the concept remains the same as for single channel devices. Different methods have been developed to facilitate channel filling and sample loading. A single pipette is often used
Separations in Multiple-Channel Microchips
9
FIGURE 1.3 Illustration of the six-channel layout of the immunoassay device referred to as SPIDY. Each reaction cell has reservoirs for sample (S1), antibody (S2), sample injection waste (SW), and running buffer (B). Expansion on the left illustrates the sample loading double-T injector design. The bottom right expansion illustrates the antibody (Ab) and antigen (Ag) mixer in the reaction cell. The lower left expansion illustrates the pattern of channels at the buffer waste (BW) reservoirs used during separations. Expansion in the upper right of the detection zone, across which a laser beam is swept, illustrates the separation of two components in the size separation channels. (From Cheng, 2001.)
to fill a single channel or multiple channels on a simple microchip; however, as the number of channels on the microchip increases, this process becomes increasingly difficult. In one approach, an 8-channel pipette has been used to load sample onto a 16-channel microchip from a microtiter plate (a second transfer is required to load all 16 samples) [45]. In another approach, a capillary array for filling 96 reservoirs (Figure 1.4) has been developed by the Mathies group [16], providing a simple method for simultaneously transferring separation medium into a radial 96-channel microchip. In this method, pressure is applied and all 96 channels are filled simultaneously. Samples are loaded onto this microchip in a similar manner. The advantage of capillary sample introduction in this method is that the rectilinear format of the microtiter plate can be coupled to the radial design of the previously described highdensity microchips.
1.5 INJECTION METHODS Sample injection is a straightforward process on a microchip but as the number of channels increases, so does the complexity of the electronics required to perform the injections. For a thorough review of sample introduction techniques for
10
Advances in Chromatography
FIGURE 1.4 Schematic of the 96-sample capillary loader. Pressurization of the rectilinear microtiter dish chamber (left) is used to transfer 96 samples to the sample reservoirs of the radial microplate (right). (From Shi, 1999.)
microfabricated separations devices, readers are directed to Roddy et al. [67] Three common methods of sample injection — T-type, optical gating, and capillary sample introduction — will be discussed briefly below.
1.5.1 T-TYPE INJECTIONS T-type injection was originally developed for single lane microchips [6, 7, 70]; however, it has also been extensively applied in the area of multiplexed microchips [31, 71]. Typically, a single channel with a side arm (sample channel) is used to perform these injections. A potential is first applied across the sample channel to load the sample, and the potential is switched and then applied across the separation channel to accomplish separation (Figure 1.5a). Various other T-type injectors have been developed to improve the control of fluid flow within the microchip channels and to increase the reproducibility of sample plug volumes, including the cross-T [24, 25, 43, 72, 73], twin-T [16, 46, 47, 49, 74], and double cross-T [30, 62]. In the cross-T injection design, as developed by Harrison and coworkers, potential is applied across the sample channel, filling the region where the channels intersect. The potential is then switched and applied across the separation channel to accomplish separation (Figure 1.5b) [27]. As shown in Figure 1.1a and Figure 1.1b, the cross-T has been applied to multiplexed microchips; however, the large number of reservoirs required for this approach quickly becomes too cumbersome to individually address. Effenhauser and coworkers designed a double-T injector that has been more recently applied to multi-channel microchips [9, 13, 30, 62, 75] and requires fewer reservoirs. The Mathies group has utilized a double-T injection scheme on their rectilinear 48-channel microchip (Figure 1.6) [30] in an attempt to minimize the number of separation channels. As a result, twice as many samples can be analyzed with half the number of separation channels. The development of radial microchips has allowed for a significant reduction of reservoirs by incorporating a single common anode at the center of the microchip [16, 31, 46, 48–50, 53, 74]. Each of the radial
Separations in Multiple-Channel Microchips
Buffer reservoir
11
Buffer reservoir +
+ Sample reservoir +
Sample reservoir Sample waste +
Waste reservoir
Buffer waste
Injection
Separation (a)
Injection
Separation (b)
FIGURE 1.5 (a) Typical microchip schematic including sample injection (left) and sample separation (right) using T-type injection. Potential is applied between positive and ground. The arrows indicate the direction of flow in the channel. (Adapted from Manz, 1992.) (b) Typical microchip schematic including sample injection (left) and separation (right) using a cross T-injector. Potential is applied between positive and ground. The arrows indicate the direction of flow in the channel. (From Harrison, 1993.)
microchips discussed in this chapter utilize various T-type injection schemes. Both the 96-straight-channel microchip and the 96-folded-channel microchip function with twin-T injectors [16, 46–49, 53, 74] (Figure 1.7a) [74]. However, the 384channel microchip utilizes the simpler, more compact T-type injector (Figure 1.7b) [31]. Additional studies have focused on investigating both the T-type and cross-T injection configurations in narrow sample channels [71, 72] to eliminate the need for leakage control during the injection process (Figure 1.8) [72]. In this design, eight parallel separation channels with different injection designs are employed with a common anode and a common cathode for the application of separation potential. The sample channels are of smaller dimension than traditional sizes used and have been studied to optimize injection on microfabricated devices. The fabrication of these microchips involves production of a glass master and subsequent pouring of poly(dimethylsiloxane) (PDMS) onto the master to form the channel structures. Each of the eight channels is individually equipped with a set of sample injection and waste reservoirs by punching holes in the PDMS layer at both ends of the sample channel. In this design, the sample channel widths are one fifth that of the separation channel. Using these narrow sample channels, a 10-fold increase in sensitivity and a 2-fold increase in column efficiency over conventional injection methods have been reported. Additionally, the narrow sample channels decrease injection volumes; thus, no leakage control is required and this narrow channel injection method provides a simplified version of traditional cross and T-type injectors [62]. Additional studies by Beard and coworkers have investigated field-amplified sample stacking of biogenic amines on microfabricated electrophoresis devices [71]. The stacking method utilizes T-type injection through narrow sample channels on a six-channel microchip where all channels share both the anodic and cathodic
12
Advances in Chromatography Sample 1
Sample 2
B
C
Cathode
Anode D
Waste
Cathode
Anode Sample 1
Sample 2
B
C (a)
Waste
Waste
Cathode
Anode
Cathode
Anode
Sample 1
Sample 2
Sample 1
Sample 2
(b)
(c)
Waste
Waste
Cathode
Anode
Cathode
Anode
Sample 1
Sample 2
Sample 1
Sample 2
(d)
(e)
FIGURE 1.6 (a) Layout of the double-T sample injector. The sample injector includes four sample reservoirs, two separation columns, and one common waste reservoir, labeled as in Figure 1.1c. (b–e) Fluorescence images illustrating the operation of the injector with fluorescein. (From Simpson, 1998).
reservoirs (Figure 1.9a) [71]. The distances between the injection intersection and the buffer inlet determine the length of the sample plug, and voltage switching on a microchip with a common anode and cathode provides a simple method of simultaneously stacking samples in multiple channels for subsequent separation (Figure 1.9b) [71].
Separations in Multiple-Channel Microchips
13
33 mm
Sample
Waste Cathode
Detection Common area anode
Twin-T injector (a)
8.0 cm
Detection point (b)
Cathode 0
S3
S2
5 S1
S4 10 (mm) (c)
FIGURE 1.7 (a) Twin-T injector design for 96-channel radial capillary array electrophoresis microplate. Two individual channels share common cathode, waste, and anode reservoirs. (From Mednitz, 2000.) (b) Single quartet of channels with their T-type injectors for 384channel radial capillary array electrophoresis microplate. All channels in a quartet share a common cathode reservoir located closest to the edge of the wafer. The T-type injector design is shown in detail below. (From Emrich, 2002.)
1.5.2 OPTICALLY GATED INJECTION Optical gating, first developed for liquid chromatography in small diameter capillaries [76, 77] and for capillary electrophoresis [78], provides an alternative to traditional T-type injection on a microchip. This injection method requires no potential switching and requires only two reservoirs per channel, lending itself nicely to multiplexed microchip electrophoresis. This injection technique, introduced on a single channel microchip by Lapos and Ewing [79], has been used to demonstrate fast, serial, and reproducible injections on a single-channel microchip. In order to
14
Advances in Chromatography
Power supplies
Platinum wire
Buffer inlet
Injectors
Detection window
Conventional cross injector
NSC cross injector
NSC tee injector
Buffer outlet
Power supplies
FIGURE 1.8 Schematic layout of a chip composed of eight channels with different injection designs. On the right are the expanded views of three typical injectors: conventional cross, narrow sample channel cross, and narrow sample channel T-injectors. (From Zhang, 2001.)
achieve optically gated sample introduction, the laser is split into two beams. The more intense gating beam is used to perform sample introduction by time-discriminated photobleaching of the sample, and the less intense beam is used for detection (Figure 1.10a). To perform an injection, potential is applied and fluorescently-labeled sample continuously flows from the sample reservoir to the waste reservoir. Initially, the gating laser beam continuously photobleaches fluorescently labeled analytes as they pass through the channel. The gating beam is then briefly blocked, thereby allowing a small, unphotobleached sample plug to move past the injection window. This plug is separated and finally detected by laser-induced fluorescence detection as it passes through the probe beam. Through the entire injection, separation, and detection process, the potential never needs to be switched or toggled; therefore, the electronics required for optical gating remain extremely simple compared to traditional injection methods. As previously mentioned, optically gated sample introduction requires fewer reservoirs than other injection methods and does not require side arms for injection. Therefore, optically gated injection should allow easier integration of a large number of separation channels onto a single microchip than traditional injection methods. In a demonstration of this principle, a schematic of a multiple-channel optical gating microchip (Figure 1.10b) shows five 8-cm channels that converge in the center of the microchip to facilitate detection [22]. Since the advent of this microchip injection method, optical gating has been applied to the analysis of amino acids [22, 79], DNA fragments [23, 80], and enzyme reactions [61].
Separations in Multiple-Channel Microchips
15
Buffer outlet (floating)
Power supply *Diagram not to scale
1
100 um
Buffer
Buffer outlet Sample reservoir
Detection window
Sample
Sample loading
Platinum wires 1
2
3
4
5
Buffer inlet (ground) Buffer outlet (ground) Buffer
6 2
100 um
Sample
Buffer inlet
Sample reservoir Stack & separate (floating)
Buffer
Buffer inlet
(a)
(b)
FIGURE 1.9 (a) Schematic layout of the six-channel capillary electrophoresis array microdevice used to investigate sample stacking on microfabricated electrophoresis devices. (b) Illustration of the stacking process within the channel network with supporting CCD footage: (1) Sample loading. (2) Sample stacking and separation. (From Beard, 2003.)
1.5.3 CAPILLARY SAMPLE INTRODUCTION In contrast to most of the injection methods reviewed here, the capillary sample introduction technique is another scheme that does not require the toggling of voltages to perform injection. The use of a capillary for sample introduction eliminates the need for individual sample and waste reservoirs, making this microchip design quite simple (Figure 1.11a). Additionally, capillary sample introduction provides an effective means to interface the microchip with the macroscopic world. In this approach, the lanes of the microchip are filled with linear polyarylamide and plugs of sample are injected into each of the lanes (Figure 1.11b) either electrokinetically or by pressure [66, 68, 69]. Without interrupting the separation potential, injections are performed via a computer-controlled micromanipulator. In one example using capillary introduction, a five-channel microchip specifically fabricated to accommodate automated capillary sample introduction in a continuous manner has been used. The microchip design involves five open-ended channels such that one common anode and one common cathode can be employed [68]. Traditional photolithographic techniques have been used to create a microchip with parallel channels 12.7 cm in length, 500 to 700 μm wide, and 250 to 350 μm deep. The etched bottom plate is thermally bonded to a flat top plate, creating five enclosed straight channels for the analysis of various biological reactions. Given reasonable parameters, it is
16
Advances in Chromatography
Gating beam
Probe beam
Sample reservoir
Waste reservoir
1
2
3
4
e
5 (a)
2 mm
3 cm 8 cm (b)
Separations in Multiple-Channel Microchips
17
FIGURE 1.10 (a) Schematic illustrating optically gated injection. Fluorescently labeled sample is continuously electrophoresed through the channel. The gating beam is used to perform sample introduction by time-discriminated photobleaching of the sample. The gating beam continuously photobleaches fluorescently labeled analytes as they pass through the channel (1). An injection is accomplished by blocking the gating for a specific time (2); thus, a plug of unphotobleached samples moves past the injection window (3), is separated (4) and detected (5) by laser-induced fluorescence detection as it passes through the probe beam. (b) Schematic of five-channel optical gating microchip. (From Xu, 2002.)
theoretically possible to carry out 550 separations in 1 hour by increasing the number of lanes on the microchip to 37 [68]. The increased throughput and the simplicity of the chip design and electronics make capillary sample introduction an attractive alternative to the traditional methods of sample introduction.
1.6 DETECTION METHODS Electrochemical detection, mass spectrometry, chemical luminescence, and laserinduced fluorescence have been the most popular methods for detecting separation analytes on single-channel microchips. Readers are directed to a review by Uchiyama et al. [81] for more information on these techniques. While these detection methods
FIGURE 1.11 (a) Chip design for capillary sample introduction with five open separation channels. (A) Microfabricated separation channel. (B) Glass partition. (b) Digital photos in chronological order (A–H) of manipulator-controlled movement of a capillary in and out of two channels of a chip. Injection of sample occurs at (C) and (H). (From Smith, 2001.)
18
Advances in Chromatography
are commonly applied to single channel microchips, detection on multi-channel microchips requires more complicated instrumentation. As a result, laser-induced fluorescence detection is by far the most common detection method for multi-channel microchips due to its inherent sensitivity and adaptability [2, 6]. Many variations of confocal laser-induced fluorescence detection, as developed by the Mathies group in the early 1990s [2, 3, 82], have been adapted for different microchip designs and have been reviewed extensively [3, 4, 26, 83, 84]. Confocal methods, scanning methods, and charge-coupled device (CCD) detection methods will be reviewed in the sections below.
1.6.1 SINGLE-POINT CONFOCAL DETECTION Single-point confocal detection has been previously described [3–5, 82, 85] and provides the basis for all other types of confocal detection. Briefly, a laser beam is collimated and subsequently reflected by a dichroic beam splitter through an objective lens. The beam is focused on the center of the microchip channel at the point of desired detection. Emitted fluoresence and scattered excitation laser light from the channel are collected and collimated by the objective lens. The light is then passed through a dichroic beam splitter and the emission wavelength is focused through an achromatic lens and pinhole that is confocal with the achromatic lens. Only fluorescence from the separation channel in the microchip is passed through the pinhole while scattered light is blocked. This increases the signal-to-noise ratio and allows sensitive detection of a variety of fluorescent molecules. Typically, photomultiplier tubes (PMT) or CCDs are used to detect the emitted fluorescence passing through the pinhole. Confocal detection has been applied to multi-channel microchips for detection in individual channels where multiple confocal detection units have been set up for each channel [45]. The adaptations for multi-channel microchip detection utilizing only a single detector will be discussed below [16, 24, 30, 31, 43, 46, 47, 49, 53, 74, 86].
1.6.2 TRANSLATION-STAGE CONFOCAL DETECTION The development of confocal detection with a translation stage was originally applied to capillary array electrophoresis and called a confocal-fluorescence capillary array scanner [2]. In this method, a computer-controlled stage is used to translate the capillary array past the optical system such that data can be collected from all capillaries in the array. Similarly, the translation stage has been adapted for use with multi-channel microchips where the microchip is moved back and forth through the excitation laser beam [22–24, 30, 61]. In one particular application, the 12-channel rectilinear microchip shown in Figure 1.1a is placed on a computer-controlled translation stage [24]. The detection laser beam is focused on the detection region of the microchip, and the microchip is translated through the beam. Fluorescence is collected by the objective, passed through a dichroic beamsplitter, and divided into two fluorescence channels with a second dichroic beamsplitter. The signal from each channel is then focused with a lens onto a confocal spatial filter prior to detection with a PMT (Figure 1.12a) [24]. This detector has been further improved to include
Separations in Multiple-Channel Microchips
19
the detection of four independent fluorophores; however, the four-color work has been completed on a capillary array [87]. In another application, the five-channel optical gating microchip shown in Figure 1.7b has been traversed through both the gating and detection (probe) beams, and fluorescence has been collected in the traditional confocal manner (Figure 1.12b) [23]. In this arrangement, the microchip is fixed to a translation stage that is attached to a voice coil actuator. The actuator is used to move the microchip back and forth at a repetition rate of 10 Hz, causing the two beams to scan repeatedly across all five channels on the microchip. This technique has been used to investigate amino acid [22] as well as oligonucleotide [23] separations. In both of these applications, the microchip is attached to a translation stage and moves through the stationary detection beam.
1.6.3 SCANNING METHODS
OF
DETECTION
A variety of detection methods exist in which the microchip remains stationary while the laser beam is manipulated such that all channels on a multi-channel device can be illuminated and thus detection of multiple channels can be achieved. Galvanometric scanner confocal, rotary scanner confocal, acousto-optical deflection (AOD) scanner, and CCD detection will be reviewed in the following sections. 1.6.3.1 Confocal Fluorescence Detection with a Galvanometric Scanner High-speed, high-throughput separations are characteristic of microchip separations. Detection methods are therefore continuously being improved to allow for the detection of entire peak profiles as a fluorescent analyte is detected. The galvanometric scanner (Figure 1.13a) [30] has been developed and used for high throughput genetic analysis on the microfabricated 96-sample capillary array electrophoresis microchip of the design shown in Figure 1.1c [30]. The laser beam is focused through a 0.33 numerical aperture scan lens system designed for quantitative field analysis to a 5-μm region [88]. The beam is scanned across 48 stationary channels at 40 Hz and data are collected through the detection optics including a dichroic beamsplitter, emission filter, lens, pinhole, and PMT in the same manner as traditional confocal detection. DNA sequencing using four-color detection on a similar instrument has also been carried out with this system [87]. 1.6.3.2 Confocal Fluorescence Detection with a Rotary Scanner The scan speed for detection with a translation stage is limited; therefore, a new confocal detection system with a rotary scanner has been developed (Figure 1.13b) [16]. This detection system consists of a rotating objective head coupled to a fourcolor confocal detection unit. A laser is directed through a series of mirrors and dichroic beamsplitters through the rotating central hollow shaft of a stepper motor. The beam is then displaced by a rhomb prism and focused into the microchip by a
20
Advances in Chromatography
Computer Amplifier
Mirror Spatial filter Lens 645 mm LP filter
PMT
Dichroic beam splitier 488 mm laser
Spectral filter
Mirror Dichrole beam splitier Objective
CAE chip
Translation stage Detection zone
Buffer reservoirs (a)
CPU
Mirrors
Lens Shutter
Argon ion laser
Motor controller
IIV
Beam splitter Gating beam
Chip fixed on a translation stage Objective
Probe beam
Collimator/expander Dichroic mirror
Keithley Voice coil actuator
Filter Photomultiplier tube
(b)
FIGURE 1.12 (a) Schematic of the laser-excited confocal-fluorescence scanner and a capillary array electrophoresis chip. (From Woolley, 1997.) (b) Multichannel system setup with optically gated sample introduction and laser-induced confocal-fluorescence detection. (From Xu, 2004.) The laser beam is stationary in both of these detection schemes while the chip is attached to a stage that translates each channel through the beam.
Separations in Multiple-Channel Microchips
21
FIGURE 1.13 (a) Schematic of the laser-excited galvanoscanner and the capillary array electrophoresis microplate assembly. (From Simpson, 1998.) (b) Schematic of the rotary confocal-fluorescence scanner used to detect separations on radial capillary array electrophoresis microplates. (From Shi, 1999.)
22
Advances in Chromatography
rotating microscope objective. This objective also serves to collect the emission wavelength derived from the sample. The fluorescence is passed along the reverse optical path through the dichroic beamsplitter to the four-color confocal fluorescence detector [16, 87]. This detection system has proven capable of scanning and collecting information at rates up to 20 revolutions per second with 5000 points collected per revolution per channel. In this detection scheme, the microchip is stationary while the microscope objective rotates and has been employed for various DNA applications as well as for an immunoassay. 1.6.3.3 Fluorescence Detection with an Acousto-Optical Deflection Scanner Thus far, all of the fluorescence detection methods described require the physical movement of either the microchip or the laser beam. Conversely, an AOD scanner does not require physical movement of either the microchip or a laser (Figure 1.14), resulting in scanning rates as high as 30 Hz [73]. This method of detection, as developed by Landers and coworkers [25, 73], allows flexible, high-speed, selfaligning scanning for fluorescence detection in multi-channel microchip electrophoresis. The scanning unit of the AOD scanner is dictated by a transparent crystal that changes the direction of laser light based on the frequency of the acoustic wave propagating through it [21, 25, 73]. In brief, the laser light is refracted as it travels through the crystal and the degree of refraction is adjusted by changing the frequency of the acoustic wave generated within the crystal using a piezoelectric transducer. A computer-controlled voltage-to-frequency converter controls this transducer, moving the laser from channel to channel. The AOD scanner is unique in that it is possible to address individual, nonadjacent channels and thus data may be collected from any number of channels in any order. Using this method, a dichroic beamsplitter, an objective, and a bandpass filter have been used to collect emission wavelength light, while a CCD camera and a video capture device have been used in place of a PMT for detection purposes for the investigation of amino acids as well as DNA digest samples.
1.6.4 FLUORESCENCE DETECTION DEVICE (CCD)
WITH A
CHARGE-COUPLED
Clearly, addressing each channel on a microchip can be challenging. Most detection methods reviewed thus far have utilized some variation of confocal detection. Confocal methods, however, are not the only way to monitor fluorescence detection on a multiple-channel microchip. In the past, CCDs have been thought to be incompatible with multi-channel microchip separations [26]; however, CCDs have been used successfully on capillary array electrophoresis systems [89–91] and are now being applied to microchip separations. Ewing and coworkers have used a detection scheme in which an argon ion laser beam is expanded and focused into a line across all channels on a microfabricated microchip (Figure 1.15) [68]. This is similar to detection schemes developed earlier for ultrathin slab gel electrophoresis [92–97].
Separations in Multiple-Channel Microchips
23
FIGURE 1.14 Acousto-optic detection system and multichannel chip. (a) Deflection of the incident laser beam by the acousto-optic crystal. Deflection angle is determined by the frequency of the acoustic wave. (b) Optical setup. (c) Eight-channel microchip configuration. (d) Detection region of the microchip for laser beam scanning. (From Ferrance, 2001.)
As sample passes through the laser line, a CCD camera perpendicular to the microchip detects the emission fluorescence [66, 68, 69]. Pictures are taken at predetermined intervals and compiled such that a complete image of the separations in all lanes is obtained. This detection method allows simultaneous detection of analytes from separations occurring in each channel, thereby providing another methodology that physically eliminates the need for translating the microchip or scanning the laser beam across all channels.
1.7 APPLICATIONS OF MULTIPLE-CHANNEL MICROCHIPS Microfluidic devices continue to be utilized for an increasing number of applications. For instance, single-channel devices have been applied to the analysis of DNA for sizing, genotyping, and sequencing as well as for the integration of PCR on a microchip. Many single-channel devices have also been used for the analysis of proteins, peptides, tryptic digests, enzyme assays, immunoassays, catecholamines, and many other biological analyses that have been reviewed by Bruin [98]. The current chapter has presented multiple-channel microchip designs, injection and detection methods, and will now discuss several bioanalytical applications of these multiplexed devices.
24
Advances in Chromatography
FIGURE 1.15 Charge-coupled device detection. Schematic of the chip setup with capillary sample introduction. Laser-induced fluorescence detection is achieved by focusing an argon ion laser into a line across the chip. A liquid nitrogen-cooled CCD is used to collect the fluoresence. (Adapted from Smith, 2001.)
1.7.1 DNA SEPARATIONS 1.7.1.1 Separations Involving Restriction Enzyme Digests — DNA Fragment Analysis Restriction enzymes are DNA-cutting enzymes found in bacteria. Because they cut within the molecule, they are often called restriction endonucleases. A restriction enzyme recognizes and cuts DNA only at a particular sequence of nucleotides, typically four to eight base pairs in length [99]. Restriction enzymes are essential for analyzing chromosome structures, mapping and sequencing very long DNA molecules, isolating genes, and creating new DNA molecules for cloning purposes. DNA fragments produced by the action of a restriction enzyme can serve as a fingerprint of an overall DNA molecule. Multiple-channel microchips provide a method for the rapid analysis of these DNA fragments. Woolley and Mathies carried out a separation of FX174 Hae III DNA restriction fragments on a 15-channel microchip, proving that microchips are capable of highresolution electrophoretic separations [43]. These separations have been completed in only 120 seconds [43]. Additionally, this process demonstrate the high-speed sizing of PCR-amplified HLA-DQa alleles [43]. This experiment laid the groundwork for high-speed, high-throughput DNA separations on multiple-channel microchips. Simultaneous separation and detection of the double-stranded DNA (dsDNA) fragments of pBR322 has been accomplished by Huang et al. on an eight-channel microchip using AOD [25]. Further studies conducted by the Mathies group employ
Separations in Multiple-Channel Microchips
25
a radial 96-channel microchip — initially using one-color detection — for the separation of 96 pBR322 MspI samples labeled with TOTO [16]. In addition to separations of short DNA fragments, there are examples in the literature of microchip separations of long DNA molecules. A nanofluidic channel device, consisting of many entropic traps, has been designed and fabricated for the separation of long DNA molecules [100]. This nongel-based method for the separation of DNA has been used to separate a T2–T7 DNA mixture [100]. The device used is composed of multiple channels with narrow constrictions and wider regions that cause size-dependent trapping of DNA at the onset of a constriction. Under these conditions, electrophoretic mobility differences are created and separation without the use of a gel matrix or pulsed electric fields could be accomplished. With this technology, samples of long DNA molecules have been efficiently separated, and multiple-channel devices operating in parallel have been demonstrated [100]. Further development of this technique has demonstrated that a microfabricated entropic trap array can be useful in separating large (5 to 200 kbp) DNA molecules efficiently (within 30 minutes) by DC electrophoresis on a microchip platform without a sieving matrix [101]. The investigators report further development of the technique, with emphasis on the optimization of separation selectivity and resolution for these DNA, and comment on the requirement and feasibility of separating megabase-pair DNA with the entropic trap array device [101]. Membrane-mediated restriction digestion and rapid analysis of cleaved DNA fragments has been accomplished using a 96-channel gel microchip electrophoresis separation system [58]. Complete digestion of an FX174 plasmid DNA by Hae III (a 101 base pair PCR product of the pUC19 cloning vector by Sma I) as well as the GFP/pABarMCS construct by Bst EII, Hind III, Bam HI and by the combination of Sac I and Sma I enzymes, have been accomplished in a few minutes on a membrane. Membrane-mediated sample loading [102] has been used to inject these samples into the separation device. The researchers report the possibility of the analysis of up to 6400 samples per day, particularly when robotic liquid handling is employed [58]. Continuous monitoring of a restriction enzyme digest of DNA on a five-channel microchip has also been demonstrated; however, this digestion was monitored dynamically and will be presented in a different application later in this review [66]. A variety of multiple channel electrophoresis devices have been utilized for the analysis of DNA fragments and as technology progresses these multiplexed microchips will be employed for various other DNA applications. 1.7.1.2 Genotyping The development of safer and more effective individualized medicine is possible through the increased understanding of how a particular genotype interacts with a potential drug. Traditional methods of genotyping, typically slab-gel electrophoresis and more recently capillary electrophoresis, are serial techniques that make them both cumbersome and labor intensive. Thus, these techniques are not well suited for large-scale genetic testing. Multi-channel microchip electrophoresis has proven to be a high-throughput, cost-effective alternative to traditional methods.
