Handbook of X-Ray Spectrometry Second Edition, Revised and Expanded
edited by René E. Van Grieken University of Antwer...
102 downloads
2270 Views
10MB Size
Report
This content was uploaded by our users and we assume good faith they have the permission to share this book. If you own the copyright to this book and it is wrongfully on our website, we offer a simple DMCA procedure to remove your content from our site. Start by pressing the button below!
Report copyright / DMCA form
Handbook of X-Ray Spectrometry Second Edition, Revised and Expanded
edited by René E. Van Grieken University of Antwerp Antwerp, Belgium
Andrzej A. Markowicz Vienna, Austria
Marcel Dekker, Inc.
New York • Basel
TM
Copyright © 2001 by Marcel Dekker, Inc. All Rights Reserved.
Copyright © 2002 Marcel Dekker, Inc.
ISBN: 0-8247-0600-5 First edition was published as Handbook of X-Ray Spectrometry: Methods and Techniques This book is printed on acid-free paper. Headquarters Marcel Dekker, Inc. 270 Madison Avenue, New York, NY 10016 tel: 212-696-9000; fax: 212-685-4540 Eastern Hemisphere Distribution Marcel Dekker AG Hutgasse 4, Postfach 812, CH-4001 Basel, Switzerland tel: 41-61-261-8482; fax: 41-61-261-8896 World Wide Web http:==www.dekker.com The publisher offers discounts on this book when ordered in bulk quantities. For more information, write to Special Sales=Professional Marketing at the headquarters address above. Copyright # 2002 by Marcel Dekker, Inc. All Rights Reserved. Neither this book nor any part may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopying, microfilming, and recording, or by any information storage and retrieval system, without permission in writing from the publisher. Current printing (last digit): 10 9 8 7 6 5 4 3 2 1 PRINTED IN THE UNITED STATES OF AMERICA
Copyright © 2002 Marcel Dekker, Inc.
Preface to the Second Edition
The positive response to the first edition of Handbook of X-Ray Spectrometry: Methods and Techniques and its commercial success have shown that in the early 1990s there was a clear demand for an exhaustive book covering most of the specialized approaches in this field. Therefore, some five years after the first edition appeared, the idea of publishing a second edition emerged. In the meantime, remarkable and steady progress has been made in both instrumental and methodological aspects of x-ray spectrometry. This progress includes considerable improvements in the design and production technology of detectors and in capillary optics applied for focusing the primary photon beam. The advances in instrumentation, spectrum evaluation, and quantification have resulted in improved analytical performance and in further extensions of the applicability range of xray spectrometry. Consequently, most of the authors who contributed to the first edition of this book enthusiastically accepted the invitation to update their chapters. The progress made during the last decade is reflected well in the chapters of the second edition, which were all considerably revised, updated, and expanded. A completely new chapter on microbeam x-ray fluorescence analysis has also been included. Chapter 1 reviews the basic physics behind x-ray emission techniques, and refers to extensive appendices for all the basic and generally applicable x-ray physics constants. New analytical expressions have been introduced for the calculation of fundamental parameters such as the fluorescence yield, incoherent scattering function, atomic form factor, and total mass attenuation coefficient. Chapter 2 outlines established and new instrumentation and discusses the performances of wavelength-dispersive x-ray fluorescence (XRF) analysis, which, with probably 15,000 units in operation worldwide today, is still the workhorse of x-ray analysis. Its applications include process control, materials analysis, metallurgy, mining, and almost every other major branch of science. The additional material in this edition covers new sources of excitation and comprehensive comparisons of the technical parameters of newly produced wavelength-dispersive spectrometers. Chapter 3 has been completely reconsidered, modified, and rewritten by a new author. The basic principles, background, and recent advances are described for the tubeexcited energy-dispersive mode, which is invoked so frequently in research on environmental and biological samples. This chapter is based on a fresh look and follows a completely different approach.
Copyright © 2002 Marcel Dekker, Inc.
Chapter 4 reviews in depth the available alternatives for spectrum evaluation and qualitative analysis. Techniques for deconvolution of spectra have enormously increased the utility of energy-dispersive x-ray analysis, but deconvolution is still its most critical step. The second edition includes discussions of partial least-squares regression and modified Gaussian shape profiles. Chapter 5 reviews quantification in XRF analysis of the classical and typical ‘‘infinitely thick’’ samples. In addition to being updated, the sections on calibration, quality control, and mathematical correction methods have been expanded. Chapter 6, on quantification for ‘‘intermediate-thickness’’ samples, now also includes the presentation of a modified version of the emission-transmission method and a discussion of both the accuracy and limitations of such methods. Chapter 7 is a completely original treatment by a new author of radioisotope-induced and portable XRF. It discusses semiconductor detectors, including the latest types, analyzes in detail the uncertainty sources, and reviews the recent and increasingly important applications. Since the appearance of the first edition, synchrotron-induced x-ray emission analysis has increased in importance. Chapter 8 was updated and modified by including a comprehensive review of the major synchrotron facilities. Although its principles have been known for some time, it is only since the advent of powerful commercial units and the combination with synchrotron sources that total reflection XRF has rapidly grown, mostly now for characterization of surfaces and of liquid samples. This is the subject of the substantially modified and expanded Chapter 9. The new authors have taken a radically different approach to the subject. Polarized-beam XRF and its new commercial instruments are treated in detail in a substantially revised and expanded Chapter 10. Capillary optics combined with conventional fine-focus x-ray tubes have enabled the development of tabletop micro-XRF instruments. The principles of the strongly growing microbeam XRF and its applications are now covered thoroughly in an additional chapter, Chapter 11. Particle-induced x-ray emission analysis has grown recently in its application types and particularly in its microversion. Chapter 12 discusses the physical backgrounds, instrumentation, performance, and applications of this technique. The sections dealing with the applications were substantially expanded. Although the practical approaches to electron-induced x-ray emission analysis— a standard technique with wide applications in all branches of science and technology— are often quite different from those in other x-ray analysis techniques, a treatment of its potential for quantitative and spatially resolved analysis is given in Chapter 13. The new and expanded sections deal with recent absorption correction procedures and with the quantitative analysis of samples with nonstandard geometries. Finally, the completely updated and revised Chapter 14 reviews the sample preparation techniques that are invoked most frequently in XRF analysis. The second edition of this book is again a multiauthored effort. We believe that having scientists who are actively engaged in a particular technique covering those areas in which they are particularly qualified outweighs any advantages of uniformity and homogeneity that characterize a single-authored book. The editors (and one coworker) again wrote three of the chapters in the new edition. For all the other chapters, we were fortunate to have the cooperation of truly eminent specialists, some of whom are new contributors (see Chapters 3, 7, 9, 10 and 11). We wish to thank all the contributors for their considerable and (in most cases) timely efforts. Copyright © 2002 Marcel Dekker, Inc.
We hope that novices in x-ray emission analysis will find this revised and expanded handbook useful and instructive, and that our more experienced colleagues will benefit from the large amount of readily accessible information available in this compact form, some of it for the first time. An effort has been made to emphasize the fields and developments that have come into prominence lately and have not been covered in other general books on x-ray spectrometry. We also hope this book will help analytical chemists and other users of x-ray spectrometry to fully exploit the capabilities of this powerful analytical tool and to further expand its applications in material and environmental sciences, medicine, toxicology, forensics, archeometry, and many other fields. Rene´ E. Van Grieken Andrzej A. Markowicz
Copyright © 2002 Marcel Dekker, Inc.
Preface to the First Edition
Scientists in recent years have been somewhat ambivalent regarding the role of x-ray emission spectrometry in analytical chemistry. Whereas no radically new and stunning developments have been seen, there has been remarkably steady progress, both instrumental and methodological, in the more conventional realms of x-ray fluorescence. For the more specialized approaches—for example, x-ray emission induced by synchrotron radiation, radioisotopes and polarized x-ray beams, and total-reflection x-ray fluorescence— and for advanced spectrum analysis methods, exponential growth and=or increasing acceptance has occurred. Contrary to previous books on x-ray emission analysis, these latter approaches make up a large portion of the present Handbook of X-Ray Spectrometry. The major milestone developments that shaped the field of x-ray spectrometry and now have widespread applications all took place more than twenty years ago. After wavelength-dispersive x-ray spectrometry had been demonstrated and a high-vacuum x-ray tube had been introduced by Coolidge in 1913, the prototype of the first modern commercial x-ray spectrometer with a sealed x-ray tube was built by Friedmann and Birks in 1948. The first electron microprobe was successfully developed in 1951 by Castaing, who also outlined the fundamental concepts of quantitative analysis with it. The semiconductor or Si(Li) detector, which heralded the advent of energy-dispersive x-ray fluorescence, was developed around 1965 at Lawrence Berkeley Laboratory. Acceleratorbased particle-induced x-ray emission analysis was developed just before 1970, mostly at the University of Lund. The various popular matrix correction methods by Lucas-Tooth, Traill and Lachance, Claisse and Quintin, Tertian, and several others, were all proposed in the 1960s. One may thus wonder whether the more conventional types of x-ray fluorescence analysis have reached a state of saturation and consolidation, typical for a mature and routinely applied analysis technique. Reviewing the state of the art and describing recent progress for wavelength- and energy-dispersive x-ray fluorescence, electron and heavy charged-particle-induced x-ray emission, quantification, and sample preparation methods is the purpose of the remaining part of this book. Chapter 1 reviews the basic physics behind the x-ray emission techniques, and refers to the appendixes for all the basic and generally applicable x-ray physics constants. Chapter 2 outlines established and new instrumentation and discusses the performances of wavelength-dispersive x-ray fluorescence analysis, which, with probably 14,000 units in operation worldwide today, is still the workhorse of x-ray analysis with applications in a wide range of disciplines including process control, materials analysis, metallurgy, Copyright © 2002 Marcel Dekker, Inc.
mining, and almost every other major branch of science. Chapter 3 discusses the basic principles, background, and recent advances in the tube-excited energy-dispersive mode, which, after hectic growth in the 1970s, has now apparently leveled off to make up approximately 20% of the x-ray fluorescence market; it is invoked frequently in research on environmental and biological samples. Chapter 4 reviews in depth the available alternatives for spectrum evaluation and qualitative analysis; techniques for deconvolution of spectra have enormously increased the utility of energy-dispersive x-ray analysis, but deconvolution is still its most critical step. Chapters 5 and 6 review the quantification problems in the analysis of samples that are infinitely thick and of intermediate thickness, respectively. Chapter 7 is a very practical treatment of radioisotope-induced x-ray analysis, which is now rapidly acquiring wide acceptance for dedicated instruments and field applications. Chapter 8 reviews synchrotron-induced x-ray emission analysis, the youngest branch, with limited accessibility but an exponentially growing literature due to its extreme sensitivity and microanalysis potential. Although its principles have been known for some time, it is only since the advent of powerful commercial units that total reflection x-ray fluorescence has been rapidly introduced, mostly for liquid samples and surface layer characterization; this is the subject of Chapter 9. Polarized beam x-ray fluorescence is outlined in Chapter 10. Particle-induced x-ray emission analysis is available at many accelerator centers worldwide; the number of annual articles on it is growing and it undergoes a revival in its microversion; Chapter 11 treats the physical backgrounds, instrumentation, performance, and applications of this technique. Although the practical approaches to electron-induced x-ray emission analysis, now a standard technique with wide applications in all branches of science and technology, are often quite different from those in other x-ray analysis techniques, a separate treatment of its potential for quantitative and spatially resolved analysis is given in Chapter 12. Finally, Chapter 13 briefly reviews the sample preparation techniques that are invoked most frequently in combination with x-ray fluorescence analysis. This book is a multi-authored effort. We believe that having scientists who are actively engaged in a particular technique covering those areas for which they are particularly qualified and presenting their own points of view and general approaches outweighs any advantages of uniformity and homogeneity that characterize a single-author book. Three chapters were written by the editors and a coworker. For all the other chapters, we were fortunate enough to have the cooperation of eminent specialists. The editors wish to thank all the contributors for their efforts. We hope that novices in x-ray emission analysis will find this book useful and instructive, and that our more experienced colleagues will benefit from the large amount of readily accessible information available in this compact form, some of it for the first time. This book is not intended to replace earlier works, some of which were truly excellent, but to supplement them. Some overlap is inevitable, but an effort has been made to emphasize the fields and developments that have come into prominence lately and have not been treated in a handbook before. Rene´ E. Van Grieken Andrzej A. Markowicz
Copyright © 2002 Marcel Dekker, Inc.
Contents
Preface to the Second Edition Preface to the First Edition Contributors 1 X-ray Physics Andrzej A. Markowicz I. Introduction II. History III. General Features IV. Emission of Continuous Radiation V. Emission of Characteristic X-rays VI. Interaction of Photons with Matter VII. Intensity of Characteristic X-rays VIII. IUPAC Notation for X-ray Spectroscopy Appendixes I. Critical Absorption Wavelengths and Critical Absorption Energies II. Characteristic X-ray Wavelengths (A˚) and Energies (keV) III. Radiative Transition Probabilities IV. Natural Widths of K and L Levels and Ka X-ray Lines (FWHM), in eV V. Wavelengths of K Satellite Lines (A˚) VI. Fluorescence Yields and Coster–Kronig Transition Probabilities VII. Coefficients for Calculating the Photoelectric Absorption Cross Sections t (Barns=Atom) Via ln–ln Representation VIII. Coefficients for Calculating the Incoherent Collision Cross Sections sc (Barns=Atom) Via the ln–ln Representation IX. Coefficients for Calculating the Coherent Scattering Cross Sections sR (Barns=Atom) Via the ln–ln Representation X. Parameters for Calculating the Total Mass Attenuation Coefficients in the Energy Range 0.1–1000 keV [Via Eq. (78)] XI. Total Mass Attenuation Coefficients for Low-Energy Ka Lines XII. Correspondence Between Old Siegbahn and New IUPAC Notation X-ray Diagram Lines References Copyright © 2002 Marcel Dekker, Inc.
56
2 Wavelength-Dispersive X-ray Fluorescence Jozef A. Helsen and Andrzej Kuczumow I. II. III. IV. V. VI. VII.
Introduction Fundamentals of Wavelength Dispersion Layout of a Spectrometer Qualitative and Quantitative Analysis Chemical Shift and Speciation Instrumentation Future Prospects References
3 Energy-Dispersive X-ray Fluorescence Analysis Using X-ray Tube Excitation Andrew T. Ellis I. II. III. IV. V.
Introduction X-ray Tube Excitation Systems Semiconductor Detectors Semiconductor Detector Electronics Summary References
4 Spectrum Evaluation Piet Van Espen I. II. III. IV. V. VI. VII. VIII. IX. X.
Introduction Fundamental Aspects Spectrum Processing Methods Continuum Estimation Methods Simple Net Peak Area Determination Least-Squares Fitting Using Reference Spectra Least-Squares Fitting Using Analytical Functions Methods Based on the Monte Carlo Technique The Least-Squares-Fitting Method Computer Implementation of Various Algorithms References
5 Quantification of Infinitely Thick Specimens by XRF Analysis Johan L. de Vries and Bruno A. R. Vrebos I. II. III. IV. V. VI.
Introduction Correlation Between Count Rate and Specimen Composition Factors Influencing the Accuracy of the Intensity Measurement Calibration and Standard Specimens Converting Intensities to Concentration Conclusion References
Copyright © 2002 Marcel Dekker, Inc.
6 Quantification in XRF Analysis of Intermediate-Thickness Samples Andrzej A. Markowicz and Rene´ E. Van Grieken I. II. III. IV.
Introduction Emission-Transmission Method Absorption Correction Methods Via Scattered Primary Radiation Quantitation for Intermediate-Thickness Granular Specimens References
7 Radioisotope-Excited X-ray Analysis Stanislaw Piorek I. II. III. IV. V. VI. VII. VIII.
Introduction Basic Equations Radioisotope X-ray Sources and Detectors X-ray and g-ray Techniques Factors Affecting the Overall Accuracy of XRF Analysis Applications Future of Radioisotope-Excited XRF Analysis Conclusions Appendix: List of Companies that Manufacture Radioisotope-Based X-ray Analyzers and Systems References
8 Synchrotron Radiation-Induced X-ray Emission Keith W. Jones I. II. III. IV. V. VI. VII. VIII. IX. X. XI. XII.
Introduction Properties of Synchrotron Radiation Description of Synchrotron Facilities Apparatus for X-ray Microscopy Continuum and Monochromatic Excitation Quantitation Sensitivities and Minimum Detection Limits Beam-Induced Damage Applications of SRIXE Tomography EXAFS and XANES Future Directions References
9 Total Reflection X-ray Fluorescence Peter Kregsamer, Christina Streli, and Peter Wobrauschek I. II. III. IV. V. VI. VII.
Introduction Physical Principles Instrumentation Chemical Analysis Surface Analysis Thin Films and Depth Profiles Synchrotron Radiation Excitation
Copyright © 2002 Marcel Dekker, Inc.
VIII. IX.
Light Elements Related Techniques References
10 Polarized Beam X-ray Fluorescence Analysis Joachim Heckel and Richard W. Ryon I. II. III. IV. V. VI. VII.
Introduction Theory Barkla Systems Bragg Systems Barkla-Bragg Combination Systems Secondary Targets Conclusion References
11 Microbeam XRF Anders Rindby and Koen H. A. Janssens I. II. III. IV. V.
Introduction and Historical Perspective Theoretical Background Instrumentation for Microbeam XRF Collection and Processing of m-XRF Data Applications References
12 Particle-Induced X-ray Emission Analysis Willy Maenhaut and Klas G. Malmqvist I. II.
III. IV. V. VI. VII. VIII.
Introduction Interactions of Charged Particles with Matter, Characteristic X-ray Production, and Continuous Photon Background Production Instrumentation Quantitation, Detection Limits, Accuracy, and Precision Sample Collection and Sample and Specimen Preparation for PIXE Analysis Applications Complementary Ion-Beam-Analysis Techniques Conclusions References
13 Electron-Induced X-ray Emission John A. Small, Dale E. Newbury, and John T. Armstrong I. II. III. IV. V.
Introduction Quantitative Analysis Microanalysis at Low Electron Beam Energy Analysis of Samples with Nonstandard Geometries Spatially Resolved X-ray Analysis References
Copyright © 2002 Marcel Dekker, Inc.
14 Sample Preparation for X-ray Fluorescence Martina Schmeling and Rene´ E. Van Grieken I. II. III. IV. V. VI. VII.
Introduction Solid Samples Fused Specimen Liquid Specimen Biological Samples Atmospheric Particles Sample Support Materials References
Copyright © 2002 Marcel Dekker, Inc.
Contributors
John T. Armstrong, Ph.D. Gaithersburg, Maryland
National Institute of Standards and Technology,
Eindhoven, The Netherlands
Johan L. de Vries, Ph.D.*
Oxford Instruments Analytical Ltd., High Wycombe,
Andrew T. Ellis, Ph.D. Buckinghamshire, England Joachim Heckel, Ph.D. many
Spectro Analytical Instruments, GmbH & Co. KG, Kleve, Ger-
Jozef A. Helsen, Ph.D.
Catholic University of Leuven, Leuven, Belgium University of Antwerp, Antwerp, Belgium
Koen H. A. Janssens, Ph.D. Keith W. Jones, Ph.D.
Brookhaven National Laboratory, Upton, New York
Peter Kregsamer, Dr. techn., Dipl. Ing.
Lublin Catholic University, Lublin, Poland
Andrzej Kuczumow, Ph.D. Willy Maenhaut, Ph.D.
Ghent University, Ghent, Belgium
Klas G. Malmqvist, Ph.D. Sweden
Lund University and Lund Institute of Technology, Lund,
Andrzej A. Markowicz, Ph.D. Dale E. Newbury, Ph.D. Maryland
Atominstitut, Vienna, Austria
Vienna, Austria
National Institute of Standards and Technology, Gaithersburg,
Copyright © 2002 Marcel Dekker, Inc.
Stanislaw Piorek, Ph.D.{ Anders Rindby, Ph.D. Go¨tebo¨rg, Sweden
Niton Corporation, Billerica, Massachusetts
Chalmers University of Technology and University of Go¨tebo¨rg,
Lawrence Livermore National Laboratory, Livermore, Califor-
Richard W. Ryon, B.A. nia
Martina Schmeling, Ph.D.
Loyola University Chicago, Chicago, Illinois
John A. Small, Ph.D. Maryland
National Institute of Standards and Technology, Gaithersburg,
Christina Streli, Ph.D.
Atominstitut, Vienna, Austria
Piet Van Espen, Ph.D.
University of Antwerp, Antwerp, Belgium
Rene´ E. Van Grieken, Ph.D.
University of Antwerp, Antwerp, Belgium
Bruno A. R. Vrebos, Dr. Ir.
Philips Analytical, Almelo, The Netherlands
Peter Wobrauschek, Ph.D.
Copyright © 2002 Marcel Dekker, Inc.
Atominstitut, Vienna, Austria
1 X-ray Physics Andrzej A. Markowicz Vienna, Austria
I.
INTRODUCTION
In this introductory chapter, the basic concepts and processes of x-ray physics that relate to x-ray spectrometry are presented. Special emphasis is on the emission of the continuum and characteristic x-rays as well as on the interactions of photons with matter. In the latter, only major processes of the interactions are covered in detail, and the cross sections for different types of interactions and the fundamental parameters for other processes involved in the emission of the characteristic x-rays are given by the analytical expressions and=or in a tabulated form. Basic equations for the intensity of the characteristic x-rays for the different modes of x-ray spectrometry are also presented (without derivation). Detailed expressions relating the emitted intensity of the characteristic x-rays to the concentration of the element in the specimen are discussed in the subsequent chapters of this handbook dedicated to specific modes of x-ray spectrometry.
II.
HISTORY
X-rays were discovered in 1895 by Wilhelm Conrad Ro¨ntgen at the University of Wu¨rzburg, Bavaria. He noticed that some crystals of barium platinocyanide, near a discharge tube completely enclosed in black paper, became luminescent when the discharge occurred. By examining the shadows cast by the rays. Ro¨ntgen traced the origin of the rays to the walls of the discharge tube. In 1896, Campbell-Swinton introduced a definite target (platinum) for the cathode rays to hit; this target was called the anticathode. For his work x-rays, Ro¨ntgen received the first Nobel Prize in physics, in 1901. It was the first of six to be awarded in the field of x-rays by 1927. The obvious similarities with light led to the crucial tests of established wave optics: polarization, diffraction, reflection, and refraction. With limited experimental facilities, Ro¨ntgen and his contemporaries could find no evidence of any of these; hence, the designation ‘‘x’’ (unknown) of the rays, generated by the stoppage at anode targets of the cathode rays, identified by Thomson in 1897 as electrons. The nature of x-rays was the subject of much controversy. In 1906, Barkla found evidence in scattering experiments that x-rays could be polarized and must therefore by waves, but W. H. Bragg’s studies of the produced ionization indicated that they were Copyright © 2002 Marcel Dekker, Inc.
corpuscular. The essential wave nature of x-rays was established in 1912 by Laue, Friedrich, and Knipping, who showed that x-rays could be diffracted by a crystal (copper sulfate pentahydrate) that acted as a three-dimensional diffraction grating. W. H. Bragg and W. L. Bragg (father and son) found the law for the selective reflection of x-rays. In 1908, Barkla and Sadler deduced, by scattering experiments, that x-rays contained components characteristic of the material of the target; they called these components K and L radiations. That these radiations had sharply defined wavelengths was shown by the diffraction experiments of W. H. Bragg in 1913. These experiments demonstrated clearly the existence of a line spectrum superimposed upon a continuous (‘‘White’’) spectrum. In 1913, Moseley showed that the wavelengths of the lines were characteristic of the element of the which the target was made and, further, showed that they had the same sequence as the atomic numbers, thus enabling atomic numbers to be determined unambiguously for the first time. The characteristic K absorption was first observed by de Broglie and interpreted by W. L. Bragg and Siegbahn. The effect on x-ray absorption spectra of the chemical state of the absorber was observed by Bergengren in 1920. The influence of the chemical state of the emitter on x-ray emission spectra was observed by Lindh and Lundquist in 1924. The theory of x-ray spectra was worked out by Sommerfeld and others. In 1919, Stenstro¨m found the deviations from Bragg’s law and interpreted them as the effect of refraction. The anomalous dispersion of x-ray was discovered by Larsson in 1929, and the extended fine structure of x-ray absorption spectra was qualitatively interpreted by Kronig in 1932. Soon after the first primary spectra excited by electron beams in an x-ray tube were observed, it was found that secondary fluorescent x-rays were excited in any material irradiated with beams of primary x-rays and that the spectra of these fluorescent x-rays were identical in wavelengths and relative intensities with those excited when the specimen was bombarded with electrons. Beginning in 1932, Hevesy, Coster, and others investigated in detail the possibilities of fluorescent x-ray spectroscopy as a means of qualitative and quantitative elemental analysis. III.
GENERAL FEATURES
X-rays, or Ro¨ntgen rays, are electromagnetic radiations having wavelengths roughly within the range from 0.005 to 10 nm. At the short-wavelength end, they overlap with g-rays, and at the long-wavelength end, they approach ultraviolet radiation. The properties of x-rays, some of which are discussed in detail in this chapter, are summarized as follows: Invisible Propagated in straight lines with a velocity of 36108 m=s, as is light Unaffected by electrical and magnetic fields Differentially absorbed while passing through matter of varying composition, density, or thickness Reflected, diffracted, refracted, and polarized Capable of ionizing gases Capable of affecting electrical properties of liquids and solids Capable of blackening a photographic plate Able to liberate photoelectrons and recoil electrons Capable of producing biological reactions (e.g., to damage or kill living cells and to produce genetic mutations) Copyright © 2002 Marcel Dekker, Inc.
Emitted in a continuous spectrum whose short-wavelength limit is determined only by the voltage on the tube Emitted also with a line spectrum characteristic of the chemical elements Found to have absorption spectra characteristic of the chemical elements
IV.
EMISSION OF CONTINUOUS RADIATION
Continuous x-rays are produced when electrons, or other high-energy charged particles, such as protons or a-particles, lose energy in passing through the Coulomb field of a nucleus. In this interaction, the radiant energy (photons) lost by the electron is called bremsstrahlung (from the German bremsen, to brake, and Strahlung, radiation; this term sometimes designates the interaction itself). The emission of continuous x-rays finds a simple explanation in terms of classic electromagnetic theory, because, according to this theory, the acceleration of charged particles should be accompanied by the emission of radiation. In the case of high-energy electrons striking a target, they must be rapidly decelerated as they penetrate the material of the target, and such a high negative acceleration should produce a pulse of radiation. The continuous x-ray spectrum generated by electrons in an x-ray tube is characterized by a short-wavelength limit lmin , corresponding to the maximum energy of the exciting electrons: lmin ¼
hc eV0
ð1Þ
where h is Planck’s constant, c is the velocity of light, e is the electron charge, and V0 is the potential difference applied to the tube. This relation of the short-wavelength limit to the applied potential is called the Duane–Hunt law. The probability of radiative energy loss (bremsstrahlung) is roughly proportional to q2 Z2 T=M20 , where q is the particle charge in units of the electron charge e, Z is the atomic number of the target material, T is the particle kinetic energy, and M0 is the rest mass of the particle. Because protons and heavier particles have large masses compared to the electron mass, they radiate relatively little; for example, the intensity of continuous x-rays generated by protons is about four orders of magnitude lower than that generated by electrons. The ratio of energy lost by bremsstrahlung to that lost by ionization can be approximated by 2 m0 ZT ð2Þ M0 1600m0 c2 where m0 the rest mass of the electron.
A.
Spectral Distribution
The continuous x-ray spectrum generated by electrons in an x-ray tube (thick target) is characterized by the following features: 1. Short-wavelength limit, lmin [Eq. (1)]; below this wavelength, no radiation is observed. Copyright © 2002 Marcel Dekker, Inc.
2.
3.
Wavelength of maximum intensity lmax , approximately 1.5 times lmin ; however, the relationship between lmax and lmin depends to some extent on voltage, voltage waveform, and atomic number. Total intensity nearly proportional to the square of the voltage and the first power of the atomic number of the target material.
The most complete empirical work on the overall shape of the energy distribution curve for a thick target has been of Kulenkampff (1922, 1933), who found the following formula for the energy distribution; ð3Þ IðvÞ dv ¼ i aZðv0 vÞ þ bZ2 dv where IðnÞ dn is the intensity of the continuous x-rays within a frequency range ðn; n þ dvÞ; i is the electron current striking the target, Z is the atomic number of the target material, n0 is the cutoff frequency ð¼ c=lmin Þ above which the intensity is zero, and a and b are constants independent of atomic number, voltage, and cutoff wavelength. The second term in Eq. (3) is usually small compared to the first and is often neglected. The total integrated intensity at all frequencies is I ¼ iða0 ZV20 þ b0 Z2 V0 Þ
ð4Þ
in which a0 ¼ aðe2 =h2 Þ=2 and b0 ¼ bðe=hÞ. An approximate value for b0 =a0 is 16.3 V; thus, I ¼ a0 iZV0 ðV0 þ 16:3ZÞ
ð5Þ
The efficiency Eff of conversion of electric power input to x-rays of all frequencies is given by Eff ¼
I ¼ a0 ZðV0 þ 16:3ZÞ V0 i
ð6Þ
where V0 is in volts. Experiments give a0 ¼ ð1:2 0:1Þ 109 (Condon, 1958). The most complete and successful efforts to apply quantum theory to explain all features of the continuous x-ray spectrum are those of Kramers (1923) and Wentzel (1924). By using the correspondence principle, Kramers found the following formulas for the energy distribution of the continuous x-rays generated in a thin target: pffiffi AZ e dv; IðvÞ dv ¼ 316p 3m0 V0 c3 IðvÞ dv ¼ 0; 2
2 5
v < v0 v > v0
ð7Þ
where A is the atomic mass of the target material. When the decrease in velocity of the electrons in a thick target was taken into account by applying the Thomson–Whiddington law (Dyson, 1973), Kramers found, for a thick target, 8pe2 h Zðv0 vÞ dv IðvÞ dv ¼ pffiffiffi 3 3lm0 c3
ð8Þ
where l is approximately 6. The efficiency of production of the x-rays calculated via Kramers’ law is given by Eff ¼ 9:2 1010 ZV0
ð9Þ
which is in qualitative agreement with the experiments of Kulenkampff (Stephenson, 1957); for example, Copyright © 2002 Marcel Dekker, Inc.
Eff ¼ 15 1010 ZV0
ð10Þ
It is worth mentioning that the real continuous x-ray distribution is described only approximately by Kramers’ equation. This is related, inter alia, to the fact that the derivation ignores the self-absorption of x-rays and electron backscattering effects. Wentzel (1924) used a different type of correspondence principle than Kramers, and he explained the spatial distribution asymmetry of the continuous x-rays from thin targets. An accurate description of continuous x-rays is crucial in all x-ray spectrometry (XRS). The spectral intensity distributions from x-ray tubes are of great importance for applying fundamental mathematical matrix correction procedures in quantitative x-ray fluorescence (XRF) analysis. A simple equation for the accurate description of the actual continuum distributions from x-ray tubes was proposed by Tertian and Broll (1984). It is based on a modified Kramers’ law and a refined x-ray absorption correction. Also, a strong need to model the spectral Bremsstrahlung background exists in electron-probe x-ray microanalysis (EPXMA). First, fitting a function through the background portion, on which the characteristic x-rays are superimposed in an EPXMA spectrum, is not easy; several experimental fitting routines and mathematical approaches, such as the Simplex method, have been proposed in this context. Second, for bulk multielement specimens, the theoretical prediction of the continuum Bremsstrahlung is not trivial; indeed, it has been known for several years that P the commonly used Kramers’ formula with Z directly substituted by the average Z ¼ i Wi Zi (Wi and Zi are the weight fraction and atomic number of the ith element, respectively) can lead to significant errors. In this context, some improvements are offered by several modified versions of Kramers’ formula developed for a multielement bulk specimen (Statham, 1976; Lifshin, 1976; Sherry and Vander Sande, 1977; Smith and Reed, 1981). Also, a new expression for the continuous x-rays emitted by thick composite specimens was proposed (Markowicz and Van Grieken, 1984; Markowicz et al., 1986); it was derived by introducing the compositional dependence of the continuum x-rays already in the elementary equations. The new expression has been combined with known equations for the self-absorption of x-rays (Ware and Reed, 1973) and electron backscattering (Statham, 1979) to obtain an accurate description of the detected continuum radiation. A third problem is connected with the description of the x-ray continuum generated by electrons in specimens of thickness smaller than the continuum x-ray generation range. This problem arises in the analysis of both thin films and particles by EPXMA. A theoretical model for the shape of the continuous x-rays generated in multielement specimens of finite thickness was developed (Markowicz et al., 1985); both composition and thickness dependence have been considered. Further refinements of the theoretical approach are hampered by the lack of knowledge concerning the shape of the electron interaction volume, the distribution of the electron within the interaction volume, and the anisotropy of continuous radiation for different x-ray energies and for different film thickness. B.
Spatial Distribution and Polarization
The spatial distribution of the continuous x-rays emitted by thin targets has been investigated by Kulenkampff (1928). The author made an extensive survey of the intensity at angles between 22 and 150 to the electron beam in terms of dependence on wavelength and voltage. The target was a 0.6-mm-thick Al foil. Figure 1 shows the continuous x-ray intensity observed at different angles for voltages of 37.8, 31.0, 24.0, and 16.4 kV filtered by 10, 8, 4, and 1.33 mm of Al, respectively (Stephenson, 1957). Curve (a) is repeated as a dotted line near each of the other curves. The angle of the maximum intensity varied from Copyright © 2002 Marcel Dekker, Inc.
Figure1 Intensity of continuous x-rays as a function of direction for different voltages. (Curve (a) is repeated as dotted line.) (From Stephenson, 1957.)
50 for 37.8 kV to 65 for 16.4 kV. Figure 2 illustrates the intensity of the continuous x-rays observed in the Al foil for different thicknesses as a function of the angle for a voltage of 30 kV (Stephenson, 1957). The theoretical curve is from the theory of Scherzer (1932). The continuous x-ray intensity drops to zero at 180 , and although it is not zero at 0 as the theory of Scherzer predicts, it can be seen from Figure 2 that for a thinner foil, a lower intensity at 0 is obtained. Summarizing, it appears that the intensity of the continuous x-rays emitted by thin foils has a maximum at about 55 relative to the incident electron beam and becomes zero at 180 . The continuous radiation from thick targets is characterized by a much smaller anisotropy than that from thin targets. This is because in thick targets the electrons are rarely stopped in one collision and usually their directions have considerable variation. The use of electromagnetic theory predicts a maximum energy at right angles to the incident electron beam at low voltages, with the maximum moving slightly away from perpendicularity toward the direction of the elctron beam as the voltage is increased. In general, an increase in the anisotropy of the continuous x-rays from thick targets is observed at the short-wavelength limit and for low-Z targets (Dyson, 1973). Copyright © 2002 Marcel Dekker, Inc.
Figure 2 Intensity of continuous x-rays as a function of direction for different thicknesses of the A1 target together with theoretical prediction. (From Stephenson, 1957.)
Continuous x-ray beams are partially polarized only from extremely thin targets; the angular region of polarization is sharply peaked about the photon emission angle y ¼ m0 c2 =E0 , where E0 is the energy of the primary electron beam. Electron scattering in the target broadens the peak and shifts the maximum to larger angles. Polarization is defined by (Kenney, 1966) Pðy; E0 ; En Þ ¼
ds?ðy; E0 ; En Þ dskðy; E0 ; En Þ ds?ðy; E0 ; En Þ þ dskðy; E0 ; En Þ
ð11Þ
where an electron of energy E0 radiates a photon of energy En at angle y; ds?ðy; E0 ; En Þ and dskðy; E0 ; En Þ are the cross sections for generation of the continuous radiation with the electric vector perpendicular (?) and parallel (k) to the plane defined by the incident electron and the radiated photon, respectively. Polarization is difficult to observe, and only thin, low-yield radiators give evidence for this effect. When the electron is relativistic before and after the radiation, the electrical vector is most probably in the ? direction. Practical thick-target Bremsstrahlung shows no polarization effects whatever (Dyson, 1973; Stephenson, 1957; Kenney, 1966).
V.
EMISSION OF CHARACTERISTIC X-RAYS
The production of characteristic x-rays involves transitions of the orbital electrons of atoms in the target material between allowed orbits, or energy states, associated with ionization of the inner atomic shells. When an electron is ejected from the K shell by electron bombardment or by the absorption of a photon, the atom becomes ionized and the ion is left in a high-energy state. The excess energy the ion has over the normal state of the atom is equal to the energy (the binding energy) required to remove the K electron to a state of rest outside the atom. If this electron vacancy is filled by an electron coming from an L level, the transition is accompanied by the emission of an x-ray line known as the Ka line. This process leaves a vacancy in the L shell. On the other hand, if the atom contains sufficient electrons, the K shell vacancy might be filled by an electron coming from an M level that is accompanied by the emission of the Kb line. The L or M state ions that remain may also give rise to emission if the electron vacancies are filled by electrons falling from further orbits. Copyright © 2002 Marcel Dekker, Inc.
A.
Inner Atomic Shell Ionization
As already mentioned, the emission of characteristic x-ray is preceded by ionization of inner atomic shells, which can be accomplished either by charged particles (e.g., electrons, protons, and a-particles) or by photons of sufficient energy. The cross section for ionization of an inner atomic shell of element i by electrons is given by (Bethe, 1930; Green and Cosslett, 1961; Wernisch, 1985) Qi ¼ pe4 ns bs
ln U UE2c;i
ð12Þ
where U ¼ E=Ec;i is the overvoltage, defined as the ratio of the instantaneous energy of the electron at each point of the trajectory to that required to ionize an atom of element i, Ec;i is the critical excitation energy, and ns and bs are constants for a particular shell: s ¼ K: s ¼ L:
ns ¼ 2; bs ¼ 0:35 ns ¼ 8; bs ¼ 0:25
The cross section for ionization Qi is a strong function of the overvoltage, which shows a maximum at U ffi 3–4 (Heinrich, 1981; Goldstein et al., 1981). The probability (or cross section) of ionization of an inner atomic shell by a charged particle is given by (Merzbacher and Lewis, 1958) ss ¼
8pr20 q2 fs Z 4 Zs
ð13Þ
where r0 is the classic radius of the electron equal to 2.81861015 m, q is the particle charge, Z is the atomic number of the target material, fs is a factor depending on the wave functions of the electrons for a particular shell, and Zs is a function of the energy of the incident particles. In the case of electromagnetic radiation (x or g), the ionization of an inner atomic shell is a result of the photoelectric effect. This effect involves the disappearance of a radiation photon and the photoelectric ejection of one electron from the absorbing atom, leaving the atom in an excited level. The kinetic energy of the ejected photoelectron is given by the difference between the photon energy hn and the atomic binding energy of the electron Ec (critical excitation energy). Critical absorption wavelengths (Clark, 1963) related to the critical absorption energies (Burr, 1974) via the equation l(nm) ¼ 1.24=E(ke V) are presented in Appendix I. The wavelenghts of K, L, M, and N absorption edges can also be calculated by using simple empirical equations (Norrish and Tao, 1993). For energies far from the absorption edge and in the nonrelativistic range, the cross section tK for the ejection of an electron from the K shell is given by (Heitler, 1954) pffiffiffi 7=2 32 2 2 Z5 m 0 c2 pr0 tK ¼ 3 ð137Þ4 hv
ð14Þ
Equation (14) is not fully adequate in the neighborhood of an absorption edge; in this case, Eq. (14) should be multiplied by a correction factor f(X ) (Stobbe, 1930): 1=2 4X arccot X D e fðXÞ ¼ 2p hv 1 e2pX where Copyright © 2002 Marcel Dekker, Inc.
ð15Þ
X¼
D hv D
1=2 ð15aÞ
with 1 m 0 c2 D ffi ðZ 0:3Þ2 2 ð137Þ2
ð15bÞ
When the energy of the incident photon is of the order m0 c2 or greater, relativistic cross sections for the photoelectric effect must be used (Sauter, 1931).
B.
Spectral Series in X-rays
The energy of an emission line can be calculated as the difference between two terms, each term corresponding to a definite state of the atom. If E1 and E2 are the term values representing the energies of the corresponding levels, the frequency of an x-ray line is given by the relation v¼
E1 E2 h
ð16Þ
Using the common notations, one can represent the energies of the levels E by means of the atomic number and the quantum numbers n, l, s, and j (Sandstro¨m, 1957): ! 2 E ðZ Sn;l Þ2 1 3 2 ðZ dn;l; j Þ ¼ þa Rh n2 n3 l þ 12 4n a2
ðZ dn;l; j Þ4 jð j þ 1Þ lðl þ 1Þ sðs þ 1Þ n3 2lðl þ 12Þðl þ 1Þ
ð17Þ
where Sn;l and dn;l; j are screening constants that must be introduced to correct for the effect of the electrons on the field in the atom, R is the universal Rydberg constant valid for all elements with Z > 5 or throughout nearly the whole x-ray region, and a is the finestructure constant given by a¼
2pe2 hc
ð17aÞ
The theory of x-ray spectra reveals the existence of a limited number of allowed transitions; the rest are ‘‘forbidden.’’ The most intense lines create the electric dipole radiation. The transitions are governed by the selection rules for the change of quantum numbers: Dl ¼ 1;
Dj ¼ 0 or 1
ð18Þ
The j transition 0 ? 0 is forbidden. According to Dirac’s theory of radiation (Dirac, 1947), transitions that are forbidden as dipole radiation can appear as multipole radiation (e.g., as electric quadrupole and magnetic dipole transitions). The selection rules for the former are Dl ¼ 0 or 2;
Dj ¼ 0; 1; or 2
The j transitions 0 ? 0, and 0 $ 1 are forbidden. The selection rules for magnetic dipole transitions are 1 1 2? 2,
Copyright © 2002 Marcel Dekker, Inc.