26
Advances in Chromatography
Hereditary hemochromatosis (HHC) is associated with an increased risk of cirrhosis, cardiomyopathy, and hepatic cancer; thus, early detection through rapid and low-cost screenings can allow early diagnosis and treatment, and a normal life span for patients. Towards this goal, genotyping of HLA-H, a marker gene for the diagnosis of HHC, has been performed on a 12-channel microchip to demonstrate the rapid analysis of biologically relevant samples [24]. Further studies have been conducted to demonstrate the functionality of a microfabricated 96-sample capillary array electrophoresis microchip for the separation and fluorescence detection of a restriction fragment marker for the diagnosis of HHC [30]. This restriction fragment marker for HHC is derived from the HFE gene and separations of 96 HFE amplicons have been performed on a 48-channel microchip [30]. A total of 19 different samples have been used for a 5-fold redundancy in sample analysis. Thus, serial injections proved to be an effective method of increasing the sample throughput with a limited number of separation channels [30]. Furthermore, this work illustrates the utility of multi-channel microchips for genetic disease screening. In an attempt to reduce the cost of high-throughput applications such as population screening for genetic diseases, Sassi and coworkers have demonstrated rapid, parallel separations of D1S80 alleles in a plastic 16-channel microchip [86]. D1S80 alleles are amplified from genomic DNA extracted from whole blood and separated. One of every four channels contains the reference alleles. Ultimately, the authors have demonstrated the feasibility to perform high-throughput genotyping in injection-molded, plastic multi-channel microchips. They report that with appropriate internal standards and optimization of microchip design and separation conditions, the accuracy of typing on these polymeric microchips could be improved to match the accuracy of conventional slab-gel electrophoresis [86]. In another genotyping experiment, the 96-channel radial microchip described earlier in this review has been used to perform an analysis of methylenetetrahydrofolate reductase (MTHFR) allelic variation. To accomplish this, PCR amplicons of the MTHFR gene are digested with Hinf I and detected in two-color mode [16]. The same 96-channel device is used to perform high-speed single nucleotide polymorphism (SNP) typing of an HHC mutation [74]. In these studies, fluorescently labeled sequence-specific primers are used to generate allele-specific products for the S65C polymorphism. The PCR product is subsequently separated and typed. Thus, the authors have demonstrated the potential of using covalently labeled sequence specific PCR amplicons with multi-channel microdevices for high-throughput SNP genotyping analyses [74]. Another study has focused on genotyping energy-transfer-cassette-labeled, short-tandem-repeat amplicons with four-color detection [46]. Covalent labeling with energy-transfer (ET) tags can be used to increase the throughput of genotyping analyses using multi-channel microchip electrophoresis. In this study, 122 samples are separated and genotyped in 96 channels in less than 8 minutes [46]. Thus, the work demonstrates the powerful combination of energy-transfer-cassette labeling and multi-channel electrophoresis for high-performance, short-tandem-repeat analysis. Similarly, demonstration of the combined use of an allele-specific polymerase chain reaction (ASPCR) with energy transfer primers and multiple-channel microchip electrophoresis has also been conducted on the 96-channel radial microchip
Separations in Multiple-Channel Microchips
27
[48]. Previous analyses have employed one- or two-color detection; however, this study demonstrates the use of four-color DNA analysis for genotyping using ET labels [48]. A different type of 96-channel gel microchip has been used for high-throughput genotyping of repeat polymorphisms in the regulatory region of the serotonin transporter gene (5-HTTLPR) [103]. To accomplish this, the injection end of the gel features a straight edge with no individual wells. Samples are injected onto the gel microchip by membrane-mediated loading technology using 96-tab membranes. Automated spotting of the loading membrane is accomplished with a robot directly from a 96-well PCR plate [103]. The researchers report separation and visualization of the PCR amplicons from either the 5-HTTLPR short allele or the 5-HTTLPR long form in a few minutes [103]. Massively parallel genetic analysis is attractive for high-throughput evaluation of genetic variation and multi-channel microchips are a valuable tool in this endeavor. A microfabricated 384-channel microchip has been developed and utilized for massively parallel genetic analysis. For instance, the simultaneous genotyping of 384 individuals for the common hemochromatosis-linked H63D mutation in the human HFE gene has been achieved in only 325 seconds (Figure 1.16) [31]. The H63D mutation is detected by digestion of a PCR product with Mbo I. Mutant amplicons remain undigested while wild-type amplicons are cleaved into two fragments [31]. This study introduces a powerful new tool for accelerating gene mapping, pharmacogenomic screening, forensics and proteomics as well as providing a massively high-throughput bioanalyzer [31]. Significant advances have been made in the area of genotyping on multi-channel microchips in the recent past. The ability to perform rapid high-throughput population screening on integrated multiplexed devices in the clinical setting can be envisioned in the near future. 1.7.1.3 Sequencing in Multiple-Channel Microchips DNA sequencing involves the determination of the precise sequence of nucleotides in a sample of DNA and thus requires excellent resolution of separated peaks such that strands that differ by only a single nucleotide can be differentiated. Single capillaries, capillary arrays, and single channel microchips have been employed for DNA sequencing and have been reviewed elsewhere [104]. Multiple-channel microchips have been applied to DNA sequencing and provide higher throughput than traditional methods [26]. Optimization of high-speed DNA sequencing has been performed in a fourchannel microchip [105]. The separation matrix, separation temperature, channel length and depth, injector dimensions, and injection parameters have been optimized such that one-color sequencing separations of single-stranded M13mp18 DNA could be accomplished [105]. Additionally, four-color sequencing separations have been performed in 7-cm long channels, and experiments completed in under 20 minutes with 99.4% accuracy to DNA sizes up to 500 bp [105]. These optimization studies have significantly advanced sequencing on microfabricated devices. A 48-channel microchip has been utilized to separate single-stranded DNA fragments up to 500 bases in length [106]. Using this separation device, a 400-base
28
Advances in Chromatography
FIGURE 1.16 DNA microchip application. Electropherograms collected for massively parallel genetic analysis on the 384-lane microcapillary array electrophoresis chip. RFLP fragments are highlighted in lanes 1 (wild type) and 193 (heterozygote). Genotyping of the samples is possible in 379 lanes (98.7% success rate). Electrophoresis is completed within 325 seconds. (From Emrich, 2002.)
read has been obtained with 97% accuracy and 440 bases are sequenced with 93% accuracy on a 10 cm channel. These experiments utilize a novel transmission imaging spectrograph for detection and provides an increase in sample throughput by color multiplexing [106]. Throughput has been significantly increased but accuracy suffers in these experiments. Additional sequencing data has been collected using a 16-channel microchip as shown in Figure 1.1b. After optimization of injection time, temperature, and template
Separations in Multiple-Channel Microchips
29
concentration, four-color DNA sequencing has been performed in 16 microfabricated channels [45]. Greater than 99% accuracy is observed for the base-called sequence of 543 bases. Separations carried out in less than 15 minutes typically resolve 457 ± 35 bases with 99% accuracy and 512 ± 44 bases with 98% accuracy [45]. DNA sequencing is highly dependent on the separation resolution; hence, several groups have employed longer channels to increase resolution. Devices containing 48 independent straight microchannels, 50 cm in length, have been fabricated [107]. Single-base resolution for read lengths greater than 600 bases has been achieved with this device and the sequence has been called to 640 bases with 98% accuracy [107]. Additionally, a 96-channel hyper-turn microchip has been developed [47, 49]. This microchip is suited for DNA sequencing where long separation lengths are required. The 16-cm-long hyper-turn channel device initially provides 96 parallel, high-quality DNA sequencing separations to approximately 500 bases in less than 30 minutes [47]. More recently, high-throughput DNA sequencing has been performed and coupled to four-color fluorescence detection [49]. Data have been successfully collected from 95 of the 96 channels where an average read length of 430 bp has been obtained. The probability of incorrectly calling or identifying a base for over 80% of the read length is 0.01% [49]. This technology is capable of producing sequencing data at a rate of 1.7 kbp/min, which the authors report is a 5-fold increase over current commercial capillary array electrophoresis instruments [49]. An interesting hybrid device consisting of a microchip containing multiple twinT injectors attached to an array of capillaries that serve as separation channels has been used to sequence over 1000 real-world samples [108]. The investigators report an average Phred20 read of 675 bases in about 70 minutes with a success rate of 91% as compared to similar samples on MegaBACE 1000 where the average Phred20 read is about 550 to 600 bases in 120 minutes with a success rate of about 80 to 90% [108]. Sequencing with a high degree of accuracy is highly desirable when searching for minute differences in DNA sequences. Thus far, multiplexed microchip electrophoresis has proven to be an efficient platform for increasing throughput for sequencing studies.
1.7.2 AFFINITY ELECTROPHORESIS MICROCHIPS
FOR IMMUNOASSAYS IN
The immunoassay is one of the most important analysis methods for biological molecules. Thus, it is clear why so many groups are interested in integrating immunoassay on a microchip, particularly for its application in point-of-care analysis. There are several methods for conducting an immunoassay on a microchip; however, most have been accomplished on single channel devices. A small-volume heterogeneous immunoassay system has been demonstrated by Hayes et al. in microchannels exploiting magnetic manipulations of small paramagnetic particles [56]. Kitamori’s group has integrated heterogeneous immunoassay systems with microfabricated separation devices as well [64, 109, 110]. Most recently, a bead-bed immunoassay system suitable for simultaneous assay of multiple samples has been constructed on a microchip [64]. The microchip used in these experiments utilized branching
30
Advances in Chromatography
channels and four reaction and detection regions. This device has been used for the simultaneous analysis of four interferon γ samples by a three-step sandwich immunoassay [64]. The assay time for the four samples is 50 minutes, a significant improvement over the 35 minutes it takes for one sample to be analyzed in the single channel assay [64]. The Harrison group has been at the forefront of immunoassay on microfabricated devices [111–116] and has recently developed a multi-channel microfluidic analysis system employing affinity capillary electrophoresis for immunoassay [62]. In this application, a six-channel immunoassay device (depicted in Figure 1.3) has been used to complete simultaneous direct immunoassays for ovalbumin and for antiestradiol. Mixing, reaction, and separation are performed in less than 60 seconds, and in 30 seconds under optimized conditions. This work demonstrates the multiplexing of mixing and reaction steps as well as injection and separation, thus providing increased throughput for complex sample processing [62]. Additional information on enzyme assays and immunoassays on microfabricated devices can be found in a review by Guijt and coworkers [57]. Recently, Bromberg and Mathies developed a homogeneous immunoassay for the detection of TNT and some of its analogues utilizing a single channel microfabricated device [117]. Following this, a high-throughput homogeneous immunoassay for the detection of TNT has been developed using 46 channels of the already described 96-channel radial microchip (Figure 1.2a) [50]. This assay is based on competition between the fluorescein-labeled trinitrobenzene derivative (TNB-Fl) and TNT for antibody binding sites. Sample preparation, reagent mixing, and incubation have been performed off the microchip, whereas injection and separation of the reactants and detection of the resulting complex have been performed on the microchip. Data collected from 46 channels showed nine different TNT sample concentrations, and each trace represented three serial injections (Figure 1.17a) [50]. A calibration curve has been constructed by plotting the ratio of signal intensity for free TNB-Fl to the signal intensity for the antibody-TNB-Fl complex as a function of the TNT concentration (Figure 1.17b) [50]. The equilibrium binding constant, binding rate constant, and dissociation rate constant of the labeled TNT have been determined. Consequently, this study demonstrates the feasibility of using multichannel microfabricated devices to simultaneously analyze multiple samples with homogeneous immunoassay methods. In the future, this technology could be used to probe for the presence of explosives in environmental samples or to perform highthroughput clinical assays [50].
1.7.3 DYNAMIC ANALYSIS OF REACTIONS WITH MULTIPLE-CHANNEL ELECTROPHORESIS 1.7.3.1 Application to DNA While the function of restriction enzymes has been explored in the laboratory, much remains to be learned about their structures, their binding characteristics, and their mechanisms of cleavage. A separations-based method of continuously sampling from a DNA restriction enzyme digest has been developed by Roddy et al., providing a
Separations in Multiple-Channel Microchips
1
5
2
10
8
15
Relative fluorescence
31
3 6 4
25
Ratio
20
4
0.2
5 0.1
2
30
6
0 0
35
7
0
100
0
1 200
2
3 300
TNT (ng/ml) (b)
40
8 45
0
9 100 Time (sec)
200
(a)
FIGURE 1.17 TNT microchip bioassay. (a) Multichannel homogeneous TNT immunoassay performed on a microfabricated capillary array electrophoresis device. Nine samples are dispensed to 46 wells (numbered to the left side of the figure). TNT sample concentrations: 0, 0.3, 1, 3, 10, 20, 50, 100, and 300 ng/mL for samples 1–9, respectively. Each trace presents three consecutive injections. The first peak marks the beginning of the 30-second injection, the second peak is the Ab-TNB-FL complex and the third peak is the free TNB-Fl. (b) Calibration curve for TNT assay obtained by plotting the ratio of signal intensity of free TNBFl to the signal intensity of the Ab-TNB-Fl complex as a function of the TNT concentration. Each point is an average of signals obtained from wells of the same TNT concentration. (From Bromberg, 2004).
means to monitor changes in substrate and product concentrations over the course of the reaction [66]. This method utilizes a capillary as a sampling device to transfer sample from a single reaction mixture to a microfabricated device with an array of separation lanes and has been initially optimized by separating Hae III-digested pBR322 [68]. More recently, this methodology has been applied to the analysis of a restriction enzyme digest during the course of digestion. Specifically, a fluorescently-labeled 62 base pair double-stranded DNA fragment containing a single Kpn
32
Advances in Chromatography
I site has been chosen to demonstrate the ability of this multi-channel separation device to monitor the progress of a dynamic reaction [66]. In brief, the reaction is initiated offline and as the sample enteres the microchip, the reaction is quenched on-line. Therefore, each injection into the microchip captures a time-point during the reaction such that all of the fragments are observed [66]. The results of a digest carried out at 45˚C using one unit of Kpn I per 13 ng of DNA have been reconstructed by plotting the fluorescence signal at the laser line collected by the CCD camera over time (Figure 1.18a) [66]. A cross-section of intensity vs. time for the separations shown in lane five yields a series of electropherograms (Figure 1.18b) [66]. A distinct change in sample composition has been observed as time progresses. This methodology has provided continuous analysis of the products of a digest reaction, thereby contributing to the fundamental understanding of restriction enzyme binding and activity as well as enhancing practical knowledge for the use of restriction enzymes as essential tools in the laboratory [66]. 1.7.3.2 Application to RNA Reactions RNA enzymes, or ribozymes, behave similarly to protein-based enzymes; thus, there is heightened interest in understanding the mechanism of cleavage of these molecules. The applications of RNA catalysts can be expanded as kinetic information is ascertained. Kinetic analysis of ribozymes typically involves the tedious process of collecting and quenching reaction time points and then fractionating by polyacrylamide gel electrophoresis (PAGE). As a way to automate and simplify this process, continuous analysis of ribozyme reactions have been demonstrated by Paxon et al. [69] Completely automated capillary sample introduction on a five-channel microchip with laser-induced fluorescence detection is employed to examine a 30-nucleotide ribozyme cleavage reaction that is sensitive to the lead ion. The ribozyme cleavage is initiated by 10 μM Pb2+ and the progress of this reaction is monitored for 60 minutes. A cross-section of a single channel is shown as a relative intensity vs. separation time (min) electropherogram in Figure 1.19a [69]. The first peak in each doublet set of peaks represents the fluorescently labeled 6-mer elution while the second peak is indicative of the 30-mer elution. Raw data from Figure 1.19a have been plotted as peak height vs. injection time and are shown in Figure 1.19b. Product formation and starting material cleavage showed mirror image behavior. The presence of only two peaks has shown to be consistent with a simple cleavage reaction as well as the absence of side reactions, RNA degradation, and alternative conformations. Fraction cleaved vs. injection time has been plotted in Figure 1.19c and kinetic information about this ribozyme has been obtained. This work demonstrated, through the use of a simple ribozyme model, the potential of multi-channel microchips with capillary sample introduction to provide valuable kinetic information for biologically relevant RNA and protein enzymes.
1.8 CONCLUSIONS AND FUTURE PERSPECTIVES Microchip separations have evolved considerably since their inception in the early 1990s, particularly in the area of multiplexing. There are numerous areas where
Separations in Multiple-Channel Microchips
33
0 500
Time (s)
1000 1500 2000 2500 3000 3500 0
200
0
400 600 Pixel number
800
1000
500 1000 1500 2000 2500 3000 3500
Intensity (a)
2500
Intensity
2000 1500 1000 500 0 0
10
20
30
40
50
60
Time (min) (b)
FIGURE 1.18 DNA digestion assay. (a) Reconstructed plot of time vs. pixel number across the width of the chip. The shaded bar represents fluorescence intensity collected by the CCD. The capillary moved right to left depositing samples into the chip, so the lane farthest to the right is lane 1. (b) Cross section through lane 5 of the plot in (a) showing intensity vs. time. (From Roddy, 2003.)
34
Advances in Chromatography
Relative intensity
7000 5250 3500 1750 0 0
10
20
40 30 Separation time (min)
50
60
70
(a) 1
7000
0.8
5000 4000
Feleaved
Peak height
6000
3000
0.6 0.4
2000 6-mer 30-mer
1000 0 0
20
40
Injection time (min) (b)
60
0.2 0
0
10 20 30 40 50 60 70 Injection time (min) (c)
FIGURE 1.19 RNA ribozyme cleavage assay. (a) Cross-section through lane two of a leadzyme reaction in the presence of 10 μM lead acetate illustrating fluorescence intensity vs. separation time. Asterisk highlights 5′ product (6-mer). (b) Plot of peak height for the 6mer and 30-mer strands vs. injection time for the experiment shown in (a). (c) Plot of fraction cleaved vs. time of injection. (From Paxon, 2004.)
multi-channel microchip separations require new advances to be made including detection methods and application to real-world samples. Indeed, a number of applications from DNA to proteins to immunoassays have been developed around this technology to date. Detection methods for multiple-channel microchips have been limited to laser induced fluorescence detection. However, work has been published on detection methods on single channel devices including sheath flow detection [118], electrochemical detection [119–125], and mass spectrometry [18, 126–134]. One can envision any number of detection methods applied to multiplexed microchips for increased sensitivity when analyzing biological samples present at very low concentrations or in limited amounts. Multiple-channel microchips hold the promise of higher throughput separations and, as capillary array electrophoresis replaced single capillary and slab-gel electrophoresis in the high throughput separations that were needed to sequence the
Separations in Multiple-Channel Microchips
35
human genome, multi-channel microchips might play an equally important role in the future of separations on chips. In addition, these microchip designs are a probable precursor to integrated systems requiring multiple channels of any kind, and the fundamental science and engineering developed in this area should be invaluable. An important part of the future of multi-channel microchips appears to be the use of these systems to follow reaction dynamics, and this will lead to more detailed and cellular applications in biology. In another context, multi-channel microchips could be used to mimic the concentration profiles in biological systems with different amounts of solutes in different channels representing spatially differing concentrations in cells, organs, or circulatory systems. In still another direction, the Orwar group has developed a multi-channel microchip for screening drugs with a cell sensor that is scanned at the outlet of multiple channels [135–138]. Although this has not yet been used for separations in the traditional sense, the use of multiple channels allows different reagents or drugs to be transported to where a cell acts as a biological detector. This is yet another example of the versatility of the multi-channel microchip concept. This chapter has presented recent advances in multi-channel microchip designs, sample introduction, and detection. Furthermore, high-throughput analyses of complex analytes have been demonstrated utilizing these multiplexed microchips. This chapter has outlined the enormous potential and current challenges for the multichannel microchip community and discussed a variety of applications that illustrate the immense growth in the field of multiplexed separations.
ACKNOWLDGMENTS The authors would like to acknowledge the NSF and NIH for funding, in part, the research presented here. We thank our co-workers past and present for work that is referenced herein. Additionally, we thank current members of the Ewing research group for helpful comments about this manuscript.
REFERENCES 1. Issaq, H.J. A decade of capillary electrophoresis. Electrophoresis, 2000, 21(10), 1921–1939. 2. Mathies, R.A. and Huang, X.C. Capillary array electrophoresis — An approach to high-speed, high-throughput DNA sequencing. Nature, 1992, 359(6391), 167–169. 3. Huang, X.H.C., Quesada, M.A., and Mathies, R.A. DNA sequencing using capillary array electrophoresis. Analytical Chemistry, 1992, 64(18), 2149–2154. 4. Huang, X.H.C., Quesada, M.A., and Mathies, R.A. Capillary array electrophoresis using laser-excited confocal fluorescence detection. Analytical Chemistry, 1992, 64(8), 967–972. 5. Clark, S.M. and Mathies, R.A. High-speed parallel separation of DNA restriction fragments using capillary array electrophoresis. Analytical Biochemistry, 1993, 215(2), 163–170.
36
Advances in Chromatography 6. Harrison, D.J., Manz, A., Fan, Z.H., Ludi, H., and Widmer, H.M. Capillary electrophoresis and sample injection systems integrated on a planar glass chip. Analytical Chemistry, 1992, 64(17), 1926–1932. 7. Manz, A., Harrison, D.J., Verpoorte, E.M.J., Fettinger, J.C., Paulus, A., Ludi, H., and Widmer, H.M. Planar chips technology for miniaturization and integration of separation techniques into monitoring systems — Capillary electrophoresis on a chip. Journal of Chromatography, 1992, 593(1-2), 253–258. 8. Harrison, D.J., Glavina, P.G., and Manz, A. Towards miniaturized electrophoresis and chemical-analysis systems on silicon — An alternative to chemical sensors. Sensors and Actuators B-Chemical, 1993, 10(2), 107–116. 9. Effenhauser, C.S., Manz, A., and Widmer, H.M. Glass chips for high-speed capillary electrophoresis separations with submicrometer plate heights. Analytical Chemistry, 1993, 65(19), 2637–2642. 10. von Heeren, F., Verpoorte, E., Manz, A., and Thormann, W. Micellar electrokinetic chromatography separations and analyses of biological samples on a cyclic planar microstructure. Analytical Chemistry, 1996, 68(13), 2044–2053. 11. Ramseier, A., von Heeren, F., and Thormann, W. Analysis of fluorescein isothiocyanate derivatized amphetamine and analogs in human urine by capillary electrophoresis in chip-based and fused-silica capillary instrumentation. Electrophoresis, 1998, 19(16-17), 2967–2975. 12. Zhang, B.L., Foret, F., and Karger, B.L. A microdevice with integrated liquid junction for facile peptide and protein analysis by capillary electrophoresis/electrospray mass spectrometry. Analytical Chemistry, 2000, 72(5), 1015–1022. 13. Effenhauser, C.S., Manz, A., and Widmer, H.M. Manipulation of sample fractions on a capillary electrophoresis chip. Analytical Chemistry, 1995, 67(13), 2284–2287. 14. Liu, Y.J., Foote, R.S., Jacobson, S.C., Ramsey, R.S., and Ramsey, J.M. Electrophoretic separation of proteins on a microchip with noncovalent, postcolumn labeling. Analytical Chemistry, 2000, 72(19), 4608–4613. 15. Liu, Y.J., Foote, R.S., Culbertson, C.T., Jacobson, S.C., Ramsey, R.S., and Ramsey, J.M. Electrophoretic separation of proteins on microchips. Journal of Microcolumn Separations, 2000, 12(7), 407–411. 16. Shi, Y.N., Simpson, P.C., Scherer, J.R., Wexler, D., Skibola, C., Smith, M.T., and Mathies, R.A. Radial capillary array electrophoresis microplate and scanner for highperformance nucleic acid analysis. Analytical Chemistry, 1999, 71(23), 5354–5361. 17. Manz, A., Graber, N., and Widmer, H.M. Miniaturized total chemical-analysis systems — A novel concept for chemical sensing. Sensors and Actuators B-Chemical, 1990, 1(1-6), 244–248. 18. Xue, Q.F., Foret, F., Dunayevskiy, Y.M., Zavracky, P.M., McGruer, N.E., and Karger, B.L. Multichannel microchip electrospray mass spectrometry. Analytical Chemistry, 1997, 69(3), 426–430. 19. Figeys, D., Ning, Y. B., and Aebersold, R. A microfabricated device for rapid protein identification by microelectrospray ion trap mass spectrometry. Analytical Chemistry, 1997, 69(16), 3153–3160. 20. Figeys, D. and Aebersold, R. High sensitivity analysis of proteins and peptides by capillary electrophoresis tandem mass spectrometry: Recent developments in technology and applications. Electrophoresis, 1998, 19(6), 885–892. 21. Huang, Z.L., Jin, L.J., Sanders, J.C., Zheng, Y.B., Dunsmoor, C., Tian, H.J., and Landers, J.P. Laser-induced fluorescence detection on multichannel electrophoretic microchips using microprocessor-embedded acousto-optic laser beam scanning. IEEE Transactions on Biomedical Engineering, 2002, 49(8), 859–866.
Separations in Multiple-Channel Microchips
37
22. Xu, X.M., Roddy, T.P., Lapos, J.A., and Ewing, A.G. Parallel analysis with optically gated sample introduction on a multichannel microchip. Analytical Chemistry, 2002, 74(21), 5517–5522. 23. Xu, H.W., Roddy, E.S., Roddy, T.P., Lapos, J.A., and Ewing, A.G. Parallel separations of oligonucleotides with optically gated sample introduction on multi-channel microchips. Journal of Separation Science, 2004, 27(1-2), 7–12. 24. Woolley, A.T., Sensabaugh, G.F., and Mathies, R.A. High-speed DNA genotyping using microfabricated capillary array electrophoresis chips. Analytical Chemistry, 1997, 69(11), 2181–2186. 25. Huang, Z., Munro, N., Huhmer, A.F.R., and Landers, J.P. Acousto-optical deflectionbased laser beam scanning for fluorescence detection on multichannel electrophoretic microchips. Analytical Chemistry, 1999, 71(23), 5309–5314. 26. Gao, Q.F., Shi, Y.N., and Liu, S.R. Multiple-channel microchips for high-throughput DNA analysis by capillary electrophoresis. Fresenius Journal of Analytical Chemistry, 2001, 371(2), 137–145. 27. Harrison, D.J., Fluri, K., Seiler, K., Fan, Z.H., Effenhauser, C.S., and Manz, A. Micromachining a miniaturized capillary electrophoresis-based chemical-analysis system on a chip. Science, 1993, 261(5123), 895–897. 28. Jacobson, S.C., Hergenroder, R., Koutny, L.B., and Ramsey, J.M. High-speed separations on a microchip. Analytical Chemistry, 1994, 66(7), 1114–1118. 29. Manz, A., Verpoorte, E., Effenhauser, C.S., Burggraf, N., Raymond, D.E., Harrison, D.J., and Widmer, H.M. Miniaturization of separation techniques using planar chip technology. HRC: Journal of High Resolution Chromatography, 1993, 16(7), 433–436. 30. Simpson, P.C., Roach, D., Woolley, A.T., Thorsen, T., Johnston, R., Sensabaugh, G.F., and Mathies, R.A. High-throughput genetic analysis using microfabricated 96-sample capillary array electrophoresis microplates. Proceedings of the National Academy of Sciences of the United States of America, 1998, 95(5), 2256–2261. 31. Emrich, C.A., Tian, H.J., Medintz, I.L., and Mathies, R.A. Microfabricated 384-lane capillary array electrophoresis bioanalyzer for ultrahigh-throughput genetic analysis. Analytical Chemistry, 2002, 74(19), 5076–5083. 32. Becker, H. and Gartner, C. Polymer microfabrication methods for microfluidic analytical applications. Electrophoresis, 2000, 21(1), 12–26. 33. Rotting, O., Ropke, W., Becker, H., and Gartner, C. Polymer microfabrication technologies. Microsystem Technologies, 2002, 8(1), 32–36. 34. Duffy, D.C., McDonald, J.C., Schueller, O.J.A., and Whitesides, G.M. Rapid prototyping of microfluidic systems in poly(dimethylsiloxane). Analytical Chemistry, 1998, 70(23), 4974–4984. 35. Ford, S.M., Kar, B., McWhorter, S., Davies, J., Soper, S.A., Klopf, M., Calderon, G., and Saile, V. Microcapillary electrophoresis devices fabricated using polymeric substrates and x-ray lithography. Journal of Microcolumn Separations, 1998, 10(5), 413–422. 36. Effenhauser, C.S., Bruin, G.J.M., Paulus, A., and Ehrat, M. Integrated capillary electrophoresis on flexible silicone microdevices: Analysis of DNA restriction fragments and detection of single DNA molecules on microchips. Analytical Chemistry, 1997, 69(17), 3451–3457. 37. McDonald, J.C., Duffy, D.C., Anderson, J.R., Chiu, D.T., Wu, H.K., Schueller, O.J.A., and Whitesides, G.M. Fabrication of microfluidic systems in poly(dimethylsiloxane). Electrophoresis, 2000, 21(1), 27–40.
38
Advances in Chromatography 38. Heckele, M., Bacher, W., and Muller, K.D. Hot embossing — The molding technique for plastic microstructures. Microsystem Technologies, 1998, 4(3), 122–124. 39. Roetting, O., Heckele, M., and Bacher, W. Reactive ion etching for the production of metal microstructures by hot embossing. Microsystem Technologies, 1999, 6(1), 11–14. 40. McCormick, R.M., Nelson, R.J., AlonsoAmigo, M.G., Benvegnu, J., and Hooper, H.H. Microchannel electrophoretic separations of DNA in injection-molded plastic substrates. Analytical Chemistry, 1997, 69(14), 2626–2630. 41. Hooper, H., McCormick, R., Alonso, G., Wang, H., and Benvegnu, D. Microchannel electrophoresis in injection molded substrates. Abstracts of Papers of the American Chemical Society, 1997, 213, PMSE-333. 42. Spesny, M. and Foret, F. Microdevice for separation and quantitative fraction collection. Electrophoresis, 2003, 24(21), 3745–3747. 43. Woolley, A.T. and Mathies, R.A. Ultra-high-speed DNA fragment separations using microfabricated capillary array electrophoresis chips. Proceedings of the National Academy of Sciences of the United States of America, 1994, 91(24), 11348–11352. 44. Woolley, A.T. and Mathies, R.A. Ultra-high-speed DNA-sequencing using capillary electrophoresis chips. Analytical Chemistry, 1995, 67(20), 3676–3680. 45. Liu, S.R., Ren, H.J., Gao, Q.F., Roach, D.J., Loder, R.T., Armstrong, T. M., Mao, Q.L., Blaga, I., Barker, D.L., and Jovanovich, S.B. Automated parallel DNA sequencing on multiple channel microchips. Proceedings of the National Academy of Sciences of the United States of America, 2000, 97(10), 5369–5374. 46. Medintz, I.L., Berti, L., Emrich, C.A., Tom, J., Scherer, J.R., and Mathies, R.A. Genotyping energy-transfer-cassette-labeled short-tandem-repeat amplicons with capillary array electrophoresis microchannel plates. Clinical Chemistry, 2001, 47(9), 1614–1621. 47. Medintz, I.L., Paegel, B.M., and Mathies, R.A. Microfabricated capillary array electrophoresis DNA analysis systems. Journal of Chromatography A, 2001, 924(1-2), 265–270. 48. Medintz, I., Wong, W.W., Berti, L., Shiow, L., Tom, J., Scherer, J., Sensabaugh, G., and Mathies, R.A. High-performance multiplex SNP analysis of three hemochromatosis-related mutations with capillary array electrophoresis microplates. Genome Research, 2001, 11(3), 413–421. 49. Paegel, B.M., Emrich, C.A., Weyemayer, G.J., Scherer, J.R. ,and Mathies, R.A. High throughput DNA sequencing with a microfabricated 96-lane capillary array electrophoresis bioprocessor. Proceedings of the National Academy of Sciences of the United States of America, 2002, 99(2), 574–579. 50. Bromberg, A. and Mathies, R.A. Multichannel homogeneous immunoassay for detection of 2,4,6-trinitrotoluene (TNT) using a microfabricated capillary array electrophoresis chip. Electrophoresis, 2004, 25(12), 1895–1900. 51. Culbertson, C.T., Jacobson, S.C., and Ramsey, J.M. Dispersion sources for compact geometries on microchips. Analytical Chemistry, 1998, 70(18), 3781–3789. 52. Paegel, B.M., Hutt, L.D., Simpson, P.C., and Mathies, R.A. Turn geometry for minimizing band broadening in microfabricated capillary electrophoresis channels. Analytical Chemistry, 2000, 72(14), 3030–3037. 53. Medintz, I.L., Paegel, B.M., Blazej, R.G., Emrich, C.A., Berti, L., Scherer, J.R., and Mathies, R.A. High-performance genetic analysis using microfabricated capillary array electrophoresis microplates. Electrophoresis, 2001, 22(18), 3845–3856.
Separations in Multiple-Channel Microchips
39
54. Martynova, L., Locascio, L.E., Gaitan, M., Kramer, G.W., Christensen, R.G., and MacCrehan, W.A. Fabrication of plastic microfluid channels by imprinting methods. Analytical Chemistry, 1997, 69(23), 4783–4789. 55. Wang, J., Chatrathi, M.P., Tian, B.M., and Polsky, R. Microfabricated electrophoresis chips for simultaneous bioassays of glucose, uric acid, ascorbic acid, and acetaminophen. Analytical Chemistry, 2000, 72(11), 2514–2518. 56. Hayes, M.A., Polson, N.A., Phayre, A.N., and Garcia, A.A. Flow-based microimmunoassay. Analytical Chemistry, 2001, 73(24), 5896–5902. 57. Guijt, R.M., Baltussen, E., and van Dedem, G.W.K. Use of bioaffinity interactions in electrokinetically controlled assays on microfabricated devices. Electrophoresis, 2002, 23(6), 823–835. 58. Guttman, A., Ronai, Z., Barta, C., Hou, Y.M., Sasvari-Szekely, M., Wang, X., and Briggs, S.P. Membrane-mediated ultrafast restriction digestion and subsequent rapid gel microchip electrophoresis of DNA. Electrophoresis, 2002, 23(10), 1524–1530. 59. Kricka, L.J. Prospects for microchips in immunoassays and immunoassay tests at the point-of-care. Journal of Clinical Ligand Assay, 2002, 25(4), 317–324. 60. Lim, T.K., Ohta, H., and Matsunaga, T. Microfabricated on-chip-type electrochemical flow immunoassay system for the detection of histamine released in whole blood samples. Analytical Chemistry, 2003, 75(14), 3316–3321. 61. Xu, H.W. and Ewing, A.G. A rapid enzyme assay for beta-galactosidase using optically gated sample introduction on a microfabricated chip. Analytical and Bioanalytical Chemistry, 2004, 378(7), 1710–1715. 62. Cheng, S.B., Skinner, C.D., Taylor, J., Attiya, S., Lee, W.E., Picelli, G., and Harrison, D.J. Development of a multichannel microfluidic analysis system employing affinity capillary electrophoresis for immunoassay. Analytical Chemistry, 2001, 73(7), 1472–1479. 63. Xue, Q.F., Wainright, A., Gangakhedkar, S., and Gibbons, I. Multiplexed enzyme assays in capillary electrophoretic single-use microfluidic devices. Electrophoresis, 2001, 22(18), 4000–4007. 64. Sato, K., Yamanaka, M., Takahashi, H., Tokeshi, M., Kimura, H., and Kitamori, T. Microchip-based immunoassay system with branching multichannels for simultaneous determination of interferon-gamma. Electrophoresis, 2002, 23(5), 734–739. 65. Jiang, X.Y., Ng, J.M.K., Stroock, A.D., Dertinger, S.K.W., and Whitesides, G.M. A miniaturized, parallel, serially diluted immunoassay for analyzing multiple antigens. Journal of the American Chemical Society, 2003, 125(18), 5294–5295. 66. Roddy, E.S., Price, M., and Ewing, A.G. Continuous monitoring of a restriction enzyme digest of DNA on a microchip with automated capillary sample introduction. Analytical Chemistry, 2003, 75(15), 3704–3711. 67. Roddy, E.S., Xu, H.W., and Ewing, A.G. Sample introduction techniques for microfabricated separation devices. Electrophoresis, 2004, 25(2), 229–242. 68. Smith, E.M., Xu, H.W., and Ewing, A.G. DNA separations in microfabricated devices with automated capillary sample introduction. Electrophoresis, 2001, 22(2), 363–370. 69. Paxon, T.L., Brown, T.S., Lin, H.N., Brancato, S.J., Roddy, E.S., Philip C. Bevilacquaa, P.C., and Ewing, A.G. Continuous monitoring of enzyme reactions on a microchip: Application to catalytic RNA self-cleavage. Analytical Chemistry, 2004, 76(23), 6921–6927.
40
Advances in Chromatography 70. Seiler, K., Harrison, D.J., and Manz, A. Planar glass chips for capillary electrophoresis — Repetitive sample injection, quantitation, and separation efficiency. Analytical Chemistry, 1993, 65(10), 1481–1488. 71. Beard, N.R., Zhang, C.X., and deMello, A.J. In-column field-amplified sample stacking of biogenic amines on microfabricated electrophoresis devices. Electrophoresis, 2003, 24(4), 732–739. 72. Zhang, C.X. and Manz, A. Narrow sample channel injectors for capillary electrophoresis on microchips. Analytical Chemistry, 2001, 73(11), 2656–2662. 73. Ferrance, J. and Landers, J.P. Exploiting sensitive laser-induced fluorescence detection on electrophoretic microchips for executing rapid clinical diagnostics. Luminescence, 2001, 16(2), 79–88. 74. Medintz, I., Wong, W.W., Sensabaugh, G., and Mathies, R.A. High speed single nucleotide polymorphism typing of a hereditary haemochromatosis mutation with capillary array electrophoresis microplates. Electrophoresis, 2000, 21(12), 2352–2358. 75. Effenhauser, C.S., Paulus, A., Manz, A., and Widmer, H.M. High-speed separation of antisense oligonucleotides on a micromachined capillary electrophoresis device. Analytical Chemistry, 1994, 66(18), 2949–2953. 76. Monnig, C.A., Dohmeier, D.M., and Jorgenson, J.W. Sample gating in open tubular and packed capillaries for high-speed liquid chromatography. Analytical Chemistry, 1991, 63(8), 807–810. 77. Monnig, C.A. and Jorgenson, J.W. On-column sample gating for high-speed capillary zone electrophoresis. Analytical Chemistry, 1991, 63(8), 802–807. 78. Moore, A.W. and Jorgenson, J.W. Study of zone broadening in optically gated highspeed capillary electrophoresis. Analytical Chemistry, 1993, 65(24), 3550–3560. 79. Lapos, J.A. and Ewing, A.G. Injection of fluorescently labeled analytes into microfabricated chips using optically gated electrophoresis. Analytical Chemistry, 2000, 72(19), 4598–4602. 80. Roddy, E.S., Lapos, J.A., and Ewing, A.G. Rapid serial analysis of multiple oligonucleotide samples on a microchip using optically-gated injection. Journal of Chromatography A, 2003, 1004(1-2), 217–224. 81. Uchiyama, K., Nakajima, H., and Hobo, T. Detection method for microchip separations. Analytical and Bioanalytical Chemistry, 2004, 379(3), 375–382. 82. Quesada, M.A., Rye, H.S., Gingrich, J.C., Glazer, A.N., and Mathies, R.A. Highsensitivity DNA detection with a laser-excited confocal fluorescence gel scanner. Biotechniques, 1991, 10(5), 616–622, 624–625. 83. Dolnik, V., Liu, S., and Jovanovich, S. Capillary electrophoresis on microchip. Electrophoresis, 2000, 21(1), 41–54. 84. Lee, T.T. and Yeung, E.S. Capillary electrophoresis detectors: Lasers. Methods in Enzymology, 1996, 270, 419–449. 85. Zhu, H.P., Clark, S.M., Benson, S.C., Rye, H.S., Glazer, A.N., and Mathies, R.A. High-sensitivity capillary electrophoresis of double-stranded DNA fragments using monomeric and dimeric fluorescent intercalating dyes. Analytical Chemistry, 1994, 66(13), 1941–1948. 86. Sassi, A.P., Paulus, A., Cruzado, I.D., Bjornson, T., and Hooper, H.H. Rapid, parallel separations of D1S80 alleles in a plastic microchannel chip. Journal of Chromatography A, 2000, 894(1-2), 203–217. 87. Kheterpal, I., Scherer, J.R., Clark, S.M., Radhakrishnan, A., Ju, J.Y., Ginther, C.L., Sensabaugh, G.F., and Mathies, R.A. DNA sequencing using a four-color confocal fluorescence capillary array scanner. Electrophoresis, 1996, 17(12), 1852–1859.
Separations in Multiple-Channel Microchips
41
88. Alexay, C., Kain, R., Hanzel, D., and Johnston, R. Fluorescence scanner employing a macro scanning objective. Proceedings of SPIE: The International Society for Optical Engineering, 1996, 2705 (Fluorescence Detection IV), 63–72. 89. Kambara, H. and Takahashi, S. Multiple-sheathflow capillary array DNA analyzer. Nature, 1993, 361(6412), 565–566. 90. Quesada, M.A. and Zhang, S.P. Multiple capillary DNA sequencer that uses fiberoptic illumination and detection. Electrophoresis, 1996, 17(12), 1841–1851. 91. Pang, H.M., Pavski, V., and Yeung, E.S. DNA sequencing using 96-capillary array electrophoresis. Journal of Biochemical and Biophysical Methods, 1999, 41(2-3), 121–132. 92. Hietpas, P.B., Bullard, K.M., Gutman, D.A., and Ewing, A.G. Ultrathin slab gel separations of DNA using a single capillary sample introduction system. Analytical Chemistry, 1997, 69(13), 2292–2298. 93. Ewing, A.G., Gavin, P.F., Hietpas, P.B., and Bullard, K.M. Continuous separations in microfabricated channels for monitoring ultrasmall biological environments. Nature Medicine, 1997, 3(1), 97–99. 94. Bullard, K.M., Hietpas, P.B., and Ewing, A.G. Capillary sample introduction of polymerase chain reaction (PCR) products separated in ultrathin slab gels. Electrophoresis, 1998, 19(1), 71–75. 95. Bullard, K.M., Gavin, P.F., and Ewing, A.G. Electrophoretic separation schemes in ultrathin channels with capillary sample introduction. Trac-Trends in Analytical Chemistry, 1998, 17(7), 401–410. 96. Hietpas, P.B., Bullard, K.M., and Ewing, A.G. Characterization of electrophoretic sample transfer from a capillary to an ultrathin slab gel. Journal of Microcolumn Separations, 1998, 10(6), 519–527. 97. Bibeau, D.R., Smith, K.B., Smith, E.M., and Ewing, A.G. Separations of DNA fragments with a coated 25-μm capillary coupled to a 25-μm high open channel. Journal of Microcolumn Separations, 1999, 11(8), 567–575. 98. Bruin, G.J.M. Recent developments in electrokinetically driven analysis on microfabricated devices. Electrophoresis, 2000, 21(18), 3931–3951. 99. Pingoud, A. and Jeltsch, A. Structure and function of type II restriction endonucleases. Nucleic Acids Research, 2001, 29(18), 3705–3727. 100. Han, J. and Craighead, H.G. Separation of long DNA molecules in a microfabricated entropic trap array. Science, 2000, 288(5468), 1026–1029. 101. Han, J.Y. and Craighead, H.G. Characterization and optimization of an entropic trap for DNA separation. Analytical Chemistry, 2002, 74(2), 394–401. 102. Guttman, A. Sample stacking during membrane-mediated loading in automated DNA sequencing. Analytical Chemistry, 1999, 71(16), 3598–3602. 103. Nemoda, Z., Ronai, Z., Szekely, A., Kovacs, E., Shandrick, S., Guttman, A., and Sasvari-Szekely, M. High-throughput genotyping of repeat polymorphism in the regulatory region of serotonin transporter gene by gel microchip electrophoresis. Electrophoresis, 2001, 22(18), 4008–4011. 104. Carrihlo, E. DNA sequencing by capillary array electrophoresis and microfabricated array systems. Electrophoresis, 2000, 21, 55–65. 105. Liu, S.R., Shi, Y.N., Ja, W.W., and Mathies, R.A. Optimization of high-speed DNA sequencing on microfabricated capillary electrophoresis channels. Analytical Chemistry, 1999, 71(3), 566–573.
42
Advances in Chromatography
106. Simpson, J.W., Ruiz-Martinez, M.C., Mulhern, G.T., Berka, J., Latimer, D.R., Ball, J.A., Rothberg, J.M., and Went, G.T. A transmission imaging spectrograph and microfabricated channel system for DNA analysis. Electrophoresis, 2000, 21(1), 135–149. 107. Backhouse, C., Caamano, M., Oaks, F., Nordman, E., Carrillo, A., Johnson, B., and Bay, S. DNA sequencing in a monolithic microchannel device. Electrophoresis, 2000, 21(1), 150–156. 108. Liu, S.R., Elkin, C., and Kapur, H. Sequencing of real-world samples using a microfabricated hybrid device having unconstrained straight separation channels. Electrophoresis, 2003, 24(21), 3762–3768. 109. Sato, K., Tokeshi, M., Odake, T., Kimura, H., Ooi, T., Nakao, M., and Kitamori, T. Integration of an immunosorbent assay system: Analysis of secretory human immunoglobulin A on polystyrene beads in a microchip. Analytical Chemistry, 2000, 72(6), 1144–1147. 110. Sato, K., Tokeshi, M., Kimura, H., and Kitamori, T. Determination of carcinoembryonic antigen in human sera by integrated bead bed immunoasay in a microchip for cancer diagnosis. Analytical Chemistry, 2001, 73(6), 1213–1218. 111. Colyer, C.L., Tang, T., Chiem, N., and Harrison, D.J. Clinical potential of microchip capillary electrophoresis systems. Electrophoresis, 1997, 18(10), 1733–1741. 112. Chiem, N.H. and Harrison, D.J. Monoclonal antibody binding affinity determined by microchip-based capillary electrophoresis. Electrophoresis, 1998, 19(16-17), 3040–3044. 113. Chiem, N.H. and Harrison, D.J. Microchip systems for immunoassay: An integrated immunoreactor with electrophoretic separation for serum theophylline determination. Clinical Chemistry, 1998, 44(3), 591–598. 114. Jiang, G.F., Attiya, S., Ocvirk, G., Lee, W.E., and Harrison, D.J. Red diode laser induced fluorescence detection with a confocal microscope on a microchip for capillary electrophoresis. Biosensors & Bioelectronics, 2000, 14(10-11), 861–869. 115. Qiu, C.X. and Harrison, D.J. Integrated self-calibration via electrokinetic solvent proportioning for microfluidic immunoassays. Electrophoresis, 2001, 22(18), 3949–3958. 116. Taylor, J., Picelli, G., and Harrison, D.J. An evaluation of the detection limits possible for competitive capillary electrophoretic immunoassays. Electrophoresis, 2001, 22(17), 3699–3708. 117. Bromberg, A. and Mathies, R.A. Homogeneous immunoassay for detection of TNT and its analogues on a microfabricated capillary electrophoresis chip. Analytical Chemistry, 2003, 75(5), 1188–1195. 118. Ertl, P., Emrich, C.A., Singhal, P., and Mathies, R.A. Capillary electrophoresis chips with a sheath-flow supported electrochemical detection system. Analytical Chemistry, 2004, 76(13), 3749–3755. 119. Woolley, A.T., Lao, K.Q., Glazer, A.N., and Mathies, R.A. Capillary electrophoresis chips with integrated electrochemical detection. Analytical Chemistry, 1998, 70(4), 684–688. 120. Martin, R.S., Gawron, A.J., Lunte, S.M., and Henry, C.S. Dual-electrode electrochemical detection for poly(dimethylsiloxane)-fabricated capillary electrophoresis microchips. Analytical Chemistry, 2000, 72(14), 3196–3202. 121. Lunte, S.M., Martin, R.S., Gawron, A., Lacher, N., Fogarty, B., and Reagan, F. Bioanalytical applications of microchip capillary electrophoresis with electrochemical detection. Abstracts of Papers of the American Chemical Society, 2001, 221, U102–U102.
Separations in Multiple-Channel Microchips
43
122. Lapos, J.A., Manica, D.P., and Ewing, A.G. Dual fluorescence and electrochemical detection on an electrophoresis microchip. Analytical Chemistry, 2002, 74(14), 3348–3353. 123. Fogarty, B.A., Lacher, N.A., Pasas, S., Vanadeever, W.R., Martin, R.S., Davies, M.I., Heppert, K,. and Lunte, S. Bioanalytical applications of microchip electrophoresis with electrochemical detection. Abstracts of Papers of the American Chemical Society, 2003, 225, U151–U151. 124. Osbourn, D.M. and Lunte, C.E. On-column electrochemical detection for microchip capillary electrophoresis. Analytical Chemistry, 2003, 75(11), 2710–2714. 125. Lacher, N.A., Lunte, S.M., and Martin, R.S. Development of a microfabricated palladium decoupler/electrochemical detector for microchip capillary electrophoresis using a hybrid glass/poly(dimethylsiloxane) device. Analytical Chemistry, 2004, 76(9), 2482–2491. 126. Ramsey, R.S. and Ramsey, J.M. Microchip devices for fluidic manipulation and esi mass spectrometry. Abstracts of Papers of the American Chemical Society, 1997, 214, ANYL-150. 127. Xue, Q.F., Dunayevskiy, Y.M., Foret, F., and Karger, B.L. Integrated multichannel microchip electrospray ionization mass spectrometry: Analysis of peptides from onchip tryptic digestion of melittin. Rapid Communications in Mass Spectrometry, 1997, 11(12), 1253–1256. 128. Meng, Z.J., Qi, S.Z., Soper, S.A., and Limbach, P.A. Interfacing a polymer-based micromachined device to a nanoelectrospray ionization Fourier transform ion cyclotron resonance mass spectrometer. Analytical Chemistry, 2001, 73(6), 1286–1291. 129. Lazar, I.M., Ramsey, R.S., Sundberg, S., and Ramsey, J.M. Subattomole-sensitivity microchip nanoelectrospray source with time-of-flight mass spectrometry detection. Analytical Chemistry, 1999, 71(17), 3627–3631. 130. Lazar, I.M., Ramsey, R.S., Jacobson, S.C., Foote, R.S., and Ramsey, J.M. Novel microfabricated device for electrokinetically induced pressure flow and electrospray ionization mass spectrometry. Journal of Chromatography A, 2000, 892(1-2), 195–201. 131. Lazar, I.M., Ramsey, R.S., and Ramsey, J.M. On-chip proteolytic digestion and analysis using “wrong-way-round” electrospray time-of-flight mass spectrometry. Analytical Chemistry, 2001, 73(8), 1733–1739. 132. Svedberg, M., Pettersson, A., Nilsson, S., Bergquist, J., Nyholm, L., Nikolajeff, F., and Markides, K. Sheathless electrospray from polymer microchips. Analytical Chemistry, 2003, 75(15), 3934–3940. 133. Ssenyange, S., Taylor, J., Harrison, D.J., and McDermott, M.T. A glassy carbon microfluidic device for electrospray mass spectrometry. Analytical Chemistry, 2004, 76(8), 2393–2397. 134. Yang, Y.N., Kameoka, J., Wachs, T., Henion, J.D., and Craighead, H.G. Quantitative mass spectrometric determination of methylphenidate concentration in urine using an electrospray ionization source integrated with a polymer microchip. Analytical Chemistry, 2004, 76(9), 2568–2574. 135. Sinclair, J., Pihl, J., Olofsson, J., Karlsson, M., Jardemark, K., Chiu, D.T., and Orwar, O. A cell-based bar code reader for high-throughput screening of ion channel-ligand interactions. Analytical Chemistry, 2002, 74(24), 6133–6138. 136. Olofsson, J., Karlsson, M., Pihl, J., Sinclair, J., and Orwar, O. Microfluidic chips for time-resolved high-throughput patch-clamp studies. Biophysical Journal, 2003, 84(2), 488a–488a.
44
Advances in Chromatography
137. Pihl, J., Sinclair, J., Olofsson, J., Sahlin, E., Karlsson, M,. and Orwar, O. A microfluidic tool for studies of multi-component concentration dependent ion channel effecters. Biophysical Journal, 2004, 86(1), 482a–482a. 138. Farre, C., Sinclair, J., Pihl, J., Olofsson, J., and Orwar, O. Microfluidic chip solutions for patch-clamp-based high-content-screening. Biophysical Journal, 2004, 86(1), 482a–482a. 139. Sinclair, J., Pihl, J., Olofsson, J., Farre, C., and Orwar, O. Microfluidic devices for high-content ion channel screening. Biophysical Journal, 2004, 86(1), 355a–355a.
2
Temperature Effects in Liquid Chromatography Elsa Lundanes University of Oslo, Norway
Tyge Greibrokk University of Oslo, Norway
CONTENTS Abstract ....................................................................................................................45 2.1 Introduction ....................................................................................................46 2.2 Temperature Effects on Retention and Selectivity ........................................46 2.3 Conventional Analytical LC...........................................................................49 2.3.1 Columns and Stationary Phases.........................................................49 2.3.2 Temperature Programming and Instrumentation ...............................51 2.3.3 Applications .......................................................................................52 2.4 Capillary LC...................................................................................................56 2.4.1 Columns and Stationary Phases.........................................................56 2.4.2 Temperature Programming and Instrumentation ...............................63 2.4.3 Applications .......................................................................................65 References................................................................................................................66
ABSTRACT Temperature has a large effect on retention, selectivity, and column efficiency and has long been accepted as an important parameter in liquid chromatography (LC). Despite this fact, temperature has not been very actively utilized in the past, mainly because of reported stability problems of the most commonly used stationary phases. However, more interest in the application of temperature for retention control has come of late because of the trend of miniaturization in chromatography and the availability of temperature-stable stationary phases. This work gives an overview of temperature effects on retention and selectivity in chromatography, especially on reversed-phase columns. Instrumental requirements, especially with respect to performing temperature gradient elution, are discussed and applications on both conventional-sized analytical columns and capillary columns are included.
45
46
Advances in Chromatography
2.1 INTRODUCTION Even though temperature influences parameters like solubility, diffusivity, mobile phase polarity, and viscosity, temperature has not been very actively used in liquid chromatography and most analyses are carried out at ambient temperature. However, in recent years there has been an increasing awareness of the potential of temperature for achieving better chromatographic resolution by either improvement in selectivity or chromatographic efficiency. At high temperatures, the viscosity is reduced and the diffusion rate is increased; thus, the mass transfer rate between the stationary and mobile phase is increased as well as the efficiency [1, 2] and it is possible to decrease the analysis time substantially [3, 4]. The temperature effect upon retention was extensively studied in reversed-phase systems as can be seen in review articles [2, 5–7] and some recent publications [4, 8–15]. However, temperature has also been actively used for retention control in normal-phase [16, 17], ion exchange [18, 19], and hydrophobic interaction chromatography [20], as well as on temperature-responsive stationary phases [21–24]. Despite the advantages of using high temperatures, the stability of the stationary phases commonly used in liquid chromatography (LC) today has hampered the development of temperature-optimized analytical methods. This review will elaborate upon the influence of temperature upon retention and selectivity as well as methods using temperature actively to optimize the separations both in conventional sized columns and smaller-diameter columns. This review will not include temperature effects in electrochromatography but information on this topic can be found elsewhere [6, 25]. Also, note that temperature regulation of retention may have been used in papers not included in this review, since the focus is on studies where temperature has been actively and explicitly used.
2.2 TEMPERATURE EFFECTS ON RETENTION AND SELECTIVITY Because reversed-phase chromatography is the separation mode used in about 80 to 90% of LC methods, the majority of studies involving temperature effects have been on reversed-phase stationary phases. The relationship between the retention factor, k, and the column temperature is well known [26, 27] and can be described by the equation: ln k = – ΔH/RT + ΔS/R + ln Φ
(2.1)
where R is the universal gas constant, T is the temperature, ΔH and ΔS are the enthalpy and entropy of a solute transfer from the mobile phase to the stationary phase, respectively, and Φ is the phase ratio of the column (the stationary to mobile phase volume ratio). The linear relationship between ln k and 1/T found in most studies with reversed-phase chromatography is also found in normal-phase as well as in cation-exchange chromatography [28]. The retention behavior in reversed-phase LC has been discussed on the basis of enthalpy−entropy compensation (EEC) between ΔH and ΔS (see [29] and references
Temperature Effects in Liquid Chromatography
47
3.5 3
In k or α
2.5 2
Benzene Toluene Selectivity
1.5 1 R2 = 0.989
0.5 0 0.0028
0.0030
84
60
0.0032 0.0034 1/T (1/K) 40
21
0.0036
0.0038
5
−10
T(ºC)
FIGURE 2.1 Van’t Hoff plots of benzene, toluene, and their difference, which is the methylene selectivity. The phase−ratio-dependent plots of benzene and toluene are non-linear while the phase-ratio-independent methylene selectivity plot is linear. Again, this suggests a phaseratio dependence in the van’t Hoff plots of benzene and possibly toluene. Experimental conditions: water-methanol (60:40) mobile phase, ODS stationary phase, at a flow-rate of 1.00 mL/min. From [41].
therein). ΔH and ΔS associated with the retention mechanism can be estimated by using the van’t Hoff equation (Equation 2.1). In reversed-phase chromatography, the value of ΔH is typically 7 to 15 kJ/mol for small solutes [2, 29–38], while larger molecules like polycyclic aromatic hydrocarbons (PAH) [28] usually have larger (negative) enthalpies. A common intersection point has been found for the ln k vs. 1/T lines of homologues in reversed-phase systems [39]. However, because of changes in the molecular conformation, plots of ln k vs. 1/T (van’t Hoff plot) often appear nonlinear for biopolymers like proteins [40]. However, Chester and Coym presented mathematical evidence and showed that phase ratio changes can also cause deviation from linearity for smaller molecules in van’t Hoff plots in the temperature range –5 to 80˚C [41] (see Figure 2.1). A wider temperature range was explored by Guillarme et al. [4], who studied the effect of temperature on the solute behavior on reversed-phase stationary phases that are thermally stable in the temperature range from 30 to 200˚C. All van’t Hoff plots were linear in the narrow temperature range 25 to 80˚C on all the types of stationary phases; however, within a larger range of temperature, a quadratic model was required. Nonlinear van’t Hoff plots were observed in the case of acetonitrile−water mixtures or water as mobile phase, while the plots of log k vs. 1/T for the Hypercarb column (20 to 180˚C) was linear whatever the organic solvent (methanol or acetonitrile) used. Jinno et al. [42] studied the effect of temperature in reversed-phase chromatography down to 50˚C and found that the retention mechanism in the lower temperature range is similar to that of the
48
Advances in Chromatography
normal temperature range when the water content of the mobile phase exceeds 10%. Mobile phases of pure water at superheated temperatures, between 100 and 200˚C, have also been explored and significant differences in the enthalpy of the retention process at high temperatures (–30 to –40 kJ/mol) were found, as compared to that at low temperatures of 15 to 55˚C (–6 to –13 kJ/mol) [43]. In a recent paper, the effect of temperature on the retention of ionizable compounds was addressed [44] and the authors pointed out that differences in ionization enthalpies of the buffers resulted in different eluent pH when the temperature was raised. Li [45] also found that buffers can play a very important role in affecting retention and selectivity with changes in temperature. The retention times of some bases increased with increasing temperature in the investigated range up to 70˚C using a mobile phase pH of about 7.8 [46]. The peak shapes also significantly improved, especially for weak bases (pKa ≈ 8). The increase in retention and improvement in efficiency were explained by the reduction in the pKa of bases with temperature, leading to reduction in protonation. Quaternary ammonium compounds showed normal behavior, i.e., reduced retention with increasing column temperature. The relationship in Equation 2.1 shows that for the same change in temperature, the retention factor of solutes with large ΔH is more strongly affected than that of solutes with small ΔH, providing different selectivity at different temperatures. Thus, a change in temperature can give a separation of solutes not obtained at the initial temperature explored. The effect of temperature upon retention is, however, rather small compared to the effect provided by an increase in mobile phase strength. A 4˚C increase in temperature was found to correspond to a 1% methanol increase [47], while another study showed that a 5˚C change in column temperature had the same effect on retention as a 1% change in acetonitrile concentration [48]. Similar values for neutral solutes were later found in another study using reversed-phase columns [49]. This result means that the elution strength window of temperature is much less than that of an organic modifier in a reversed-phase system and temperature programs cannot generally replace solvent gradients. In a series of papers, Snyder et al. [50–53] optimized reversed-phase separations of different solutes by changing the temperature or the solvent gradient to improve the selectivity, while Dolan et al. varied the temperature and solvent gradient steepness for maximizing the resolution using computer simulations [54]. These researchers showed that simultaneous changes in temperature and either gradient time (tG) or solvent strength (% B) is effective for control of selectivity in reversed-phase LC. When a solvent gradient is undesirable, isocratic separation can be optimized by varying the temperature and the mobile phase strength [55]. Neue and Mazzeo performed a theoretical study on optimization of solvent gradients at elevated temperatures and showed that the reduction in viscosity and improvement in diffusivity — and hence the mass transfer — made it possible to increase the speed of the separation [56]. The temperature dependence on retention using different mobile phases to assess the validity of hydrophobic theory as the driving force for retention has been addressed [57], as well as the influence of the temperature (23 to 73˚C) on the equilibrium isotherm [58].
Temperature Effects in Liquid Chromatography
49
2.3 CONVENTIONAL ANALYTICAL LC 2.3.1 COLUMNS
AND
STATIONARY PHASES
In almost all the papers mentioned above, silica-based reversed-phase columns have been used, mostly C18 columns. The mechanism of separation using C1, C4, C8, and C18 has been shown to be the same in the temperature range 0 to 77˚C [59] and changes in relative retention with temperature were approximately constant for different values of the gradient time tG for different C18 and C18 reversed-phase columns [60]. Raman spectroscopy has been used to study the temperature-induced surface changes in high-density C18 stationary phases [61]. Conformations and interchain coupling of the alkyl components of these phases were found to be sensitive to temperature and surface coverage, and each of the high-density C18 stationary phases exhibited subtle temperature-dependent phase changes. Linear van’t Hoff plots were produced for various monomeric and polymeric C18 phases over temperature ranges of about 30 K using water-containing mobile phases [27, 35, 62–64]. ΔS˚ values were provided only in a few studies [63, 64]; however, in the temperature range –5 to 80˚C, the entropic contribution to retention becomes more significant with respect to the enthalpic contribution when the stationary phase bonding density is increased for monomeric C18 phases [65] and the stationary phase structure gives an important contribution to selectivity, particularly at subambient temperatures [66]. In an early study, Chmielowiec and Sawatzky also showed that entropy effects can be large for molecules like polycyclic aromatic hydrocarbons on these phases [67]. The influence of temperature upon separation on C30 phases has been addressed in several studies [68, 69] and Albert et al. [68, 70] found that the strongly temperature-dependent separation characteristics of polymeric C30 phases correlated with the alkyl chain arrangement at the silica surface. By FTIR studies, titania-based self-assembled monolayer C30 phases were found to possess a higher degree of conformational order than zirconia- and silica-based C30 [71]. The stability of the commonly used silica-based reversed-phase materials has been a limiting factor for exploiting temperature for retention and selectivity control and the benefits of using higher temperatures for separation have not been achieved. Silica-based materials have been used up to 100˚C [72, 73] as can be seen in Figure 2.2, even though temperatures above 60˚C are usually not recommended [4, 15]. In addition to the temperature, the mobile phase composition, including the type of buffer used and the pH, influences the column lifetime of silica-based phases. In a recent paper, Claessens and van Straten [74] reviewed the chemical and thermal stability of reversed-phase stationary phases. Different silica-based, zirconia-based, and polymer columns are compared and they concluded that for many separation needs, specific silica reversed phases offered sufficient thermal stability from subambient and up to 90˚C. However, they also pointed out that reversed-phase stationary phases based on other inorganic substances and polymers may be needed for methods requiring higher thermal stability. In the recent study by Guillarme et al. [4], silica-based columns were used only in the temperature range 20 to 100˚C due to thermal instability, while the porous carbon, PS-DVB, and zirconia-based columns
50
Advances in Chromatography
mAU 80 70
(a)
60 50 40 30 20 10 0
0
5
10
15
20
25
30
min
10
15
20
25
30
min
10
15
20
25
30
min
mAU 80 70
(b)
60 50 40 30 20 10 0 0
5
mAU 80 70
(c)
60 50 40 30 20 10 0 0
5
FIGURE 2.2 Separation of polymers of average mass (A) 30,000, (B) 15,000, and (C) 8,000 on a non-porous Hytach C18 column operated at 100˚C with a gradient of 0 to 100% B in 60 min, where A is 400 mM NaClO4 and solvent B is 400 mM NaClO4 in acetonitrile-water (20:80). From [73].
could be used up to 180˚C. That the carbon columns and polymeric columns display better thermal stability has also been reported by others [4, 15, 75] and the influence of temperature on retention and selectivity on the polymeric columns has also been studied using the solvation parameter model [76]. An alumina-based C18 material was used up to 150˚C [77] but these materials have not been widely used. As mentioned above, zirconia-based stationary phases have been reported to be stable at high temperatures (see also [3, 78, 79]). The effect of temperature on the thermodynamic and dynamic properties of the zirconia-based stationary phases were investigated by Carr et al. [80] who found that, depending on the solute, the temperature can be a very important variable to also improve the selectivity on these
Temperature Effects in Liquid Chromatography
51
3
400
(b) 4 300
2 1
200
5
Absorbance (mAU)
100
0 500
0.0
0.2
3
0.4
0.6
0.8
1.0 (a)
400
4
2 1
300 200
5 100 0 0
3
6
9
12
15
Time (min)
FIGURE 2.3 Chromatograms of a reversed-phase mixture on PBD-coated zirconia. Plot A is the chromatogram at 30˚C and 1 mL/min and plot B is the chromatogram at 100˚C and 5 mL/min. Solutes: 1, uracil; 2, p-nitroaniline; 3, methyl benzoate; 4, phenetole; and 5, toluene. From [3].
phases. By using high temperatures, very fast separations using high flow rates are possible [3]. An 18-fold improvement relative to that at lower temperatures and normal flow rates was reported on these phases (see Figure 2.3).