ð19Þ
Dl ¼ 0;
Dj ¼ 0 or 1
ð20Þ
The j transition 0 ? 0 is forbidden. The commonly used terminology of energy levels and x-ray lines is shown in Figure 3. A general expression relating the wavelength of an x-ray characteristic line with the atomic number of the corresponding element is given by Moseley’s law (Moseley, 1914): 1 ¼ kðZ sÞ2 l
ð21Þ
where k is a constant for a particular spectral series and s is a screening constant for the repulsion correction due to other electrons in the atom. Moseley’s law plays an important role in the systematizing of x-ray spectra. Appendix II tabulates the energies and wavelengths of the principal x-ray emission lines for the K, L, and M series with their approximate relative intensities, which can be defined either by means of spectral line peak intensities or by area below their intensity distribution curve. In practice, the relative
Figure 3
Commonly used terminology of energy levels and x-ray lines. (From Sandstro¨m, 1957.)
Copyright © 2002 Marcel Dekker, Inc.
intensities of spectral lines are not constant because they depend not only on the electron transition probability but also on the specimen composition. Considering the K series, the Ka fraction of the total K spectrum is defined by the transition probability pKa, which is given by (Schreiber and Wims, 1982) pKa ¼
IðKa1 þ Ka2 Þ IðKa1 þ Ka2 Þ þ IðKb1 þ Kb2 Þ
ð22Þ
Wernisch (1985) proposed a useful expression for the calculation of the transition probability pKa for different elements: 8 11 Zi 19 < 1:052 4:39 104 Z2i ; ð23Þ pKa;i ¼ 0:896 6:575 104 Zi ; 20 Zi 29 : 1:0366 6:82 103 Zi þ 4:815 105 Z2i ; 30 Z2i 60 For the L level, split into three subshells, several electron transitions exist. The transition probability pLa, defined as the fraction of the transitions resulting in La1 and La2 radiation from the total of possible transitions into the L3 subshell, can be calculated by the expression (Wernisch, 1985) 8 39 Zi 44 < 0:944; ð24Þ pLa;i ¼ 4:461 101 þ 5:493 102 Zi : 7:717 104 Z2i þ 3:525 106 Z3i ; 45 Zi 82 Radiative transition probabilities for various K and L x-ray lines (West, 1982–83) are presented in detail in Appendix III. The experimental results, together with the estimated 95% confidence limits, for the relative intensity ratios for K and L x-ray lines for selected elements with atomic number from Z ¼ 14 to 92 have been reported by Stoev and Dlouhy (1994). The values are in a good agreement with other published experimental data. Because the electron vacancy created in an atom by charged particles or electromagnetic radiation has a certain lifetime, the atomic levels E1 and E2 [Eq. (16)] and the characteristic x-ray lines are characterized by the so-called natural widths (Krause and Oliver, 1979). The natural x-ray linewidths are the sums of the widths of the atomic levels involved in the transitions. Semiempirical values of the natural widths of K, L1 , L2 and L3 levels, Ka1 and Ka2 x-ray lines for the elements 10 Z 110 are presented in Appendix IV. Uncertainties in the level width values are from 3% to 10% for the K shell and from 8% to 30% for the L subshell. Uncertainties in the natural x-ray linewidth values are from 3% to 10% for Ka1;2 . In both cases, the largest uncertainties are for low-Z elements (Krause and Oliver, 1979).
C.
X-ray Satellites
A large number of x-ray lines have been reported that do not fit into the normal energylevel diagram (Clark, 1955; Kawai and Gohshi, 1986). Most of the x-ray lines, called satellites or nondiagram lines, are very weak and are of rather little consequence in analytical x-ray spectrometry. By analogy to the satellites in optical spectra, it was supposed that the origin of the nondiagram x-ray lines is double or manyfold ionization of an atom through electron impact. Following the ionization, a multiple electron transition results in emission of a single photon of energy higher than that of the characteristic x-rays. The majority of nondiagram lines originate from the dipole-allowed deexcitation of multiply Copyright © 2002 Marcel Dekker, Inc.
ionized or excited states and are called multiple-ionization satellites. A line where the initial state has two vacancies in the same shell, notably the K shell, is called a hypersatellite. In practice, the most important nondiagram x-ray lines occur in the Ka series; they are denoted as the Ka3 ;a4 doublet, and their origin is a double electron transition. The probability of a multiple-electron transition resulting in the emission of satellite x-ray lines is considerably higher for low-Z elements than for heavy and medium elements. For instance, the intensity of the AlKa3 satellite line is roughly 10% of that of the AlKa1 ; a2 characteristic x-rays. Appendix V tabulates wavelengths of the K satellite lines. A new class of satellites that are inside the natural width of the parent lines was observed by Kawai and Gohshi (1986). The origin of these satellites, called parasites or hidden satellites, is multiple ionization in nonadjacent shells. D.
Soft X-ray Emission-Band Spectra
In the soft x-ray region, the characteristic emission spectra of solid elements include continuous bands of width varying from 1 to 10 electron volts (eV); the same element in vapor form produces only the usual sharp spectral lines. The bands occur only when an electron falls from the outermost or valency shell of the atom, the levels of which are broadened into a wide band when the atoms are packed in a crystal lattice. Investigation of the emission-band spectra is of great significance in understanding the electronic structure of solid metals, alloys, and complex coordination compounds. E.
Auger Effect
It has already been stated that the excess of energy an atom possesses after removing one electron from an inner shell by whatever means may be emitted as characteristic radiation. Alternatively, however, an excited atom may return to a state of lower energy by ejecting one of its own electrons from a less tightly bound state. The radiationless transition is called the Auger effect, and the ejected electrons are called Auger electrons (Auger, 1925; Burhop, 1952). Generally, the probability of the Auger effect increases with a decrease in the difference of the corresponding energy states, and it is the highest for the low-Z elements. Because an excited atom already has one electron missing (e.g., in the K shell) and another electron is ejected in an Auger process (e.g., from the L shell), the atom is left in a doubly-ionized state in which two electrons are missing. This atom may return to its normal state by single- or double-electron jumps with the emission of diagram or satellite lines, respectively. Alternatively, another Auger process may occur in which a third electron is ejected from the M shell. The Auger effect also occurs after capture of a negative meson by an atom. As the meson changes energy levels in approaching the nucleus, the energy released may be either emitted as a photon or transferred directly to an electron that is emitted as a high-energy Auger electron (in the keV range for hydrogen and the MeV range for heavy elements). Measurements of the energy and intensity of the Auger electrons are applied extensively in surface physics studies (Auger electron spectroscopy). F.
FluorescenceYield
An important consequence of the Auger effect is that the actual number of x-ray photons produced from an atom is less than expected, because a vacancy in a given shell might be Copyright © 2002 Marcel Dekker, Inc.
filled through a nonradiative transition. The probability that a vacancy in an atomic shell or subshell is filled through a radiative transition is called the fluorescence yield. The application of this definition to the K shell of an atom is straightforward, and the fluorescence yield of the K shell is oK ¼
IK nK
ð25Þ
where IK is the total number of characteristic K x-ray photons emitted from a sample and nK is the number of primary K shell vacancies. The definition of the fluorescence yield of higher atomic shells is more complicated, for the following two reasons: 1. Shells above the K shell consist of more than one subshell; the average fluorescence yield depends on how the shells are ionized. 2. Coster–Kronig transitions occur, which are nonradiative transitions between the subshells of an atomic shell having the same principal quantum number (Fink, 1974; Bambynek et al., 1972). In case Coster–Kronig transitions are absent, the fluorescence yield of the ith subshell of a shell, whose principal quantum number is indicated by XðX ¼ L; M; . . .Þ, is given as oX i ¼
IX i nX i
ð26Þ
X for the shell X is defined as An average or mean fluorescence yield o X ¼ o
k X
ð27Þ
X NX i oi
i¼1
where NX i is the relative number of primary vacancies in the subshell i of shell X: nX i NX i ¼ Pk
X i¼1 ni
;
k X
NX i ¼1
ð28Þ
i¼1
The summations in Eqs. (27) and (28) extend over all k subshells of shell X. For the definition of the average fluorescence yield, the primary vacancy distribution must be fixed; X generally is not that is, Coster–Kronig transitions must be absent. It is noteworthy that o a fundamental property of the atom, but depends both on the atomic subshell fluorescence X yields oX i and on the relative number of primary vacancies Ni characteristic of the method used to ionize the atoms. In the presence of Coster–Kronig transitions, which modify the primary vacancy distribution by the transfer of ionization from one subshell with a given energy to a subshell with less energy, the average fluorescence yields can be calculated by using two X is regarded as a alternative approaches. In the first, the average fluorescence yield o with a vacancy distribution linear combination of the subshell fluorescence yields oX i modified by Coster–Kronig transitions: X ¼ o
k X i¼1
X VX i oi ;
k X i¼1
Copyright © 2002 Marcel Dekker, Inc.
VX i >1
ð29Þ
where VX i is the relative number of vacancies in the subshell i of shell X, including vacancies shifted to each subshell by Coster–Kronig transitions. The VX i values can be expressed in terms of the relative numbers NX i of primary vacancies and the Coster–Kronig transition probability for shifting a vacancy from a subshell Xi to a higher subshell Xj , denoted as f X ij (Bambynek et al., 1972): X VX 1 ¼ N1
ð30Þ
X X X VX 2 ¼ N2 þ f 12 N1
VX 3
¼
NX 3
þ
X fX 23 N2
þ
ðfX 13
þ
fX 12
X fX 23 ÞN1
X is a linear combination of In an alternative approach, the mean fluorescence yield o : the relative numbers of primary vacancies NX i X ¼ o
k X
X NX i ni
ð31Þ
i¼1
where nX i represents the total number of characteristic x-rays that result per primary vacancy in the Xi subshell. The transformation relations between the coefficients nX i and the subshell fluorescence yields oX i follow from Eqs. (29) through (31) and are given in Fink (1974) and Bambynek et al. (1972). X Among the fluorescence yield oX i , the Auger yield ai , and the Coster–Kronig X transition probabilities fij , the following relationship must hold (Krause, 1979): X oX i þ ai þ
k X
fX ij ¼ 1
ð32Þ
i¼1
The mean Auger yield aX is given by ax ¼
k X
X VX i ai
ð33Þ
i¼1
The values of the K, L, and M shell fluorescence yields, the Coster–Kronig transition probabilities, as well as the Auger yields are given in Appendix VI. Although, in principle, the K shell fluorescence yield ok can be calculated theoretically, experimental data are applied in practice. The following semiempirical equation, due to Burhop (1952), gives values correct to a few percent between Z ¼ 23 and Z ¼ 57 and less accurate values outside these limits: 1=4 oK ¼ 6:4 102 þ 3:40 102 Z 1:03 106 Z3 ð34Þ 1 oK The fluorescence yield for the K series can also be calculated from a different equation: 1=4 oK ¼ 0:217 þ 0:03318Z 1:14 106 Z3 ð35Þ 1 oK which gives quite good agreement with the experimental values for almost all elements. Based on a critical review of the available experimental and theoretical data, a recom L ; and o M values together with the following analytical expressions mended set of oK ; o were given by Hubbell (1989) and Hubbell et al. (1994): Copyright © 2002 Marcel Dekker, Inc.
" oK ð1 Z 100Þ ¼
3 X
Ci Zi
i¼0
#4 8 < :
" 1þ
3 X i¼0
Ci Zi
#4 91 = ;
ð36Þ
with C0 C1 C2 C3
¼ 0:0370 0:0052 ¼ 0:03112 0:00044 ¼ ð5:44 0:11Þ 105 ¼ ð1:25 0:07Þ 106
L ð3 Z 36Þ ¼ 1:939 108 Z3:8874 o " #4 8 " #4 91 < = 3 3 X X L ð37 Z 100Þ ¼ Ci Zi 1þ Ci Zi o : ; i¼0 i¼0
ð37Þ
with C0 C1 C2 C3
¼ 0:17765 ¼ 0:00298937 ¼ 8:91297 105 ¼ 2:67184 107
M ð13 Z 100Þ ¼ 1:29 109 ðZ 13Þ4 o
ð38Þ
Other useful expressions for the calculation of the fluorescence yields oK ð12 Z 42Þ and oL3 ð38 Z 79Þ have been proposed by Hanke et al. (1985), based on literature and experimental data: oK ¼ 3:3704 101 6:0047 102 Z þ 3:3133 103 Z2 3:9215 105 Z3 oL3 ¼ 4:41 102 4:7559 103 Z þ 1:1494 104 Z2 1:8594 107 Z3
ð39Þ ð40Þ
N data is a theoretical work of McGuire (1974) For the N shell, the best source of o which provides oN1 ; oN2 , and oN3 values for 25 elements over the range 38 Z 103 and oN4 ; oN5 , and oN6;7 values for 20 elements over the range 50 Z 103. N can be calculated from (Hubbell, 1989) The average fluorescence yield o N ¼ o
7 X 1 NN oN 32 i i i¼1
ð41Þ
where NNi are the numbers of electrons in each Ni subshell. A comparison of the total x-ray yields for bulk samples (including both the probability of ionization and the fluorescence yield) in terms of photons per steradian per incident quantum for electrons, protons, and x-ray photons is shown in Figure 4. Copyright © 2002 Marcel Dekker, Inc.
Figure 4 Total x-ray yields for excitation by electrons, protons, and primary x-ray photons as a function of energy of the exciting quantum. (From Birks, 1971a.)
G.
Fine Features of X-ray Emission Spectra (Valence or Chemical Effects)
Because characteristic x-ray emission is a process in which the innermost electrons in the atom are concerned, it is reasonable to suppose that the external, or valence, electrons have little or no effect on the x-ray emission lines. However, this is not fully true for K lines of low-Z elements and L or M lines of higher-Z elements, where the physical state and chemical combination of the elements affect the characteristic x-rays (Clark, 1955). The changes in fine features of x-ray emission spectra with chemical combination can be classified into three groups: (1) shifting in wavelength (Kallithrakas-Kontos, 1996), (2) distortion of line shape, and (3) intensity changes (Kawai et al., 1993; Rebohle et al., 1996). Wavelength shifts to both longer and shorter wavelengths result from energy-level changes due to electrical shielding or screening of the electrons when the valence electrons are drawn into a bond. Generally, the so-called last or highest-energy member of a given series is most affected by chemical combination; maximum energy shifts are of the order of a few electron volts. Distortion of an x-ray emission line shape gives some indication of the energy distribution of the electrons occupying positions in or near the valence shell. The changes in the characteristic x-ray intensity are a result of alterations in excitation Copyright © 2002 Marcel Dekker, Inc.
probabilities of the electrons undergoing transitions. Certain x-ray lines or bands appear or disappear with chemical combinations. In the case of the K series, the most noticeable chemical effects on x-ray emission are seen in spectra from low-Z elements (4 Z 17). The L series shows as large or even larger changes with chemical combination of the elements than K series. The valence effects in L spectra have been observed for elements of the first transition series and others nearby in the periodic table. Because the fine features of x-ray emission spectra may be applied to determine how each element is chemically combined in the sample (speciation), the valence effects found numerous applications in such fields as physics of solids and surface or near-surface characterization. VI.
INTERACTION OF PHOTONS WITH MATTER
Interactions of photons with matter, by which individual photons are removed or deflected from a primary beam of x or g radiation, may be classified according to the following: The kind of target, such as electrons, atoms or nuclei, with which the photon interacts The type of event, such as absorption, scattering, or pair production, that takes place These interactions are thought to be independent of the origin of the photon (nuclear transition for g-rays versus electronic transition for x-rays); hence, we use the term ‘‘photon’’ to refer to both g- and x-rays here. Possible interactions are summarized in Table 1 (Hubbell, 1969), where t is the total photoelectric absorption cross section per atom (t ¼ tK þ tL þ ) and sR and sC are Rayleigh and Compton collision cross sections, respectively. The probability of each of these many competing independent processes can be expressed as a collision cross section per atom, per electron, or per nucleus in the absorber. The sum of all these cross sections, normalized to a per atom basis, is then the probability stot that the incident photon will have an interaction of some kind while passing through a very thin absorber that contains one atom per square centimeter of area normal to the path of the incident photon: stot ¼ t þ sR þ sC þ
ð42Þ
The total collision cross section per atom stot , when multiplied by the number of atoms per cubic centimeter of absorber, is then the linear attenuation coefficient m per centimeter of travel in the absorber: 1 cm2 g N0 atoms m r ¼ stot ð43Þ cm cm3 g atom A where r is the density of the medium and N0 is Avogadro’s number (6.0225261023 atoms=g atom). The mass attenuation coefficient m (cm2=g) is the ratio of the linear attenuation coefficient and the density of the material. It is worth mentioning that the absorption coefficient is a much more restricted concept than the attenuation coefficient. Attenuation includes the purely elastic process in which the photon is merely deflected and does not give up any of its initial energy to the absorber; in this process, only a scattering coefficient is involved. In a photoelectric interaction, the entire energy of the incident photon is absorbed by an atom of the medium. Copyright © 2002 Marcel Dekker, Inc.
Table 1
Classification of Photon Interactions Scattering
Type of interaction Interaction with atomic electrons
Interaction with nucleus or bound nucleons
Interaction with electrical field surrounding charged particles
Interactions with mesons
Absorption
Elastic (coherent)
Inelastic (incoherent)
Multiphoton effects
Photoelectric effecta Z4 low energy t Z5 high energy Nuclear photoelectric effect: reactions (g, n) (g, p), photofission Z (E 10 MeV) 1. Electron–positron pair production in field of nucleous, Z2 (E 1.02 MeV) 2. Electron–positron pair production in electron field Z (E 2.04 MeV) 3. Nucleon–antinucleon pair production (E 3 GeV) Photomeson production (E 150 MeV)
Rayleigh scatteringa sR Z2
Compton scatteringa sC Z
Two-photon Compton scattering Z
Nuclear coherent scattering (g, g) Z2
Nuclear Compton scattering (g, g0 ) Z
Delbru¨ck scattering Z4
Coherent resonant scattering (g, g)
a Major effects of photon attenuation in matter, which are of great importance in practical x-ray spectrometry. Source: From Hubbell, 1969.
Copyright © 2002 Marcel Dekker, Inc.
In the Compton effect, some energy is absorbed and appears in the medium as the kinetic energy of a Compton recoil electron; the balance of the incident energy is not absorbed and is present as a Compton-scattered photon. Absorption, then, involves the conversion of incident photon energy into the kinetic energy of a charged particle (usually an electron), and scattering involves the deflection of incident photon energy. For narrow, parallel, and monochromatic beams, the attenuation of photons in homogeneous matter is described by the exponential law: I ¼ I0 em
t
ð44Þ
where I is the transmitted intensity, I0 is the incident intensity, and t is the absorber thickness in centimeters. If the absorber is a chemical compound or a mixture, its mass attenuation coefficient can be approximately evaluated from the coefficients mi for the constituent elements according to the weighted average: m¼
n X
Wi mi
ð45Þ
i¼1
where Wi is the weight fraction of the ith element and n is the total number of the elements in the absorber. The ‘‘mixture rule’’ [Eq. (45)] ignores changes in the atomic wave function resulting from changes in the molecular, chemical, or crystalline environment of an atom. Above 10 keV, errors from this approximation are expected to be less than a few percent (except in the regions just above absorption edges), but at very low energies (10–100 eV), errors of a factor of 2 can occur (Deslattes, 1969). For situations more complicated than the narrow-beam geometry, the attenuation is still basically exponential, but it is modified by two additional factors. The first of these, sometimes called a geometry factor, depends on the source absorber geometry. The other factor, often called the buildup factor, takes into account secondary photons produced in the absorber, mainly as the result of one or more Compton scatters, which finally reach the detector. The determination of the buildup factor, defined as the ratio of the observed effect to the effect produced only by the primary radiation, constitutes a large part of g-ray transport theory (Evans, 1963). In subsequent sections, only major effects of photon attenuation are discussed in detail.
A.
Photoelectric Absorption
In the photoelectric absorption described partially in Sec. V.A, a photon disappears and an electron is ejected from an atom. The K shell electrons, which are the most tightly bound, are the most important for this effect in the energy region considered in XRS. If the photon energy drops below the binding energy of a given shell, however, an electron from that shell cannot be ejected. Hence, a plot of t versus photon energy exhibits the characteristic ‘‘absorption edges.’’ The mass photoelectric absorption coefficient tN0 =A at the incident energy E (keV) can approximately be calculated based on Walter’s equations (Compton and Allison, 1935): Copyright © 2002 Marcel Dekker, Inc.