2.3.2 TEMPERATURE PROGRAMMING
AND INSTRUMENTATION
A majority of the studies on the influence of temperature in analytical-sized columns has been performed using conventional equipment and isothermal conditions. The column thermostatting devices used vary widely in their design, and the actual temperature in the column may be quite different when two different LC systems are used [81, 82]. In most cases, mobile phase preheating is performed by either placing a preheating tube into a column liquid bath or an air circulation oven. A water or oil bath is more efficient than the air bath [83]; however, a recent study
52
Advances in Chromatography
concludes that it is quite possible to use an air system with analytical sized columns up to 150˚C and with columns with I.D. of 1 mm or less up to 200˚C [4]. Spearman et al. [82] monitored the effective column temperature by exploiting the shape selectivity and hydrophobicity of selected solutes, and suggested a method to distinguish between temperature and mobile phase composition differences that cause nonreproducibility in transferring methods between instruments and laboratories. They also found a marked increase in efficiency as the mobile phase temperature was reduced below the column temperature, in contrast to claims that thermal equilibration is essential [81, 83, 84]. The use of temperature gradients for retention control and selectivity optimization has not been widely used in analytical LC even though a temperature-programmed separation on alumina was published in 1966 [85], as well as a temperature-programmed (15 to 50˚C) separation of PAHs on a 4.6 mm I.D. C18 column in 1979 [67]. One of the reasons for the near-absence of temperature programming in analytical sized systems is that this methodology has been associated with radial temperature gradients and subsequent band broadening [81]. Poppe and Kraak [83] showed that the difference in temperature between the incoming mobile phase and the column was detrimental to the efficiency, a problem confirmed by others [81, 84]. Wolcott et al. suggested that the temperature difference between the entering mobile phase and the column oven should be within ±6˚C [81], and Djordjevic et al. [86] showed that analytical sized columns can be used with temperature programming if a mobile phase preheater is installed. In a recent study, Molander et al. [87] demonstrated that the column I.D. is not a critical limiting factor when performing temperature programming in LC, at least for columns narrower than 4.6 mm I.D. in the temperature range 30 to 90˚C. However, the relative apparent efficiency was inversely proportional to the column I.D., making the capillary columns generally more attractive for use with temperature gradients with respect to efficiency, as will be elaborated upon later in this review. Others have also found that with narrow-bore columns, the thermal mismatch effects are reduced significantly and can be neglected due to the fast heat transfer [47, 84]. However, despite these reports, temperature gradient elution was used in a few studies using microbore [88, 89] and conventional sized columns [11, 15, 90]. A thermal gradient technique that combined flow and temperature programming was suggested by Moore and Synovec [91] for microbore LC.
2.3.3 APPLICATIONS Even though most published studies utilizing temperature in LC have been of more or less fundamental character, temperature has also been actively used in conventional LC to solve separation needs, the goal of most analysts. In the majority of studies, C18 or C8 reversed-phase materials were used despite the fact that the largest influence of temperature has been found for long alkyl chain length stationary phases [7]. By optimizing the temperature, isothermal solvent gradient separation of 11 carotenoid isomers was obtained on a C30 column in the temperature range 15 to 25˚C [9]. The use of C30 stationary phases for determination of carotenoids, retinoids, and other related compounds in food has been reviewed [92].
Temperature Effects in Liquid Chromatography
53
Snyder et al. actively used temperature to optimize the separation of tryptic digests as well as other peptide and protein samples on C18 columns in the temperature range 20 to 60˚C [93, 94]. In a series of papers, Hodges et al. also studied the temperature selectivity effects of peptides in reversed-phase LC [13, 95–97] using both silica-based C8, CN [13, 95, 97] and PV-DVB [96] columns and a temperature range (isothermal) from 5 to 80˚C. They argued that variations in temperature can be used to affect significant changes in selectivity among peptide analogues and that a temperature-based approach to reversed-phase LC can be used to distinguish varying amino acid substituents at the same site of the peptide sequence. Lee et al. [12] studied the effect of temperature on the retention behavior of 14-residue cyclic and linear cationic peptides on a C18 column in the range 5 to 80˚C as a method to detect and measure self-association of small molecules. Temperature-controlled (35 to 60˚C) reversed-phase LC has also been used to obtain separation of phenylthiohydantoin-amino acids [98]. In this case, a temperature step program was used simultaneously with a solvent gradient. In a study on the effect of temperature (20 to 60˚C) and flow rate on the chromatography of basic compounds on a C18 column [8], the efficiency was improved at higher temperatures. However, the retention of bases often showed an anomalous increase with temperature, and the complex retention mechanisms were also dependent upon the buffer cation concentration. Guiochon et al. [10] studied the effect of temperature and pressure on protein adsorption equilibrium in reversed-phase C8 columns and found that the retention time increased substantially (up to 300%) when the pressure or the temperature increased in the pressure range 47 to 147 bar, and temperature range 25 to 50˚C. The relation between temperature and ln k is nonlinear with a parabolic shape. Dolfinger and Locke [99] studied the effect of temperature and eluent composition on the selectivity on different fluorinated and hydrocarbon reversed-phase columns for separation of 15 taxanes and found an increase of k for the xylosyl taxanes with increasing temperature in the range 25 to 55˚C on the C8 column, while the other taxanes behaved more normally. The same pattern was observed on one of the fluorinated phases. The separation of DNA fragments (21 to 587 bp) on nonporous C18 alkylated PS-DVB and alkylated wide-pore silica columns was studied in the temperature range 25 to 60˚C, and the optimum temperature was found between 40 and 50˚C using a solvent gradient [100]. Increasing the temperature changes the solvent polarity [101], and superheated water [102] and superheated deuterium oxide [11] have also been used as mobile phases in reversed-phase chromatography. Superheated water (170 to 200˚C) was used for the separation of steroids on a polymer-coated zirconia column and a complementary retention and retention mechanism compared to that on a traditional silica-based C18 and acetonitrile-water mobile phase at 40˚C was found; however, unique selectivity was obtained. Nuclear magnetic resonance (NMR) analysis of a ginger extract was obtained, on-line and off-line, by temperature programmed elution (50 to 130˚C) on a silica-based C18 column using superheated deuterium oxide (see Figure 2.4). The authors reported the efficiency was much better on the silica-based reversed-phase column than on a PBD-zirconia phase using superheated deuterium oxide as mobile phase. Dimethyl sulfoxide-modified subcritical water has been used for the separation of polar and nonpolar analytes such as polyhydroxybenzenes,
2.70:2
15.0
3.10:1
30.0
45.0
60.0
mU
Advances in Chromatography
72.5
54
min 47.0
40.0
35.0
30.0
25.0
20.0
15.0
10.0
5.0
−1.7 0.0
Zingerone
FIGURE 2.4 Separation of a 30-μL ginger extract on Xterra RP18 column using superheated deuterium oxide as mobile phase. Conditions: temperature 50 to 130˚C at 4˚C/min; eluent, deuterium oxide at 0.5 mL/min; detection, 280 nm. From [11].
phenol derivatives, alkylated benzenes, and PAHs using a silica-based C18 column, which was found to be stable in the temperature range 25 to 125˚C [103]. Methanol− water mixtures at elevated temperatures up to 140˚C have been used for the separation of nonpolar analytes like PAHs and polychlorinated biphenyls (PCB) on a silica-based C18 column [104]. By using ion-pair reversed-phase chromatography on a C8 column and optimizing the temperature in the range 35 to 55˚C, an almost complete separation of 23 trinitrotoluene metabolites and EPA Method 8338-compounds was achieved [14]. Increased temperature resulted in shorter retention time and better resolution, especially for the amino aromatics. The influence of temperature on the separation of steroids on a C18 column with a mobile phase modified by β-cyclodextrin was studied in the temperature range 5 to 80˚C [105] and the plot of ln k vs. 1/T was nonlinear in the investigated range. Ion-pair chromatography of aminoalcohol enantiomers on a carbon column using a mobile phase with N-derivatized peptides as the chiral selector was studied in the temperature range from –5 to 25˚C [106]. The
Temperature Effects in Liquid Chromatography
55
plots of ln k vs. 1/T were almost linear and the best separation was obtained at the lower temperatures. The effect upon retention in hydrophobic interaction chromatography (HIC) has also been studied, and nonlinear van’t Hoff plots are found for aromatic alcohol homologues on a silica-based polyethylene glycol column in the temperature range 0 to 80˚C [20]. The effect of temperature on the separation of polar compounds like urea, sucrose, and glycine on a diol column used in HIC has also been studied and ln k was found to decrease linearly with increasing temperature in the range 30 to 60˚C [107]. The influence of temperature in ion chromatography has not been widely examined. However, a study on the separation of aluminum and its fluoro, oxalate, and citrate species, where the temperature was varied in the range –10 to 55˚C on a cation exchanger, revealed that the distribution of the species varied with temperature and that temperatures below 10˚C could be used to suppress disintegration of the fluoride and oxalate ligands [18]. The retention and resolution of alkali metal on chelating stationary phases increased with decreasing temperature in the investigated range 1 to 75˚C and the dependence of ln K on 1/T could be approximated by second-order polynomials [108]. The temperature-responsive phases are made with the intention that the temperature should influence the retention. Hosoya et al. [109] developed a poly(N-isopropylacrylamide) polymer-based stationary phase that provided temperature-dependent separation selectivity in drug separation in the temperature range 30 to 50˚C. Separations of peptides and proteins were also explored on an N-isopropylacrylamide polymer-modified column in the temperature range 5 to 10˚C [21]. An N-isopropylacrylamide copolymer-modified silica column was developed and applied for the separation of five steroids having various hydrophobicities [110]. The retention increased with increasing temperature (5 to 50˚C); however, the van’t Hoff plots were not linear. The retention of some proteins was also studied at temperatures of 5 and 30˚C, the latter giving the highest retention. The poly(N-isopropylacrylamide)bonded stationary phase showed an increase in retention and column efficiency with an increase in temperature between 30 and 40˚C using benzene derivatives and alkyl phenyl ketones as model substances [22]. The temperature-responsive properties were reduced with increasing methanol content of the mobile phase and lost at 40% or higher. The temperature-responsive poly(N-isopropylacrylamide)-modified silica was also examined for its suitability to provide separation of steroids, alkaloids and substituted anilines [23]. In this case, the temperature-responsive phase showed changes in surface properties with an increase in temperature (5 to 55˚C) with a methanol content up to 20%. The temperature-responsive chiral polymer column composed of L-valine diamide derivatives provided enantiomeric separation of amino acid derivatives in the temperature range 0 to 70˚C in aqueous mobile phases (containing 1% methanol) [24]. The retention increased with increasing temperature and the enantioselectivity was also enhanced with temperature up to a critical temperature. The effect of temperature on the enantioseparation of 71 chiral compounds was studied on four macrocyclic glycopeptide chiral selectors — teicoplanin, its aglycone, ristocetin A and vancomycin — using both reversed-phase, normal-phase, and polar ionic mode
56
Advances in Chromatography
(PIM) mobile phases [111]. In the majority of cases, the enantioseparation decreased or even vanished when the temperature was increased. All van’t Hoff plots were linear, showing that the selector did not change in the investigated temperature range (5 to 45˚C). The retention of enantiomers on silica-based tris(1,3-dimethylphenylcarbamate)-derivatized cellulose was studied in the temperature range 12 to 40˚C in the reversed-phase mode using various mobile phase compositions [112]. The van’t Hoff plots were linear, the retention increased with decreasing temperature and the thermodynamic data revealed that the enantio separation was enthalpy-controlled. A human serum albumin (HSA) chiral stationary phase has also been explored for its potential to give separation of dansyl amino acids in the temperature range 3 to 27˚C [113]. Different van’t Hoff plot shapes were observed with different mobile phase pH, indicating a change in retention mechanism.
2.4 CAPILLARY LC The main advantage of narrow I.D. columns is the possibility to achieve lower detection limits when using concentration sensitive detectors, and this is especially advantageous when the sample amount is limited. The first attempts to study the effects of temperature with narrow-bore columns was in the 1980s [114–118] but it is the advent of packed capillary columns that has spurred the interest in using temperature as a parameter in liquid chromatography. In the following sections, the recent achievements in capillary LC utilizing the temperature for retention control and separation optimization are presented.
2.4.1 COLUMNS
AND
STATIONARY PHASES
The low thermal mass of the capillary columns results in a quick response to temperature changes, making rapid temperature programs possible. Thus, capillary columns with temperature gradients have recently been demonstrated in several studies (see Table 2.1). Most of the recent studies on temperature in capillary chromatography have been carried out using packed columns; however, both monolithic columns [119–121] and open capillary columns [122–126] have been utilized. So far, open tubular columns have not gained much interest because of either the requirement of a narrow I.D. or of very high temperatures. Monolithic polystyrene/divinylbenzene (PS-DVB) capillary columns have been used for separation of nucleic acids, proteins, and tryptic digests. Because of the polymerization-related variation in surface area of these monolithic columns, temperature regulation has been applied to optimize the resolution and obtain harmonized elution profiles for columns in parallel [119, 120] (see Figure 2.5). The inner diameter of these columns were 0.2 mm or less and they had a relatively short length of about 6 cm. Generally, temperatures above 40˚C are used isothermally [119, 120, 127–131] but temperaturegradient programming has also been used [121]. A methacrylate monolithic column intended for electrochromatography has also been used in the LC mode and showed a decrease in retention with increasing temperature (12 to 50˚C) [132]. The most common inner diameters of the packed capillary columns are 0.25 and 0.32 mm, but 0.50 mm I.D. has also been used (see Table 2.1). The column body
ACN/phosphate buffer (pH 7) 25–90 (I) MeOH/0.8% H3PO4 (56/44) 20–65 (G)
30–120 (G) 68–55 (inv. G) 30–80 (I)
0.05 × 150 mm nonporous PB-zirconia 0.30 × 150 mm Hypersil C18 BDS (3)
30–90 (G) 50–100 (G+I)
0.32+ × 150 mm Hypersil ODS (3) 0.50 × 100 mm PBD-ZrO2 (3)
0.25 × 200 mm Kromasil C18 (3.5) 0.50 × 100 mm Kromasil C18 (3.5) 0.20 × 100 mm Hypercarb (5)
15–70 (I) 30–150 (G+I) UV ELSD + ESITOFMS UV UV
ELSD
UV
UV and MS
Detector
Lab-made cLC-oven
UV UV
“Tailor-made” UV cLC-oven ELSD Col. oven UV
GC-oven GC-oven
cLC-oven GC-oven
30–120 (G)
ACN/water ACN/20 mM phosphate pH 7 (50/50) ACN ACN/THF/water (40/5/55) ACN/water; pH 2.2 and 12
GC-oven
50–120 (I+G)
22–70 (I)
Heating System
0.18 × 60 mm PLRP-S 300 PS-PVB (5) ACN/water/TFA 0.18 × 60 mm monolithic PS-PVB 0.32 × 350 mm Hypersil ODS (3) Ethyl acetate/ACN/TEA/AA (45.0/44.5/10.0/0.1) 0.32 × 100 mm PS-PVB SAX (10) pH-gradient of amine buffers 0.32 × 600 mm Kromasil C18 (3.5) Acetone/ACN
ACN/water/additive
Mobile Phase(s)
Temperature Range (°°C) (G = Gradient, I = Isothermal) Col. oven or circulating water bath GC-oven
0.20 × 60 mm monolithic PS-PVB
Dimension and Packing Particle Diameter (µm)
TABLE 2.1 Temperature Studies on Capillary Columns
Split 20
500a 100b 0.06
0.05 0.05
100 0.05–0.5
0.5
0.05
0.5
Inj. Vol. (µL)
Ref.
Polymer additives PEG 4000-3500 Sulfated glycosaminoglycan disaccharides Test compounds Chlorophenoxy acid herbicides
Hindered amine light stabilizers Proteins Polyglycerol fatty acid esters Alkylbenzenes Test compounds
Proteins
[140] [156]
[138] [155] [144]
[87] [139]
[19] [146]
[153]
[121]
Peptides, proteins [162]
Analyte(s)
Temperature Effects in Liquid Chromatography 57
0.32 0.32 0.32 0.32
× × × ×
400 400 300 250
mm mm mm mm
Kromasil C18 (3.5) Hypersil ODS (3) Hypersil ODS (3) Hypersil ODS (3)
0.30 × 150 mm Spherisorb ODS2 (3)
Temperature Range (°°C) (G = Gradient, I = Isothermal)
ACN/ammonium acetate pH 7 (69/31) ACN/DCM, THF/water, ACN/water ACN/water (30/70) MeOH/citric acid pH 3.1 (5/95)
Col. oven cLC-oven
51–57 (I) 10–90 (I+G)
80–30 (inv. G) 30–100 (G)
8–75 (G)
cLC-oven, GC-oven cLC-oven GC-oven
cLC-oven
GC-oven
80–95 (G)
15–70 (G)
GC-oven
cLC-oven GC-oven
Col. oven
Heating System
110–170 (G)
54–57 (I) ACN/(TEA+EDTA; pH 7) gradient 5–90 (G) ACN added TEA+FA ACN/ammonium formate pH 35–100 (G) 4 (25/75)
Mobile Phase(s)
0.32 × 1000 mm 4000Å Nucleosil silica TCB/isododecane/BuOH (5) 0.32 × 250 mm Hypersil ODS (3) ACN/ammonium formate pH 6 (40/60) 0.20 × 60 mm monolithic PS-DVB × 4 ACN/TEEA pH 7 0.32 × 200 mm C18 columnsc (3) MeOH and ACN/water pH 2–7
0.25 × 250 mm Kromasil C18 (3.5) 0.32 × 300 mm Kromasil C18 (3.5)
0.20 × 60 mm monolithic PS-DVB
Dimension and Packing Particle Diameter (µm)
TABLE 2.1 (CONTINUED) Temperature Studies on Capillary Columns
ELSD Electrochemical
UV
UV
LIF UV
UV
ELSD
UV UV
UV
Detector
0.55 0.06
0.2
5
1 0.05
1000b
0.06
500b at 5˚C 10
1
Inj. Vol. (µL)
PEG 1000 Tyrosines, DOPA
Diisocyanate derivatives DNA fragments Acidic, basic, neutral compounds Triazine herbicides Polystyrenes
Nucleic acids, proteins Irganox 1076 Citalopran, fluoxetine, paroxetine + metabolites Polymer blend
Analyte(s)
[163] [151]
[149]
[158]
[119] [49]
[135]
[17]
[136] [157]
[120]
Ref.
58 Advances in Chromatography
cLC-oven GC-oven GC-oven GC-oven
GC-oven
10–70 (G) 30–150 (G) 25–70 (I) 40–150 (I+G)
50–150 (G)
ACN/dichloromethane (70/30) 0.32 × 400–700 mm Hypersil ODS (3/5) ACN/DMF and other nonaqueous solvents 0.32 × 250 mm Suplex pKb-100 (5) ACN/5% ammonium acetate/AA (94.9/5/0.075) 0.32 × 230–435 mm Hypersil ODS (5) ACN or 0.32 × 230 mm Kromasil C18 (5) MeOH/AA/pyridine/water (85/4/1/10) 0.32 × 700 mm Hypersil ODS or BDS ACN/DMF (90/10) or (3/5) or Kromasil C18 (5) ACN/pyridine (95/5)
0.32 × 400 mm YMC C30 (5)
GC-oven
70–130 (G)
cLC-oven GC-oven
7–90 (G) 30–160 (G)
ACN ACN or ethyl acetate/ACN (10/90)
0.32 × 350–1000 mm Kromasil C18 (5) ACN/MIBK (75/25)
GC-oven GC-oven GC-oven
50–120 (G) 30–130 (G) 30–130 (G)
ACN ACN and ACN/MIBK ethyl acetate/ ACN/TEA (40/50/10)
cLC-oven cLC-oven GC-oven
0.32 × 300 mm Kromasil C18 (3.5) 0.32 × 350–1000 mm Kromasil C18 (5) 0.32 × 350 mm Hypersil ODS (3) 0.32 × 350 mm Hypersil BDS (5) 0.32 × 350 mm Kromasil C18 (5) 0.32 × 280 mm Kromasil C18 (5) 0.32 × 280 mm Hypersil ODS (3) 0.32 × 700 mm Hypersil BDS (5)
6–90 (G+I) 5–90 (G) 28–120 (G)
ACN/water ACN/(TEA+FA) ACN/TEA (95/5)
0.32 × 250 mm Hypersil C18 (3) 0.32 × 500 mm Kromasil C18 (3.5) 0.32 × 400 mm Hypersil ODS (3)
ELSD
ICP-MS
UV-VIS
UV or ELSD
UV-VIS
ELSD
UV ELSD
UV + FT-IR ELSD ELSD
UV ELSD ELSD
0.06
0.06
100
0.06
20 at 10˚C
0.06 at 100˚C
100 at 7˚C 0.05
0.06 0.06 0.50
0.05 100 at 5˚C 0.05
[159]
[168]
[134] [150]
[166] [167] [147]
[164] [165] [148]
[171]
[170]
Fatty amides, [172] glycerylmonoste arate
Retinol, retinoic acids Tetreethyl- and tetramethyl lead
Polymer additives [169]
Irganox 1076 Polymer additives and glyceryl monostearate Hydrocarbon waxes Retinyl esters
Iodixanol Ceramides Hindered amine stabilizers Polymer additives Technical waxes Hindered amine stabilizers
Temperature Effects in Liquid Chromatography 59
c
b
a
MeOH/acetate buffer pH 5 (30/70) ACN/water ACN/water (72/28) MeOH/water ACN/DMF or ACN/N,Ndimethylacetamide ACN/water or MeOH/water ACN/water or MeOH/water MeOH or ACN/water
Mobile Phase(s)
Lab. made Lab. made Lab. made
GC-oven SFC-oven GC-oven GC-oven
30–120 (G) 24–80 (I) 25–150 (I+G) 50–150 (G) 100–150 (I) 100–200 (I) 30–80 (G)
Lab.-made
Heating System
150 (I)
Temperature Range (°°C) (G = Gradient, I = Isothermal)
cold zone at 10˚C using a column switching system columns used: Zorbax C18 (XDB, SB and Extend), Hypersil ODS, XTerra MS 18 and YMC C30
AA = acetic acid FA = formic acid TCB = trichlorobenzene
0.05 × 1500 mm SB-methyl-100 0.05–0.1 × 8000mm octyl polysiloxane 0.08–0.25 × 1000–5000 mm Develosil ODS (3/5/7/10)
0.05 × 2000 mm cyano and octyl polysiloxane 0.18 × 250 mm Zorbax SB ODS (6) 0.2 × 800 mm Phenomenex C18 (5) 0.05 × 1500 mm octyl polysilixane 0.32 ×700 mm Hypersil ODS (3/5)
Dimension and Packing Particle Diameter (µm)
TABLE 2.1 (CONTINUED) Temperature Studies on Capillary Columns
UV UV UV
UV UV UV + MS UV
UV
Detector
0.07 0.005 0.1–0.2
0.06 0.06 0.01 0.06
10
Inj. Vol. (µL)
Test compounds Test compounds p-nitrobenzyl esters of fatty acids
Alkylbenzenes Test compounds Test compounds Polymer additives
Test compounds
Analyte(s)
[123] [122] [175]
[48] [173] [125] [174]
[126]
Ref.
60 Advances in Chromatography
Temperature Effects in Liquid Chromatography
61
Signal intensity
Column 1 55ºC
Column 1 55ºC
Column 2 56ºC
Column 2 55ºC
Column 3 57ºC
Column 3 55ºC
Column 4 55ºC 0
2
12
4 Time (min)
6
13 14 % acetonitrile (a)
8
15
Column 4 55ºC 0
2
10
4 6 Time (min) 11
8
10
12 13 % acetonitrile
14
(b)
FIGURE 2.5 Adjustment of the chromatographic elution profiles of different columns by means of variation of column temperature. (a) The same sample was injected onto four different columns kept at the same temperature in a commercially available enforced air circulation oven with Peltier cooling. Because of differences in surface area, retention of DNA fragments varies between columns. (b) Variation of the individual column temperatures allows harmonization of chromatographic profiles. Columns, 4 × monolithic PS/DVB, 60 × 0.2 mm i.d., mobile phase, (A) 100 mM TEAA at pH 7.0, (B) 100 mM TEAA at pH 7.0, 25% acetonitrile; linear gradient, (a) 45–61% B in 8.0 min, (b) 38–56% B in 10 min; flow rate, 2.0–3.0 μL/min; temperature, (a) 55˚C, (b) 55–57˚C; detection, LIF, emission monitored at 525 nm; injection volume, 1 μL each; sample, equimolar mixture of two FAM-labeled 209bp Y chromosome alleles differing in a single base (168 A > G). From [119].
62
Advances in Chromatography
material is most often fused silica — PEEK shielded fused silica and glass-lined steels have also been employed in some commercial columns, but these are less suited for temperature-gradient programming. The length of the packed columns varies from 100 to 1000 mm, depending on the particle size and the desired resolution. A particle size of 3 or 3.5 μm is most common and the columns are relatively short. When the inner diameter of the column is reduced, the injection volume must be reduced accordingly (the down-scale factor) to maintain high efficiency. Typical injection volumes in capillary LC should be in the range 0.05 to 0.1 μL (see Table 2.1). However, packed capillary columns can be used advantageously for trace determination of analytes in samples of limited sample size [133] by using larger sample volumes that can be introduced either by on-column focusing techniques [134] or by column switching techniques [135, 136]. Injection of large volumes without detrimental effect on the column efficiency can be achieved by solute focusing with a solvent with low elution strength [137], or by temperature [138]. Up to 1.0 mL of sample has been applied utilizing solvent-promoted focusing in a 0.32 mm I.D. column switching system [135, 137] (see Figure 2.6). In reversedAUFS
0.02
3 0.01
1 4 2
0.00
20
40
60
min
FIGURE 2.6 Packed capillary column-switching LC separation of 12 ng of each of the 12MP derivatives of 2,6-TDI (1), HDI (2), 2,4-TDI (3) and MDI (4) from a spiked air sampling filter on a 3 μm Hypersil ODS (0.32 × 250 mm) column. The injection volume was 1.0 mL and the mobile phase was acetonitrile: 10 mM ammonium formate (pH 6.0) (40:60, v/v). A temperature program from 80˚C (37 min) to 95˚C (25 min) using a temperature ramp of 5.0˚C/min was used. From [135].
Temperature Effects in Liquid Chromatography
63
phase systems, the retention is usually lower at low temperatures and up to 500 μL of a sample containing a hydrophobic analyte has been introduced at 5˚C using a column switching system [136]. In this case both the sample solvent and eluting mobile phase consisted of 100% acetonitrile and the analyte Irganox 1076 was completely retained at the 27 mm-long C18 enrichment column at 5˚C. A subsequent temperature program from 5 to 90˚C provided optimal separation of the analyte from the matrix components, which was coextracted from the polymer sample using 100% acetonitrile as extraction solvent. Of the packed capillary columns, silica-based reversed-phase column materials have been most extensively studied (see Table 2.1). Only a limited number of studies have been published using other materials like silica and ion exchange. In an early study, Jinno and Hirata investigated the (low) temperature effect in normal phase micro LC [16], and lately Molander et al. [17] resolved an extruded polymer blend by using temperature-programmed (110 to 170˚C) elution on large-pore silica stationary phase. The separation was obtained according to both size exclusion and normal phase interaction. Isoelectric point (pI) separation of proteins by pH-gradient ion-exchange with temperature control was recently demonstrated on a packed capillary column [19]. The temperature tolerance of the zirconia materials has promoted their use for high temperature separations [15, 79, 139, 140]; however, only one paper on their use in capillary LC has been published [139]. Polybutadiene-modified zirconia packed in 0.1 mm I.D. glass-lined stainless steel columns was found to be stable at 100˚C using acetonitrile/phosphate buffer (pH 7) as mobile phase. On the other hand, in temperature-programming mode (not exceeding 100˚C), a rapid decrease in both column efficiency and retention factors was observed. Even though carbon materials have shown high stability under temperature-programmed high-temperature LC conditions [15], carbon materials have been used mostly at ambient temperature in capillary LC [141–143]. So far, only one study has used the temperature for retention control [144]. Linear van’t Hoff plots were obtained for sulphated glycosamin glycan disaccharides, however, of different sign at acidic (2.2) and alkaline (12) pH. The effect of temperature on retention on the cellulose triacetate enantioselective stationary phase has been studied [145], but so far no separation using the temperature-responsive stationary phases [21–24] have been demonstrated on capillary columns.
2.4.2 TEMPERATURE PROGRAMMING
AND INSTRUMENTATION
The temperature range studied in capillary LC has been limited on the high end mainly by the stability of the packing material but, in some cases, also by the stability of the solutes. Temperatures below 5˚C have seldom been explored due to the lower efficiency at low temperature and by the available ovens (see Table 2.1). Only two commercially available LC ovens with the option for temperature programming exist with temperature ranges of 5 to 150˚C and < 0 to 200˚C, respectively. The former oven (cLC-oven) has been used in a number of publications, as can be seen in Table 2.1, but the drawback is the slow cooling back to 5˚C after performing a run. For
64
Advances in Chromatography
temperature programming, the ovens should provide rapid return to the start of the temperature program; however, the Peltier elements in the larger ovens are too slow and cryo-cooling is required for a rapid return to start. If temperatures below 0˚C are needed (e.g., for focusing of solutes), an oven with two separate temperature zones is advantageous [138]. For temperatures above ambient, GC-ovens are suitable both for isothermal and temperature-programmed elution. Other laboratory-made devices for column temperature control have also been used, as shown in Table 2.1. The repeatability of retention time with temperature programs has been found to be comparable to that obtained with solvent gradient elution. The relative standard deviations (%RSDs) were in the range 0.5 to 1.4 for a 30 to 150˚C (at 1.5 K/min) gradient [146], 0.4 to 1.2 for a 7 to 90˚C (at 3 K/min) gradient [134], and 2 for a 30 to 120˚C (at 5 K/min) gradient [138]. Temperature programming has been used in a number of applications (Table 2.1) to obtain separation in a reasonable time with a good efficiency. Temperatures up to 90˚C have been used on silica-based C18 columns with aqueous mobile phases, even though temperatures above 60˚ are not recommended [15]. When nonaqueous mobile phases are used, higher temperatures may be utilized. Andersen et al. separated hindered amine stabilizers using nonaqueous mobile phases and temperature programming from 30 to 120/130˚C on C18 columns [146–148]. Temperatures up to 200˚C have been used on the open tubular columns by Liu et al., obtaining over 1 mill theoretical plates in 50 minutes using a 20 m-long column [122]. The majority of studies on capillary LC with temperature control have utilized UV detection or evaporative light scattering detection (ELSD) (see Table 2.1). When temperature−gradient programming is used in combination with UV detection, a rising baseline is often observed due to refractive index changes but a rising baseline can also be due to column bleeding [15]. The rising baseline may be corrected mathematically by a commercial software package [149] or by dual wavelength measurements and corrections [77]. The nebulizer of the ELSD has been modified for the low flow rates used in capillary LC to obtain improved limits of detection and increased linearity [150] and baseline drift with temperature−gradient programming is normally not a problem. The laboratory-made electrochemical detector (ECD) for packed capillary columns also showed a rising baseline with temperature programming but the drift could be corrected mathematically [151]. Mass spectrometry (MS), especially with electrospray ionization, benefits from the low flows used in capillary LC and the compatibility between temperature programming in LC and MS was shown more than a decade ago [152]. However, MS has only been applied in a few studies where the temperature has been used actively for retention control in capillary LC [146, 153]. Since temperature programming can be expected to affect the electrospray ionization process less than solvent gradients, temperature gradient programming in LC with MS detection can in many cases be a better choice than solvent gradient elution. Low-volume flow cells for NMR have now become commercially available and the combination of capillary LC and NMR seems very promising compared to conventional-sized systems when the sample amount is limited [154]. Resonance shifts due to changes in solvent composition are avoided by isocratic elution and much better detection limits can be achieved.
Temperature Effects in Liquid Chromatography
65
A small overpressure is necessary to keep the mobile phase in the liquid phase when temperatures above the boiling point of the mobile phase are applied. This is normally done by connecting a fused silica capillary with small I.D. (15 to 30 μm) between the column and the detector when MS or ELSD is used and by using a similar capillary from the detector to waste when UV detection is utilized.
2.4.3 APPLICATIONS Even though most of the studies on temperature effects in capillary columns have utilized test compounds (see Table 2.1) for exploring the separation power of packed capillary column systems, a limited number of papers have also shown the applicability of this methodology for real-life samples [19, 134, 136, 155–159]. Capillary LC with large-volume injection and temperature-programmed elution was used for the determination of Irganox 1076 that was extracted from low-density polyethylene by microwave-aided extraction with acetonitrile [134]. In this case, acetonitrile was used both as injection solvent (100 μL) and as the mobile phase. By utilizing temperature-promoted, large-volume solute focusing, sample volumes up to 500 μL could be used to improve the detection limit to 0.6 ppm [136]. The concentration of retinyl palmitate in Arctic seal liver was determined using a nonaqueous temperature-programmed reversed-phase separation with large-volume oncolumn injection [159]. A method for the simultaneous determination of three selective serotonin reuptake inhibitors and their metabolites in solid phase extracts from plasma, using temperature-programmed aqueous reversed-phase separation with on-column focusing of large injection volumes, provided quantification limits in the concentration range 0.05 to 0.26 μM [157]. The presence of eight triazines in ground water was determined using temperature-programmed elution on a reversed-phase column after large volume injection [158], while chlorophenoxy acid herbicides and their esters were determined in extracts from soil using large-volume injection and temperature-programmed elution on a C18 reversed-phase column [156] (see Figure 2.7). The detection limits reported were between 0.3 and 0.5 μg/g of soil. Low levels of polyethylene glycols (PEG) in extracts from low-density polyethylene could be determined by large-volume injection, temperature-programmed elution on a C18 reversed-phase column using a column switching system [155]. In this case, inverse temperature programming from 68 to 45˚C was used since the retention of PEGs increases with increasing column temperature, a fact explained by conformational changes of PEGs at different temperatures. One special application is the isoelectric point separation of proteins by capillary pH-gradient ion exchange chromatography, where the temperature was used to optimize the separation of skimmed milk proteins [19]. Temperature has been actively used for high-resolution chromatographic determination of nucleic acids on PS-DVB monolithic columns as well as on packed PSDVB columns [160]. In addition to their use in genomic research, the PS-DVB monolithic columns have also been used in proteomics [161]. Huber et al. recently published a study on the effect of temperature and mobile phase additives on peptide and protein analysis on PS-DVB monoliths [162] where they showed that the efficiency improved significantly upon increasing the temperature up to 70˚C. The
66
Advances in Chromatography
2, 4-D
2, 4-D methyl ester
2, 4-DB 2, 4-D butyl ester
b a
0
10 Time, min
20
FIGURE 2.7 LC chromatogram of (a) unspiked soil; (b) soil spiked with 4.0 μg/g of chlorophenoxy acids. Column: LC Packing Hypersil C18 BDS, 3 μm, 150 mm × 0.3 mm I.D. Mobile phase: MeOH/H3PO4 0.8% aqueous solution (56/44, v/v). Flow rate in column: 10 μL/min. Temperature gradient from 20 to 65 ˚C at 9˚C/min. Injection volume: 20 μL. From [156].
selectivity of peptide elution was significantly modulated by the temperature, whereas the effect on protein analysis was only minor. The construction of monolithic capillary arrays, fine-tuned by using the temperature (see also Figure 2.5), allows high-throughput analyses and their use will probably increase in the years to come.
REFERENCES 1. Ooms, B. Temperature control in high performance liquid chromatography. LC-GC, 1996, 14, 306–313. 2. Greibrokk, T. and Andersen, T. High-temperature liquid chromatography. J. Chromatogr. A, 2003, 1000, 743–755. 3. Li, J.W., Hu, Y., and Carr, P.W. Fast separations at elevated temperatures on polybutadiene-coated zirconia reversed-phase material. Anal. Chem., 1997, 69, 3884–3888.
Temperature Effects in Liquid Chromatography
67
4. Guillarme, D., Heinisch, S., and Rocca, J.L. Effect of temperature in reversed phase liquid chromatography. J. Chromatogr. A, 2004, 1052, 39–51. 5. Dolan, J.W. Temperature selectivity in reversed-phase high performance liquid chromatography. J. Chromatogr. A, 2002, 965, 195–205. 6. Greibrokk, T. and Andersen, T. Temperature programming in liquid chromatography. J. Sep. Sci., 2001, 24, 899–909. 7. Sander, L.C. and Wise, S.A. The influence of column temperature on selectivity in reversed-phase liquid chromatography for shape-constrained solutes. J. Sep. Sci., 2001, 24, 910–920. 8. McCalley, D.V. Effect of temperature and flow-rate on analysis of basic compounds in high-performance liquid chromatography using a reversed-phase column. J. Chromatogr. A, 2000, 902, 311–321. 9. Böhm, V. Use of column temperature to optimize carotenoid isomer separation by C30 high performance liquid chromatography. J. Sep. Sci., 2001, 24, 955–959. 10. Szabelski, P., Cavazzini, A., Kaczmarski, K., Liu, X., Van Horne, J., and Guiochon, G. Experimental studies in pressure/temperature dependence of protein adsorption equilibrium in reversed-phase high-performance liquid chromatography. J. Chromatogr. A, 2002, 950, 41–53. 11. Saha, S., Smith, R.M., Lenz, E., and Wilson, I.D. Analysis of ginger extract by highperformance liquid chromatography coupled to nuclear magnetic resonance spectroscopy using superheated deuterium oxide as the mobile phase. J. Chromatogr. A, 2003, 991, 143–150. 12. Lee, D.L., Mant, C.T., and Hodges, R.S. A novel method to measure self-association of small amphipathic molecules. J. Biol. Chem., 2003, 278, 22918–22927. 13. Chen, Y., Mant, C.T., and Hodges, R.S. Temperature selectivity effects in reversed phase liquid chromatography due to conformation differences between helical and non-helical peptides. J. Chromatogr. A, 2003, 1010, 45–61. 14. Borch, T. and Gerlach, R. Use of reversed-phase high-performance liquid chromatography-diode array detection for complete separation of 2,4,6-trinitrotoluene metabolites and EPA Method 8330 explosives: Influence of temperature and ion-pair reagent. J. Chromatogr. A, 2004, 1022, 83–94. 15. Marin, S.J., Jones, B.A., Felix, W.D., and Clark, J. Effect of high-temperature on high-performance liquid chromatography column stability and performance under temperature-programmed conditions. J. Chromatogr. A, 2004, 1030, 255–262. 16. Jinno, K. and Hirata, Y. Investigation of the low temperature effect in normal phase micro high performance liquid chromatography. J. HRC & CC, 1982, 5, 85–90. 17. Molander, P., Greibrokk, T., Iveland, A., and Ommundsen, E. Simultaneous dualmechanism separation of a semicrystalline polymer blend on large-pore silica by temperature-programmed packed capillary liquid chromatography. J. Sep. Sci., 2001, 24, 136–140. 18. Busch, M. and Seubert, A. Influence of column temperature on the ion chromatographic separation of aluminum species. Anal. Chim. Acta, 1999, 399, 223–235. 19. Andersen, T., Pepaj, M., Trones, R., Lundanes, E., and Greibrokk, T. Isoelectric point separation of proteins by capillary pH-gradient ion-exchange chromatography. J. Chromatogr. A, 2004, 1025, 217–226. 20. Wei, Y., Yao, C., Zhao, J., and Geng, X. Influences of the mobile phase composition and temperature on the retention behaviour of aromatic alcohol homologues in hydrophobic interaction chromatography. Chromatographia, 2002, 55, 659–665.
68
Advances in Chromatography 21. Kanazawa, H., Yamamoto, K., Kashiwase, Y., Matsushima, Y., Takai, N., Kikuchi, A., Sakurai, Y., and Okano, T. Analysis of peptides and proteins by temperatureresponsive chromatographic system using N-isopropylacrylamide polymer-modified columns. J. Pharm. Biomed. Anal., 1997, 15, 1545–1550. 22. Go, H., Sudo, Y., Hosoya, T., Ikegami, T., and Tanaka, N. Effects of mobile-phase composition and temperature on the selectivity of poly(N-isopropylacrylamide)bonded silica gel in reversed-phase liquid chromatography. Anal. Chem., 1998, 70, 4086–4093. 23. Song, Y.-X., Wang, J.-Q., Su, Z.-X., and Chen, D.-Y. High-performance liquid chromatography on silica modified with temperature-responsive polymers. Chromatographia, 2001, 54, 208–212. 24. Kurata, K., Shimoyama, T., and Dobashi, A. Enantiomeric separation using temperature-responsive chiral polymers composed of L-valine diamide derivatives in aqueous liquid chromatography. J. Chromatogr. A, 2003, 1012, 47–56. 25. Jiskra, J., Claessens, H.A., and Cramers, C.A. Thermodynamic behaviour in capillary electrochromatography. J. Sep. Sci., 2002, 25, 569–576. 26. Snyder, L.R. Temperature-induced selectivity in separations by reversed-phase liquid chromatography. J. Chromatogr., 1979, 179, 157–172. 27. Grushka, E., Colin, H., and Guiochon, G. Retention behavior of alkylbenzenes as a function of temperature and mobile phase composition in reversed-phase chromatography. J. Chromatogr., 1982, 248, 325–339. 28. Takeuchi, T., Kumaki, M., and Ishii, D. Role of column temperature in open-tubular microcapillary liquid chromatography. J. Chromatogr., 1982, 235, 309–322. 29. Miyabe, K. and Guiochon, G. Thermodynamic interpretation of retention equilibrium in reversed-phase liquid chromatography based on enthalpy-entropy compensation. Anal. Chem., 2002, 74, 5982–5992. 30. Colin, H., Diez-Masa, J.C., Guiochon, G., Czajkowska, T., and Miedziak, I. The role of the temperature in reversed-phase high-performance liquid chromatography using pyrocarbon-containing adsorbents. J. Chromatogr., 1978, 167, 41–65. 31. Knox, J.H. and Vasvari, G. The performance of packings in high-speed liquid chromatography. III. Chemically bonded pellicular materials. J. Chromatogr., 1973, 83, 181–194. 32. Colin, H. and Guiochon, G. Introduction to reversed-phase high-performance liquid chromatography. J. Chromatogr., 1977, 141, 289–312. 33. Colin, H. and Guiochon, G. Comparison of some packings for reversed-phase highperformance liquid — solid chromatography. II. Some theoretical considerations. J. Chromatogr., 1978, 158, 183–205. 34. Horvath, C. and Melander, W. Liquid chromatography with hydrocarbonaceous bonded phases, theory and practice of reversed phase chromatography. J. Chromatogr. Sci., 1977, 15, 393–404. 35. Issaq, H.J. and Jaroniec, M. Enthalpy and entropy effects for homologous solutes in HPLC with alkyl chain bonded phases. J. Liq. Chromatogr., 1989, 12, 2067–2082. 36. Alvarez-Zepeda, A., Barman, B.N., and Matire, D.E. Thermodynamic study of the marked differences between acetonitrile/water and methanol/water mobile-phase systems in reversed-phase liquid chromatography. Anal. Chem., 1992, 64, 1978–1984. 37. Guillaume, Y.C. and Guinchard, C. Retention mechanism of weak polar solutes in reversed phase liquid chromatography. Anal. Chem., 1996, 68, 2869–2873. 38. Guillaume, Y.C. and Guinchard, C. ACN clusters in a water/ACN mixture, with implications for the RPLC weak polar solute retention. Anal. Chem., 1997, 69, 183–189.
Temperature Effects in Liquid Chromatography
69
39. FuAn, W., JianChi, S., YingChen, G., YiHong, W., and Yan, Z. Rule of intersection point between retention of homologues and column temperature in reversed-phase liquid chromatography. Microchem. J., 1995, 52, 200–204. 40. Bai, Q. and Geng, X. The temperature convergence of biopolymers in reversed-phase liquid chromatography. Chromatographia, 2001, 54, 335–338. 41. Chester, T.L. and Coym, J.W. Effect of phase ratio on van ‘t Hoff analysis in reversedphase liquid chromatography, and phase-ratio-independent estimation of transfer enthalpy. J. Chromatogr. A, 2003, 1003, 101–111. 42. Jinno, K., Ohshima, T., and Hirata, Y. Investigation of the low temperature effect in reversed phase micro high performance liquid chromatography. J. High Resolut. Chromatogr., 1982, 5, 621–624. 43. Coym, J.W. and Dorsey, J.G. Reversed-phase retention thermodynamics of pure-water mobile phases at ambient and elevated temperature. J. Chromatogr. A, 2004, 1035, 23–29. 44. Castells, C.B., Gagliardi, L.G., Rafols, C., Roses, M., and Bosch, E. Effect of temperature on the chromatographic retention of ionizable compounds. I. Methanol-water mobile phases. J. Chromatogr. A, 2004, 1042, 23–35. 45. Li, J. Effect of temperature on selectivity in reversed-phase liquid chromatography, a thermodynamic analysis. Anal. Chim. Acta, 1998, 369, 21–37. 46. Buckenmaier, S.M.C., McCalley, D.V., and Euerby, M.R. Rationalisation of unusual changes in efficiency and retention with temperature shown for bases in reversedphase high-performance liquid chromatography at intermediate pH. J. Chromatogr. A, 2004, 1060, 117–126. 47. Bowermaster, J. and McNair, H.M. Temperature programmed microbore HPLC-Part I. J. Chromatogr. Sci., 1984, 22, 165–170. 48. Chen, M.H. and Horvath, C. Temperature programming and gradient elution in reversed-phase chromatography with packed capillary columns. J. Chromatogr. A, 1997, 788, 51–61. 49. Tran, J.V., Molander, P., Greibrokk, T., and Lundanes, E. Temperature effects on retention in reversed phase liquid chromatography. J. Sep. Sci., 2001, 24, 930–940. 50. Zhu, P.L., Snyder, L.R., Dolan, J.W., Djordjevic, N.M., Hill, D.W., Sander, L.C., and Waeghe, T.J. Combined use of temperature and solvent strength in reversed-phase gradient elution. I. Predicting separation as a function of temperature and gradient conditions. J. Chromatogr. A, 1996, 756, 21–39. 51. Zhu, P.L., Dolan, J.W., and Snyder, L.R. Combined use of temperature and solvent strength in reversed-phase gradient elution. II. Comparing selectivity for different samples and systems. J. Chromatogr. A, 1996, 756, 41–50. 52. Zhu, P.L., Dolan, J.W., Snyder, L.R., Hill, D.W., VanHeukelem, L., and Waeghe, T.J. Combined use of temperature and solvent strength in reversed-phase gradient elution. III. Selectivity for ionizable samples as a function of sample type and pH. J. Chromatogr. A, 1996, 756, 51–62. 53. Zhu, P.L., Dolan, J.W., Snyder, L.R., Djordjevic, N.M., Hill, D.W., Lin, J.T., Sander, L.C., and Van Heukelem, L. Combined use of temperature and solvent strength in reversed-phase gradient elution. IV. Selectivity for neutral (non-ionized) samples as a function of sample type and other separation conditions. J. Chromatogr. A, 1996, 756, 63–72.
70
Advances in Chromatography 54. Dolan, J.W., Snyder, L.R., Djordjevic, N.M., Hill, D.W., Sounders, D.L., Van Heukelem, L., and Waeghe, T.J. Simultaneous variation of temperature and gradient steepness for reversed-phase high-performance liquid chromatography method development. I. Application to 14 different samples using computer simulation. J. Chromatogr. A, 1998, 803, 1–31. 55. Wolcott, R.G., Dolan, J.W., and Snyder, L.R. Computer simulation for the convenient optimization of isocratic reversed-phase liquid chromatographic separations by varying temperature and mobile phase strength. J. Chromatogr. A, 2000, 869, 3–25. 56. Neue, U.D. and Mazzeo, J.R. A theoretical study of the optimization of gradients at elevated temperature. J. Sep. Sci., 2001, 24, 921–929. 57. Cole, L.A., Dorsey, J.G., and Dill, K.A. Temperature dependence of retention in reversed-phase liquid chromatography. 2. Mobile-phase considerations. Anal. Chem., 1992, 64, 1324–1327. 58. Kim, H., Gritti, F., and Guiochon, G. Effect of the temperature on the isotherm parameters of phenol in reversed-phase liquid chromatography. J. Chromatogr. A, 2004, 1049, 25–36. 59. Issaq, H.J., Fox, S.D., Lindsey, K., McConnell, J.H., and Weiss, D.E. Effect of temperature on HPLC separations using C1, C4, C8, and C18 alkyl chain bonded silica columns. J. Liq. Chromatogr., 1987, 10, 49–70. 60. Dolan, J.W., Snyder, L.R., Blanc, T., and Van Heukelem, L. Selectivity differences for C18 and C8 reversed-phase columns as a function of temperature and gradient steepness. I. Optimizing selectivity and resolution. J. Chromatogr. A, 2000, 897, 37–50. 61. Ducry, M.W.J., Orendorff, C.J., Pemberton, J.E., and Sander, L.C. Structure-function relationships in high-density octadecylsilane stationary phases by Raman spectroscopy. 1. Effects of temperature, surface coverage, and preparation procedure. Anal. Chem., 2002, 74, 5576–5584. 62. Tchapla, A., Heron, S., Colin, H., and Guiochon, G. Role of temperature in the behavior of a homologous series in reversed phase liquid chromatography. Anal. Chem., 1988, 60, 1443–1448. 63. Yamamoto, F.M., Rokushika, S., and Hatano, H. Comparison of thermodynamic retention behaviors on various C18 columns different in their hydrophobicity. J. Chromatogr. Sci., 1989, 27, 704–709. 64. Sander, L.C. and Field, L.R. Effect of eluent composition on thermodynamic properties in high-performance liquid chromatography. Anal. Chem., 1980, 52, 2009–2013. 65. Cole, L.A. and Dorsey, J.G. Temperature dependence of retention in reversed-phase liquid chromatography. 2. Stationary-phase considerations. Anal. Chem., 1992, 64, 1317–1323. 66. Sentell, K.B., Ryan, N.I., and Henderson, A.N. Temperature and solvation effects on homologous series selectivity in reversed phase liquid chromatography. Anal. Chim. Acta, 1995, 307, 203–215. 67. Chmielowiec, J. and Sawatzky, H. Entropy dominated high performance liquid chromatographic separations of polynuclear aromatic hydrocarbons. Temperature as a separation parameter. J. Chromatogr. Sci., 1979, 17, 245–252. 68. Albert, K. Correlation between chromatographic and physiochemical properties of stationary phases in HPLC: C30 bonded reversed-phase silica. Trends Anal. Chem., 1998, 17, 648–658. 69. Bell, C.M., Sander, L.C., and Wise, S.A. Temperature dependence of carotenoids on C18, C30 and C34 bonded stationary phases. J. Chromatogr. A, 1997, 757, 29–39.
Temperature Effects in Liquid Chromatography
71
70. Pursch, M., Strohschein, S., Händel, H., and Albert, K. Temperature-dependent behaviour of C30 interphases. A solid-state NMR and LC-NMR study. Anal. Chem., 1996, 68, 386–393. 71. Srinivasan, G., Pursch, M., Sander, L.C., and Müller, K. FTIR studies of C30 selfassembled monolayers on silica, titania and zirconia. Langmuir, 2004, 20, 1746–1752. 72. Yang, Y. Stationary phases for high-temperature liquid chromatography. LC-GC Europe, 2003, 16, 37–41. 73. Bullock, J. High-temperature reversed-phase high-performance liquid chromatographic analysis of a synthetic copolymer on a non-porous support. J. Chromatogr. A, 1995, 694, 415–423. 74. Claessens, H.A. and van Straten, M.A. Review on the chemical and thermal stability of stationary phases for reversed-phase liquid chromatography. J. Chromatogr. A, 2004, 1060, 23–41. 75. He, P. and Yang, Y. Studies on the long-term thermal stability of stationary phases in subcritical water chromatography. J. Chromatogr. A, 2003, 989, 55–63. 76. Bolliet, D. and Poole, C.F. Influence of temperature on retention and selectivity in reversed phase liquid chromatography. Analyst, 1998, 123, 295–299. 77. Renn, C.N. and Synovec, R.E. Thermal gradient microbore liquid chromatography with dual-wavelength absorbance detection. Anal. Chem., 1991, 63, 568–574. 78. Li, J.W. and Carr, P.W. Effect of temperature on the thermodynamic properties, kinetic performance, and stability of polybutadiene-coated zirconia. Anal. Chem., 1997, 69, 837–843. 79. Dunlap, C.J., Carr, P.W., McNeff, C.V., and Stoll, D. Zirconia stationary phases for extreme separations. Anal. Chem., 2001, 73, 598A–607A. 80. Li, J.W. and Carr, P.W. Evaluation of temperature effects on selectivity in RPLC separations using polybutadiene-coated zirconia. Anal. Chem., 1997, 69, 2202–2206. 81. Wolcott, R.G., Dolan, J.W., Snyder, L.R., Bakalyar, S.R., Arnold, M.A., and Nichols, J.A. Control of column temperature in reversed-phase liquid chromatography. J. Chromatogr. A, 2000, 869, 211–230. 82. Spearman, L., Smith, R.M., and Dube, S. Monitoring effective column temperature by using shape selectivity and hydrophobicity and the effects of mobile phase temperature. J. Chromatogr. A, 2004, 1060, 147–151. 83. Poppe, H. and Kraak, J.C. Influence of thermal conditions on the efficiency of highperformance liquid chromatographic columns. J. Chromatogr., 1983, 282, 399–412. 84. Thompson, J.D., Brown, J.S., and Carr, P.W. Dependence of thermal mismatch broadening on column diameter in high-speed liquid chromatography at elevated temperatures. Anal. Chem., 2001, 73, 3340–3347. 85. Hesse, G. and Engelhardt, H. Temperatureprogrammierung bei der adsorptionschromatographie von lösungen. J. Chromatogr., 1966, 21, 228–238. 86. Djordjevic, N.M., Fowler, P.W.J., and Houdiere, F. High temperature and temperature programming in high-performance liquid chromatography: Instrumental considerations. J. Microcol. Sep., 1999, 11, 403–413. 87. Molander, P., Olsen, R., Lundanes, E., and Greibrokk, T. The impact of inner diameter on chromatographic performance in temperature gradient liquid chromatography. Analyst, 2003, 128, 1341–1345. 88. Biggs, W.R. and Fetzer, J.C. Thermal gradient liquid chromatography: Application to selective element detection by inductively coupled plasma emission spectrometry. Anal. Chem., 1989, 61, 236–240.
72
Advances in Chromatography
89. Kim, W.-S., Lee, J.Y., Lee, S.K., Park, Y.H., and Lee, D.W. Investigation of retention behavior for herbicides in micro-HPLC with temperature changes. J. Liq. Chromatogr., 2003, 26, 3357–3369. 90. Hu, W., Hasebe, K., and Haddad, P.R. Temperature effects on retention in reversedphase liquid chromatography of nucleosides and their bases using water as the mobile phase. Anal. Commun., 1997, 34, 322–324. 91. Moore, L.K. and Synovec, R.E. Axial thermal gradient microbore liquid chromatography by flow programming. Anal. Chem., 1993, 65, 2663–2670. 92. Sander, L.C., Sharpless, K.E., and Pursch, M. C30 stationary phases for the analysis of food by liquid chromatography. J. Chromatogr. A, 2000, 880, 189–202. 93. Hancock, W.S., Chloupek, R.C., Kirkland, J.J., and Snyder, L.R. Temperature as a variable in reversed-phase high-performance liquid chromatographic separations of peptide and protein samples. I. Optimizing the separation of a growth hormone tryptic digest. J. Chromatogr. A, 1994, 686, 31–43. 94. Chloupek, R.C., Hancock, J.R., Marchylo, B.A., Kirkland, J.J., Boyes, B.E., and Snyder, L.R. Temperature as a variable in reversed-phase high-performance liquid chromatographic separations of peptide and protein samples. II. Selectivity effects observed in the separation of several peptide and protein mixtures. J. Chromatogr. A, 1994, 686, 45–59. 95. Mant, C.T., Chen, Y., and Hodges, R.S. Temperature profiling of polypeptides in reversed-phase liquid chromatography. I. Monitoring of dimerization and unfolding of amphipathic γ-helical peptides. J. Chromatogr. A, 2003, 1009, 29–43. 96. Mant, C.T., Tripet, B., and Hodges, R.S. Temperature profiling of polypeptides in reversed-phase liquid chromatography. II. Monitoring of folding and stability of twostranded γ-helical coilded-coils. J. Chromatogr., 2003, 1009, 45–59. 97. Chen, Y., Mehok, A.R., Mant, C.T., and Hodges, R.S. Optimum concentration of trifluoroacetic acid for reversed-phase liquid chromatography of peptides revisited. J. Chromatogr. A, 2004, 1043, 9–18. 98. Hayakawa, K., Hirano, M., Yoshikawa, K., Katsumata, N., and Tanaka, N. Separation of phenylthiohydantoin-amino acids by temperature-controlled reversed-phase highperformance liquid chromatography. J. Chromatogr. A, 1999, 846, 73–82. 99. Dolfinger, R. and Locke, D.C. Dependence of selectivity on eluent composition and temperature in the HPLC separation of taxanes using fluorinated and hydrocarbon phases. Anal. Chem., 2003, 75, 1355–1364. 100. Issaq, H.J., Xu, H., Chan, K.C., and Dean, M.C. Effect of temperature on the separation of DNA fragments by high-performance liquid chromatography and capillary electrophoresis: A comparative study. J. Chromatogr. B, 2000, 738, 243–248. 101. Yang, Y., Belghazi, M., Lagadec, A., Miller, D.J., and Hawthorne, S.B. Elution of organic solutes from different polarity sorbents using subcritical water. J. Chromatogr. A, 1998, 810, 149–159. 102. Fields, S.M., Ye, C.Q., Zhang, D.D., Branch, B.R., Zhang, X.J., and Okafo, N. Superheated water as eluent in high-temperature high-performance liquid chromatographic separations of steroids on a polymer-coated zirconia column. J. Chromatogr. A, 2001, 913, 197–204. 103. Kondo, T., Yang, Y., and Lamm, L. Separation of polar and non-polar analytes using dimethyl sulfoxide-modified subcritical water. Anal. Chim. Acta, 2002, 460, 185–191. 104. Jones, A. and Yang, Y. Separation of nonpolar analytes using methanol-water mixtures at elevated temperatures. Anal. Chim. Acta, 2003, 485, 51–55.