8 30:3Z3:94 > < AE3
tN0 4:30 ¼ 0:978Z AE3 > A : 0:78Z3:94 AE3
for E > EK for EL1 < E < EK for EM1 < E
ð46Þ
Based on available experimental and theoretical information for approximately 10,000 combinations of Z and E covering 87 elements and the energy range 1 keV to 1 MeV, the following ln–ln polynomials for the photoeffect cross section tj have been fitted in incident photon energy between each absorption-edge region (Hubbell et al., 1974): ln tj ¼
1;2;or P3
Aij fln½E ðkeVÞgi
ð47Þ
i¼0
In this polynomial, the total photoeffect cross section tj represents one of the following sums: t1 ¼ tM þ tN þ tO þ
or
tN þ tO þ ;
EM1 < E < EL3 or E < EM5 t2 ¼ tL þ tM þ tN þ tO þ ; EL1 < E < EK t3 ¼ tK þ tL þ tM þ tN þ tO þ ;
ð48Þ
E > EK
The values of the fitted coefficients Aij for the ln–ln representation are given in Appendix VII (McMaster et al., 1969). In multiple-edge regions (e.g., between L1 and L3 edge energies), the photoelectric absorption cross sections are also obtained via Eq. (47) by using the following constant ‘‘jump ratios’’ j (t just above an absorption edge divided by t just below that absorption edge): j ¼ 1:16 j ¼ 1:64 ¼ 1:16 1:41 j ¼ 1:1 j ¼ 1:21 ¼ 1:1 1:1 j ¼ 1:45 ¼ 1:1 1:1 1:2 j ¼ 2:18 ¼ 1:1 1:1 1:2 1:5
for for for for for for
EL2 < E < EL1 EL3 < E < EL2 EM2 < E < EM1 EM3 < E < EM2 EM4 < E < EM3 EM5 < E < EM4
ð49Þ
Simple expressions for calculating the values of the energies of all photoabsorption edges are given in Sec. VI.D. The experimental ratio of the total photoelectric absorption cross section t to the K shell component tK can be fitted with an accuracy of 2–3% by the equation (Hubbell, 1969) t ffi 1 þ 0:01481 ln2 Z 0:000788 ln3 Z tK
ð50Þ
Based on the tables of McMaster et al. (1969), Poehn et al. (1985) found a useful approximation for the calculation of the jump ratios (called also jump factors) for the K shell ( jK ) and L3 subshell ( jL3 ): jK ¼ 1:754 10 6:608 101 Z þ 1:427 102 Z2 1:1 104 Z3
for 11 Z 50
jL3 ¼ 2:003 10 7:732 101 Z þ 1:159 102 Z2 5:835 105 Z3 Copyright © 2002 Marcel Dekker, Inc.
for 30 Z 83
ð51Þ
As already mentioned [Eq. (49)], the values of the jump factors at the L2 and L1 absorption edges are constant for all elements and equal to 1.41 and 1.16, respectively. Tabulated values for the photoelectric absorption cross sections for the elements 1 Z 100 in the energy range of 1 keV to 100 MeV are also available in the work of Storm and Israel (1970), which provides the photon cross sections for all major interaction processes as well as the atomic energy levels, K and L x-ray line energies, weighted average energies for the K and L x-ray series, and relative intensities for K and L x-ray lines. When the apparently sharp x-ray absorption discontinuities are examined at high resolution, they are found to contain a fine structure that extends in some cases to about a few hundred electron volts above the absorption edge. The fine structure very close to an absorption edge (less than or equal to 50 eV above the edge) is generally referred to as the Kossel structure and is designated as XANES (x-ray absorption near-edge structure). Peaks and trenches in this region, which can differ by a factor of 2 or more from the smoothly extrapolated data, can be described in terms of transitions of the (very low energy) ejected electrons to unfilled discrete energy states of the atom (or molecule), rather than to the continuum of states beyond a characteristic energy (Sandstro¨m, 1957; Koningsberger and Prins, 1988; Behrens, 1992b). Superimposed on the Kossel structure is the so-called Kronig structure [extended x-ray absorption fine structure (EXAFS)], which usually extends to about 300 eV above the absorption edge (occasionally to nearly 1 keV above an edge). The Kronig structure can be described in terms of interference effects on the de Broglie waves of the ejected electrons by the molecular or crystalline spatial ordering of neighboring atoms (Hasnain, 1991; Behrens, l992a). The oscillations of the absorption coefficient are of the order of 50% in the energy region 50–60 eV above an absorption edge and of the order of 15% in the region beyond 200 eV above the edge. Modulations of the absorption coefficient in the energy region above an absorption edge can be described theoretically in terms of the electronic parameters (Lee and Pendry, 1975). Through a Fourier transform relationship, the modulations are closely related to the radial distribution function around the element of interest (Sayers et al., 1970). Because both the Kossel and the Kronig fine structures can vary in magnitude and in energy displacement of the features, depending on the molecular, crystalline, or thermal environment of the atom, they can be applied for local structural analysis of various materials, including powders, disordered solids, and liquid and amorphous substances (Lagarde, 1983, Behrens, 1992a, 1992b; Koningsberger and Prins, 1988).
B.
Compton Scattering
Compton scattering (Compton, 1923a, 1923b) is the interaction of a photon with a free electron that is considered to be at rest. The weak binding of electrons to atoms may be neglected, provided the momentum transferred to the electron greatly exceeds the momentum of the electron in the bound state. Considering the conservation of momentum and energy leads to the following equations: hn ¼
hn0 1 þ gð1 cos yÞ
gð1 cos yÞ T ¼ hn0 hn ¼ hn0 1 þ gð1 cos yÞ 1 y tan j ¼ cot 1þg 2 Copyright © 2002 Marcel Dekker, Inc.
ð52Þ ð53Þ ð54Þ
with g¼
hn0 m 0 c2
where hn0 and hn are the energies of the incident and scattered photon, respectively, y is the angle between the photon directions of travel before and following a scattering interaction, and T and f are the kinetic energy and scattering angle of the Compton recoil electron, respectively. For f ¼ 180 , Eqs. (52) and (53) reduce to ðhnÞmin ¼ hn0
1 1 þ 2g
ð55Þ
and Tmax ¼ hn0
2g 1 þ 2g
ð56Þ
The differential Klein–Nishina collision cross section dsKN =dO (defined as the ratio of the number of photons scattered in a particular direction to the number of incident photons) for unpolarized photons striking unbound, randomly oriented electrons is given by (Klein and Nishina, 1929) dsKN r20 hn 2 hn0 hn ¼ þ sin2 y dO 2 hn0 hn hn0
cm2 electron sr
ð57Þ
where sr is an abbreviation for steradian. Substitution of Eq. (52) for Eq. (57) gives the differential cross section as a function of the scattering angle y: dsKN r20 1 þ cos2 y ¼ dO 2 ½1 þ gð1 cos yÞ2 ( ) g2 ð1 cos yÞ2 1þ ð1 þ cos2 yÞ½1 þ gð1 cos yÞ
cm2 electron sr
ð58Þ
For very small energies hn0 m0 c2 , the expression reduces to the classic Thompson scattering cross section for electromagnetic radiation on an electron: dsTh r20 ¼ ð1 þ cos2 yÞ dO 2
cm2 electron sr
ð59Þ
For low energies of incident photons (approximately less than a few tens of a kiloelectron volt, the angular distribution of Compton-scattered photons is symmetrical about y ¼ 90 ; at higher incident photons energies, the Compton scattering becomes predominantly forward. The differential Klein–Nishina scattering cross section dsSKN =dO for unpolarized radiation, defined as the ratio of the amount of energy scattered in a particular direction to the energy of incident photons, is given by dssKN hn dsKN ¼ hn0 dO dO
cm2 electron sr
Copyright © 2002 Marcel Dekker, Inc.
ð60Þ
The average (or total) collision cross section sKN gives the probability of any Compton interaction by one photon while passing normally through a material containing one electron per square centimeter: Zp
dsKN 2p sin y dy dO 0 ( ) lnð1 þ 2gÞ lnð1 þ 2gÞ 1 þ 3g 2 1 þ g 2ðg þ 1Þ ¼ 2pr0 þ g2 1 þ 2g g 2g ð1 þ 2gÞ2
sKN ¼
cm2 electron ð61Þ
Again, at the low-energy limit, this cross section reduces to the classic Thomson cross section: 8 sTh ¼ pr20 ¼ 0:6652 1024 3
cm2 electron
ð62Þ
At extremely high energies hn0 m0 c2 , Eq. (61) reduces to 1 1 cm2 ln 2g þ sKN ¼ pr20 electron g 2
ð63Þ
The average (or total) scattering cross section, defined as the total scattered energy in photons of various energies hn, is given by Zp
ssKN
dssKN 2p sin y dy dO 0 " # 2ð1 þ gÞð2g2 2g 1Þ 8g2 2 lnð1 þ 2gÞ ¼ pr0 þ þ g3 3ð1 þ 2gÞ3 g2 ð1 þ 2gÞ2
¼
cm2 electron
ð64Þ
The usual Klein–Nishina theory that assumes that the target electron is free and at rest cannot be directly applicable in some cases. Departures from it occur at low energies because of electron-binding effects and, at high energies, because of the possibility of emission of an additional photon (double Compton effect) and radiative corrections associated with emission and reabsorption of virtual photons; these corrections are discussed in the work of Hubbell (1969). The total incoherent (Compton) collision cross section per atom sC, involving the binding corrections by applying the so-called incoherent scattering function S(x, Z ), can be calculated according to Z1 ( 1 2 ½1 þ gð1 cos yÞ2 sC ¼ r0 2 1 " # ) g2 ð1 cos yÞ2 cm2 2 1 þ cos y þ ð65Þ ZSðx; ZÞ 2pdðcos yÞ 1 þ gð1 cos yÞ atom where x ¼ sinðy=2Þ=l is the momentum transfer parameter and l is the photon wavelength (in angstroms). Copyright © 2002 Marcel Dekker, Inc.
The values of the incoherent scattering function S(x, Z ) and the incoherent collision cross section sC are given by Hubbell et al. (1975). A useful combination of analytical functions for calculating S(x, Z ) has recently been proposed by Szalo´ki (1996): 8 3 X > < s1 ðx; ZÞ ¼ di ½expðxÞ 1i ; 0 x xi Sðx; ZÞ ¼ i¼1 > : s2 ðx; ZÞ ¼ ½Z s1 ðx; ZÞ t2 g1 ðxÞ þ t2 g2 ðxÞ þ s1 ðx1 ; ZÞ x1 < x ð66Þ where g1 ðxÞ ¼ 1 exp½t1 ðx1 xÞ and g2 ðxÞ ¼ 1 exp½t3 ðx1 xÞ The parameters for the calculation of the S(x, Z ), including the critical value of x1, are given by Szalo´ki (1996) for all elements (Z ¼ 1–100) and any values of x. The average deviation between the calculated [Eq. (66)] and tabulated data (Hubbell et al., 1975) is slightly above 1%. The incoherent collision cross sections sC can also be calculated by using ln–ln polynomials already defined by Eq. (47) (by simply substituting tj with sC and taking i ¼ 3). The values of the fitted coefficients for the ln–ln representation for sC valid in the photon energy range 1 keV to 1 MeV are given in Appendix VIII. To complete this subsection, it is worth mentioning the Compton effect for polarized radiation. The differential collision cross section ðdsKN =dOÞpp for the plane-polarized radiation scattered by unoriented electrons has also been derived by Klein and Nishina. It represents the probability that a photon, passing through a target containing one electron per square centimeter, will be scattered at an angle y into a solid angle dO in a plane making an angle b with respect to the plane containing the electrical vector of the incident wave: dsKN r20 hn 2 hn0 hn 2 2 þ ¼ 2 sin y cos b dO pp 2 hn0 hn hn0
cm2 electron sr
ð67Þ
The cross section has its maximum value for b ¼ 90 , indicating that the photon and electron tend to be scattered at right angles to the electrical vector of the incident radiation. The scattering of circularly polarized (cp) photons by electrons with spins aligned in the direction of the incident photon is described by 2 dsKN 2 hn ¼ r0 dO cp hn0 hn0 hn hn0 hn cm2 þ ð68Þ sin2 y cos y hn hn0 hn hn0 electron sr The first term is the usual Klein–Nishina formula for unpolarized radiation. The þ sign for the additional term applies to right circularly polarized photons. Copyright © 2002 Marcel Dekker, Inc.
C.
Rayleigh Scattering
Rayleigh scattering is a process by which photons are scattered by bound atomic electrons and in which the atom is neither ionized nor excited. The incident photons are scattered with unchanged frequency and with a definite phase relation between the incoming and scattered waves. The intensity of the radiation scattered by an atom is determined by summing the amplitudes of the radiation coherently scattered by each of the electrons bound in the atom. It should be emphasized that, in Rayleigh scattering, the coherence extends only over the Z electrons of individual atoms. The interference is always constructive, provided the phase change over the diameter of the atom is less than one-half a wavelength; that is, whenever 4p y ra sin <1 l 2
ð69Þ
where ra is the effective radius of the atom. Rayleigh scattering occurs mostly at the low energies and for high-Z materials, in the same region where electron binding effects influence the Compton scattering cross section. The differential Rayleigh scattering cross section for unpolarized photons is given by (Pirenne, 1946) dsR 1 2 ¼ r0 ð1 þ cos2 yÞ j Fðx; ZÞj2 2 dO
cm2 atom sr
ð70Þ
where Fðx; ZÞ is the ‘‘atomic form factor,’’ Z1 Fðx; ZÞ ¼
rðrÞ4pr
sin½ð2p=lÞrs dr ð2p=lÞrs
ð71Þ
0
where rðrÞ is the total density, r is the distance from the nucleus, and s ¼ 2 sinðy=2Þ. The atomic form factor has been calculated for Z < 26 using the Hartree electronic distribution (Pirenne, 1946) and for Z > 26 using the Fermi–Thomas distribution (Compton and Allison, 1935). At high photon energies, Rayleigh scattering is confined to small angles; at low energies, particularly for high-Z materials, the angular distribution of the Rayleigh-scattered radiation is much broader. A useful simple criterion for judging the angular spread of Rayleigh scattering is given by (Evans, 1958).
0:0133Z1=3 yR ¼ 2 arcsin E ðMevÞ
ð72Þ
where yR is the opening half-angle of a cone containing at least 75% of the Rayleighscattered photons. In the forward direction, jFðx; ZÞj2 ¼ Z2 , so that Rayleigh scattering becomes appreciable in magnitude and must be accounted for in any g- or x-ray scattering experiments. The total coherent (Rayleigh) scattering cross section per atom sR can be calculated from Copyright © 2002 Marcel Dekker, Inc.
1 sR ¼ r20 2
Z1 ð1 þ cos2 yÞjFðx; ZÞj2 2p dðcos yÞ 1
3 ¼ sTh 8
ð73Þ
Z1 ð1 þ cos2 yÞjFðx; ZÞj2 dðcos yÞ 1
2
cm atom
The values of the atomic form factor F(x, Z ) and the coherent scattering cross section sR are given in the work of (Hubbell et al., 1975). Recently, Szalo´ki (1996) proposed a useful combination of analytical functions to calculate F(x, Z): 8 0 x x1 f11 ðx; ZÞ ¼ a expðb1 xÞ þ ðZ aÞ expðcxÞ; > > > f ðx; ZÞ ¼ f ðx ; ZÞ expðb ðx xÞ; < x 1 x x2 12 11 1 2 1 ð74Þ F1 ðx; ZÞ ¼ f13 ðx; ZÞ ¼ f12 ðx2 ; ZÞ exp½b3 ðx2 xÞ; x 2 x x3 > h ib4 > > : f14 ðx; ZÞ ¼ f13 ðx3 ; ZÞ x ; x 3 x x4 x3 where 1 Z 7 8 < f21 ðx; ZÞ ¼ a expðb1 xÞ þ ðZ aÞ expðcxÞ; 0 x x1 x 1 x x2 F2 ðx; ZÞ ¼ f22 ðx; ZÞ ¼ f21 ðx1 ; ZÞ exp½b2 ðx1 xÞ; : f23 ðx; ZÞ ¼ f22 ðx2 ; ZÞ exp½b3 ðx2 xÞ; x 2 x x3
ð75Þ
where 8 Z 100 The parameters for the calculation of F(x, Z ), including the critical values of x1, x2, and x3, are given by Szalo´ki (1996) for all elements (Z ¼ 1–100) and the momentum transfer x from 0 to 15 A˚71. The average deviation between the calculated [Eqs. (74) and (75)] and tabulated data (Hubbell et al., 1975) is less than 2%. The simplest method for calculating the coherent scattering cross section sR consists in applying the ln–ln representation [see Eq. (47) with sR instead of tj and i ¼ 3]. The values of the fitted coefficients for ln–ln polynomials for calculating sR in the photon energy range 1 keV to 1 MeV are given in Appendix IX.
D. Total Mass Attenuation Coefficient An extensive review of current tabulations of x-ray attenuation coefficients has been given by Hubbell (1984). Differences between various compilations of total mass attenuation coefficients result from uncertainties in our knowledge of partial cross sections for the interaction of photons with matter as a function of elemental atomic number Z and photon energy E. Present discrepancies are disturbing, to say the least, frequently amounting to 5–10% in the photon energy region below 10 keV and rising to as much as 30% near an absorption edge. Hubbell (1982) has tabulated mass attenuation coefficients and mass energy absorption coefficients for photon energies from 1 keV to 20 MeV for 40 elements ranging from hydrogen (Z ¼ 1) to uranium (Z ¼ 92) and for 45 mixtures and compounds of dosimetric interest. The uncertainty ranges for the total mass attenuation coefficient values in the tabulation of McMaster et al. (1969) have been estimated by Hubbell et al. (1974). These ranges of uncertainties fall into four categories. Category I (uncertainty below 2%) applies over the energy region 6–40 keV (except near absorption edges) for the following elements: C, Copyright © 2002 Marcel Dekker, Inc.
Mg, Al, Ti, Fe, Ni, Cu, Zn, Zr, Mo, Pd, Ag, Cd, Sn, La, Gd, Ta, W, Pt, Au, Pb, Th, and U. In this category, the photon energy region above 100 keV is also included for all elements in which incoherent scattering comprises more than 90% of the total cross section. Category II (uncertainty of 2–5%) applies to the energy region 2–6 keV for all elements, 6–40 keV for all elements not specified in category I, and above 40 keV except for the scattering-dominated region specified in category I. In category III (uncertainty of 5–15%), the authors (Hubbell et al., 1974) included (1) the elements hydrogen, helium, and lithium, (2) the energy region 1–2 keV for elements, and (3) the regions containing K, L, M, and N absorption edges, and the fine-structure regions extending from 200 eV to 1 keV above each of these regions. The experimental uncertainties in cases 1 and 2 greatly exceed 15%. Category IV (uncertainty above 15%) applies to the photon energy region about 200 eV above an absorption edge (Kossel and Kronig fine-structure regions) for all elements. Based on the tables published by McMaster et al. (1969), Wernisch et al. (1984) developed an algorithm for the calculation of the total mass attenuation coefficient valid for the photon energy range from 1 to 50 keV and for 73 elements (11 Z 83). The authors have applied the simple expression m
cm2 ¼ Hedþk g
ð76Þ
ln E
Values of H, d, and k have been obtained from least-squares fits applied to the data published by McMaster et al. (1969); they are given in Figure 5. The values of the edge energies EK, EL1, EL2, EL3, EM1, EM2, EM3, EM4, and EM5 can simply be calculated from (Wernisch et al., 1984) Ei ¼ ri þ si Z þ ti Z2 þ ni Z3
keV
ð77Þ
the parameters ri, si, ti, and ni for various absorption edges i are given in Table 2. Another flexible semiempirical scheme to calculate the total mass attenuation coefficient m for a very wide photon energy range (0.1–1000 keV) has been proposed by Orlic et al. (1993) m ¼ exp½ p1 þ p2 ðln lÞ þ p3 ðln lÞ2 þ p4 ðln lÞ3 þ sKN ZN0 A1
ð78Þ
where sKN is the average incoherent collision cross section gives by Eq. (61). The values of the fitting parameters p1, p2, p3, and p4 are constant for each element and within the energy regions defined by two adjacent absorption edges or for energies beyond the K absorption edge; the parameters are given in Appendix X. The experimental and theoretical values of total interaction cross sections [stot, Eq. (42)] and the mass attenuation coefficients m for the elements (1 Z 98) in the energy range 4.9–24.9 keV have been collected and compared by Arndt et al. (1992). Most of the available tabulations of x-ray attenuation coefficients do not include the photon energy region below 1 keV that corresponds to the energies of characteristic K x-rays of light elements (Z < 11). Experimental data in this energy region are incomplete and it should not be assumed that the accuracy of the available tabulated values is better than 15% (Veigele, 1974; Appendix XI). E.
Diffraction, Refraction, and Dispersion
When a beam of monochromatic x-rays falls onto a crystal lattice, a regular periodic arrangement of atoms, a diffracted beam only results in definite directions. The phenomenon of x-ray diffraction at an ordered array of atoms (or molecules) can also be Copyright © 2002 Marcel Dekker, Inc.
Figure 5 Definition of the energy ranges (a) and values of the parameters H, d, and k (b) applied for the calculation of the total mass attenuation coefficients according to Eq. (76). (From Wernisch et al., 1984.) Copyright © 2002 Marcel Dekker, Inc.