Temperature Effects in Liquid Chromatography
73
105. Zarzycki, P.K., Wierzbowska, M., and Lamparczyk, H. The influence of temperature on the high performance liquid chromatographic separation of steroids using mobile phases modified by β-cyclodextrin. J. Pharm. Biomed. Anal., 1996, 14, 1305–1311. 106. Karlsson, A. and Charron, C. Reversed-phase ion-pair chromatography at a column temperature below 0˚C using three generations of Hypercarb as solid-phase. J. Chromatogr. A, 1996, 732, 245–253. 107. Tanaka, H., Zhou, X., and Masayoshi, O. Characterization of a novel diol column for high-performance liquid chromatography. J. Chromatogr. A, 2003, 987, 119–125. 108. Haidekker, A. and Huber, C.G. Ion chromatography on chelating stationary phases: Separation of alkali metals. J. Chromatogr. A, 2001, 921, 217–226. 109. Hosoya, K., Kimata, K., Araki, T., Tanaka, N., and Frechet, J.M.J. Temperaturecontrolled high-performance liquid chromatography using a uniformly sized temperature-responsive polymer-based packing material. Anal. Chem., 1995, 67, 1907–1911. 110. Kanazawa, H., Kashiwase, Y., Yamamoto, K., Matsushima, Y., Kikuchi, A., Sakurai, Y., and Okano, T. Temperature-responsive liquid chromatography. 2. Effects of hydrophobic groups in N-isopropylacrylamide copolymer-modified silica. Anal. Chem., 1997, 69, 823–830. 111. Berthod, A., He, B.L., and Beesley, T.E. Temperature and enantioseparation by macrocyclic glycopeptide chiral stationary phases. J. Chromatogr. A, 2004, 1060, 205–214. 112. Kazusaki, M., Shoda, T., Kawabata, H., and Matsukura, H. Influence of mobile phase composition on thermodynamic properties in high-performance liquid chromatography. J. Liq. Chromatogr. R.T., 2001, 24, 141–151. 113. Peyrin, E., Guillaume, Y.C., and Guinchard, C. Interactions between dansyl amino acids and human serum albumin using high-performance liquid chromatography: Mobile-phase pH and temperature considerations. Anal. Chem., 1997, 69, 4979–4984. 114. Takeuchi, T., Watanabe, Y., and Ishii, D. Role of column temperature in micro high performance liquid chromatography. J. High Resolut. Chromatogr., 1981, 4, 300–302. 115. Hirata, Y. and Sumiya, E. Temperature-programmed reversed-phase liquid chromatography with packed fused-silica columns. J. Chromatogr., 1983, 267, 125–131. 116. Bowermaster, J. and McNair, H. Microbore high-performance liquid chromatographic columns: Speed, efficiency, sensitivity and temperature programming. J. Chromatogr., 1983, 279, 431–438. 117. Jinno, K., Phillips, J.B., and Carney, D.P. Temperature-controlled high-speed microcolumn liquid chromatography. Anal. Chem., 1985, 57, 574–576. 118. McNair, H. and Bowermaster, J. HPLC column performance and temperature programming capabilities. J. High Resolut. Chromatogr., 1987, 10, 27–31. 119. Premstaller, A., Xiao, W., Oberacher, H., O’Keefe, M., Stern, D., Willis, T., Huber, C.G., and Oefner, P.J. Temperature-modulated array high-performance liquid chromatography. Genome Res., 2001, 11, 1944–1951. 120. Premstaller, A., Oefner, P.J., Oberacher, H., and Huber, C.G. Capillary array highperformance liquid chromatography of nucleic acids and proteins. Anal. Chem., 2002, 74, 4688–4693. 121. Andersen, T., Nguyen, Q.-N.T., Trones, R., and Greibrokk, T. Studies on the longterm thermal stability of particulate and monolithic poly(styrene-divinylbenzene) capillaries in reversed-phase chromatography. Analyst, 2004, 129, 191–197. 122. Liu, G., Djordjevic, N.M., and Erni, F. High-temperature open-tubular capillary column liquid chromatography. J. Chromatogr., 1992, 592, 239–247.
74
Advances in Chromatography
123. Djordjevic, N.M., Stegehuis, D., Liu, G., and Erni, F. Improved ultraviolet detection in high-temperature open-tubular liquid chromatography. J. Chromatogr., 1993, 629, 135–141. 124. Ryan, K., Djordjevic, N.M., and Erni, F. Temperature programming in open tubular liquid chromatography. J. Liq. Chromatogr. R.T., 1996, 19, 2089–2099. 125. Nyholm, L.M., Sjöberg, P.J.R., and Markides, K.E. High-temperature open tubular liquid chromatography coupled to atmospheric pressure chemical ionisation mass spectrometry. J. Chromatogr. A, 1996, 755, 153–164. 126. Nyholm, L.M. and Markides, K.E. Column preparation for reversed-phase hightemperature open tubular column liquid chromatography. J. Chromatogr. A, 1998, 813, 11–20. 127. Premstaller, A., Oberacher, H., and Huber, C.G. High-performance liquid chromatography-electrospray ionization mass spectrometry of single- and double-stranded nucleic acids using monolithic capillary columns. Anal. Chem., 2000, 72, 4386–4393. 128. Huber, C.G., Premstaller, A., Xiao, W., Oberacher, H., Bonn, G.K., and Oefner, P.J. Mutation detection by capillary denaturing high-performance liquid chromatography using monolithic columns. J. Biochem. Biophys. Methods, 2001, 47, 5–19. 129. Oberacher, H., Oefner, P.J., Hölzl, G., Premstaller, A., Davis, K., and Huber, C.G. Re-sequencing of multiple single nucleotide polymorphisms by liquid chromatography-electrospray ionization mass spectrometry. Nucleic Acids Res., 2002, 30, e67/1–e67/6. 130. Premstaller, A., Oberacher, H., Rickert, A., Huber, C.G., and Oefner, P.J. Multiplex analysis of single-nucleotide extension products on a 16-capillary, denaturing, highperformance liquid chromatography array. Genomics, 2002, 79, 793–798. 131. Oberacher, H., Premstaller, A., and Huber, C.G. Characterization of some physical and chromatographic properties of monolithic poly(styrene-co-divinylbenzene) columns. J. Chromatogr. A, 2004, 1030, 201–208. 132. Dulay, M.T., Quirino, J.P., Bennett, B.D., Kato, M., and Zare, R.N. Photopolymerized sol-gel monoliths for capillary electrochromatography Anal. Chem., 2001, 73, 3921–3926. 133. Vissers, J.P.C., Claessens, H.A., and Cramers, C.A. Microcolumn liquid chromatography: Instrumentation, detection and applications. J. Chromatogr. A, 1997, 779, 1–28. 134. Molander, P., Haugland, K., Hegna, D.R., Ommundsen, E., Lundanes, E., and Greibrokk, T. Determination of low levels of an antioxidant in polyolefins by large volume injection temperature-programmed packed capillary liquid chromatography. J. Chromatogr. A, 1999, 864, 103–109. 135. Molander, P., Thomassen, A., Lundanes, E., Fladseth, G., Thorud, S., Thomassen, Y., and Greibrokk, T. Determination of 1-(2-methoxyphenyl)-piperazine derivatives of airborne diisocyanates by packed capillary liquid chromatography with pre-column large-volume enrichment. J. Sep. Sci., 2001, 24, 947–954. 136. Molander, P., Holm, A., Lundanes, E., Hegna, D.R., Ommundsen, E., and Greibrokk, T. Temperature-promoted large-volume solute enrichment in column-switching miniaturized liquid chromatography: Determination of an antioxidant. Analyst, 2002, 127, 892–897. 137. Holm, A., Molander, P., Lundanes, E., and Greibrokk, T. Determination of rotenone in river water utilizing packed capillary column switching liquid chromatography with UV and time-of-flight mass spectrometric detection. J. Chromatogr. A, 2003, 983, 43–50.
Temperature Effects in Liquid Chromatography
75
138. Holm, A., Molander, P., Lundanes, E., and Greibrokk, T. Novel column oven concept for cold spot large volume enrichment in high throughput temperature gradient capillary liquid chromatography. J. Sep. Sci., 2003, 26, 1147–1153. 139. Andersen, T., Nguyen, Q.-N.T., Trones, R., and Greibrokk, T. Mesoporous polybutadiene-modified zirconia for high-temperature packed capillary liquid chromatography: Column preparation and temperature programming stability. J. Chromatogr. A, 2003, 1018, 7–18. 140. Xiang, Y., Yan, B., Yue, B., McNeff, C.V., Carr, P.W., and Lee, M.L. Elevatedtemperature ultrahigh-pressure liquid chromatography using very small polybutadiene-coated nonporous zirconia particles. J. Chromatogr. A, 2003, 983, 83–89. 141. Holm, A., Solbu, K., Molander, P., Lundanes, E., and Greibrokk, T. Sensitive biomonitoring of phthalate metabolites in human urine using packed capillary column switching liquid chromatography coupled to electrospray ionization ion-trap mass spectrometry. Anal. Bioanal. Chem., 2004, 378, 1762–1768. 142. Törnkvist, A., Sjöberg, P.J.R., Markides, K.E., and Bergquist, J. Analysis of catecholamines and related substances using porous graphitic carbon as separation media in liquid chromatography-tandem mass spectrometry. J. Chromatogr. A, 2004, 801, 323–329. 143. Törnkvist, A., Nilsson, S., Amirkhani, A., Nyholm, L.M., and Nyholm, L. Interference of the electrospray voltage on chromatographic separations using porous graphitic carbon columns. J. Mass Spectrom., 2004, 39, 216–222. 144. Koivisto, P. and Stefansson, M. Retention studies of sulphated glycosaminoglycan disaccharides on porous graphitic carbon capillary columns. Chromatographia, 2003, 57, 37–45. 145. Chen, J., Steenackers, D., and Sandra, P. Micro LC on cellulose triacetate (CTA): Influence of particle diameter and column temperature. J. Microcol. Sep., 1993, 7, 259–263. 146. Andersen, T., Holm, A., Skuland, I.L., Trones, R., and Greibrokk, T. Characterization of complex mixtures of polyglycerol fatty acid esters using temperature and solvent gradients in packed capillary LC. J. Sep. Sci., 2003, 26, 1133–1140. 147. Trones, R., Andersen, T., Greibrokk, T., and Hegna, D.R. Hindered amine stabilizers investigated by the use of packed capillary temperature-programmed liquid chromatography I. Poly((6-((1,1,3,3-tetramethylbutyl)-amino)-1,3,5-triazine-2,4-diyl)(2,2,6, 6-tetramethyl-4-piperidyl)imino)-1,6-hexanediyl((2,2,6,6-tetramethyl-4-piperidyl)imino). J. Chromatogr. A, 2000, 874, 65–71. 148. Trones, R., Andersen, T., Hegna, D.R., and Greibrokk, T. Hindered amine stabilizers investigated by the use of packed capillary temperature programmed liquid chromatography II. Poly-(N-β-hydroxyethyl-2,2,6,6-tetramethyl-4-hydroxypiperidyl succinate). J. Chromatogr. A, 2000, 902, 421–426. 149. Bruheim, I., Molander, P., Theodorsen, M., Ommundsen, E., Lundanes, E., and Greibrokk, T. Temperature-programmed packed-capillary liquid chromatographic separation of polystyrenes. Chromatographia, 2001, 53, S266–S270. 150. Trones, R., Andersen, T., and Greibrokk, T. Improved modification of a laser lightscattering detector for use in packed capillary high temperature liquid chromatography. J. High Resolut. Chromatogr., 1999, 22, 283–286. 151. Nordstrom, O., Molander, P., Greibrokk, T., Blomhoff, R., and Lundanes, E. Evaluation of temperature programming for gradient elution in packed capillary liquid chromatography coupled to electrochemical detection. J. Microcol. Sep., 2001, 13, 179–185.
76
Advances in Chromatography
152. Yoo, J.S., Watson, J.T., and McGuffin, V.L. Temperature-programmed microcolumn liquid chromatography/mass spectrometry. J. Microcol. Sep., 1992, 4, 349–362. 153. Andersen, T., Skuland, I.L., Holm, A., Trones, R., and Greibrokk, T. Temperatureprogrammed packed capillary liquid chromatography coupled to evaporative lightscattering detection and electrospray ionization time-of-flight mass spectrometry for characterization of high-molecular-mass hindered amine light stabilizers. J. Chromatogr. A, 2004, 1029, 49–56. 154. Sandvoss, M., Roberts, A.D., Ismail, I.M., and North, S.E. Direct on-line hyphenation of capillary liquid chromatography to nuclear magnetic resonance spectroscopy: Practical aspects and application to drug metabolite identification. J. Chromatogr. A, 2004, 1028, 259–266. 155. Skuland, I.L., Andersen, T., Trones, R., Eriksen, R.B., and Greibrokk, T. Determination of polyethylene glycol in low-density polyethylene by large volume injection temperature gradient packed capillary liquid chromatography. J. Chromatogr. A, 2003, 1011, 31–36. 156. Rosales-Conrado, N., Leon-Gonzalez, M.E., Perez-Arribas, L.V., and Polo-Diez, L.M. Determination of chlorophenoxy acid herbicides and their esters in soil by capillary high performance liquid chromatography with ultraviolet detection, using large volume injection and temperature gradient. Anal. Chim. Acta, 2002, 470, 147–154. 157. Molander, P., Thomassen, A., Kristoffersen, L., Greibrokk, T., and Lundanes, E. Simultaneous determination of citalopram, fluoxetine, paroxetine and their metabolites in plasma by temperature-programmed packed capillary liquid chromatography with on-column focusing of large injection volumes. J. Chromatogr. B, 2001, 766, 77–87. 158. Leon-Gonzalez, M.E., Perez-Arribas, L.V., Polo-Diez, L.M., Panis, C., and San Andres, M.P. Determination of triazine herbicides by capillary liquid chromatography with on-column focusing and temperature gradient. Anal. Chim. Acta, 2001, 445, 29–34. 159. Molander, P., Thommesen, S.J., Bruheim, I.A., Gundersen, T.E., Trones, R., Greibrokk, T., and Lundanes, E. Temperature-programmed packed capillary liquid chromatography separation with large volume on-column focusing injection of retinyl esters. J. High Resolut. Chromatogr., 1999, 22, 490–494. 160. Oefner, P.J. and Huber, C.G. A decade of high-resolution liquid chromatography of nucleic acids on styrene-divinylbenzene copolymers. J. Chromatogr. B, 2002, 782, 27–55. 161. Walcher, W., Oberacher, H., Troiani, S., Hölzl, G., Oefner, P.J., Zolla, L., and Huber, C.G. Monolithic capillary columns for liquid chromatography — Electrospray ionization mass spectrometry in proteomic and genomic research. J. Chromatogr. B, 2002, 782, 111–125. 162. Walcher, W., Toll, H., Ingendoh, A., and Huber, C.G. Operational variables in highperformance liquid chromatography-electrospray ionization mass spectrometry of peptides and proteins using poly(styrene-divinylbenzene) monoliths. J. Chromatogr. A, 2004, 1053, 107–117. 163. Andersen, T., Molander, P., Trones, R., Hegna, D.R., and Greibrokk, T. Separation of polyethylene glycol oligomers using inverse temperature programming in packed capillary liquid chromatography. J. Chromatogr. A, 2001, 918, 221–226. 164. Molander, P., Theodorsen, M., Lundanes, E., Soerenssen, D.M., and Greibrokk, T. Temperature effects on packed-capillary liquid chromatography of the x-ray contrast agent Iodixanol. J. Chromatogr. Sci., 2000, 38, 157–161.
Temperature Effects in Liquid Chromatography
77
165. Molander, P., Holm, A., Lundanes, E., Greibrokk, T., and Ommundsen, E. Separation of ceramides by sub-ambient temperature-assisted large volume injection in temperature-programmed packed capillary liquid chromatography. J. High. Resolut. Chromatogr., 2000, 23, 653–655. 166. Bruheim, I., Molander, P., Lundanes, E., Greibrokk, T., and Ommundsen, E. Temperature-programmed packed capillary liquid chromatography coupled to Fouriertransform infrared spectroscopy. J. High Resolut. Chromatogr., 2000, 23, 525–530. 167. Molander, P., Ommundsen, E., and Greibrokk, T. Separation of technical waxes by temperature-programmed packed-capillary liquid chromatography. Chromatographia, 2000, 51, 349–352. 168. Molander, P., Ommundsen, E., and Greibrokk, T. High temperature injection in capillary high temperature liquid chromatography. J. Microcol. Sep., 1999, 11, 612–619. 169. Molander, P., Trones, R., Haugland, K., and Greibrokk, T. Aspects and applications of non-aqueous high temperature packed capillary liquid chromatography. Analyst, 1999, 124, 1137–1141. 170. Molander, P., Gundersen, T.E., Haas, C., Greibrokk, T., Blomhoff, R., and Lundanes, E. Determination of retinoids by packed-capillary liquid chromatography with largevolume on-column focusing and temperature optimization. J. Chromatogr. A, 1999, 847, 59–68. 171. Trones, R., Tangen, A., Lund, W., and Greibrokk, T. Packed capillary high-temperature liquid chromatography coupled to inductively coupled plasma mass spectrometry. J. Chromatogr. A, 1999, 835, 105–112. 172. Trones, R., Andersen, T., Hunnes, I., and Greibrokk, T. Modified laser light-scattering detector for use in high temperature micro liquid chromatography. J. Chromatogr. A, 1998, 814, 55–61. 173. Sheng, G., Shen, Y., and Lee, M.L. Elevated temperature liquid chromatograpy using reversed-phase packed capillary columns. J. Microcol. Sep., 1997, 9, 63–72. 174. Trones, R., Iveland, A., and Greibrokk, T. High-temperature liquid-chromatography on packed capillary columns with nonaqueous mobile phases. J. Microcol. Sep., 1995, 7, 505–512. 175. Hirata, Y. and Jinno, K. High resolution liquid chromatography with a packed micro glass capillary column. J. High Resolut. Chromatogr., 1983, 6, 196–199.
3
Lipophilicity Measurements by Liquid Chromatography Simion Gocan Babes-Bolyai University, Cluj-Napoca, Romania
Gabriela Cimpan Sirius Analytical Instruments Ltd., Forest Row, U.K.
John Comer Sirius Analytical Instruments Ltd., Forest Row, U.K.
CONTENTS 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9
Introduction ....................................................................................................80 Lipophilicity: Concept and Definitions .........................................................81 The Uses of Lipophilicity ..............................................................................81 logP, logD, and Other Terms for Expressing Lipophilicity ..........................82 The Effect of Chemical Reactions on Partitioning .......................................83 Octanol and Other Partition Solvents ............................................................87 Calculations Involving LOGP ..........................................................................90 logP Measurement by Shake-Flask ...............................................................91 Lipophilicity Measurement by High-Performance Liquid Chromatography (HPLC)...............................................................................94 3.9.1 HPLC with Octanol-Coated Columns ...............................................96 3.9.2 Other Methods using Octanol-Coated Columns .............................100 3.9.3 ELogP Method .................................................................................100 3.9.4 Reversed-Phase HPLC (RP-HPLC).................................................102 3.9.5 The extrapolation method, logkw .....................................................102 3.9.6 Isocratic logk Values ........................................................................105 3.9.7 The Significance of Slope S ............................................................106 3.9.8 Correction for Ionization .................................................................107 3.9.9 The Influence of Stationary Phase on Lipophilicity Measurements...................................................................................108 3.9.10 Mobile Phase Composition and Lipophilicity.................................112 79
80
Advances in Chromatography
3.9.10.1 Organic Modifier...............................................................112 3.9.10.2 Buffer in the Mobile Phase ..............................................114 3.9.10.3 Ionic Strength.................................................................... 114 3.9.10.4 Masking Agents ................................................................115 3.9.10.5 Association in Water.........................................................116 3.9.10.6 Association in the Organic Phase.....................................116 3.9.10.7 The Effect of Temperature ...............................................116 3.9.10.8 Gradient RP-HPLC...........................................................117 3.9.11 Chromatographic Hydrophobicity Index (CHI) ..............................117 3.9.12 Abraham-Type Equations for RP-HPLC .........................................125 3.10 Lipophilicity Estimation by Thin-Layer Chromatography .........................127 3.10.1 Relationship between RM and Molar Fraction X of the Organic Modifier in the Mobile Phase ............................................128 3.10.2 Influence of pH ................................................................................131 3.10.3 Influence of Mobile Phase Composition on RM0 Value..........................................................................................132 3.10.4 Influence of Stationary Phase on RM0 Value ...................................133 3.10.5 Specific Hydrophobic Surface Area of the Solute ..........................134 3.10.6 Relationship between Intercepts (RM0) and Slopes (a1) ..................135 3.10.7 RM Determination and Correlation with Partition Coefficients, logP .............................................................................143 3.11 Chromatographic Determination of Hydrophilic–Lipophilic Balance........145 3.11.1 Drug Lipophilicity............................................................................146 3.11.2 Miscellaneous...................................................................................148 3.11.3 Studies using Both RP-HPLC and RP-HPTLC ..............................149 3.12 Other Chromatographic Methods ................................................................154 3.12.1 Micellar Liquid Chromatography (MLC)........................................154 3.12.2 Reversed-Phase Ion Pair Chromatography (RP-IPC)......................155 3.12.3 Countercurrent Chromatography (CCC) .........................................156 Acknowledgments..................................................................................................157 Glossary .................................................................................................................157 References..............................................................................................................158
3.1 INTRODUCTION Many methods have been studied in an attempt to find a suitable model for the difficult, classical partition experiment in a shake-flask vessel. Liquid chromatography provided a particular approach due to the possibility of using a liquid mobile phase and a liquid or liquid-like stationary phase. This chapter will review different liquid chromatographic methods for measuring lipophilicity, including liquid-liquid chromatography, reversed-phase liquid chromatography (isocratic, gradient, and micellar methods) and counter current chromatography. This chapter will not review liquid chromatography with stationary phases that mimic the biological environment such as affinity chromatography or immobilized artificial membrane chromatography, as these form a distinct group of methods and
Lipophilicity Measurements by Liquid Chromatography
81
comprehensive reviews have been published recently [1–5]. Neither will it review micellar electrokinetic chromatography (MECK) or microemulsion electrokinetic chromatography (MEECK), as these techniques are based on capillary electrophoresis and details can also be found in a recent review [6].
3.2 LIPOPHILICITY: CONCEPT AND DEFINITIONS Lipophilicity is a general concept that characterizes the affinity of molecules for lipid-like structures. A related term, hydrophobicity, refers to water’s “dislike” of a molecule. The two terms are often used interchangeably in the literature but it has been agreed that hydrophobicity relates to the forces between the nonpolar groups in an aqueous environment. Water’s internal molecular organization excludes nonpolar moieties; therefore, hydrophobicity is only one component of the more general term lipophilicity [7]. For many years, confusion existed between lipophilicity and hydrophobicity and both terms were used to characterize molecules that show a preference for a nonpolar phase. The definitions agreed by IUPAC for these terms are [8]: lipophilic, adj., -ity, n.: Having an affinity for fat and high lipid solubility: a physicochemical property which describes a partitioning equilibrium of solute molecules between water and an immiscible organic solvent, favouring the latter, and which correlates with bioaccumulation. hydrophobic, adj., -ity n.: Describing the character of a molecule or atomic group which is insoluble in water, or resistant to wetting or hydration. Hydrophobic interaction: the tendency of hydrocarbons (or of lipophilic hydrocarbon-like groups in solutes) to form intermolecular aggregates in an aqueous medium, and analogous intramolecular interactions. The name arises from the attribution of the phenomenon to the apparent repulsion between water and hydrocarbons. However, the phenomenon ought to be attributed to the effect of the hydrocarbon-like groups on the water-water interaction. Hydrophobicity contributes towards lipophilicity but the two terms are not synonymous. Hydrophobic interaction is a different process and can be found as a driving force in lipophilicity studies.
3.3 THE USES OF LIPOPHILICITY Since the pioneering work of Overton and Meyer at the beginning of the 20th century [9, 10], when compound partition in olive oil was found to correlate well with biological activity, much work has been done to standardize experimental conditions and to understand molecular influences on lipophilicity. Applications of lipophilicity can now be found in the pharmaceutical industry, in environmental applications, toxicology, and any other field involving a potential biological effect on a living
82
Advances in Chromatography
organism. It has been widely used in quantitative structure-activity relationships (QSAR) and in drug and pesticide design as well as in toxicology studies either in the pharmaceutical or environmental sciences. Lipophilicity has been used in the following areas of study: •
• • • •
• • • •
Estimation of solubility — lipophilicity is used (in conjunction with other parameters) to estimate the aqueous solubility by a range of semi-empirical equations [11] Environmental applications [12] General [13] Pesticide occurrence in the environment [14] Drug discovery in the pharmaceutical industry. When creating a new molecule with potential biological activity, knowledge of its lipophilicity helps to build an understanding of its absorption, distribution, metabolism, excretion, and toxicity (ADMET) [15] Oral absorption [16] Serum albumin binding [17] Permeability [18] CNS penetration [19]
According to EU regulations, lipophilicity (expressed as logPo/w) is a required value for any new commercially available chemical and a liquid chromatographic method for this measurement has been approved [20, 21].
3.4 logP, logD, AND OTHER TERMS FOR EXPRESSING LIPOPHILICITY Lipophilicity is generally expressed by the partition between water and a waterimmiscible solvent. Traditionally, the partition coefficient was defined as the ratio between the equilibrium concentration of a compound (C) in a water-immiscible solvent (s) and its concentration in water (w) [22, 23]. In 1891 Walther H. Nernst was appointed as Extraordinary Professor at Göttingen University. During the same year, he developed the distribution law: If two liquids (or solids) a and b are partially immiscible and if there is a third component i present in both phases that behaves individually as an ideal solute (if it is sufficiently dilute), the ratio of its concentrations (x) is independent of the individual values of x. xia = constant xib
(3.1)
Nernst’s law of distribution is described today by the equation for the partition coefficient of a solute between two immiscible phases:
Lipophilicity Measurements by Liquid Chromatography
P=
[C ]s [C ]w
83
(3.2)
The term C refers to the neutral form of the molecule and the square brackets that enclose it denote that its concentration is at equilibrium. According to IUPAC recommendations from 1993, the partition coefficient Po/w or Ko/w is a “measure of lipophilicity by determination of the equilibrium distribution between octan-1-ol and water, as used in pharmacological studies and in the assessment of environmental fate and transport of organic chemicals.” From the same source, the partition coefficient is the ratio of the distribution of a substance between two phases when the heterogeneous system (of two phases) is in equilibrium; the ratio of concentrations (or, strictly speaking, activities) of the same molecular species in the two phases is constant at constant temperature. The partition coefficients most frequently used in acute toxicology are lipid/water and octan-1-ol/water distributions. From now on in this review the term “octanol” will refer to octan-1-ol unless otherwise stated [8]. Measurable P values are usually in the range 10–2 to 107. As these values extend over several orders of magnitude, they may by expressed logarithmically (as log10) and logP is often used instead of P for practical reasons. If the molecule is ionisable, the pH of the aqueous phase will influence the concentrations of the ionised and neutral forms of the molecule.
D=
∑ [C ] ∑ [C ]
s
(3.3)
w
D is the distribution coefficient of a compound at a defined pH and will take into account both neutral and ionized species present in the aqueous and in the organic phase. Partition coefficients have been widely used to define the lipophilic character of compounds [24] and logP and logD have been extensively included in quantitative structure-activity relationships (QSAR) [25].
3.5 THE EFFECT OF CHEMICAL REACTIONS ON PARTITIONING The dissociation of ionizable compounds in the aqueous phase is the most common reaction that accompanies the partition in an octanol-water system. Two simple cases, a monoprotic acid and monoprotic base, will illustrate how dissociation influences the partition. The ionization of a monoprotic acid in solution may be written as, HA ⇔ H+ + A–
(3.4)
84
Advances in Chromatography
Organic solvent HA
Water
HA
H+ + A−
FIGURE 3.1 Solvent–water system and the species partition for a weak acid.
The neutral species, HA, will be able to partition in the lipophilic solvent, while the ionized species will stay in water. Therefore, the organic solvent will contain only the HA form, while the aqueous phase will contain both HA and A– forms. In this case, the distribution coefficient can be expressed as follows:
D=
[ HA ]0 [ HA ]w + ⎡⎣ A – ⎤⎦w
(3.5)
where the indices “o” and “w” stand for organic solvent and water, respectively, and the square brackets symbolize that the concentrations of species at equilibrium is taken into consideration. The ionization process in the aqueous phase is governed by the ionization constant,
Ka =
⎡⎣ A – ⎤⎦ ⎡⎣ H + ⎤⎦ [ HA ]w w w ⇒ ⎡⎣ A – ⎤⎦ = K a w ⎡⎣ H + ⎦⎤ [ HA ]w w
(3.6)
The distribution coefficient will have the expression: P ⎡⎣ H + ⎤⎦ P ⎡⎣ H + ⎤⎦ P = = D= ≈ Ka Ka ⎛ ⎡⎣ H + ⎤⎦ + K a Ka ⎞ 1 + [ HA ]w ⎜⎜ 1 + ⎡ + ⎤ ⎟⎟ ⎡⎣ H + ⎤⎦ ⎝ ⎣H ⎦ ⎠
[ HA ]0
(3.7)
The logarithm of this approximation is, logD = logP + log[H+] – logKa
(3.8)
logD = logP + pKa – pH
(3.9)
which may be written as:
S
O
NH3+
N
H
XH2+
H H
O
XH0
H
XH
N CH3
±
COO−
NH3+
N
CH3
HO
O
O−
O
H H NH
S
CH3
O
H
X−
NH2
N
N CH3
COO−
−
CH3 X
0
25
50
75
100
0
25
50
75
100
1
1
2
2
3
XH2+
3
XH2+
4
4 6 7 8
9
6
7
8
9 pH (concentration scale)
5
XH0
Distribution of species
pH (concentration scale)
5
XH±
Distribution of species
FIGURE 3.2 The distribution of species for Ampicillin (a zwitterion) and for morphine (an ampholyte).
HO
O
HO
S
CH3
NH
O
CH3
NH+
Morphine (an ampholyte) pKas = 9.26, 8.17
HO
O
HO
Ampicillin (a zwitterion) pKas = 7.14, 2.55
O
NH
H H
COOH
CH3 XH2+
% Species % Species
CH3
10
10
11
11
12
X−
12
−
X
13
13
Lipophilicity Measurements by Liquid Chromatography 85
86
Advances in Chromatography
A similar equation can be derived for monoprotic bases, whose ionization may be written as, B + H+ ⇔ BH+
(3.10)
Considering that the only reaction is the base ionization in water, the organic solvent will contain only the neutral base, B, while the aqueous phase will contain both neutral and ionized base, B and BH+, and the following approximation may be derived: logD = logP – pKa + pH
(3.11)
In many applications concerned with biological activity of different organic compounds (e.g., drug discovery, toxicology studies), it is considered that the distribution coefficient at pH = 7.4, the pH of the blood, is sufficient to give an estimate for the in vivo human absorption of compounds. However, the distribution of species can change dramatically with the pH, as in the gastrointestinal tract where pH can vary from 1 to 8 depending on the fed–fasted state and the part of the tract considered (Figure 3.3 and Figure 3.4). The relationship between logP and logD has been derived for amphoteric compounds as a function of the two dissociation constants, pKa1 and pKa2 [26]: logP = logD + log (1 + 10pH – pKa1 + 10pKa2 – pH)
(3.12)
Figure 3.5 shows the lipophilicity profiles for partitioning of diprotic molecules. The curves were drawn using the “What-if” feature of Sirius’ RefinementPro software for simulated samples with aqueous pKa values of 8 and 5. Although the pKa values of real samples may differ, the shapes and relative positions of the curves will follow the examples below for different classes of molecule. A detailed description of lipophilicity-pH profiles for acids, bases, and ampholytes can be found in [27]. 2 1
LogD
0 −1 −2 −3 −4 −5
0
1
2
3
4
5
6 7 pH
8
9 10 11 12 13
FIGURE 3.3 Lipophilicity profile for morphine, pKas = 9.26, 8.17; logP = 0.90.