Table 2 Values of the Parameters ri,si,ti, and ni Applied for Calculating the Energies of Absorption Edges via Eq. (77) Applicability range i K L1 L2 L3 M1 M2 M3 M4 M5
ri
si 1
1.304610 4.5066101 6.0186101 3.3906101 8.645 7.499 6.280 4.778 2.421
ti 3
2.633610 1.5666102 1.9646102 4.9316102 3.9776101 3.4596101 2.8316101 2.1846101 1.1726101
ni 3
9.718610 7.5996104 5.9356104 2.3366103 5.9636103 5.2506103 4.1176103 3.3036103 1.8456103
5
4.144610 1.7926105 1.8436105 1.8366106 3.6246105 3.2636105 2.5056105 2.1156105 1.3976105
%a
Zmin
Zmax
3.5 2.2 2.3 1.9 0.4 0.4 0.4 0.4 0.4
11 28 30 30 52 55 55 60 61
63 83 83 83 83 83 83 83 83
a Standard deviation of calculated energies [Eq. (77)] relative to the energies from the tables of McMaster et al. (1969). Source: From Wernisch et al., 1984.
interpreted as a reflection of an incident x-ray beam by the interior planes of a crystal (Bragg reflection). By elementary calculation of the difference in path between two coherent rays, W. L. Bragg found the reinforcement condition for reflection (known as Bragg’s equation or law): yn nl ¼ 2d sin ð79Þ 2 where n is the order of reflection, d is the interplanar spacing, and yn=2 is the angle of reflection (or Bragg angle) defined as the angle between the reflecting plane of the crystal and the incident or reflected beam. The first-order reflection (n ¼ 1) is normally strongest, and the reflected intensity decreases as n increases. Bragg’s law as given in Eq. (79) is only a first approximation, as the refraction in the crystal interferes with the angle of reflection. Because the refractive index of x-rays is slightly less than unity, the deviations from Bragg’s law, Eq. (79), were not observed in the early years until methods were found for precise measurements of x-ray wavelengths. The refraction is accounted for by ascribing a slightly different value dn of the lattice constant to each order of reflection; the simple Bragg’s law [Eq. (79)] can thus be written as (Sandstro¨m, 1957) 4d2 d yn nl ¼ 2d 1 2 2 sin ð80Þ n l 2 where d ¼ 1 n0 for small photon absorption (l < 1 A˚) and n0 is the refractive index for x-rays. d is a small positive number of the order 105 for heavy elements and 106 for light elements at l ¼ 1 A˚ and is proportional to l2. Because the value of d is positive, total reflection occurs back into air when an x-ray beam meets a surface at a large enough angle of incidence. Provided no absorption occurs, the critical glancing angle ytr (tr, total reflection) is defined by (Sandstro¨m, 1957) pffiffiffiffiffi sin ytr ¼ 2d ð81Þ Copyright © 2002 Marcel Dekker, Inc.
For l ¼ 0.1 nm, the value of ytr is of the order 103 for light elements and 56103 for heavy elements, increasing in proportion to l The refractive index n0 for a medium containing one type of atom can be calculated from (Hirsch, 1962) n0 ¼ 1
Nl2 e2 Fð0Þ 2p m0 c2
ð82Þ
where N is the number of atoms per unit volume and Fð0Þ is the atomic scattering factor at zero scattering angle. Equation (82) shows that n0 depends on the wavelength; this phenomenon is called dispersion. The anomalous dispersion causes the quantity d=l2 , Eq. (80), to vary slightly with the wavelength. The variations become important only in the neighborhood of the absorption edges of the constituents of the crystal. Dividing the quantity d=l2 into one normal part ½ðd=l2 Þn and one anomalous part ½ðd=l2 Þa , the theory of anomalous dispersion leads to an expression of Bragg’s law that can be written as (Sandstro¨m, 1957) 4d2 d 4d2 d yn nl ¼ 2d 1 2 sin ð83Þ 2 2 2 n n 2 l n l a Combined with Bragg’s law in its uncorrected form, Eq. (79), this expression becomes (Sandstro¨m, 1957) 4d2 d yn =2 4d2 d l ¼ 2d 1 2 l ð84Þ sin n n n2 l 2 a l2 n The values of ðd=l2 Þa giving a correction for anomalous dispersion can be determined experimentally. The theory of anomalous dispersion has been applied by Sparks (1974, 1975) to explain the inelastic angular-independent scattering from elements having an absorptionedge energy just above the energy of the incident x-rays. The observed intensity of the inelastically scattered radiation was found to be dependent on the nearness of the energy of an absorption edge to the energy of the incident x-rays. The energy of the inelastic peaks is shifted from the incident energy by the binding energy of the most tightly bound shell from which electrons could be photoejected by the incident radiation. F.
X-ray Raman Scattering
Immediately after the discovery of Raman scattering in the visible-wavelength region, a similar effect concerning x-ray radiation was experimentally examined (Davis and Mitchell, 1928; Krishnan, 1928). X-ray Raman scattering appears as a band spectrum having a short-wavelength edge corresponding to a definite energy loss equal to the K electron-binding energy EK of the element. This inelastic effect was observed, for example, when CrKa and CuKa radiation was scattered by solids of light elements, such as lithium, beryllium , boron, and by graphite (Suzuki, 1966; Suzuki et al., 1970; Suzuki and Nagasawa, 1975). The shape of the Raman band is similar to that of the soft x-ray K absorption spectrum of the solids. Mizuno and Ohmura (1967) have found the following two conditions for x-ray Raman scattering: 4p a <1 l0 Copyright © 2002 Marcel Dekker, Inc.
ð85Þ
and hv0 EK
ð86Þ
where a is the mean radius of charge distribution of the K electrons and l0 is the wavelength of the incident x-rays. The intensity of the x-ray Raman scattering I(y, l) is given by (Suzuki and Nagasawa, 1975) ( 2 2 ) 4p a y 4p a y 2 ð87Þ sin þT2 ðlÞ sin Iðy; lÞ ffi ð1 þ cos yÞ T1 ðlÞ l0 2 l0 2 where T1(l) and T2(l) factors are related to the dipole and multipole transitions, respectively. Although according to this equation a slight displacement of the peak position of the Raman band is expected with the scattering angle, the peak position does not shift in most experiments. In general, x-ray Raman scattering gives information about the unoccupied states above the Fermi level of the relevant solids. Moreover, this kind of inelastic scattering by electrons in solids should sometimes be taken into account in x-ray spectrum evaluation.
VII.
INTENSITY OF CHARACTERISTIC X-RAYS
This section provides some of the necessary background information for subsequent chapters dealing with quantitative x-ray analysis. Derivation of any relationship between excitation source intensity and measured characteristic x-rays is sometimes complex and is presented in detail in many relevant books, such as those by Jenkins et al. (1981) and Tertian and Claisse (1982) on XRF analysis and by Goldstein et al. (1981) and Heinrich (1981) on EPXMA. A.
Photon Excitation
When continuous (polychromatic) radiation is used to excite the characteristic x-rays of element i in a completely homogeneous sample of thickness T (cm) and when enhancement effects are neglected, the intensity of the fluorescent radiation Ii (Ei) is described by Ii ðEi Þ dO1 dO2 ¼
dO1 dO2 eðEi Þ 4p sin C1 E Zmax 1 exp½rTðmðE0 Þ csc C1 þ mðEi Þ csc C2 Þ ai ðE0 Þ I0 ðE0 Þ dE0 mðE0 Þ csc C1 þ mðEi Þ csc C2
ð88Þ
Ec;i
with ai ðE0 Þ ¼
Wi t0i ðE0 Þoi pi
1 1 ji
ð89Þ
where dO1 and dO2 are the differential solid angles for the incident (primary) and emerging (characteristic) radiation, respectively; eðEi Þ is the intrinsic detector efficiency for recording a photon of energy Ei ; Ec;i and Emax are the critical absorption energy of element i and the maximum energy in the excitation spectrum; r is the density of the specimen (in g=cm3); C1 and C2 are the effective incidence and takeoff angles, respectively; mðE0 Þ and mðEi Þ are the total mass attenuation coefficients (in cm2=g) for the whole specimen [Eq. (45)] at Copyright © 2002 Marcel Dekker, Inc.
energies E0 and Ei , respectively; I0 ðE0 Þ dE0 is the number of incident photons per second per steradian in the energy interval E0 to E0 þ dE0 ; Wi is the weight fraction of the ith element; and t0i ðE0 Þ is the total photoelectric mass absorption coefficient for the ith element at the energy E0 (in cm2=g). Because large solid angles O1 and O2 are used for the excitation and characteristic x-rays in practical spectrometers, Eq. (88) should also be integrated over these finite solid angles. Such calculations can often be omitted; however, and for a given measurement geometry, an experimentally determined geometry factor G can be applied. As seen from Eq. (88), the intensity of characteristic x-rays is modified by the effects of primary [mðE0 Þ] and secondary [mðEi Þ] absorption in the specimen; this is a major source of the so-called matrix effects in XRF analysis. If the excitation source is monochromatic (emits only one energy), Eq. (88) simplifies to eðEi Þai ðE0 ÞI0 ðE0 Þ sin C1 1 exp½rTðmðE0 Þ csc C1 þ mðEi Þ csc C2 Þ mðE0 Þ csc C1 þ mðEi Þ csc C2
Ii ðEi Þ ¼ G
ð90Þ
The enhancement effect, consisting of an extra excitation of the element of interest by the characteristic radiation of some matrix elements, modifies the equations for the intensity Ii ðEi Þ. In the case of monochromatic photon excitation, a factor, 1 þ Hi , should be included in Eq. (90), where Hi is the enhancement term defined as Hi ¼
m X 1 1 Wk o k 1 m ðEk Þmk ðE0 Þ 2mi ðE0 Þ k¼1 jk i lnð1 þ mðE0 Þ=½mðEk Þ sin C1 Þ lnð1 þ mðEi Þ=½mðEk Þ sin C2 Þ þ mðE0 Þ= sin C1 mðEi Þ= sin C2
ð91Þ
where mi ðE0 Þ and mi ðEk Þ are the total mass attenuation coefficients for the ith element at the energy of incident radiation ðE0 Þ and characteristic radiation of the element k ðEk Þ, respectively, mk ðE0 Þ is the total mass attenuation coefficient for the element k at the energy E0, and mðEk Þ is the total mass attenuation coefficient for the whole specimen at the energy Ek . 1. Thin-SampleTechnique If the total mass per unit area of a given sample m ¼ rT is small, Eq. (90) simplifies to Iithin ðEi Þ ¼
G eðEi Þai ðE0 ÞI0 ðE0 Þm sin C1
ð92Þ
The relative error resulting from applying Eq. (92) instead of Eq. (90) does not exceed 5% when the total mass per unit area satisfies the condition m thin
0:1 mðE0 Þ csc C1 þ mðEi Þ csc C2
ð93Þ
A major feature of the thin-sample technique is that the intensity of the characteristic x-rays, Iithin , depends linearly on the concentration of the ith element (or on its mass per unit area); it is equivalent to the fact that, in the thin-sample technique, matrix effects can safely be neglected. Copyright © 2002 Marcel Dekker, Inc.
2. Thick-SampleTechnique By the term ‘‘thick sample,’’ we mean a sample whose mass per unit area (thickness) is greater than the so-called saturation mass (thickness). The saturation thickness is defined as a limiting value above which practically no further increase in the intensity of the characteristic radiation is observed as the sample thickness is increased. If the total mass per unit area of a given sample is sufficiently large, Eq. (90) simplifies to Iithick ðEi Þ ¼
GeðEi Þai ðE0 ÞI0 ðE0 Þ mðE0 Þ þ sin C1 =ðsin C2 ÞmðEi Þ
ð94Þ
The relative error resulting from applying Eq. (94) instead of Eq. (90) does not exceed 1% when the total mass per unit area satisfies the condition m thick
B.
4:61 mðE0 Þ csc C1 þ mðEi Þ csc C2
ð95Þ
Electron Excitation
For a thin foil (defined as a foil of thickness such that the beam electron undergoes only one scattering act), the intensity of the characteristic x-rays Iitf (in photons per second) is given by Iitf ¼ Ki I0 Qi oi N0
1 rTWi Ai
ð96Þ
where Ki is a factor depending on the measurement geometry and detection efficiency and the other symbols have the following dimensions: I0 ¼ electrons=s, Qi ¼ ionizations per electron for 1 atom=cm2, oi ¼ photons=ionization. N0 ¼ atoms=mol, 1=Ai ¼ mol=g, r ¼ g=cm3, and T ¼ cm. For a bulk target, Eq. (96) must be integrated over the electron path, taking into account the loss of energy by the electron beam:
Iibulk
Wi ¼ I0 oi Ki N0 Ai
ZEc;i
Qi ðEÞ dE dE=dðrxÞ
ð97Þ
E0
where dE=dðrxÞ is the electron energy loss per unit of distance traveled in a material, given by the Bethe equation (Heinrich, 1981; Bethe and Ashkin, 1953): n dE 1 X Wj Z j E ¼ 78; 500 ln 1:166 dðrxÞ E j¼1 Aj Jj
keV=g cm2
ð98Þ
where Jj is the mean ionization potential of element jðJj ¼ 9:76Zj þ 58:5Zj0:19 eVÞ (Goldstein et al., 1981). Strictly speaking, the Bethe equation is only valid for the electron energy E > 6:3Jj . Below this limit, the energy loss of the electrons should be described either by the modification of Rao-Sahib and Wittry (1972) [Eq. (99)] or by the modification of Love et al. (l978) [Eq. (l00)]: Copyright © 2002 Marcel Dekker, Inc.
n dE 1 X Wj Z j pffiffiffiffi keV=g cm2 ¼ 0:6236 105 pffiffiffiffi dðrxÞ A E j¼1 j Jj n X dE Wj Z j 1 ¼ 1=2 5 A J dðrxÞ j j 1:18 10 ðE=Jj Þ þ 1:47 106 E=Jj j¼1
ð99Þ keV=g cm2
ð100Þ
However, to describe the intensity of characteristic x-rays emitted, three effects should be considered additionally (Heinrich, 1981; Goldstein et al., 1981; Love and Scott, 1981): 1. 2. 3.
C.
Absorption of characteristic x-rays within the specimen Electron backscattering Secondary fluorescence by characteristic x-rays and=or bremsstrahlung continuum produced by an electron beam
Particle Excitation
For a thin, uniform, homogeneous target, the intensity of characteristic x-rays Itti is given by the simple formula I tti ¼ Ki I0 N0 ss;i ðE0 Þoi
1 mi Ai
ð101Þ
where mi is the areal density of the element with atomic number Zi and atomic mass Ai . For a thick, homogeneous target, the intensity of characteristic x-rays, Iithick , from the element i of concentration Wi can be calculated by (Campbell and Cookson, 1984) Iithick ¼
0 X Ki oi N0 ss;i ðEÞTi ðEÞ dE I 0 Wi Ai SðEÞ E
ð102Þ
0
where S(E) is the stopping power and Ti ðEÞ is the photon attenuation factor. The latter is given by 0 1 ZE sin C1 dE C B Ti ðEÞ ¼ exp@mðEi Þ ð103Þ A SðEÞ sin C2 E0
More details on various x-ray analytical techniques are provided in subsequent chapters.
VIII.
IUPAC NOTATION FOR X-RAY SPECTROSCOPY
The nomenclature commonly used in XRS to describe x-ray emission spectra was introduced by M. Sieghahn in the l920s and is based on the relative intensity of lines from different series. A new and more systematic notation for x-ray emission lines and absorption edges, based on the energy-level designation, was developed by the International Union of Pure and Applied Chemistry (Jenkins et al. 1991). Because the new notation, called the IUPAC notation, replaces the Siegbahn’s notation, some characteristic features of the new nomenclature must be mentioned. The IUPAC notation prescribes Arabic numerals for subscripts; the original notation uses Roman numerals (e.g., L2 and L3 , Copyright © 2002 Marcel Dekker, Inc.
instead of LII and LIII ). In the IUPAC notation, states with double or multiple vacancies should be denoted by, for example, K2 ; KL1 , and Ln2;3 , which correspond to the electron configurations 1s2 ; 1s1 2s1 , and 2pn , respectively. X-ray transitions and x-ray emission diagram lines are denoted by the initial (placed first) and final x-ray levels separated by a hyphen. To conform with the IUPAC notation of x-ray spectra, the hyphen separating the initial and final state levels should also be introduced into the notation for Auger electron emission process. The IUPAC notation is compared with the Sieghahn notation in Appendix XII.
Copyright © 2002 Marcel Dekker, Inc.
APPENDIX I: CRITICAL ABSORPTION WAVELENGTHS AND CRITICAL ABSORPTION ENERGIES K edge Atomic number Element 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29
H He Li Be B C N O F Ne Na Mg Al Si P S Cl Ar K Ca Sc Ti V Cr Mn Fe Co Ni Cu
L1 edge
L2 edge
L3 edge
M4 edge
A˚
keV
A˚
keV
A˚
keV
A˚
keV
918 504 226.953 106.9 64.6 43.767 31.052 23.367 18.05 14.19 11.48 9.512 7.951 6.745 5.787 5.018 4.397 3.871 3.437 3.070 2.757 2.497 2.269 2.070 1.896 1.743 1.608 1.488 1.380
0.014 0.025 0.055 0.116 0.192 0.283 0.399 0.531 0.687 0.874 1.08 1.303 1.559 1.837 2.142 2.470 2.819 3.202 3.606 4.037 4.495 4.963 5.462 5.987 6.535 7.109 7.707 8.329 8.978
258 225 197 143 105 81.0 64.2 52.1 43.2 36.4 30.7 26.8 23.4 20.5 18.3 16.3 14.6 13.3 12.22 11.27
0.048 0.055 0.063 0.087 0.118 0.153 0.193 0.238 0.287 0.341 0.399 0.462 0.530 0.604 0.679 0.762 0.849 0.929 1.015 1.100
564 365 248 170 125 96.1 75.6 61.1 50.2 41.8 35.2 30.2 27.0 23.9 21.3 19.1 17.2 15.6 14.2 13.0
0.022 0.034 0.050 0.073 0.099 0.129 0.164 0.203 0.247 0.297 0.352 0.411 0.460 0.519 0.583 0.650 0.721 0.794 0.871 0.953
564 365 253 172 127 96.9 76.1 61.4 50.6 42.2 35.5 30.8 27.3 24.2 21.6 19.4 17.5 15.9 14.5 13.3
0.022 0.034 0.049 0.072 0.098 0.128 0.163 0.202 0.245 0.294 0.349 0.402 0.454 0.512 0.574 0.639 0.708 0.779 0.853 0.933
Copyright © 2002 Marcel Dekker, Inc.
A˚
keV
M5 edge A˚
keV
30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63
Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te I Xe Cs Ba La Ce Pr Nd Pm Sm Eu
1.283 1.196 1.117 1.045 0.980 0.920 0.866 0.816 0.770 0.728 0.689 0.653 0.620 0.589 0.561 0.534 0.509 0.486 0.464 0.444 0.425 0.407 0.390 0.374 0.359 0.345 0.331 0.318 0.307 0.295 0.285 0.274 0.265 0.256
Copyright © 2002 Marcel Dekker, Inc.
9.657 10.365 11.100 11.860 12.649 13.471 14.319 15.197 16.101 17.032 17.993 18.981 19.996 21.045 22.112 23.217 24.341 25.509 26.704 27.920 29.182 30.477 31.800 33.155 34.570 35.949 37.399 38.920 40.438 41.986 43.559 45.207 46.833 48.501
10.33 9.54 8.73 8.107 7.506 6.97 6.46 5.998 5.583 5.232 4.867 4.581 4.298 4.060 3.83 3.626 3.428 3.254 3.085 2.926 2.777 2.639 2.511 2.389 2.274 2.167 2.068 1.973 1.889 1.811 1.735 1.665 1.599 1.536
1.200 1.30 1.42 1.529 1.651 1.78 1.92 2.066 2.220 2.369 2.546 2.705 2.883 3.054 3.24 3.418 3.616 3.809 4.018 4.236 4.463 4.695 4.937 5.188 5.451 5.719 5.994 6.282 6.559 6.844 7.142 7.448 7.752 8.066
11.87 10.93 9.94 9.124 8.416 7.80 7.21 6.643 6.172 5.755 5.378 5.026 4.718 4.436 4.180 3.942 3.724 3.514 3.326 3.147 2.982 2.830 2.687 2.553 2.429 2.314 2.204 2.103 2.011 1.924 1.843 1.767 1.703 1.626
1.045 1.134 1.248 1.358 1.473 1.59 1.72 1.865 2.008 2.153 2.304 2.467 2.627 2.795 2.965 3.144 3.328 3.527 3.726 3.938 4.156 4.380 4.611 4.855 5.102 5.356 5.622 5.893 6.163 6.441 6.725 7.018 7.279 7.621
12.13 11.10 10.19 9.39 8.67 8.00 7.43 6.89 6.387 5.962 5.583 5.223 4.913 4.632 4.369 4.130 3.908 3.698 3.504 3.324 3.156 3.000 2.855 2.719 2.592 2.474 2.363 2.258 2.164 2.077 1.995 1.918 1.845 1.775
1.022 1.117 1.217 1.32 1.43 1.55 1.67 1.80 1.940 2.079 2.220 2.373 2.523 2.677 2.837 3.001 3.171 3.351 3.537 3.728 3.927 4.131 4.340 4.557 4.780 5.010 5.245 5.488 5.727 5.967 6.213 6.466 6.719 6.981
15.56
0.7967
15.89
0.7801
13.122 12.459
0.9448 0.9951
13.394 23.737
0.9257 0.9734
11.288 10.711
1.0983 1.1575
11.552 11.013
1.0732 1.1258
Appendix I
Continued L1 edge
K edge Atomic number 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92
L2 edge
L3 edge
Element
A˚
keV
A˚
keV
A˚
keV
A˚
keV
Gd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn Fr Ra Ac Th Pa U
0.247 0.238 0.231 0.223 0.216 0.209 0.202 0.196 0.190 0.184 0.178 0.173 0.168 0.163 0.158 0.153 0.149 0.415 0.141 0.137 0.133 0.130 0.126 0.123
50.215 51.984 53.773 55.599 57.465 59.319 61.282 63.281 65.292 67.379 69.479 71.590 73.856 76.096 78.352 80.768 83.046 85.646 88.037 90.420 93.112 95.740 98.418 101.147
0.116 0.113 0.110 0.108
106.759 109.741 112.581 115.610
1.477 1.421 1.365 1.317 1.268 1.222 1.182 1.140 1.100 1.061 1.025 0.990 0.956 0.923 0.893 0.863 0.835 0.808 0.782 0.757 0.732 0.709 0.687 0.665 0.645 0.625 0.606 0.588 0.56
8.391 8.722 9.081 9.408 9.773 10.141 10.487 10.870 11.271 11.681 12.097 12.524 12.968 13.427 13.875 14.354 14.837 15.338 15.858 16.376 16.935 17.490 18.058 18.638 19.229 19.842 20.458 21.102 21.764
1.561 1.501 1.438 1.390 1.338 1.288 1.243 1.199 1.155 1.114 1.075 1.037 1.001 0.967 0.934 0.903 0.872 0.843 0.815 0.789 0.763 0.739 0.715 0.693 0.671 0.650 0.630 0.611 0.592
7.938 8.256 8.619 8.918 9.260 9.626 9.972 10.341 10.732 11.128 11.533 11.953 12.380 12.817 13.266 13.731 14.210 14.695 15.205 15.713 16.244 16.784 17.337 17.904 18.478 19.078 19.677 20.311 20.938
1.710 1.649 1.579 1.535 1.482 1.433 1.386 1.341 1.297 1.255 1.216 1.177 1.140 1.106 1.072 1.040 1.008 0.979 0.950 0.923 0.897 0.872 0.848 0.825 0.803 0.782 0.761 0.741 0.722
7.250 7.517 7.848 8.072 8.361 8.650 8.941 9.239 9.554 9.874 10.196 10.529 10.867 11.209 11.556 11.917 12.3 12.655 13.041 13.422 13.817 14.215 14.618 15.028 15.439 15.865 16.293 16.731 17.160
Copyright © 2002 Marcel Dekker, Inc.
M4 edge
M5 edge
A˚
keV
A˚
keV
8.601
1.4415
8.847 8.487
1.4013 1.4609
6.87 6.59 6.33 6.073 5.83 5.59 5.374 5.157 4.952 4.757 4.572
1.804 1.880 1.958 2.042 2.126 2.217 2.307 2.404 2.504 2.606 2.711
7.11 6.83 6.560 6.30 6.05 5.81 5.584 5.36 5.153 4.955 4.764
1.743 1.814 1.890 1.967 2.048 2.133 2.220 2.313 2.406 2.502 2.603
3.557 3.436 3.333
3.485 3.608 3.720
3.729 3.618 3.497
3.325 3.436 3.545
93 94 95 96 97 98 99 100
Np Pu Am Cm Bk Cf Es Fm
0.105 0.102 0.099 0.097 0.094 0.092 0.090 0.088
118.619 121.720 124.816 128.088 131.357 134.683 138.067 141.510
Source: From Clark, 1963 and Burr, 1974.
Copyright © 2002 Marcel Dekker, Inc.
0.553 0.537 0.521 0.506 0.491 0.477 0.464 0.451
22.417 23.097 23.793 24.503 25.230 25.971 26.729 27.503
0.574 0.557 0.540 0.525 0.509 0.494 0.480 0.466
21.596 22.262 22.944 23.640 24.352 25.080 25.824 26.584
0.704 0.686 0.669 0.653 0.637 0.622 0.607 0.593
17.614 18.066 18.525 18.990 19.461 19.938 20.422 20.912
APPENDIX II: CHARACTERISTIC X-RAY WAVELENGTHS (A) AND ENERGIES (keV) Table 1
K Series Diagram Lines (A˚)a
Line
a1,2
a1
a2
Approximate intensity
150
100
50
Li B B C N O F Ne Na Mg Al Si P S Cl Ar K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc Ru Rh Pd
3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46
230 113 67 44 31.603 23.707 18.307 14.615 11.909 9.889 8.339 7.126 6.155 5.373 4.729 4.192 3.744 3.360 3.032 2.750 2.505 2.291 2.103 1.937 1.791 1.659 1.542 1.437 1.341 1.256 1.177 1.106 1.041 0.981 0.927 0.877 0.831 0.788 0.748 0.710 0.676 0.644 0.614 0.587
Copyright © 2002 Marcel Dekker, Inc.
8.338 7.125 6.154 5.372 4.728 4.191 3.742 3.359 3.031 2.749 2.503 2.290 2.102 1.936 1.789 1.658 1.540 1.435 1.340 1.255 1.175 1.105 1.040 0.980 0.926 0.875 0.829 0.786 0.747 0.709 0.675 0.643 0.613 0.585
8.341 7.127 6.157 5.375 4.731 4.194 3.745 3.362 3.034 2.753 2.507 2.294 2.105 1.940 1.793 1.661 1.544 1.439 1.344 1.258 1.179 1.109 1.044 0.984 0.930 0.880 0.833 0.791 0.751 0.713 0.679 0.647 0.617 0.590
b1
b3 15
14.460 11.574 9.559 7.960
b2
b4
5
<1
11.726 9.667 8.059 6.778
5.804 5.032 4.403 3.886 3.454 3.089 2.780 2.514 2.285 2.085 1.910 1.757 1.621 1.500 1.392 1.296 1.207 1.129 1.057 0.992 0.933 0.879 0.829 0.783 0.740 0.701 0.665 0.632
1.393 1.208 1.129 1.058 0.993 0.933 0.879 0.830 0.784 0.741 0.702 0.666 0.633 0.601
0.572 0.546 0.521
0.573 0.546 0.521
1.489 1.381 1.284 1.196 1.117 1.045 0.980 0.921 0.866 0.817 0.771 0.728 0.690 0.654 0.621 0.590 0.562 0.535 0.510
0.866 0.815 0.770 0.727 0.689 0.653 0.620 0.561 0.534
Table 1
Continued
Line
a1,2
a1
a2
Approximate intensity
150
100
50
0.559 0.535 0.512 0.491 0.470 0.451 0.433 0.416 0.401 0.385 0.371 0.357 0.344 0.332 0.321 0.309 0.299 0.289 0.279 0.270 0.261 0.253 0.244 0.236 0.229 0.222 0.215 0.209 0.202 0.196 0.191 0.158 0.180 0.175 0.170 0.165 0.161 0.156 0.152 0.148 0.144 0.144 0.140 0.133 0.131 0.126
0.564 0.539 0.517 0.495 0.475 0.456 0.438 0.421 0.405 0.390 0.376 0.362 0.349 0.337 0.325 0.314 0.304 0.294 0.284 0.275 0.266 0.258 0.250 0.241 0.234 0.227 0.220 0.213 0.207 0.201 0.196 0.190 0.185 0.180 0.175 0.170 0.165 0.161 0.157 0.153 0.149 0.149 0.145 0.138 0.136 0.131
Ag Cs In Sn Sb Te I Xe Cs Ba La Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn Fr Ra Ac Th Pa U
47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92
0.561 0.536 0.514 0.492 0.472 0.453 0.435 0.418 0.402 0.387 0.373 0.359 0.346 0.334 0.322 0.311 0.301 0.291 0.281 0.272 0.263 0.255 0.246 0.238 0.231 0.224 0.217 0.211 0.204 0.198 0.193 0.187 0.182 0.177 0.172 0.167 0.162 0.185
0.135 0.128
a Conversion equation: E (keV) ¼ 12.4=l (A˚). Source: Copyright ©From 2002 Clark, Marcel1963. Dekker, Inc.
b1
b3
b2
b4
5
<1
0.498 0.476 0.455 0.436 0.418 0.401 0.385 0.360 0.355 0.342 0.329 0.317 0.305 0.294
0.487 0.465 0.445 0.426 0.408 0.391 0.376
0.486
0.274 0.265 0.256 0.246 0.238 0.231 0.223 0.216 0.208 0.203 0.196 0.191 0.185 0.179 0.174 0.169 0.164 0.160 0.155 0.151 0.147 0.143
0.267 0.258 0.249 0.239 0.231
15 0.497 0.475 0.455 0.435 0.417 0.400 0.384 0.369 0.355 0.341 0.328 0.316 0.305 0.294
0.346 0.333 0.320 0.309 0.297 0.287
0.444 0.425 0.407
0.319 0.307
0.283 0.274 0.264 0.255 0.246 0.237 0.230 0.222 0.215 0.208 0.202 0.195 0.190 0.184 0.179 0.173 0.168 0.163 0.159 0.154 0.150 0.146 0.142 0.138 0.134 0.131 0.127 0.127 0.124 0.117 0.115 0.111
0.135 0.132 0.128 0.128 0.125 0.118 0.116 0.112
0.217 0.203 0.197 0.190 0.185 0.179 0.174 0.169 0.164 0.159 0.155 0.150 0.146 0.147 0.138 0.133
0.184 0.179 0.174 0.168 0.163 0.159 0.154 0.150 0.146 0.141 0.138
0.114
0.114
0.108
0.108
Table 2
L Series Diagram Lines (A˚)
Line Approximate intensity Na Mg Al Si P S Cl Ar K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te I Xe Cs
11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55
a
a1
a2
b1
b2
b3
b4
b5
b6
b7
b9
110
100
10
50
20
6
4
1
<1
<1
<1
36.393 31.393 27.445 24.309 21.713 19.489 17.602 16.000 14.595 13.357 12.282 11.313 10.456 9.671 8.990 8.375
36.022 31.072 27.074 23.898 21.323 19.158 17.290 15.698 14.308 13.079 12.009 11.045 10.194 9.414 8.735 8.126
7.318 6.863 6.449 6.070 5.725 5.406
7.325 6.870 6.456 6.077 5.732 5.414
7.075 6.623 6.211 5.836 5.492 5.176
4.846 4.597 4.368 4.154 3.956 3.752 3.600 3.439 3.290 3.148
4.854 4.605 4.376 4.162 3.965 3.781 3.609 3.448 3.299 3.157
2.892
2.902
Copyright © 2002 Marcel Dekker, Inc.
19.429 17.757 15.742 14.269 13.167 12.115 11.225
8.930
5.586 5.238 4.923
6.788 6.367 5.983 5.632 5.310 5.013
6.821 6.403 6.018 5.668 5.346 5.048
6.984 6.519 6.094 5.710 5.361 5.048
4.620 4.374 4.146 3.935 3.739 3.555 3.385 3.226 3.077 2.937
4.372 4.130 3.909 3.703 3.514 3.339 3.175 3.023 2.882 2.751
4.487 4.253 4.034 3.834 3.644 3.470 3.306 3.152 3.009 2.874
4.532 4.289 4.071 3.870 3.681 3.507 3.344 3.190 3.046 2.912
4.487 4.242 4.016 3.808 3.614 3.436 3.270 3.115 2.971 2.837
3.155 3.005 2.863 2.730
3.792 3.605 3.430 3.268 3.115 2.973 2.839 2.713
2.683
2.511
2.628
2.666
2.593
2.485
2.478
b10
b15
b17
g1
g2
g3
g4
g5
g6
g8
‘
Z
s
t
<1
<1
<1
10
1
2
<1
<1
<1
<1
3
1
<1
<1
410 260 180
5.384 5.036 4.726
3.799 3.611 3.437 3.274 3.121 2.979 2.847 2.720
4.182 3.944 3.725 3.523 3.336 3.162 3.001 2.852 2.712 2.582
2.492
2.348
67.84 56.212 47.835 41.042 36.671 31.423 27.826 24.840 22.315 20.201 18.358 16.693 15.297 14.081 12.976 11.944 11.069 10.293 9.583
67.25 56.813 47.325 40.542 35.200 30.942 27.375 24.339 21.864 19.73 17.86 16.304 14.940 13.719 12.620 11.608 10.732 9.959 9.253
6.754 6.297 5.875 5.497 5.151 4.387
8.363 7.836 7.356 6.918 6.517 6.150
8.042 7.517 7.040 6.606 6.210 5.847
2.026 2.778 2.639 2.511 2.391
4.288 4.045 3.822 3.616 3.426 3.249 3.085 2.932 2.790 2.657
5.503 5.217 4.952 4.707 4.480 4.269 4.071 3.888 3.716 3.557
5.204 4.922 4.660 4.418 4.193 3.983 3.789 3.607 3.438 3.280
2.174
2.417
3.267
2.994
6.045 5.644 5.283 4.953 4.654 4.380 3.897 3.685 3.489 3.307 3.137 2.980 2.835 2.695 2.567 2.447 2.237
Copyright © 2002 Marcel Dekker, Inc.
2.233
Table 2
Continued
Line Approximate intensity Ba La Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn Fr Ra Ac Th Pa U Np Pu Am
56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95
a
a1
a2
b1
b2
b3
b4
b5
b6
b7
b9
110
100
10
50
20
6
4
1
<1
<1
<1
2.776 2.665 2.561 2.463 2.370 2.283 2.199 2.120 2.046 1.976 1.909 1.845 1.785 1.726 1.672 1.619 1.569 1.522 1.476 1.433 1.391 1.352 1.313 1.277 1.242 1.207 1.175 1.144 1.114
2.785 2.674 2.570 2.473 2.382
2.404 2.303 2.208 2.119 2.035
2.516 2.410 2.311 2.216 2.126
2.555 2.449 2.349 2.255 2.166
2.482 2.379 2.282 2.190 2.103
2.382 2.275 2.180 2.091 2.009
2.376 2.282 2.188 2.100 2.016
2.210 2.131 2.057 1.986 1.920 1.856 1.796 1.738 1.682 1.630 1.580 1.533 1.487 1.444 1.402 1.363 1.325 1.288 1.253 1.218 1.186 1.155 1.126
2.567 2.458 2.356 2.259 2.166 2.081 1.998 1.920 1.847 1.777 1.710 1.647 1.587 1.530 1.476 1.424 1.374 1.327 1.282 1.238 1.197 1.158 1.120 1.083 1.049 1.015 0.982 0.952 0.921
1.882 1.812 1.746 1.682 1.623 1.567 1.514 1.463 1.416 1.370 1.327 1.285 1.245 1.206 1.169 1.135 1.102 1.070 1.040 1.010 0.983 0.955 0.929
1.962 1.887 1.815 1.747 1.681 1.619 1.561 1.505 1.452 1.402 1.353 1.307 1.263 1.220 1.179 1.141 1.104 1.068 1.034 1.001 0.969 0.939 0.908
2.000 1.926 1.853 1.785 1.720 1.658 1.601 1.544 1.491 1.441 1.392 1.346 1.302 1.260 1.218 1.179 1.142 1.106 1.072 1.039 1.007 0.977 0.948
1.779
1.856 1.788 1.723 1.659 1.599
1.862 1.792
1.494
1.485
1.387 1.342 1.298 1.256 1.215 1.177 1.140 1.106 1.072 1.040 1.010 0.981 0.953 0.926 0.900
1.946 1.875 1.807 1.742 1.681 1.622 1.567 1.515 1.466 1.419 1.374 1.331 1.290 1.252 1.213 1.179 1.143 1.111 1.080 1.050 1.021 0.993 0.967
1.395 1.350 1.306 1.264 1.224 1.186 1.149 1.115 1.082 1.050 1.019 0.990 0.962 0.935
1.384 1.336 1.291 1.247 1.204 1.165 1.126 1.090 1.054 1.021 0.986 0.957 0.927 0.898
1.030 1.005
1.017
0.840 0.814
0.858 0.836
0.803
0.841
0.807
0.871
0.817
0.769
0.956 0.933 0.911 0.890 0.868 0.849
0.968 0.945 0.923 0.901 0.880 0.860
0.766 0.742 0.720 0.698 0.678 0.658
0.794 0.774 0.755 0.735 0.719 0.701
0.755 0.732 0.710
0.793 0.770 0.748
0.765 0.746 0.726
0.828 0.803 0.789
0.775 0.755 0.736
0.723 0.701 0.681
0.669
0.707
0.691
Source: From Clark, 1963.
Copyright © 2002 Marcel Dekker, Inc.
1.577
b10
b15
b17
g1
g2
g3
g4
g5
g6
g8
‘
Z
s
t
<1
<1
<1
10
1
2
<1
<1
<1
<1
3
1
<1
<1
2.387 2.290 2.195 2.107 2.023
2.442 2.141 2.048 1.961 1.878
2.138 2.046 1.960 1.879 1.801
2.134 2.041 1.955 1.874 1.797
2.075 1.983 1.899 1.819 1.745
1.870 1.800 1.731 1.667
1.726 1.657 1.592 1.530 1.473 1.417 1.364 1.316 1.268 1.222 1.179 1.138 1.098 1.061 1.025 0.991 0.958 0.927 0.897 0.868 0.840 0.814 0.786
1.659 1.597 1.534 1.477 1.423 1.371 1.321 1.274 1.228 1.185 1.144 1.105 1.068 1.032 0.998 0.966 0.934 0.905 0.876 0.848 0.822 0.796 0.771
1.655 1.591 1.529 1.471 1.417 1.364 1.315 1.268 1.222 1.179 1.138 1.099 1.062 1.026 0.992 0.959 0.928 0.898 0.869 0.842 0.815 0.790 0.764
1.606 1.544 1.485 1.427 1.374 1.323 1.276
1.494 1.392 1.343 1.299 1.254 1.212 1.172 1.133 1.097 1.062 1.028 0.996 0.964 0.934 0.905
1.372 1.328 1.437 1.287 1.247 1.339 1.208 1.293 1.171 1.137 1.166 1.072 1.128 1.041 1.090 1.012 1.056 0.984 1.022 0.957 0.989 0.931
1.185 1.143 1.103 1.065 1.028 0.993 0.959 0.928 0.897 0.867 0.839 0.812 0.761
2.309 2.222 3.135 2.862 2.205 3.006 2.740 2.110 2.023 2.892 2.620 2.020 1.936 2.784 2.512 1.935 1.855 2.675 2.409
1.708
1.518 1.462 1.406 1.355 1.307 1.260 1.215 1.173 1.132 1.094 1.057 1.022 0.988 0.956 0.925 0.895 0.867 0.840
1.243 1.198 1.155 1.114 1.074 1.037 1.001 0.967 0.934 0.903 0.873 0.845 0.817 0.791 0.765
2.482 1.632 2.395 2.312 2.234 2.158 2.086 2.019 1.955 1.250 1.894 1.204 1.836 1.161 1.782 1.120 1.728 1.081 1.678 1.044 1.630 1.008 1.585 0.974 1.541 0.941 1.499 0.010 1.460 0.880 1.422 0.852 1.385 0.824 1.350 0.799 1.317 1.283
2.218 2.049 1.898 1.826 1.757 1.695 1.635 1.478 1.523 1.471 1.421 1.374 1.328 1.285 1.243 1.202 1.164 1.127 1.092 1.058
1.831 1.776 1.663 1.723 1.612 1.672
1.352 1.414 1.279 1.342 1.244 1.308 1.210
0.716 0.776 0.838 0.844 0.694 0.682 0.675 0.649 0.717 0.673 0.680 1.167 0.908 0.730 0.708 0.687
0.653 0.642 0.635 0.611 0.675 0.632 0.640 1.115 0.855 1.011 1.080 0.634 0.624 0.617 0.594 0.655 0.613 1.091 0.830 0.615 0.605 0.598 0.577 0.635 0.595 0.601 1.067 0.806 0.904 1.035 0.597 0.579 0.562
Copyright © 2002 Marcel Dekker, Inc.
Table 3
M Series Diagram Lines (A˚)
Line K Cu Ru Rh Pd Ag Cd In Sn Sb Te Ba La Ce Pr Nd Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Th Pa U
a1 19 29 44 45 46 47 48 49 50 51 52 56 57 58 59 60 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 90 91 92
a2
b
g
l 680 170
26.85 25.00 21.80 20.46
14.88 14.06
14.51 13.78
12.675 Band Band Band Band Band Band Band 8.460
Band Band 10.744 10.253 9.792 9.364 8.965 8.593 8.246 7.909 7.600 7.304 7.022 6.756 6.504 6.267 6.037 5.828 5.623 5.452 5.250 5.075 4.909 3.942 3.827 3.715
8.139
8.155 7.840
7.539 7.251 6.983 6.528
7.546 7.258 6.990 6.490
6.261 6.046 5.840
6.275 6.057 5.854 5.666
5.461 5.285 5.118 4.138 4.022 3.910
Source: From Clark, 1963.
Copyright © 2002 Marcel Dekker, Inc.