Lipophilicity Measurements by Liquid Chromatography
5
87
Desipramine pKa = 10.14 Diclofenac pKa = 3.99
4
Diphenhydramine pKa = 8.26
3 Phenobarbital pKa = 7.43
LogD
2
Triamterene pKa = 3.92
Nifuroxime pKa = 10.56
1
0 −1 −2 0
2
4
6
8
10
12
14
pH
FIGURE 3.4 logD = f(pH) for acids (negative slope) and bases (positive slope).
If a compound is suspected to form dimers in one of the phases, this process has to be taken into consideration when defining logD [28]. Applications in the pharmaceutical industry tend to correlate logP/logD values with drug absorption in vivo. In many cases, a good correlation is obtained if octanol is used for the partition measurements and the logD values measured at pH 5 to 7.4 (physiological pH values) provide more relevant information compared with logP [29]. Although the IUPAC recommended notation for partition coefficient is K, the large majority of publications still use logP and logD to differentiate between the partition of the neutral and charged molecule. For this reason, we will keep the same notations throughout this chapter.
3.6 OCTANOL AND OTHER PARTITION SOLVENTS Octanol was not the first partition solvent used as a model for the in vivo process. Olive oil was one of the first investigated lipophilic solvents but differences between batches and difficult standardization led to the use of other solvents such as paraffin oil. Hansch later proposed the use of octanol and the largest database in the world for partition coefficients is based on the octanol–water system [30–32]. Today, octanol–water partition coefficients (logPo/w) are established as a significant physical property correlated with biological activity. An important reason is the ability of octanol to form clusters in an aqueous environment, thus providing a
10
H2Z = H2Z* Z = Z* logPH2Z > logPZ
H2Z = H2Z* Z = Z* logPH2Z = logPZ
Less octanol, low ionic strength
HX = HX*
Less octanol, low ionic strength
B = B*
Less octanol, low ionic strength
H2A = H2A*
H2Z = H2Z* Z = Z* logPZ > logPH2Z
More octanol, high ionic strength
HX = HX* X = X* logPHX > logPX
Special case-very rare
B = B* BH = BH* logPB = logPBH
Special case-very rare
H2A = H2A* HA = HA* logPH2A = logPHA
Most zwitterionic molecules are hydrophilic, logD values are usually negative
More octanol, high ionic strength
H2X = H2X* HX = HX* logPHX > logPH2X
More octanol, high ionic strength
B = B* BH = BH* logPB > logPBH
More octanol, high ionic strength
2 0 H2A = H2A* −2 HA = HA* −4 logP H2A > logPHA −6 pH −8 2 4 6 8
FIGURE 3.5 Lipophilicity profiles for diprotic molecules.
4: Zwitterion One acid, one base group (acid pKa < basic pKa), HZ is the “neutral” species but is zwitterionic: HZ± pKa equations: H + Z = HZ, H + HZ = H2Z
3: Ordinary ampholyte One acid, one base group (acid pKa > basic pKa), HX is the neutral species pKa equations: H + X = HX, H + HX = H2X
2: Base Two base groups, B is the neutral species pKa equations: B + H = BH, BH + H = BH2
1: Acid Two acid groups, H2A is the neutral species pKa equations: H + A = HA, H + HA = H2A
88 Advances in Chromatography
Lipophilicity Measurements by Liquid Chromatography
89
model for the partition into a real biological membrane. Octanol has an amphiprotic character and, as well as displaying lipophilic interaction, it can form hydrogen bonds. It can therefore interact with with neutral or ionized molecules and with ionpairs [33, 34]. The partition of a solute between water and octanol must take into account the interactions between the solute and the water present in the organic phase, as octanol can dissolve a significant amount of water (2.3 to 2.5 M). Experimental evidence exists that amphiprotic and aprotic solutes can either “drag” water molecules or “exclude” them from the octanol phase. It appears that besides octanol–water clusters containing a variable number of octanol molecules around water, there also exist an octanol–water complex having a highly organized structure with a water molecule in the center and four octanol molecules around it bound by hydrogen bonds. Any interaction between the solute and octanol is likely to disturb this arrangement [35, 36]. Octanol and water can both participate in hydrogen bonding because both have hydroxyl groups and, therefore, a considerable amount of water can dissolve in octanol. The distribution of a compound between octanol and water is mainly dependent on the differences between the interactions of the lipophilic part of the molecule with water and with the organic solvent rather than the interaction of the polar and hydrogen bonding groups [37]. The extent to which the octanol model is the right choice to model human, animal, or plant membranes, as well as environmental distribution of toxins, still has to be established. Nevertheless, a very large number of correlations between the logPo/w value and biological activity have been reported during the last 50 years. Although octanol is commonly used in lipophilicity studies, the partition in other solvents is sometimes a better model for the in vivo process. For example, it is known that hydrogen bonding plays little role in transport across the blood–brain barrier, which is better modelled by alkane–water systems. Many alkanes have been used including hexane, cyclohexane, heptane, isooctane, dodecane, and hexadecane. Other investigated systems include chloroform–water (H-donor) and di-n-butylether (H-acceptor) [38]. A correlation between the partition coefficient in heptane/water and absorption from the cerebrospinal fluid was reported in 1959, where it was noted that the oxidative dealkylation of foreign N-alkylamines by rabbit liver microsomes appeared to be limited to compounds that were lipid soluble, as shown by high chloroform to water partition coefficients at physiological pH. Since the microsomal hydroxylation of aromatic compounds also appeared to be limited to lipid soluble substances, an intracellular fat-like boundary was suggested that separates normally occurring polar substances from the highly nonspecific microsomal enzymes [39]. The choice of four solvents has been proposed for measuring partition coefficients, dependent upon the specific application of the data obtained. These are octanol (amphiprotic), alkanes (neutral), chloroform (proton donor), and propylene glycol dipelargonate (PGDP) (proton acceptor) [40]. Octanol-water systems have been used to model serum protein binding as well as the interaction with membranes formed mainly of proteins. Biological membranes can have very different hydrogen bonding properties ranging from neutral — when
90
Advances in Chromatography
the partition in an alkane may be a better model — to structures comprising mainly proton donor groups, modelled by chloroform [41]. They can also comprise mainly proton acceptor groups as in phospholipid membranes, where the partition in propylene glycol dipelargonate may be a better model [41, 42]. According to Buchwald and Bodor, the main reasons for using octanol-water for lipophilicity estimations are: • • • • • •
Large database of experimental data Octanol has low vapor pressure at room temperature and is therefore suitable for different experimental methods Octanol has low UV absorption and therefore does not interfere in the measurement of solute concentration Octanol retains some of the hydrogen-bonding ability of water and is able to dissolve a significant amount of water There are many methods available that reasonably model the partition in the octanol-water system (e.g., RP-HPLC) There are many correlations between the logP value of a compound and its biological activity and therefore it is reasonable to assume that compounds with certain logP values will be absorbed in vivo [43]
Other systems, like oil-water or hexadecane-water, may be better models for the blood–brain barrier than octanol-water. However, the chromatographic applications are limited and the rest of this chapter will describe methods that model the partition in an octanol-water system.
3.7 CALCULATIONS INVOLVING
LOGP
There are a number of methods for calculating logP from chemical structures and for using logP values in other mathematical models and predictions. These have been reviewed [44] and are summarized below: • • • • •
Group contribution models: Hansch [45, 46], Rekker [47, 48], Fujita [31] Atomic contribution or surface area models [49] Molecular methods [50] Linear solvation energy relationships (LSER) [51] Commercially available software for logP calculation: CLogP [52–54], ACD [55], ADME Boxes [56], KowWin, ALogP, IAlogP, XLOGP [57]
The link between the partition in different water-solvent systems can be achieved by applying the Collander equation or by free energy solvation relationships. In 1951, Collander observed the linear relationship between the partition coefficients obtained in different alkanol-water systems, 1 and 2 [58]: logP1 = a logP2 + b
(3.13)
Lipophilicity Measurements by Liquid Chromatography
91
where a and b are coefficients. The Collander equation looked very promising in the beginning because it allowed a transfer from a measured partition coefficient in a defined partitioning system to another one. In practice, the Collander equation has limited application to solutes with similar hydrogen bonding characteristics [34]. Another attempt to generalize the theory of lipophilicity is seen in the linear solvation energy relationship (LSER) defined by Abraham et al. (see RP-HPLC section). LSER equations have been used to compare published partition coefficients in octanol-water (613 compounds), hexadecane-water (370 compounds), alkanewater (200 compounds) and cyclohexane-water (170 compounds) [59]. The main factors governing the partitioning of a compound in the octanol-water system are its dipolarity/polarizability and hydrogen bond basicity, which orient the compound towards water, and its molecular size, which orients the compound towards octanol. In alkane-water systems, polarizability and all forms of hydrogen bonding orient the compound towards water while molecular size orients the compound towards alkane.
3.8 logP MEASUREMENT BY SHAKE-FLASK The shake-flask method has changed very little over the decades and is still the reference for any new method of measuring lipophilicity. The compound is mixed with two immiscible solvents until equilibrium is reached, then the two phases are separated as well as possible and the solute concentration is measured in both of them. If the concentration measurement is difficult in one of the phases, the compound concentration can be measured in the other phase and the concentration in the unmeasured phase obtained by the difference from the total compound quantity. It is essential to mutually presaturate both phases for a successful experiment. Water is usually the lower layer (e.g., when using octanol or cyclohexane) but sometimes can be the upper layer (e.g., when using chloroform). The subscript “o/w” refers to octanol-water partitioning, as in Po/w. In many cases, the equilibrium is established in 5 to 30 minutes but sometimes a longer time is required. Although, in theory, the equilibration time is infinite, the system is considered as reaching equilibrium when two consecutive measurements from both phases show no detectable concentration change. However, this approach requires care as any sampling from one phase will modify the compound concentration and, therefore, the partition process. Also, the shaking of the two immiscible solvents can produce stable emulsions, making the two phases very difficult to separate. Shake-flask experiments can be prone to serious errors and the published logP data should be reviewed carefully, as up to 1 unit span range is indicated for many compounds with reported logPo/w values in the range 1.2 to 5.6 [20]. Although the need to measure the partition coefficients exists in many industrial areas, the pharmaceutical industry has the largest number of samples. The last decade has seen the development of the high-throughput methods for screening compound libraries. Several automated approaches have been developed since the shake-flask approach was decided to be the reference method in industry for logP/logD measurements. An automated shake-flask method for high-throughput logD measurements in an octanol-water system has been described by Valkó [60]. This micro
92
Advances in Chromatography
shake-flask method is performed in capped vials and the dual phase system is sampled by a needle piercing the vial cap. The method was developed for the measurement of logD values at pH = 7.4. A similar system has been described for samples in 96 well plates, where the entire liquid handling is performed by a robotic system. Samples are taken from the octanol and aqueous phase and are consequently analysed by reversed-phase highperformance liquid chromatography (RP-HPLC). The system was validated for a set of chemically diverse compounds as 1 mg/mL stock solutions in octanol in 96 well plates. Liquid handling was performed by a Beckman Biomek 2000 robot (Beckman Instruments, Fullerton, CA, USA). The aqueous phase was 400 μL phosphate buffer (pH = 7.4) and the organic phase was 200 μL octanol. A quantity of 200 μL of the solute stock solutions was transferred into each of the wells. Different methods for creating the dual-phase system were investigated, all of them leading to the same analyte concentration and the same final volumes. The plates were sealed and placed in a reciprocal shaker for 30 minutes, then centrifuged. Aliquots were taken from both phases by sampling at different heights and analysed by isocratic and gradient RP-HPLC on an ODS column. Similar samples were taken from both phases and analysed in the same RP-HPLC conditions from the solution resulting from the other two procedures, the Eppendorf shake-flask and the manual plate method. Details of the methods are shown in Table 3.1. This automated method was fully validated for logD values in the range 2 to 4 against the logD values obtained from the Eppendorf shake-flask method and against the manual plate method [61]. The system described above can measure 24 to 96 samples per day. The methods were validated for equilibration time and reproducibility within the same method. Very good correlations (r2 = 0.996 to 1.000) were obtained between the data produced by the three methods. A commercially available instrument using automated shake-flask method is manufactured by Analiza Inc. [62] Octanol–buffer solutions at different pH values are preprepared and the sample is injected into the dual-phase system. When the equilibrium is established, the vials are centrifuged for the phase separation. Aliquots are sampled from both phases and analysed by UV, nitrogen detector, or LS-MS. This system can measure up to 200 samples per day with a typical logD range from 3 to 4. A micro shake-flask experiment was reported for derivatives of purine and pyrimidine nucleosides. The compounds were stored as 0.5 mg/mL DMSO stock solutions. For experiments, 20 μL aliquots of the stock solutions were dissolved in 1 mL buffer saturated octanol and then the compound was partitioned between the octanol and 1 mL aqueous phosphate buffer phase (pH = 7.0) in a Mixxor apparatus (Genex Corp., Gaithersburg, MD, USA). Instead of shaking, this apparatus uses piston strokes. The mass transfer after 6 piston strokes is equivalent to 40 shakes in a separatory funnel. The phases were separated by centrifugation and aliquots from both phases were analysed by HPLC on an ODS column and acetonitrile-phosphate buffer (pH 7.0) mobile phase [63]. Compiling the data from personal experience and from the numerous published papers, the drawbacks are summarized here for the shake-flask method:
Lipophilicity Measurements by Liquid Chromatography
93
TABLE 3.1 Experimental Settings for Three Methods for Measuring the Distribution Coefficients, logD Eppendorf Method Method
shake-flask
Sampling Samples (stock solutions)
manual 1 mg/mL analyte in octanol (1.8 mL Eppendorf vials) octanol-phosphate buffer pH 7.4 100, 300, 400 μL stock solution mixed with 400, 200 or 100 μL octanol, then 500 μL aqueous phosphate pH 7.4 was added. The procedure was done with the same volumes, but starting with aqueous solutions for the hydrophilic compounds, then adding octanol. 3
Partition system Experimental
Manual Plate Method modified shakeflask in 96 well plates manual
octanol-phosphate buffer pH 7.4 200, 300 or 400 μL stock solution plus 200, 100 or 0 μL octanol (or phosphate buffer pH 7.4 solutions), then 400 μL of the other phase was added.
Automated Method modified shake-flask in 96 well plates automated 1 mg/mL analyte in octanol (96 well plates) octanol–phosphate buffer pH 7.4 400 μL phosphate buffer (100 mM pH 7.4) were pipetted in each well; 0, 100, 200 μL octanol added to each well, then 400, 200 and 300 μL respectively, analyte stock solution was also added
Repeat experiment/ 3 3 compound Compound partition reciprocal shaker, 5 min reciprocal shaker, 30 min Phase separation centrifugation centrifugation centrifugation Amount of sample 10 μL of organic and 80 10 μL of organic various injected in the μL of aqueous phase and 80 μL of HPLC system aqueous phase Analyte concentration measured by: Isocratic RP-HPLC: 100 mm long column, packed with 2 μm TKS gel Super-ODS (TosoHaas, Stuttgart, Germany); mobile phase: acetonitrile in 25 mM phosphate buffer with 0.2% triethylamine, pH 3.0; acetonitrile concentration adjusted to obtain retention times about 5 min; flow rate 1 mL/min UV detection, 260 nm Fast Gradient RP-HPLC: 50 mm long column packed with 2 μm TKS gel Super-ODS (TosoHaas, Stuttgart, Germany); mobile phase A: 0.07% (v/v) formic acid in acetonitrile/water (95:5) mobile phase B: 0.1% (v/v) formic acid in water The gradient starts with 100% B up to 100 A in 3.5 min and held there for 1 min. The total cycle time was 5.5 min
94
Advances in Chromatography
1. The formation of extremely stable emulsions that cannot be separated can prevent the accurate measurement of the analyte concentration in both phases. Complete separation is desirable as any droplets of octanol in water can contain significant quantities of the investigated compound. Even when octanol/water emulsions are quite stable, the two phases usually separate after 10 minutes of centrifugation. It takes a longer time to separate the phases when using propylene glycol dipelargonate (PGDP) as the partition solvent. 2. Measurement of extreme logP/logD values (e.g., for very hydrophilic or very hydrophobic compounds) requires a disproportionate volume ratio of the aqueous and solvent phase. In this situation, it is very difficult to sample a reasonable amount of solution from the phase with the smaller volume and automatic sampling is almost impossible. 3. The sample concentration must be less than the critical micelle concentration. 4. The sample concentration must be set below the aqueous solubility limit. 5. Some compounds, particularly the lipophilic compounds, can adhere to the surface of the vessel. 6. Some compounds can form micelles in the aqueous phase, act as surfactants that concentrate at the interface between the two liquids, or form foams. These compounds cannot be measured by a shake-flask method or any other method requiring agitation. 7. Compounds must have high purity. Measuring the exact concentration of compound in both phases is crucial. If a separation method is used for the quantification, the method is also applicable for impure compounds. When the volume ratio between the two solvent phases is high, an aliquot may be sampled from the larger volume before a suitable solvent is added to homogenously mix the two formerly immiscible phases, after which an aliquot is sampled from the resulting solution.
3.9 LIPOPHILICITY MEASUREMENT BY HIGHPERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) Chromatography has been intensively studied because of its remarkable capacity for separating compounds from the most complex mixtures. Assuming constant temperature, chromatographic separation is determined by three main properties: the chemical structure of the solute, the physicochemical properties of the mobile phase and of the stationary phase, respectively [64]. Of all chromatographic methods, liquid chromatography was of particular interest for lipophilicity measurement because of the similarity between the solute partition in a chromatographic system, having a liquid mobile phase and a liquid-like stationary phase, and the solute partition in a dual liquid phase environment.
Lipophilicity Measurements by Liquid Chromatography
95
What determines whether compound will spend more time in the mobile phase or in the stationary phase of a chromatographic system? The answer lies in the interactions between the solute and the two phases. These interactions include van der Waals forces, hydrogen bonding, interactions between dipoles, electron donor or acceptor interactions, and solvophobic interactions. Liquid chromatography is a more complex system than gas chromatography (GC), where the intermolecular interactions in the mobile phase are minimized by the gaseous state of the mobile phase. Therefore, creating model equations for the partition in a liquid chromatographic system is a complex task that has to take into account many variables. Reversed-phase liquid chromatography (RP-LC) is the generic name of a range of liquid chromatographic methods in which the stationary phase is less polar than the mobile phase. RP-LC is usually employed for lipophilicity estimations as it is easier to immobilize a nonpolar, viscous phase on the stationary phase than to use it as a liquid. RP-LC has other advantages over the traditional shake-flask method: • • •
• • • •
The method requires a much smaller amount of sample It is simple and rapid It is relatively insensitive to impurities, as it is a separation method — however, the impurity peak can have a larger area than the main compound if the impurity has a better UV absorption than the sample It can be used for a larger lipophilicity scale than is practically possible in shake-flask experiments Other practical problems encountered in shake-flask experiments are eliminated, like the formation of stable emulsions Universal detection methods can be used for the large majority of samples There is a linear relationship between the chromatographic retention parameter and the organic solvent concentration in the mobile phase
Liquid–liquid partition chromatography was pioneered by Martin and Synge, who received a Nobel prize for the separation of amino acids on a silica stationary phase with an aqueous mobile phase [65]. They also provided the first theoretical explanation of chromatographic separation by introducing the concept of theoretical plates, based on the plates used in column fractionation by distillation. Solvophobic interactions are an important part of the retention in RP-LC. Retention in liquid–liquid chromatography is a complex process that must take into account three major equilibria: the solute partition between the two liquid phases, one stationary and the other one mobile; the solute adsorption on the solid support; and the solute absorption at the liquid–liquid interface. Secondary processes such as diffusion into the pores of the solid support and flow patterns through the column are also present and can be a source for retention problems [66]. Thus, the retention is a function of the partition coefficients and the volumes of the two liquid phases. This is reflected by the retention factor (k) in an isocratic system,
96
Advances in Chromatography
k=
CSVS V =K S C M VM VM
(3.14)
where CS and CM are the solute concentration in the stationary and mobile phase, respectively, and VS and VM are the volume of stationary phase and mobile phase, respectively. The value K is the partition coefficient of the solute between the two phases. The retention in a chromatographic system can be expressed by the retention volume (VR) or by the retention time (tR): VR = VM (1 + k)
(3.15)
tR = tM (1 + k)
(3.16)
where tM is the dead time and can be measured for every column as the retention time of a completely unretained compound. Combining Equation 3.15 and Equation 3.16, the retention factor (k) can be expressed as, k=K
VS VR − VM = VM VM
(3.17)
K=
VR – VM VR′ = VS VS
(3.18)
and
Therefore, the partition coefficient between the two phases can be calculated from the retention data and the column parameters. This direct calculation involves errors in estimating the accurate volumes and is much easier to correlate k values with the partition coefficients in a true liquid dual-phase system. The retention in liquid–liquid chromatography is based on the different solvation of the analysed compound in the two phases. By contrast, the retention in reversedphase liquid chromatography is based on the hydrophobic effect caused by the exclusion of the solute by the water molecules in the mobile phase. As a consequence, the compound will orientate with the polar groups towards the aqueous stationary phase and the nonpolar, bulky part of the structure towards the hydrocarbonaceous stationary phase [67, 68].
3.9.1 HPLC
WITH
OCTANOL-COATED COLUMNS
Chromatography with octanol-coated columns is an interesting option for measuring the partition coefficient between an aqueous buffer and octanol. Usually, octanol is physically retained on a silica stationary phase and the mobile phase is an aqueous buffer containing different amounts of methanol or acetonitrile. The method was
Lipophilicity Measurements by Liquid Chromatography
97
described over 25 years ago [69]. McCall showed that logk values for benzene derivatives are in good agreement with logP values (shake-flask), especially for simple hydrocarbons, but significant deviations appear with increasing polarity of compounds [70]. HPLC methods using octanol-coated columns were reported as early as the 1970s. The stationary phase in the column was C18 Corasil, 37 to 50 μm particle size, in a 50 cm long column and 2 mm I.D. The column was coated with octanol following McCall’s method [70]. High purity octanol was injected directly into the column and the mobile phase was aqueous buffer and 25% (v/v) methanol in aqueous buffer (pH 7). The baseline normally stabilized in 30 to 45 minutes. Pyridine Noxide, formamide or simply methanol were used to measure the dead time and the peaks were detected by UV. Flow rate was 1 to 2 mL/min. The correlation with logD measured by the shake-flask method was highly linear for neutral derivatives of N-(-styryl)formamide. However, no linear correlation was seen for compounds with hydrogen-bonding capabilities, leading to the conclusion that methanol present in the mobile phase has a crucial influence [71]. In Mirrlees’ method, silanized kieselguhr was packed in columns (L = 10 to 30 cm, 4.6 mm I.D.) that were then dynamically coated with octanol. The mobile phase was water or pH adjusted buffer saturated with octanol. Pyridine N-oxide or tyrosine (both with logP < 1.5) were added for t0 estimation. Retention times of standards were run for each logP range and elution times up to 2 hours could be measured accurately. Values of logP in the range 0.3 to 3.7 were measured by varying the column length and flow rate. The column could be recoated with octanol after 50 hours of use. A remarkable slope equal to unity was obtained for the correlation between logP values and logarithm of retention times. The method seems to have two limitations. The first concerns compounds with poor solubility that tended to give lower logP values than expected, probably because the elution time was too short to reach equilibrium. The second limitation concerns compounds such as propanolol, where both charged and neutral species were extracted into octanol. This occurred because the pH of the mobile phase (7.4) was rather lower than the pKa of propranolol (9.56). In this case, the logD result obtained was lower than logP for the unionized compound. The authors also reported that a high salt concentration (above 0.1 M) in the buffer used for the mobile phase can influence the resulting logP/logD value. If the coating is not uniform, the resulting peaks will show a degree of splitting, making the choice of the correct retention time difficult. The recommended time for coating and octanol elution is 6 to 24 hours and recoating is normally completed in 2 hours. The flow rate was 0.2 to 8 mL/min and the volume of injected sample was typically 10 μL [69]. Although the chromatographic method using liquid octanol is promising, the authors stated that “it cannot be too strongly emphasized that, on all the available evidence, the only true model for octanol is octanol itself.” Octanol-coated HPLC silica-based columns were used by Slater et al. to measure the partition values for over 20 compounds. The results were compared with the corresponding logP/logD values obtained by potentiometric titration. The correlation between logPHPLC and logPliterature was very good (r = 0.991). Similar results were obtained for the correlation between logPpH-metric and logPliterature, where the intercept
98
Advances in Chromatography
was very close to zero (0.06), the slope close to 1 (0.9928) and a correlation coefficient of r = 0.994 [72]. Diatomaceous earth (Hyflosupercel, Johns Manville Inc., U.S.) has also been used as a support for nonpolar liquid coatings. This porous material is irregularly shaped and has different particle sizes but absorbs nonpolar liquids such as octanol and PGDP for long periods of time. No silanization is required if the material is treated with HCl, washed and dried. Mobile phase degassing is recommended to avoid spiking from the air bubbles trapped in the UV cell but it is not always necessary [34]. A modified HPLC method using octanol-coated columns is now commercially available from Sirius Analytical Instruments Ltd. on the instrument called ProfilerLDA [73]. The columns have a proprietary stationary phase and are dynamically coated with octanol. When the column is equilibrated, the samples are injected and the elution is continuously UV monitored. The software will automatically convert the retention times in logD values at the pH of the mobile phase. Usually, pH = 7.4 is used for the assays as it is the physiological pH value. The pH of the mobile phase can be modified for other physiologically relevant values (e.g., pH = 6.5). The instrument was designed for high-throughput screening of new chemical entities, mostly in the pharmaceutical industry, and can measure one sample in about 15 minutes. Samples are taken directly from 10 mM DMSO stock solutions from 96well plates. Typically, a 5 μL sample is injected into a 20 μL injection loop, filled with methanol, then injected into the column. The effluent is monitored by multi-wavelength UV spectroscopy. To maximize peak height vs. background noise, a maximum of 75 wavelengths are summed together and plotted against a logarithmic time scale. The range of available methods and logD range is shown in Table 3.2. The lifetime of an octanol-coated column is about 300 injections, although up to 600 injections have been reported without a significant change in the stationary phase uniformity. These numbers may seem low when compared with typical HPLC columns, which can withstand thousands of injections before showing the first degradation signs,
TABLE 3.2 Sirius ProfilerLDA
LogD range Column length Flow rate Eluent
Method 1 (low logD range)
Method 2 (medium logD range)
Method 3 (high logD range)
1.0 to 2.0 15 cm 3.5 mL/min 10 mM phosphate buffer; pH 7.4; 0.15 M ionic strength
1.5–3.5 1 cm 12 mL/min 10 mM phosphate buffer; pH 7.4; 0.15 M ionic strength
3.0–4.5 1 cm 12 mL/min 10 mM phosphate buffer containing 30% (v/v) methanol; pH 7.4; 0.15 M ionic strength
All methods use the same column type, coated with octanol; average run is 15 min/sample.
Lipophilicity Measurements by Liquid Chromatography
99
TABLE 3.3 logD7.4 Values Measured on GLpKa and ProfilerLDA
Compound
Predicted logD7.4 Value (ACD) [55]
logD7.4 Measured on ProfilerLDA (HPLC)
logD7.4 Measured on GLpKa (potentiometric)
3-chlorophenol Carazolol Carbendazim Chlorpromazine Desipramine Diclofenac Flurbiprofen Ibuprofen Indomethacin Lidocaine Nifuroxime Nitrazepam Papaverine Propanolol Quinine Tetracaine Tramadol Triamterene
2.42 1.58 1.42 3.36 1.23 0.11 0.92 0.77 –0.07 1.20 0.73 2.84 3.38 1.36 1.72 2.59 0.77 1.27
2.47 1.64 1.51 3.54 2.08 1.15 1.05 1.21 0.95 1.72 1.17 2.07 2.98 1.38 2.27 2.50 0.77 1.08
2.59 1.58 1.89 3.69 1.74 1.11 0.71 1.21 1.02 1.79 1.34 2.31 2.89 1.24 2.31 2.29 0.56 1.21
Compound Type acid base ampholyte base base acid acid acid acid base acid ampholyte base base base base base base
All measurements were performed at the same ionic strength, 0.15 M KCl. Sirius Analytical Instruments Ltd., Forest Row, U.K.
but this liquid-liquid chromatographic system works in a fragile thermodynamic equilibrium and care has been taken to provide a specific flow pattern that extends the column lifetime. The linearity of the relationship between the retention parameter and logD values is periodically checked with standards [74, 75]. Data in Table 3.3 show good correlation between the experimentally measured logD values at a pH of 7.4 by two different methods. ProfierLDA is based on the method described above, while GLpKa is based on a potentiometric titration method of the compound in aqueous solution in the presence of a layer of octanol. The potentimetric method for measuring pKa and logP/logD values is well documented [76–80]. Calibration is required at least once a day for octanol-coated columns to assess the coating uniformity. Column lifetime ranges from 10 to 600 injections, depending on the coating method and instrument used. Very lipophilic compounds have long retention times, often up to a couple of hours. The peaks will be broad and the peak shape nonsatisfactory for a usual HPLC experiment but the retention times can still be assessed with reasonable accuracy and the resulting logP values have a better confidence range than in shake-flask experiments. The lifetime of PGDP coated columns is usually longer that of octanol-coated columns [40, 42].
100
Advances in Chromatography
One of the difficulties of using octanol-coated columns is keeping the octanol in the column — usually by physical forces — and maintaining its initial volume constant during the elution. To the best of our knowledge, there is only one commercially available modified HPLC instrument using immobilized octanol on a solid packing [73]. Another difficulty in using octanol as stationary phase is related to the low number of theoretical plates compared with other HPLC columns and, consequently, a low capacity to separate impurities.
3.9.2 OTHER METHODS
USING
OCTANOL-COATED COLUMNS
One of the first methods for measuring partition coefficients used generator columns with a diluted solution of propylbezene in octanol to cover a silica-based column packing [81]. The octanol–water partition coefficients can be calculated from the measured propylbenzene concentrations in the equilibrated octanol and aqueous phase. The columns can be reused by pumping methanol to remove the organic phase and dried by flowing through a warm stream of nitrogen. The dried column is ready for recoating. The partition coefficient values obtained from the generator column method were used to evaluate six methods for Po/w estimation of highly hydrophobic aromatic hydrocarbons, one a measurement by RP-HPLC and the others computational methods using fragments or molecular descriptors [82]. The solvent-generated technique was designed to overcome the problems of poor modelling of the octanol–water partition by liquid chromatography with a “solid” stationary phase consisting of a long alkyl chain, bonded to a solid support, with similar properties with a liquid phase. The reported chromatographic systems used C8 and C18 chemically bonded silica as stationary phase: LiChrospher Si RP-8 and RP-18 (Merck, Darmstadt, Germany), and Nucleosil Si C8 (Macherey-Nagel, Düren, Germany). The mobile phase was 40% methanol with aqueous buffer (phosphate buffer pH = 2.0 for acids and pH = 7.5 for bases), both saturated with octanol following a special procedure to avoid the formation of stable emulsions and allowing a few drops of octanol on top of the mobile phase reservoir at all times to ensure saturation [83]. The formation of the octanol stationary phase was generated by flowing octanolsaturated phosphate buffer through the column. A group of neutral, acidic and basic aromatic compounds was analyzed by this method. The retention factor (logk) was measured on all columns by RP-HPLC and by the solvent generation method and then compared with logPo/w. The column intercomparison showed that the retention (logk values) were more reproducible when using the solvent generation method than when using traditional RP-HPLC. Also, the estimated logPo/w values were in good agreement with experimental logPo/w values. Other applications of the method are described in [84, 85].