5.472 5.299 5.129 4.151 4.035 3.924
17.94 16.92 15.93 12.700 12.064 11.534 10.997 10.504 9.599 9.211 8.844 8.485 8.144 7.865 7.545
18.38
14.22 13.57 12.98 12.43 11.86 11.37
7.023 6.761 6.543 6.312 6.088 5.887 5.681 5.501 5.320 5.145
10.48 10.07 9.69 9.32 8.96 8.63
4.825 4.674 4.531 3.679 3.577 3.480
6.97 6.74 6.52 5.24 5.08 4.95
8.02 7.74 7.47
Table 4 Atomic number 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50
Energies of Principal K and L X-ray Emission Lines (keV) Element Li Be B C N C F Ne Na Mg Al Si P S Cl Ar K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn
Kb2
8.328 8.976 9.657 10.365 11.100 11.863 12.651 13.465 14.313 15.184 16.083 17.011 17.969 18.951 19.964 21.012 22.072 23.169 24.297 25.454 26.641 27.859 29.106
Copyright © 2002 Marcel Dekker, Inc.
Kb1
1.067 1.297 1.553 1.832 2.136 2.464 2.815 3.192 3.589 4.012 4.460 4.931 5.427 5.946 6.490 7.057 7.647 8.264 8.904 9.571 10.263 10.981 11.725 12.495 13.290 14.112 14.960 15.834 16.736 17.666 18.621 19.607 20.585 21.655 22.721 23.816 24.942 26.093 27.274 28.483
Ka1
Ka2
0.052 0.110 0.185 0.282 0.392 0.523 0.677 0.851 1.041 1.254 1.487 1.486 1.40 1.739 2.015 2.014 2.309 2.306 2.622 2.621 2.957 2.955 3.313 3.310 3.691 3.688 4.090 4.085 4.510 4.504 4.952 4.944 5.414 5.405 5.898 5.887 6.403 6.390 6.930 6.915 7.477 7.460 8.047 8.027 8.638 8.615 9.251 9.234 9.885 9.854 10.543 10.507 11.221 11.181 11.923 11.877 12.648 12.597 13.394 13.335 14.164 14.097 14.957 14.882 15.774 15.690 16.614 16.520 17.478 17.373 18.410 18.328 19.278 19.149 20.214 20.072 21.175 21.018 22.162 21.988 23.172 22.982 24.207 24.000 25.270 25.042
Lg1
2.302 2.462 2.623 2.792 2.964 3.144 3.328 3.519 3.716 3.920 4.131
Lb2
Lb1
2.219 2.367 2.518 2.674 2.836 3.001 3.172 3.348 3.528 3.713 3.904
0.344 0.399 0.458 0.519 0.581 0.647 0.717 0.790 0.866 0.943 1.032 1.122 1.216 1.517 1.419 1.526 1.638 1.752 1.872 1.996 2.124 2.257 2.395 2.538 2.683 2.834 2.990 3.151 3.316 3.487 3.662
La1
La2
0.341 0.395 0.492 0.510 0.571 0.636 0.704 0.775 0.849 0.928 1.009 1.096 1.166 1.282 1.379 1.480 1.587 1.694 1.692 1.806 1.805 1.922 1.920 2.042 2.040 2.166 2.163 2.293 2.290 2.424 2.420 2.558 2.554 2.696 2.692 2.838 2.833 2.994 2.978 3.133 3.127 3.287 3.279 3.444 3.436
Table 4
Continued
Atomic number
Element
Kb2
51 Sb 30.387 52 To 31.698 53 I 33.016 54 Xe 34.446 55 Cs 35.819 56 Ba 37.255 57 La 38.728 58 Ce 40.231 59 Pr 41.772 60 Nd 43.298 61 Pm 44.955 62 Sm 46.553 63 Eu 48.241 64 Gd 49.961 65 Tb 51.737 66 Dy 53.491 67 Ho 55.292 68 Er 57.088 69 Tm 58.969 70 Yb 60.959 71 Lu 62.946 72 Hf 64.936 73 Ta 66.999 74 W 69.090 75 Re 71.220 76 Os 73.393 77 Ir 75.605 78 Pt 77.866 79 Au 80.165 80 Hg 82.526 81 Tl 84.904 82 Pb 87.343 83 Bi 89.833 84 Po 92.386 85 At 94.976 86 Rn 97.616 87 Fr 100.305 88 Ra 103.048 89 Ac 105.838 90 Th 108.671 91 Pa 111.575 92 U 114.549 93 Np 117.533 94 Pu 120.592 95 Am 123.706 96 Cm 126.875 97 Bk 130.101 98 Cf 133.383 99 Es 136.724 100 Fm 140.122 Copyright © 2002 Marcel Dekker, Inc.
Kb1
Ka1
Ka2
Lg1
Lb2
Lb1
La1
La2
29.723 30.993 32.292 33.644 34.984 35.376 37.799 39.255 40.746 42.269 43.945 45.400 47.027 48.718 50.391 52.178 53.934 55.690 57.576 59.352 61.282 63.209 65.210 67.233 69.298 71.404 73.549 75.736 77.968 80.258 82.558 94.922 87.335 89.809 92.319 94.877 97.483 100.136 102.846 105.592 108.408 111.289 114.181 117.146 120.163 123.235 126.362 129.544 132.781 136.075
26.357 27.471 28.610 29.802 30.970 32.191 33.440 34.717 36.023 37.359 38.649 40.124 41.529 42.983 44.470 45.985 47.528 49.099 50.730 52.360 54.063 55.757 57.524 59.310 61.131 62.991 64.886 66.820 68.794 70.821 72.860 74.957 77.097 79.296 81.525 83.800 86.119 88.485 90.894 93.334 95.851 98.428 101.005 103.653 106.351 109.098 111.896 114.745 118.646 120.598
26.109 27.200 28.315 29.485 30.623 31.815 33.033 34.276 35.548 36.845 38.160 39.523 40.877 42.280 43.737 45.193 46.686 48.205 49.762 51.326 52.959 54.579 56.270 57.973 59.707 61.477 63.278 65.111 66.980 68.894 70.320 72.794 74.805 76.868 78.956 81.080 83.243 85.446 87.681 89.942 92.271 94.648 97.023 99.457 101.932 104.448 107.023 109.603 112.244 114.926
4.347 4.570 4.800 5.036 5.280 5.531 5.789 6.052 6.322 6.602 6.891 7.180 7.478 7.788 8.104 8.418 8.748 9.089 9.424 9.779 10.142 10.514 10.892 11.283 11.684 12.094 12.509 12.939 13.379 13.828 14.288 14.762 15.244 15.740 16.248 16.768 17.301 17.845 l8.405 18.977 19.559 20.163 20.774 21.401 22.042 22.699 23.370 24.056 24.758 25.475
4.100 4.301 4.507 4.720 4.936 5.156 5.384 5.613 5.850 6.090 6.336 6.587 6.842 7.102 7.368 7.638 7.912 8.188 8.472 8.758 9.048 9.346 9.649 9.959 10.273 10.596 10.918 11.249 11.582 11.923 12.268 12.620 12.977 13.338 13.705 14.077 14.459 14.839 l5.227 15.620 16.022 16.425 16.837 17.254 17.667 18.106 18.540 18.980 19.426 19.879
3.543 4.029 4.220 4.422 4.620 4.828 5.043 5.262 5.489 5.722 5.956 6.206 6.456 6.714 6.979 7.249 7.528 7.810 8.103 8.401 8.709 9.021 9.341 9.670 10.008 10.354 10.706 11.069 11.439 11.823 12.210 12.611 13.021 13.441 13.873 14.316 14.770 15.233 15.712 16.200 16.700 17.218 17.740 18.278 18.829 19.393 19.971 20.562 21.l66 21.785
3.605 3.769 3.937 4.111 4.286 4.467 4.651 4.840 5.034 5.230 5.431 5.636 5.846 6.039 6.275 6.495 6.720 6.948 7.181 7.414 7.654 7.898 8.145 8.396 8.651 8.910 9.173 9.441 9.711 9.987 10.266 10.549 10.836 11.128 11.424 11.724 12.029 12.338 12.650 12.966 13.291 13.613 13.945 14.279 14.618 14.961 15.309 15.661 16.018 16.379
3.595 3.758 3.926 4.098 4.272 4.451 4.635 4.823 5.014 5.208 5.408 5.609 5.816 6.027 6.241 6.457 6.680 6.904 7.135 7.367 7.604 7.843 8.087 8.333 8.584 8.840 9.098 9.360 9.625 9.896 10.170 10.448 10.729 11.014 11.304 11.597 11.894 12.194 12.499 12.808 13.120 13.438 12.758 14.082 14.411 14.743 15.079 15.420 15.764 16.113
APPENDIX III: RADIATIVE TRANSITION PROBABILITIES Radiative Transition Probabilities for K X-ray Lines
20 22 24 26 28 30 32 34 36 38 40 42 44 46 48 50 52 54 56 58 60 62 64 66 68 70 72 74 76 78 80 82 84 86 88 90 92 94 96 98 100
Ca Ti Cr Fe Ni Zn Ge Sc Kr Sr Zr Mo Ru Pd Cd Sn Te Xe Ba Ce Nd Sm Gd Dy Er Yb Hf W Os Pt Hg Pb Po Rn Ra Th U Pu Cm Cf Em
Ka2 =Ka1 0.505 0.505 0.506 0.506 0.507 0.509 0.511 0.513 0.515 0.518 0.520 0.523 0.526 0.528 0.531 0.533 0.536 0.537 0.542 0.545 0.549 0.551 0.556 0.560 0.565 0.568 0.572 0.576 0.580 0.585 0.590 0.595 0.600 0.605 0.612 0.619 0.624 0.631 0.638 0.646 0.652
e e
Kb1 ¼ KM2 þ KM3 þ KM4;5 : Kb2 ¼ KN2;3 þ KO2;3 : c Kb ¼ Kb1 þ Kb2 : d Ka ¼ Ka1 þ Ka2 : Source: From West, 1982–83. e
b
e
a
Copyright © 2002 Marcel Dekker, Inc.
e
Atomic number Element
Kb3 =Kb1
ðKb1 þ Kb3 Þ=Ka1
Kb1 a =Ka1
0.516 0.517 0.518 0.519 0.521 0.522 0.523 0.525 0.526 0.527 0.529 0.531 0.532 0.534 0.535 0.537 0.539 0.541 0.542 0.544 0.546 0.548 0.550 0.552 0.554 0.556
0.116 0.137 0.155 0.172 0.189 0.202 0.215 0.225 0.235 0.244 0.251 0.258 0.264 0.270 0.275 0.280 0.285 0.290 0.294 0.298 0.303 0.307 0.310 0.314 0.317 0.320 0.324 0.326 0.330 0.333 0.336 0.339 0.342 0.345 0.348 0.351 0.354 0.356 0.359 0.362 0.364
0.116 0.137 0.156 0.171 0.187 0.202 0.215 0.225 0.235 0.244 0.252 0.259 0.265 0.271 0.277 0.282 0.287 0.292 0.297 0.301 0.306 0.311 0.314 0.318 0.322 0.325 0.329 0.332 0.336 0.339 0.343 0.346 0.350 0.353 0.356 0.360 0.363 0.366 0.370 0.374 0.377
e
Table 1
Kb2 b =Ka1
Kbc =Kad
0.006 0.013 0.022 0.034 0.043 0.048 0.051 0.054 0.056 0.060 0.064 0.670 0.076 0.082 0.085 0.088 0.089 0.090 0.090 0.091 0.092 0.094 0.097 0.100 0.103 0.106 0.110 0.113 0.118 0.123 0.125 0.130 0.134 0.138
0.069 0.095 0.114 0.128 0.133 0.137 0.142 0.153 0.164 0.175 0.185 0.193 0.201 0.209 0.216 0.222 0.226 0.232 0.240 0.244 0.247 0.250 0.253 0.256 0.259 0.261 0.263 0.267 0.270 0.274 0.277 0.282 0.285 0.288 0.291 0.295 0.299 0.301 0.305 0.309 0.312
Table 2 Atomic number 36 38 40 42 44 46 48 50 52 54 56 58 60 62 64 66 68 70 72 74 76 78 80 82 84 86 88 90 92 94
Radiative Transition Probabilities for L1 X-ray Lines Normalized to Lb3 ¼ 100 Element
Lb3
Lb4
Lg3
Lg2
Kr Sr Zr Mo Ru Pd Cd Sn Te Xe Ba Ce Nd Sm Gd Dy Er Yb Hf W Os Pt Hg Pb Po Rn Ra Th U Pu
100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100
— — — 71.0 64.9 61.9 61.0 61.4 62.6 64.1 65.5 66.7 67.5 69.7 68.0 67.8 67.6 67.5 67.6 68.3 69.6 71.7 74.8 78.8 83.9 89.8 95.0 102.0 110.0 120.0
10.8 14.7 18.1 21.0 23.4 25.5 27.2 28.5 29.5 30.3 30.8 31.2 31.4 31.5 31.5 31.4 31.4 31.4 31.4 31.6 31.9 32.3 33.0 34.0 35.2 36.8 38.7 41.0 43.8 47.5
— — — — — — — — — — — — — — — — — 18.5 20.0 21.5 23.2 25.0 26.9 28.8 30.9 33.1 35.3 37.7 40.2 44.0
Copyright © 2002 Marcel Dekker, Inc.
Table 3 Atomic number 30 32 34 36 38 40 42 44 46 48 50 52 54 56 58 60 62 64 66 68 70 72 74 76 78 80 82 84 86 88 90 92 94
L2 X-ray Lines Normalized to Lb1 ¼ 100 Element
Lb1
LZ
Lg1
Lg6
Zn Ge Se Kr Sr Zr Mo Ru Pd Cd Sn Te Xe Ba Ce Nd Sn Gd Dy Er Yb Hf W Os Pt Hg Pb Po Rn Ra Th U Pu
100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100
12.3 10.4 8.75 7.40 6.25 5.25 4.35 3.63 3.0 2.6 2.35 2.25 2.2 2.16 2.12 2.10 2.10 2.10 2.10 2.10 2.12 2.13 2.16 2.20 2.23 2.28 2.33 2.40 2.45 2.50 2.60 2.80 2.30
— — — — — 0.91 6.71 10.6 13.1 14.5 15.3 15.4 15.6 15.6 15.7 15.8 15.9 16.1 16.5 17.0 17.6 18.4 19.2 20.1 20.9 21.7 22.3 22.8 23.2 23.4 23.7 24.0 24.2
— — — — — — — — — — — — — — — — — — — — — — 0.375 1.73 2.42 2.98 3.45 3.88 4.29 4.74 5.25 5.88 6.65
Copyright © 2002 Marcel Dekker, Inc.
Table 4
L3 X-ray Lines Normalized to La1 ¼ 100
Atomic number 22 24 26 28 30 32 34 36 38 40 42 44 46 48 50 52 54 56 58 60 62 64 66 68 70 72 74 76 78 80 82 84 86 88 90 92 94
Element Ti Cr Fe Ni Zn Ge Se Kr Sr Zr Mo Ru Pd Cd Sn Te Xe Ba Ce Nd Sm Gd Dy Er Yb Hf W Os Pt Hg Pb Po Rn Ra Th U Pu
Source: From West, 1982–83.
Copyright © 2002 Marcel Dekker, Inc.
La1
Lb2,15
La2
Lb5
Lb6
Ll
100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100
— — — — — — — — — 2.43 6.40 9.55 12.1 13.9 15.4 16.4 17.2 17.8 18.2 18.5 18.8 19.2 19.6 20.0 20.5 21.2 21.9 22.7 23.5 24.4 25.3 26.2 26.8 27.3 27.5 27.5 27.0
— — — — — — — — — — 12.5 12.2 12.1 11.9 11.7 11.5 11.3 11.2 11.1 11.1 11.0 11.1 11.1 11.2 11.2 11.3 11.3 11.4 11.4 11.5 11.5 11.4 11.4 11.3 11.1 11.0 10.5
— — — — — — — — — — — — — — — — — — — — — — — — — — 0.242 0.873 1.74 2.62 3.24 3.85 4.28 4.69 4.94 5.20 5.40
— — — — — — — — — — — — — — — — — — — — — — — — — 1.15 1.28 1.38 1.46 1.55 1.59 1.65 1.70 1.75 1.80 1.85 1.89
40.37 26.13 15.35 10.29 7.56 5.96 4.98 4.36 3.98 3.75 3.65 3.58 3.55 3.56 5.59 3.62 3.67 3.73 3.79 3.86 3.92 3.99 4.07 4.15 4.23 4.32 4.42 4.53 4.65 4.78 4.93 5.09 5.27 5.46 5.69 5.93 6.18
APPENDIX IV: NATURAL WIDTHS OF K AND L LEVELS AND Ka X-RAY LINES (FWHM), IN eV Level Element
X-ray line
K
L1
L2
L3
Ka1
Ka2
(0.000) (0.000) 0.00l 0.004 0.015 0.032 0.054 0.083 0.126 0.152
(0.000) (0.000) 0.001 0.004 0.014 0.033 0.054 0.087 0.128 0.156
0.24 0.30 0.36 0.43 0.49 0.57 0.65 0.72 0.81 0.89
0.24 0.30 0.36 0.43 0.49 0.56 0.64 0.72 0.80 0.89
10 11 12 13 14 15 16 17 18 19
Ne Na Mg Al Si P S Cl Ar K
0.24 0.30 0.36 0.42 0.48 0.53 0.59 0.64 0.68 0.74
( < 0.1) (0.2) 0.41 0.73 1.03 1.26 1.49 1.58 1.63 1.92
20 21 22 23 24 25 26 27 28 29
Ca Sc Ti V Cr Mn Fe Co Ni Cu
0.81 0.86 0.94 1.01 1.08 1.16 1.25 1.33 1.44 1.55
2.07 2.21 2.34 2.41 2.54 2.62 2.76 2.79 2.89 3.06
0.17 0.19 0.24 0.26 0.29 0.34 0.37 0.43 0.52 0.62
0.17 0.19 0.22 0.24 0.27 0.32 0.36 0.43 0.48 0.56
0.98 1.05 1.16 1.26 1.35 1.48 1.61 1.76 1.94 2.11
0.98 1.06 1.18 1.28 1.37 1.50 1.62 1.76 1.96 2.17
30 31 32 33 34 35 36 37 38 39
Zn Ga Ge As Se Br Kr Rb Sr Y
1.67 1.82 1.96 2.14 2.33 2.52 2.75 2.99 3.25 3.52
3.28 3.38 3.53 3.79 3.94 4.11 4.28 4.44 4.67 4.71
0.72 0.83 0.95 1.03 1.13 1.21 1.31 1.43 1.54 1.65
0.65 0.76 0.82 0.94 1.00 1.08 1.17 1.27 1.39 1.50
2.32 2.59 2.78 3.08 3.33 3.60 3.92 4.26 4.63 5.02
2.39 2.66 2.92 3.17 3.46 3.73 4.06 4.92 4.79 5.18
40 41 42 43 44 45 46 47 48 49
Zr Nb Mo Tc Ru Rh Pd Ag Cd In
3.84 4.14 4.52 4.91 5.33 5.77 6.24 6.75 7.28 7.91
4.78 3.94 4.25 4.36 4.58 4.73 4.93 4.88 4.87 5.00
1.78 1.87 1.97 2.08 2.23 2.35 2.43 2.57 2.62 2.72
1.57 1.66 1.78 1.91 2.00 2.13 2.25 2.40 2.50 2.65
5.40 5.80 6.31 6.82 7.33 7.90 8.49 9.16 9.79 10.56
5.62 6.01 6.49 6.99 7.56 8.12 8.67 9.32 9.91 10.63
50 51 52 53
Sn Sb Te I
8.49 9.16 9.89 10.6
2.97 3.13 3.32 3.46
2.84 3.00 3.12 3.25
2.75 2.87 2.95 3.08
11.2 12.0 12.8 13.7
11.3 12.2 13.0 13.8
Copyright © 2002 Marcel Dekker, Inc.
Appendix IV
Continued Level
Element
X-ray line
K
L1
L2
L3
Ka1
Ka2
54 55 56 57 58 59
Xe Cs Ba La Ce Pr
11.4 12.3 13.2 14.1 15.1 16.2
3.64 3.78 3.92 4.06 4.21 4.34
3.40 3.51 3.57 3.68 3.80 3.89
3.13 3.25 3.32 3.41 3.48 3.60
14.6 15.6 16.5 17.6 I8.6 19.8
14.8 15.8 16.8 17.8 18.9 20.1
60 61 62 63 64 65 66 67 68 69
Nd Pm Sm Eu Gd Tb Dy Ho Er Tm
17.3 18.5 19.7 21.0 22.3 23.8 25.2 26.8 28.4 30.1
4.52 4.67 4.80 4.91 5.05 5.19 5.25 5.33 5.43 5.47
3.97 4.06 4.15 4.23 4.32 4.43 4.55 4.66 4.73 4.79
3.65 3.75 3.86 3.91 4.01 4.12 4.17 4.26 4.35 4.48
20.9 22.2 23.6 24.9 26.4 27.9 29.4 31.1 32.7 34.6
21.3 22.5 23.8 25.2 26.7 28.2 29.8 31.5 33.1 34.9
70 71 72 73 74 75 76 77 78 79
Yb Lu Hf Ta W Re Os Ir Pt Au
31.9 33.7 35.7 37.7 39.9 42.1 44.4 46.8 49.3 52.0
5.53 5.54 5.63 5.58 5.61 6.18 7.25 8.30 9.39 10.5
4.82 4.92 5.02 5.15 5.33 5.48 5.59 5.69 5.86 6.00
4.60 4.68 4.80 4.88 4.98 5.04 5.16 5.25 5.31 5.41
36.5 38.4 40.5 42.6 44.9 47.2 49.6 52.1 54.6 57.4
36.7 38.7 40.7 42.9 45.2 47.6 50.0 52.5 55.2 58.0
80 81 82 83 84 85 86 87 88 89
Hg Tl Pb Bi Po At Rn Fr Ra Ac
54.6 57.4 60.4 63.4 66.6 69.8 73.3 76.8 80.4 84.1
11.3 12.0 12.2 12.4 12.6 12.8 13.1 13.3 13.4 13.6
6.17 6.32 6.48 6.67 6.83 7.01 7.20 7.47 7.68 7.95
5.50 5.65 5.81 5.98 6.13 6.29 6.41 6.65 6.82 6.98
60.1 63.1 66.2 69.4 72.7 76.1 79.7 83.4 87.2 91.1
60.8 63.8 66.8 70.1 73.4 76.8 80.5 84.2 88.1 92.0
90 91 92 93 94 95 96 97 98 99
Th Pa U Np Pu Am Cm Bk Cf Es
88.0 91.9 96.1 100 105 109 114 119 124 129
13.7 14.3 14.0 14.0 13.5 13.3 13.6 13.8 14.0 14.3
8.18 8.75 9.32 9.91 10.5 10.9 11.4 11.8 12.2 12.7
7.13 7.33 7.43 7.59 7.82 8.04 8.26 8.55 8.75 9.04
95.2 99.3 103.5 108 113 117 122 127 132 138
96.2 100.7 105.4 110 115 120 125 130 136 141
Copyright © 2002 Marcel Dekker, Inc.