3.9.2 ELogP METHOD Classic C8 and C18 silica-based stationary phases and octanol-enriched methanolwater mobile phases were tested by Minick et al. for lipophilicity measurements [86]. If methanol in the aqueous mobile phase is mixed with a small amount of
Lipophilicity Measurements by Liquid Chromatography
101
octanol, the new system will have similar characteristics with the dual phase octanolwater because octanol is significantly more lipophilic than methanol and should concentrate in the stationary phase, competing with methanol molecules and producing a more octanol-like stationary phase by minimizing hydrogen-bond discrimination. The aqueous part of the mobile phase was buffered with phosphate buffer (0.005 M) or with 4-morpholinepropanesulfonic acid (MOPS, 0.02 M) buffer to pH = 7.4. MOPS is an organic salt and should be less susceptible to precipitation at higher methanol concentration. The nature of the salt used to buffer the mobile phase had no significant influence on the retention of compounds. Amino modifiers such as n-decylamine (0.15%, v/v) or N,N-diethylamine (0.075%, v/v) were added to the deionized/distilled water used for the mobile phase to mask silanophilic interactions with the stationary phase. N-decylamine provided better correlations between logkw and logPo/w than N,N-diethylamine for both C8 and C18 stationary phases. This research work concluded that 0.25% octanol added to the methanol used for the mobile phase is the optimum amount to mimic a true octano–water environment. Higher concentrations were also tested but resulted in octanol separation when methanol was less than 50% (v/v). The method should provide a dynamic coating with octanol of the stationary phase, producing similar results to those obtained with a column coated with liquid octanol by direct injection [69]. Lombardo et al. used Minick’s method with octanol-enriched mobile phase and silica-based columns for the lipophilicity estimation of neutral compounds [15]. In Lombardo’s chromatographic system, the octanol from the mobile phase formed a fine molecular coating on the silica support of the column [87]. In this case, the presence of the amine masking agent was not necessary in the mobile phase. The retention parameter was called ELogP and was compared with logP and logD values, respectively. The ELogP values for neutral compounds [15] and for neutral and basic compounds [88] were reported in Lombardo’s work. The column used in Lombardo’s studies was deactivated, electrostatically coated Supelcosil LC-ABZ. The mobile phase was 15 to 70% (v/v) methanol in 20 mM MOPS buffer, pH = 7.4. The aqueous buffer was saturated with octanol and methanol contained 0.25% (v/v) methanol. A quantity of 0.15% (v/v) n-decylamine was added to the buffer for logD estimations to minimise the interactions between the charged compounds and the stationary phase. Extrapolated logkw values were used for correlations. The difference between the obtained ElogD values and experimental logD values were in the range 0.10 to 0.58 for 90 structurally unrelated drugs. Equation 3.19 describes the correlation between experimental logD values (obtained by other methods or reported in the literature) and extrapolated logkw values for the 90 drugs, covering the lipophilicity range from 1.49 for terbutaline sulphate (experimental logD value was 1.35) to 5.95 for amiodarone (experimental logD value was 6.1) [88]: logD = 0.207 + 1.127 log kw
(3.19)
n = 90, r2 = 0.964 The intercept and the slope are reasonably close to zero and unity, respectively, offering a good correlation between the shake-flask and RP-HPLC experiments. This
102
Advances in Chromatography
method is suitable for accurate lipophilicity estimation of neutral and basic compounds with unrelated structures but it is not recommendable for acids. The authors believe that specific interactions take place between anions and the stationary phase, despite adding different masking agents (some of them being zwitterions or anions) to the mobile phase [15, 88]. Even when great care is taken to model octanol–water partition, a compound can follow different partition patterns in the chromatographic separation and in a shake-flask experiment. The larger the surface between the two phases, the greater the chances that other factors such as surface activity and shape will also influence the partition [89].
3.9.4 REVERSED-PHASE HPLC (RP-HPLC) In reversed-phase high-performance liquid chromatography (RP-HPLC), the stationary phase is nonpolar (or less polar than the mobile phase) and the mobile phase is polar (or more polar than the stationary phase). Octadecylsilane (ODS) is one of the most popular packings used in RP-HPLC and the alkane chains form a molecular “fur” on the silica gel granule. RP-HPLC with a purely aqueous mobile phase is not the best experimental design for separation. The alkane “fur” needs a small amount of organic solvent in the mobile phase to help the alkane chains raise from a flat position and allow the free circulation of the solvent molecules. Methanol and acetonitrile are the most popular solvents used as organic modifiers in the mobile phase for RP-HPLC experiments related to lipophilicity estimations. However, the use of ethanol has also been reported [90]. Any organic compound that has a lipophilic and a hydrophilic part of the molecule will interact with both chromatographic phases with the lipophilic part positioned towards the stationary phase and the hydrophilic groups towards the mobile phase. Many authors believe that the retention in a reversed-phase chromatographic system with a C18 or C8 stationary phase is due to the hydrophobic exclusion of the solute molecule from the structural organization of water molecules in the mobile phase rather than on a specific interaction between the solute and the stationary phase.
3.9.5 THE
EXTRAPOLATION METHOD, LOGKW
The retention factor k, normally expressed as logk (log10k), is used for correlations with logPo/w. The relationship between logk — sometimes called capacity factor — and the fraction of organic modifier in the aqueous mobile phase (ϕ) is one of the most studied in RP-HPLC. The relationship is often linear (Equation 3.20) but, in many cases. there is parabolic dependence, especially at low organic solvent concentration (Equation 3.21): logk = a1 + b1 ϕ
(3.20)
logk = a2 + b2 ϕ + c2 ϕ2
(3.21)
Lipophilicity Measurements by Liquid Chromatography
103
where a1, b1, a2, b2 and c2 are coefficients. logk = logkw + S ϕ
(3.22)
This relationship is not linear for the entire range of organic modifier concentration in the mobile phase but is a good approximation for 0.5 < logk < 1.5 [89]. The extrapolated retention factor value (logkw) shows the retention of a compound in a chromatographic environment where only buffered water is used as a mobile phase. The value of logkw cannot usually be measured directly because of extremely long elution times. Therefore, the solute partition between the stationary phase and water could be a good approximation for the partition between an organic phase (e.g., octanol) and water if there are physicochemical similarities between the organic solvent and the stationary phase. All extrapolation methods rely on the linearity of Equation 3.22, assuming that the linearity is preserved even at low concentrations of organic solvent in the mobile phase. Sometimes, the extrapolation is performed from very high concentrations of organic solvent in the mobile phase (40 to 50%, v/v) without having an idea about how retention changes at lower concentrations. In practice, it is not always possible to measure retention times for very lipophilic compounds at low organic solvent concentration in the mobile phase because of very long elution times and distorted peaks The linearity of Equation 3.22 changes when water exceeds 90% (v/v) in the mobile phase. The logk values of nonpolar compounds measured in methanol-water systems are lower than the values predicted by extending the linear correlation; therefore, any prediction will provide a higher value than the real one. For polar compounds, however, the real logk values are higher than the extrapolated ones and any prediction will underestimate the real logk value [91]. A similar trend in the sign of the error between the measured and the extrapolated logkw values was found for acetonitrile-water mixtures (Figure 3.6). Accurate determination of the dead time (tM or t0) is very important, especially for logk values lower that 1 (e.g., for polar compounds) [92]. Errors in measuring the dead time will affect the k values and, ultimately, the correlation with logP values.
(a)
(b)
FIGURE 3.6 ODS structure in water and in water with methanol.
104
Advances in Chromatography
Most RP-HPLC applications for lipophilicity measurements use ODS stationary phases and methanol–buffer mobile phases. Standard procedures are still to be established as the methods are quite diverse. A compound’s capacity to donate or accept hydrogen bonds influences its partition between the alkyl-bonded stationary phase and the mobile phase. To eliminate this influence, some methods require the extrapolation of logk values to logkw (mobile phase containing 100% water). The extrapolation to 0% organic modifier in the mobile phase is based on the assumption that the linearity of the relationship between the retention factor logk and the concentration of the organic solvent in the aqueous mobile phase is maintained for the entire concentration range, even in the range 0 to 5%. In many cases, it is not possible to measure logkw because of the very high elution times. A linear relationship between logP and logkw for over 50 monosubstituted heteroaromatic series — derivatives of pyridine, pyrazine, prymidine, pyridazine, thiazole, and furane — was reported with a very good correlation coefficient, r2 = 0.95 [93]. According to Yamagami [93], logP values correlate well with logkw values for most substituents but, in some cases, usually for strong hydrogen-bond acceptors, the error in estimating logP is quite high and the obtained logkw value is usually higher than the real one. These conclusions may strongly depend on the type of stationary phase, particularly the amount of free silanol groups [94]. There are many studies reporting the use of logkw for lipophilicity studies. Most of them use methanol–water or methanol–aqueous buffer binary mobile phases with logkw obtained from extrapolations against methanol content [95–100]. The linearity of the relationship between logk and Φ depends on the solvent concentration in the mobile phase. Linearity is maintained for methanol–water mobile phases if the methanol concentration is in the range 30 to 70% (v/v). Outside this range, the relationship can show a pronounced curvature dependent on the retained compound [101]. When acetonitrile is used in the binary mobile phase, the relationship is almost always quadratic [102]. Values of logkw should be the same regardless of the column type and the organic solvent in the mobile phase but often this is not the case [103]. Different organic modifiers used in the mobile phase have different capacities to wet the same stationary phase. Moreover, the same mobile phase will not cover the same type of stationary phases from different manufacturers or from different batches in the same way. As a consequence, the intercept value (logkw) will be dependent on the mobile phase composition and the type of stationary phase used. Very good correlations of logkw to logP have been reported for benzene, furan, benzofuran, 1-Me-pyrrole, and 1-Me-indole derivatives in a chemically bonded, reversed-phase HPLC system with methanol-water mobile phase (r = 0.998 to 1.000). Most of these compounds do not form hydrogen bonds or are weak hydrogenbond acceptors. Pyrrole and indole derivatives are hydrogen-bond donors, so the logkw values have the tendency to underestimate the logP values. The corresponding ester derivatives are strong hydrogen-bond acceptors; therefore, the logkw values will overestimate the corresponding logP values. Good correlations between the isocratic logk50 measured at 50% methanol in the aqueous mobile phase were also obtained for these type of compounds. However, it is well known that the isocratic retention parameters in HPLC usually give good correlations with logP values for
Lipophilicity Measurements by Liquid Chromatography
105
congeneric series of compounds. The correlation is not so good when considering a group of chemically unrelated compounds [98]. A good correlation between isocratic logk and experimental logP was obtained for a noncongeneric but closely related group of acidic heterocycles. The chromatographic measurements were performed on a Waters μBondapak C18 column (10 μm packing particle size) with a mobile phase formed by 90% (v/v) distilled water adjusted to a pH of 3.5 with acetic acid, 5% isopropanol, and 5% methanol [104]. logk = 0.675 logP + 0.236
(3.23)
r = 0.964 It has been suggested that logkw is not a suitable parameter to be used in correlations and contributes to poor prediction performance. The value of logkw seems to correlate better with logP in homologous series of compounds and the correlation is rather poor when considering a chemically diverse set of compounds [6].
3.9.6 ISOCRATIC
LOGK
VALUES
The isocratic retention factor, logk, obtained on C18 silica-bonded phase was investigated in an interlaboratory study to determine its suitability for predicting partition coefficients in octanol-water [105]. The results showed that RP-HPLC is a reliable method to replace the reference method, shake-flask partitioning. However, there are a number of conditions to be met: at least six reference substances should be used, the minimum water content of the eluent should be 25% (v/v) and the extrapolation beyond the calibration range of logPo/w should only be carried out for logPo/w > 6. Many studies report isocratic logk values correlating reasonably well with logP values. Usually, logk50 is taken into consideration, obtained for 50% organic modifier in the mobile phase. Yamagami reported [93] an extremely good correlation between logP and logk50 (aqueous mobile phase containing 50% methanol) for series of lipophilic pyrazines and monosubstituted pyrazines covering six orders of magnitude. Measuring only logk50 values saves time compared with measuring logk at different concentrations of the organic solvent in the mobile phase and extrapolating to a hypothetical value of logkw that would be obtained with 100% water in the mobile phase. Why does 50% water in the mobile phase seem to provide a better correlation with logP values than the extrapolated chromatographic parameter, logkw? One possible answer is that when methanol is the organic modifier, the stationary phase is dynamically coated with methanol and gains octanol-like properties. If this hypothesis is true, the chromatographic system with octadecyl silane as stationary phase and equal parts of methanol and water (buffer) in the mobile phase could be more similar to the octanol-water system than any other methanol concentrations in the mobile phase [93]. Other studies also mention an isocratic logk [106–108].
106
Advances in Chromatography
3.9.7 THE SIGNIFICANCE
OF
SLOPE S
Because many experimental results from literature show that logkw values do not always model octanol–water partition correctly, some researchers have investigated the significance of the slope (S) with a view to using it as a second independent variable together with logkw: logP = aS + b logkw + c
(3.24)
where a, b, and c are coefficients [89]. A chromatographic hydrophobicity index (Φ0) was suggested to better model the octanol-water partition in a RP-HPLC system. The index (Φ0) was defined as the organic phase concentration (methanol or acetonitrile) in the mobile phase required to produce a capacity factor of logk = 0 [109] (see also 3.9.11). In this case, the retention time is double the dead time (tR = 2t0) and the molar fraction of the solute is the same in the mobile and the stationary phase (see Equation 3.15 and Equation 3.16): k=K
VS CV = S S VM C M VM
(3.25)
where CS and CM are the solute concentration in the stationary and mobile phase, respectively, and VS and VM are the volume of stationary and mobile phases. The value K is the partition coefficient of the solute between the stationary phase and the mobile phase at a defined composition. The logarithm of this equation can be expressed as: log k = log K + log
VS = Sϕ + log kw VM
(3.26)
A similar relationship can be written for Kw, the partition coefficient between the stationary phase and a purely aqueous mobile phase: log kw = K w + log
VS VM
(3.27)
Taking into account Equation 3.22, Φ0 can be expressed as: ϕ0 = −
log kw S
(3.28)
The Chromatographic Hydrophobicity Index concept was validated on a large number of compounds structurally related or not: nicotinate derivatives, hydroxyl aromatics, herbicides, substituted aromatics, aromatic hydrocarbon derivatives, 143
Lipophilicity Measurements by Liquid Chromatography
107
unrelated drug molecules, morphine derivatives, tricyclic drug molecules, deoxyuridine derivatives, benzodiazepines, aniline and phenol derivatives, adenosine monophosphate derivatives, β-lactam antibiotics, and barbiturate derivatives. The comprehensive study has proved that Φ0 is characteristic for every compound and depends only on the type of organic modifier, temperature, and pH. It is independent of the type of reversed-phase packing in the chromatographic column, the column dimensions, flow rate, and the mobile phase composition of the measurements. The Φ0 values obtained for acetonitrile and for methanol are in excellent correlation with each other, r = 0.96 for 72 compounds. The correlation between Φ0, ACN and logP for 140 compounds (r = 0.88) was slightly better than the correlation between Φ0, MeOH and logP for 448 compounds (r = 0.787) [109] where ACN is acetonitrile and MeOH is methanol.
3.9.8 CORRECTION
FOR IONIZATION
If compounds are ionized at the mobile phase pH, a correction for the right logkw must be applied to the apparent extrapolated logk value (logkwapp): logkw = logkwapp + log (1 + 10a(pH – pKa))
(3.29)
where a = 1 for monoprotic acids and a = –1 for monoprotic bases. More complex equations can be derived for multiprotic acid and bases [38]. Horvath has reported an equation between the capacity factor of a neutral molecule (kn) and the capacity factor of the same molecule but ionized (ki): Ka ⎡⎣ H + ⎤⎦ Ka 1+ ⎡⎣ H + ⎤⎦
kn + k j k=
(3.30)
where k is the capacity factor at a pH where the molecule is partially ionized and Ka is the ionisation constant [110, 111]. The retention of amphoteric compounds is directly dependent on the mobile phase pH, organic modifier in the mobile phase, and the presence of ion pairs. The chromatographic retention for seven ampholytes (niflumic acid, pyridoxine, norfloxacin, pefloxacin, morphine, nitrazepam, sulfadimine, and 11-amino undecanoic acid) was investigated on a Hypersil 5 ODS, 5-μm packing particle size, from Bioseparation Technique Ltd. (Budapest, Hungary) with 50-50 (v/v) methanol–aqueous phosphate buffer solutions, pH = 3 to 8 [26]. The pH of the aqueous buffer was shifted up 1 unit after mixing with equal parts of methanol. Two different categories were investigated, one involving zwitterionic amphoteric compounds when only the zwitterionic species exist at the pH of the isoelectric point (e.g., 11-amino undecanoic acid) and the other involving true ordinary amphoterics when the solution of the pH of the isoelectric point contains both ionized and
108
Advances in Chromatography
neutral species (e.g., nitrazepam and sulfadimine). The retention of the compounds that have a neutral species in the solution is definitely increased around the pH of the isoelectric point. It is reasonable to assume that the maximum retention of a compound will take place at a pH close to the isoelectric point and that the retention at the pH of the isoelectric point will be dramatically different for the compounds in the second category compared with the first category. The ion pairing process could explain why retention increases in pH regions where only ionized species are expected. The results of this study suggest there are enough similarities between the partition in a chromatographic system C18/methanol–water and the octanol–water dual-phase system to support the use of chromatographic retention parameters to model the partition between octanol and water [26]. A detailed study of ionizable compound retention in HPLC can be found in the literature [112, 113].
3.9.9 THE INFLUENCE MEASUREMENTS
OF
STATIONARY PHASE
ON
LIPOPHILICITY
The stationary phase most widely used in lipophilicity studies by RP-HPLC comprises hydrocarbon chains, usually C8 or C18, bonded on a silica-based support. Depending on the manufacturing process, the resulting stationary phase can be monomeric, where only one bond is formed per silane molecule, or polymeric, where multiple bonds are formed between the chains. The monomeric layer has a brushlike structure, activated by the presence of an organic modifier in the mobile phase, while the polymeric structure is not completely known. Both phases have unreacted, free silanol groups but their accessibility to passing molecules from the mobile phase or from the sample, is higher in the monomeric type. Different C18 silica-based stationary phases were studied with methanol-water eluents to investigate the reliability of the extrapolated logkw value. The variability of obtained logkw values was within 2 to 4%, showing that, if correctly determined, it is a solute property and independent of the stationary phase [114]. However, this result should be taken with caution because the practice shows poor reproducibility between similar types of stationary phases of different origin. One of the main problems of silica-based columns is their poor reproducibility when produced by different manufacturers. Although the structure seems identical, there are large differences in the chromatographic properties [115]. Values of logk obtained on a typical RP-HPLC system do not encode the same factors contributing to logPo/w values (e.g., logk values are influenced by the solute hydrogen-bond acidity while logPo/w values are not) [116]. Also, the alkyl-bonded stationary phase is not a bulk liquid and the solute has to penetrate it vertically, unlike the solvation in octanol [117]. Retention properties for 18 RP-HPLC columns were studied in an attempt to find similarities between the retention and the column type [118]. The correlations between logkw and logP were also investigated, showing that the slopes of the linear relationships are less than 1, ranging from 0.79 to 0.86. End-capping of silica-based stationary phases provide more reliable lipophilic correlation because of the blocking of some of the free silanols [87].
Lipophilicity Measurements by Liquid Chromatography
109
Silica-based stationary phases, whether derivatized or not, are suitable for dynamic coating with octanol and provide a system similar to the octanol-water dual phase. However, they are not suitable for coating with other solvents that may be of interest for correlations with biogical properties. A C18 stationary phase coderivatized with styrene divinylbenzene copolymer was reported and could be suitable for alkane coating [96]. Because of the limited chemical stability of silica-based stationary phases used in RP-HPLC, these phases cannot be used for important biologically active weak bases, such as alkaloids, with pKa values above 7.0. Other stationary phases (nonsilica based) were therefore investigated in an attempt to find correlations between the retention factor and lipophilicity. Alkyl-bonded alumina stationary phases have similar retention mechanisms as alkyl-bonded silica phases but have better stability at high pH. Alumina has a couple of advantages over silica gel — it is stable over a wide pH range and has no active silanol groups. An experimental study investigated the retention of alkaloids on three octadecyl-bonded stationary phases, octadecyl alumina (ODA), octadecyl silica (ODS) and octadecyl-modified polystyrene-divinylbenzene (ACT-1), and also on polybutadiene-coated alumina (PBD) [119]. Mixtures of methanol-buffer were used in all cases with different concentrations of methanol to obtain a retention factor in the range of 4 to 6. The aqueous phase was buffered at pH = 7.4 with phosphate buffers in all cases except ODA columns, where 4-morpholinopropanesulfonate (MOPS) was used because of the column instability in phosphate environment. The ionization of bases can be suppressed on ODA columns by working with high pH values in the mobile phase, up to 11.0. The correlations between logk and calculated logPo/w values for the studied alkaloids have better correlation coefficients for ODA columns (r = 0.953 to 0.980) than for ODS (r = 0.938 to 0.970), PBD (r = 0.881 to 0.985) and ACT-1 columns (r = 0.686 to 0.804). Poly(styrene-divinylbenzene) stationary phases are stable over the pH range 1 to 14 but the correlation with logPo/w for structurally unrelated compounds is not very good [120]. Shrinking and welling has been reported for these stationary phases [121]. Other stationary phases used in lipophilicity studies are: • • •
•
Monolithic silica columns [122, 123] Monolithic columns based on polystyrenes, polymethacrylates and polyacrylamides [124] Octadecylpolyvinyl copolymer — this stationary phase can provide good correlations with logP values but can also strongly retain specific compounds [125] Alkylamide silica (alkyl chains attached to the polar alkylamide moiety) [126, 127]
Good correlations for congeneric compounds are reported in O’Gara et al. and Czajkowska et al. [128, 129] and not so good in Buszewski et al. [130] Nonsilicabased stationary phases would not have the residual silanol groups and will be
110
Advances in Chromatography
sufficiently stable at high pH values. Hydrocarbon-coated alumina, zirconia, titania and ceria are also resistant to high pH values [131–134]. Two polyethylene-coated columns, one silica-based (PECSiO2) and the other zirconia based (PECZrO2), were investigated for their hydrophobic properties by testing 25 simple, structurally diverse organic compounds [135]. The mobile phase was 10% methanol in phosphate buffer, pH = 7.2. In some cases, the capacity factor had to be measured with 80% methanol in the mobile phase and the corresponding logk values for 10% methanol obtained by extrapolation against the concentration of the organic modifier in the mobile phase. The comparison between the capacity factors for the same experimental conditions on the two investigated columns has a correlation coefficient r = 0.976 for the 25 investigated compounds (Equation 3.31): logkPECZrO2 = 0.224 + 0.988 logkPECSiO2
(3.31)
n = 25, r = 0.976 The intercept and the slope are reasonably close to zero and unity, respectively, suggesting that the columns have similar properties with stronger retention on the silica-based column. Although the studied chromatographic systems looked promising, the LSER studies showed that a direct correlation with logPo/w values is not possible in this case. However, a good lipophilicity scale and quantitative-structureretention relationships (QSRR) prediction models can be easily obtained. Poly(butadiene)-coated alumina columns, Spherisorb A5Y (ES Industries, Marlton, NJ, USA) or home-made, were also used for lipophilicity evaluation of basic, neutral and acidic compounds. The mobile phase was methanol–buffer (1:1, v/v) with the pH adjusted to keep the compounds in the neutral form (a pH of 1.65 for acids and up to 11.5 for bases). The relationship between logk and logP for 24 unrelated organic compounds was linear with a good correlation coefficient (r = 0.96) but, more importantly, a high pH could be used in the mobile phase to measure the retention of basic compounds in their neutral form [136, 137]. The chromatographic retention data on the diol stationary phase was found to provide better correlations with logPo/w than RP-HPLC for series of polar polycyclic aromatic compounds (N-PAC) [138]. The study used two columns: an ODS (Phenomenex Prodigy ODS-2) and a Diol type (Nucleosil 7 OH, Macherey Nagel). Experiments with n-hexane as eluent showed no difference in retention times of benzene and nitro-benzene on the ODS column, confirming a negligible amount of residual silanol groups. The mobile phase was a mixture of methanol and water, 35 and 50% (v/v) methanol for the Diol column and 65, 75, and 85% (v/v) methanol for the ODS column. The retention data in the RP-HPLC systems were compared with partition coefficient values obtained from generator column experiments, on Chromosorb W generator columns (dimethyl chlorosilane, treated, and acid washed) and an extractor column packed with polygosil 35-42 octadecyl silica. Octadecyl polyvinylalcohol copolymer (ODP) columns can be used with 100% water in the mobile phase where no masking agent is required and the packing is stable at pH from 1 to 14 [139–141].
Lipophilicity Measurements by Liquid Chromatography
111
The influence of organic modifiers in the formation of the stationary phase in RP-HPLC was investigated for methanol, acetonitrile, and tetrahydrofuran [142]. The formation of the stationary phase for C8 and C18 bonded silica involves the adsorption of molecules from the aqueous mobile phase. The organic modifiers used in the mobile phase can selectively enrich the stationary phase while the residual silanol groups will prefer molecules with hydrogen-bonding capabilities like water, methanol, acetonitrile, and tetrahydrofuran. Methanol was widely used in aqueous mixtures and was particularly studied [143]. The concentration of methanol in the aqueous mobile phase was varied from 0 to 100% and the amount of methanol and water that bonds to the C8 or C18 stationary phase was measured by gas chromatography with flame ionization detection (for methanol) and thermal conductivity detection (for water). The increase of methanol fraction in the mobile phase leads to an increase of the amount of methanol bonded to the stationary phase. The effect is more pronounced for C18 phases than for C8. On solvation with methanol, C8 forms a “brush-like” structure on the surface of the solid packing, allowing direct access of the solvent molecules to the free silanol groups. There are structural differences between the C8 and C18 stationary phases, resulting in C18 chains being closer to each other than C8. As a consequence, a lower methanol concentration (less than 20%, v/v) will not open the C18 chains in the same way it does with C8, so the methanol layer on the stationary phase will be less on C18 than on C8. At high methanol concentrations, C18 will also show a “brush-like” structure, allowing the solvent molecules access to all binding sites in the stationary phase. Due to the stationary liquid phase formed on the C8/C18 silica support, the retention mechanism is governed by partition. What happens when the mobile phase is 100% aqueous in a C8/C18 chromatographic environment? Water molecules have been found on the stationary phase and their presence can be explained either by adsorption on the active sites or by being trapped under the hydrocarbon chains. Either way, the retention mechanism will be governed by adsorption rather than partition [143]. A nonporous, ultra-small (1.5 μm), C18 stationary phase was investigated for its suitability in RP-HPLC lipophilicity studies. The column was NPS RP-18 (Micrascientific, Northbrook, IL) and the mobile phases were binary aqueous mixtures, methanol-phosphoric acid (0.03 M) and acetonitrile-phosphoric acid (0.03 M) with various concentrations of organic modifier including 0% (v/v) [144]. Very good correlations between logk and Φ were found in all cases (r2 = 0.99) for the investigated structurally diverse compounds. The following equations were obtained for logPo/w to logkw correlations: logPo/w = 0.87 + 1.18 logkw,ACN,
r2 = 0.977
(3.32)
logPo/w = 1.11 + 0.90 logkw,MeOH,
r2 = 0.961
(3.33)
where logkw,ACN and logkw,MeOH are the extrapolated values for zero content of acetonitrile and methanol, respectively, in the mobile phase. The slopes in the above equations are a measure of the mobile phase sensitivity to changes in the
112
Advances in Chromatography
hydrophobicity of the studied compounds. Both slopes are close to unity, showing a good correlation between logPo/w and logkw with acetonitrile being more sensitive than methanol to structural changes in the solute. Specially formulated stationary phases allow the use of pure aqueous mobile phases in a reversed-phase mechanism. Trifluorpropyl siloxane, commonly used in gas chromatography, has been applied on nonporous glass microspheres [145]. The benefits of a RP-HPLC column that works with pure water as a mobile phase range from low solvent consumption to the possibility of directly measuring the logkw value without the problems faced by the extrapolation method. The retention mechanism on the trifluorpropyl siloxane stationary phase has been proven to be a reversed phase mechanism, similar to the RP-HPLC environment on silica-C18. The relationship between logk and logPo/w has an excellent correlation coefficient (r = 0.998) for ten hydrophobic simple organic compounds.
3.9.10 MOBILE PHASE COMPOSITION
AND
LIPOPHILICITY
3.9.10.1 Organic Modifier Pure water cannot be used on most chromatographic columns, and thus an organic modifier is usually added. The retention on hydrocarbon stationary phases with water as eluent never correlates with logP and many efforts have been made to find the most suitable organic modifier in the mobile phase [34]. Methanol, acetonitrile, and tetrahydrofuran are popular organic modifiers in RPHPLC. In theory, the extrapolated logkw values should be the same regardless of the organic modifier used in the mobile phase. Significant differences exist in practice, making the extrapolation method unreliable. Also, the linear relationship between the retention factor, logk, and the concentration of organic solvent in the mobile phase, tends to curve at high water concentration. The errors depend on the nature of the organic modifier (see related text on the extrapolation method) and can be as much as 10% for methanol-water mixtures and 20% for acetonitrile-water mobile phases [91]. The hydrocarbon-bonded silica-based stationary phase has a brush-like structure at high concentration of organic solvent in the mobile phase. This structure collapses when high water content mobile phases are used, significantly reducing the active surface area of the stationary phase. This could be one of the reasons why the extrapolation method fails to accurately predict the logkw values. Methanol is by far the most popular organic modifier used in RP-HPLC. Good correlations are obtained between logP and logk/logkw for the large majority of studied congeneric compounds: logP = a logk + b
(3.34)
In many cases, the slope (a) is close to unity and the intercept (b) is close to zero, indicating a solid correlation between the forces involved in the two methods. Methanol has hydrogen-bond donor and acceptor properties, changing the cohesion of the water mo