APPENDIX IV
Continued Level
Element 100 101 102 103 104 105 106 107 108 109 110
Fm Md No Lr
X-ray line
K
L1
L2
L3
Ka1
Ka2
135 140 145 150 156 162 168 174 181 187
14.4 14.8 15.1 15.9 16.2 16.5 16.8 17.0 17.2 17.6
13.1 13.6 14.0 14.4 14.9 15.5 15.9 16.4 17.0 17.5
9.33 9.61 9.90 10.1 10.5 10.8 11.2 11.6 12.0 12.3
144 150 155 161 167 173 179 185 193 200
148 153 159 165 171 177 184 190 198 205
193
18.1
18.1
12.7
206
211
Source: From Krause and Oliver, 1979.
Copyright © 2002 Marcel Dekker, Inc.
APPENDIX V: WAVELENGTHS OF K SATELLITE LINES (A) Atomic number 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47
Element Na Mg Al Si P S Cl Ar K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag
aII
aI
a3
11.860 11.830 9.845 9.820 8.302 8.283 7.100 7.095 7.082 6.140 6.131 6.117 5.370 5.354 5.341 4.725 3.739 3.356 3.029 2.747 2.502
3.728 3.347 3.021 2.740 2.496 2.282 2.095 1.930 1.784 1.653 1.536 1.431 1.138 1.250 1.173
Source: From Clark, 1963.
Copyright © 2002 Marcel Dekker, Inc.
a3I
a3II
a4
11.810
5.339
9.804 8.267 7.073 6.109 5.344 5.334
a4I
a5
a6
a7
b0
11.742 11.711 9.749 9.724 8.226 8.206 7.069 7.035 7.017 7.026 6.075 6.063
bI
6.816 6.838 4.415
3.721 3.340 3.015 2.733 2.491 2.279 2.093 1.928 1.782 1.651 1.535 1.430 1.335 1.250 1.172 1.101 1.036
3.718 3.338 3.013 2.731 2.489 2.277 2.091 1.926 1.780 1.650 1.533 1.428 1.334 1.248 1.170 1.100 1.035
3.724 3.343 3.018 2.737 2.493 2.280 2.094 1.929 1.783 1.652
3.717 3.337 3.013 2.731 2.490 2.278 2.091 1.927 1.781 1.650 1.534 1.429 1.335 1.249 1.171 1.101 1.036
0.8727
0.8646
0.8721
0.7836
0.7828 0.7432 0.7065
0.7832 0.7436 0.7070
3.496 3.133 2.819 2.551 2.320 2.118 1.783 1.645 1.522
1.061 0.9958 0.9367 0.8832 0.8835 0.7881 0.7457 0.7056 0.6706 0.6369 0.5766 0.5493
0.5833
0.5837 0.5007
3.101 2.789 2.522 2.291 2.090 1.914 1.760 1.623 1.502 1.394 1.296 1.209
b1x
bx
5.800 5.028
6.753 5.792 5.023
bV
bII
b IIIII
5.712 4.400 3.882 3.449 3.087
bIII
bIV
b5
b6
b7
b8
b9
b10
1.116 1.044 0.9786 0.9194
1.064 0.9996 0.9403
1.050 0.9854 0.9629
1.054 0.9889 0.9297
1.047 0.9228
0.8155 0.7699 0.7218 0.6890 0.6531
0.8365 0.7911 0.7492 0.7105 0.6741
0.8234 0.7778 0.7362 0.6973 0.6619 0.6288
0.8259 0.7802 0.7382 0.6993 0.6636 0.6303
0.8193 0.7736 0.7315 0.6924 0.6568 0.6235
0.6256
0.6519
0.5693 0.5427
0.5708 0.5439 0.5189 0.4954
0.5639 0.5374
0.5399
0.5607 0.5452
0.4902
0.4920
5.691
4.395 3.441 3.082 2.772 2.506 2.277 2.078
3.412 3.054 2.749 2.489 2.262 2.066 1.895 1.742 1.607 1.487 1.380 1.282
1.749 1.614 1.494 1.391 1.295 1.207 1.128 1.042 0.9770 0.9177 1.8745 0.8135 0.7681 0.7261 0.6870 0.6510
0.5589
3.404 3.048 2.744
0.5808
0.4943
Copyright © 2002 Marcel Dekker, Inc.
APPENDIX VI: FLUORESCENCE YIELDS AND COSTER^KRONIG TRANSITION PROBABILITIES Table 1
K Shell Fluorescence Yield oK
Atomic number 6 7 8 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44
Element C N O Ne Na Mg Al Si P S Cl Ar K Sc Ca Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc Ru
oK
Atomic number
Element
oK
0.0009 0.0015 0.0022 0.0100 0.020 0.030 0.040 0.055 0.070 0.090 0.105 0.125 0.140 0.165 0.190 0.220 0.240 0.26 0.285 0.32 0.345 0.375 0.41 0.435 0.47 0.50 0.53 0.565 0.60 0.635 0.665 0.685 0.71 0.72 0.755 0.77 0.785 0.80
45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 82 92
Rh Pd Ag Cd In Sn Sb Te I Xe Cs Ba La Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb In Hf Ta W Re Os Ir Pd Au Hg Pb U
0.81 0.82 0.83 0.84 0.85 0.86 0.87 0.875 0.88 0.89 0.895 0.90 0.905 0.91 0.915 0.92 0.925 0.93 0.93 0.935 0.94 0.94 0.945 0.945 0.95 0.95 0.95 0.955 0.955 0.96 0.96 0.96 0.96 0.965 0.965 0.965 0.97 0.97
Source: From Birks, 1971b and Fink, 1974.
Copyright © 2002 Marcel Dekker, Inc.
Table 2
Experimental L Subshell Fluorescence Yields oi
Atomic number
Element
o1
o2
54 56 65 67
Xe Ba Tb Ho
68
Er
70
Yb
71
Lu
72
Hf
73
Ta
74
W
75 76 77
Re Os Ir
78
Pt
79
Au
80
Hg
81
Tl
0.07 0.02
0.319 0.010 0.373 0.025
82
Pb
0.07 0.02 0.09 0.02
0.363 0.015
83
Bi
0.12 0.01 0.095 0.005
0.32 0.04 0.38 0.02
0.06 0.18
0.165 0.018 0.170 0.055 0.185 0.060 0.188 0.011
Copyright © 2002 Marcel Dekker, Inc.
0.25 0.02 0.257 0.013
0.331 0.021
0.39 0.03 0.319 0.010
o3 0.10 0.01 0.05 0.01 0.188 0.016 0.22 0.03 0.169 0.030 0.21 0.03 0.172 0.032 0.20 0.02 0.183 0.011 0.22 0.03 0.251 0.035 0.22 0.03 0.228 0.025 0.27 0.01 0.25 0.03 0.191 0.228 0.013 0.254 0.025 0.207 0.272 0.037 0.284 0.043 0.290 0.030 0.244 0.262 0.036 0.262 0.31 0.04 0.317 0.029 0.291 0.018 0.276 0.31 0.04 0.317 0.025 0.40 0.02 0.32 0.05 0.367 0.050 0.300 0.010 0.37 0.07 0.386 0.053 0.306 0.010 0.330 0.021 0.337 0.315 0.013 0.32 0.35 0.05 0.354 0.028 0.367 0.36 0.37 0.05 0.362 0.029 0.40 0.05
Table 2
Continued
Atomic number
Element
90
Th
91 92
Pa U
96
Cm
o1
0.28 0.06
o2
o3
0.552 0.032 0.55 0.02
0.340 0.018 0.42 0.517 0.042 0.46 0.05 0.44 0.500 0.040 0.515 0.034 0.63 0.02
f13
f23
Source: From Bambynek et al., 1972.
Table 3
Measured L Shell Coster–Kronig Yields
Atomic number
Element
f12 0.66 0.07 0.41 0.36
56 65 67 68 70 73
Ba Tb Ho Er Yb Ta
74 75 77 78 79
W Re Ir Pt Au
80
Hg
81
Tl
0.17 0.05 0.14 0.03
82
Pb
83
Bi
0.15 0.04 0.17 0.05 0.19 0.05
0.76 0.10 0.57 0.10 0.56 0.07 0.56 0.05 0.57 0.03 0.61 0.08 0.58 0.05
0.18 0.02
0.58 0.02
92 93
U Np
0.10 0.04
0.55 0.09
94
Pu
96
Cm
0.038 0.022
0.68 0.04
Source: From Bambynek et al., 1972.
Copyright © 2002 Marcel Dekker, Inc.
< 0.14
0.25 0.13
0.43 0.28
0.19 < 0.36 0.27 0.03 0.30 0.04 0.46 0.06 0.50 0.05 0.51 0.13 0.61 0.07
0.74 0.04
0.066 0.014 0.205 0.034 0.255 0.025 0.142 0.009 0.148 0.010 0.20 0.04
0.22 0.22 0.04 0.08 0.02 0.188 0.010 0.25 0.13 0.169 0.010 0.159 0.013 0.164 0.016 0.156 0.010 þ0.14 0.06 7 0.06 0.164 0.23 0.12 þ0.05 0.02 70.02 0.22 0.08 0.24 0.08 0.188 0.019
Table 4
Theoretical L Subshell Fluorescence Yields oi and Coster–Kronig Yields fij a
Atomic number
Element
o1
13 14 15 16 17 18 19 20 22 24 26
Al Si P S Cl Ar K Ca Ti Cr Fe
3.05-6 9.77-6 2.12-5 3.63-5 5.60-5 8.58-5 1.15-4 1.56-4 2.80-4 2.97-4 3.84-4
28 29 30 32 33 34 35 36
Ni Cu Zn Ge As Se Br Kr
4.63-4
37 38 40
Rb Sr Zr
42
Mo
44 47
Ru Ag
50
Sn
51 54 56 60
Sb Xe Ba Nd
65 67
Tb Ho
70 74
Yb W
79 80 83 85 90 93
Au Hg Bi At Th Np
a
5.23-4 7.70-4 1.40-3 1.30-3 1.85-3 2.19-3 1.32-2 3.00-3 3.97-3 3.96-3 5.75-3 6.34-3 7.74-3 1.02-2 1.01-2 1.30-2 1.30-2 3.11-2 5.84-2 4.46-2 7.46-2 6.00-2 0.112 0.094 0.112 0.115 0.138 0.105 0.098 0.120 0.129 0.197
o2
1.43-3 2.69-3 3.57-3 7.72-3 8.85-3 9.94-3 1.09-2 2.20-2 1.19-2
o3 2.40-3 1.08-3 4.1-4 2.9-4 2.3-4 1.9-4 2.1-4 2.1-4 1.18-3 3.29-3 5.59-3 1.49-3 8.02-3 3.83-3 1.08-2 1.44-2 9.74-3 1.78-2
f12
f13
0.629 0.636 0.652
0.325
0.622
0.322 0.266 0.282 0.302
0.624 0.671 0.547 0.616
2.36-2 1.23-2
0.230 0.225
0.686 0.585
2.43-2 2.95-2 2.01-2 3.73-2 2.59-2 4.50-2 6.02-2 4.49-2 7.37-2
0.249 0.236 0.271 0.166 0.048 0.057 0.052 0.064 0.052 0.072 0.164 0.179 0.168 0.207 0.165
0.646 0.648 0.522 0.689 0.692 0.779 0.786 0.695 0.784 0.693 0.316 0.274 0.336 0.303 0.332
0.202 0.178 0.180 0.195 0.160 0.083 0.101 0.069 0.082 0.069
0.309 0.317 0.316 0.332 0.324 0.644 0.618 0.656 0.612 0.575
6.33-2 9.70-2 8.99-2 0.135 0.120 0.160 0.201
0.287 0.271 0.357 0.352 0.417 0.422 0.529 0.460
0.268 0.253 0.327 0.321 0.389 0.380 0.461 0.472
f23
0.982 0.975 0.971 0.968 0.964 0.965 0.962 0.955 0.313 0.317 0.302
2.24-2 2.94-2 1.89-2 3.50-2 2.45-2 4.18-2 5.47-2 4.30-2 6.56-2 5.67-2 6.16-2 9.12-2 9.07-2 0.133 0.120 0.166 0.203
f12 þ f13
Figures following a sign indicate powers of 10. For example, 3.05-6 means 3.0561076. Copyright 2002 Bambynek Marcel Dekker, Source:©From et al., Inc. 1972.
7.24-2 9.97-2 0.109 2.49-2 4.13-2 5.95-2 7.64-2 8.97-2 9.22-2 0.107 0.115 0.118 0.123 0.124 0.126 0.136 0.152 0.130 0.162 0.136 0.138 0.173 0.151 0.141 0.142 0.131 0.138
0.123 0.117 0.132 0.108 0.101 0.100 0.102 0.209
Table 5 Atomic number 76 79 79 82 82 83 83 83 92 93
96
Measured M Shell Fluorescence Yields and Coster–Kronig Probabilities Element Os Au Au Pb Pb Bi Bi Bi U Np
M o
oLMa
oLMb
0.013 0.0024
0.016 0.003
0.024 0.005
0.030 0.006
0.026 0.005
0.032 0.006
0.030 0.006
0.037 0.005
o1 þ f12o2
ni
oi
0.023 0.001 0.029 0.002 0.037 0.007 0.035 0.002 0.06
Cm
a
Corrected for a 20% contribution from double M shell vacancies. Uncorrected values. Source: From Bambynek et al., 1972. b
Copyright © 2002 Marcel Dekker, Inc.
þ 0.003 0.002 70.002
þ 0.0089 0.0075 70.0075
n1 ¼ 0.065 0.014 n2 ¼ 0.0080 0.029 n3 ¼ 0.062 0.005 n4 ¼ 0.065 0.012 n4 ¼ 0.065 0.012 n4,5 ¼ 0.081 0.016
o5 ¼ 0.06 0.012
þ 0.0051 n1 ¼ 0.068 0.023 o2 ¼ 0.0046 n2 ¼ 0.062 0.019 70.0046 n3 ¼ 0.080 0.006 n4,5 ¼ 0.075 0.012 o5 ¼ 0.075 0.012
Table 6
Adopted Values of Fluorescence and Coster-Kronig Yieldsa ok
Z 5 6 7 8 9
o1
o2
o3
f1
f12
f13
f23
B C N O F
1.7E-03 2.8E-03 5.2E-03 8.3E-03 0.013
10 11 12 13 14
Ne Na Mg Al Si
0.018 0.023 0.030 0.039 0.050
2.9E-05 2.6E-05 3.0E-05
1.2E-03 7.5E-04 3.7E-04
1.2E-03 7.5E-04 3.8E-04
0.962 0.965 0.959
0.32 0.32 0.32
0.64 0.64 0.64
15 16 17 18 19
P S Cl Ar K
0.063 0.078 0.097 0.118 0.140
3.9E-05 7.4E-05 1.2E-04 1.8E-04 2.4E-04
3.1E-04 2.6E-04 2.4E-04 2.2E-04 2.7E-04
3.1E-04 2.6E-04 2.4E-04 2.2E-04 2.7E-04
0.951 0.944 0.939 0.934 0.929
0.32 0.32 0.32 0.31 0.31
0.63 0.62 0.62 0.62 0.62
20 21 22 23 24
Ca Sc Ti V Cr
0.163 0.188 0.214 0.243 0.275
3.1E-04 3.9E-04 4.7E-04 5.8E-04 7.1E-04
3.3E-04 8.4E-04 1.5E-03 2.6E-03 3.7E-03
3.3E-04 8.4E-04 1.5E-03 2.6E-03 3.7E-03
0.920 0.912 0.902 0.894 0.885
0.31 0.31 0.31 0.31 0.31
0.61 0.60 0.59 0.58 0.57
25 26 27 28 29
Mn Fe Co Ni Cu
0.308 0.340 0.373 0.406 0.440
8.4E-04 1.0E-03 1.2E-03 1.4E-03 1.6E-03
5.0E-03 6.3E-03 7.7E-03 8.6E-03 0.010
5.0E-03 6.3E-03 7.7E-03 9.3E-03 0.011
0.877 0.868 0.856 0.847 0.839
0.30 0.30 0.30 0.30 0.30
0.58 0.57 0.56 0.55 0.54
0.028 0.028
30 31 32 33 34
Zn Ga Ge As Se
0.474 0.507 0.535 0.562 0.589
1.8E-03 2.1E-03 2.4E-03 2.8E-03 3.2E-03
0.011 0.012 0.013 0.014 0.016
0.012 0.013 0.015 0.016 0.018
0.831 0.822 0.815 0.809 0.804
0.29 0.29 0.28 0.28 0.28
0.54 0.53 0.53 0.53 0.52
0.026 0.032 0.050 0.063 0.076
35 36 37 38 39
Br Kr Rb Sb Y
0.618 0.643 0.667 0.690 0.710
3.6E-03 4.1E-03 4.6E-03 5.1E-03 5.9E-03
0.018 0.020 0.022 0.024 0.026
0.020 0.022 0.024 0.026 0.028
0.800 0.797 0.794 0.790 0.785
0.28 0.27 0.27 0.27 0.26
0.52 0.52 0.52 0.52 0.52
0.088 0.100 0.109 0.117 0.126
40 41 42 43 44
Zr Nb Mo Tc Ru
0.730 0.747 0.765 0.780 0.794
6.8E-03 9.4E-03 0.010 0.011 0.012
0.028 0.031 0.034 0.037 0.040
0.031 0.034 0.037 0.040 0.043
0.779 0.713 0.712 0.711 0.709
0.26 0.10 0.10 0.10 0.10
0.52 0.61 0.61 0.61 0.61
0.132 0.137 0.141 0.144 0.148
45 46 47 48 49
Rh Pd Ag Cd In
0.808 0.820 0.831 0.843 0.853
0.013 0.014 0.016 0.018 0.020
0.043 0.047 0.051 0.056 0.061
0.046 0.049 0.052 0.056 0.060
0.705 0.700 0.694 0.688 0.681
0.10 0.10 0.10 0.10 0.10
0.60 0.60 0.59 0.59 0.59
0.150 0.151 0.153 0.155 0.157
Copyright © 2002 Marcel Dekker, Inc.
Table 6
Continued
Z
ok
o1
o2
o3
f1
f12
f13
f23
50 51 52 53 54
Sn Sb Te I Xe
0.862 0.870 0.877 0.884 0.891
0.037 0.039 0.041 0.044 0.046
0.065 0.069 0.074 0.079 0.083
0.064 0.069 0.074 0.079 0.085
0.439 0.448 0.495 0.461 0.466
0.17 0.17 0.18 0.18 0.19
0.27 0.28 0.28 0.28 0.28
0.157 0.156 0.155 0.154 0.154
55 56 57 58 59
Cs Ba La Ce Pr
0.897 0.902 0.907 0.912 0.917
0.049 0.052 0.055 0.058 0.061
0.090 0.096 0.103 0.110 0.117
0.091 0.097 0.104 0.111 0.118
0.470 0.474 0.478 0.482 0.485
0.19 0.19 0.19 0.19 0.19
0.28 0.28 0.29 0.29 0.29
0.154 0.153 0.153 0.153 0.153
60 61 62 63 64
Nd Pm Sm Eu Gd
0.921 0.925 0.929 0.932 0.935
0.064 0.066 0.071 0.075 0.079
0.124 0.132 0.140 0.149 0.158
0.125 0.132 0.139 0.147 0.155
0.488 0.490 0.492 0.493 0.493
0.19 0.19 0.19 0.19 0.19
0.30 0.30 0.30 0.30 0.30
0.152 0.151 0.150 0.149 0.147
65 66 67 68 69
Tb Dy Ho Er Tm
0.938 0.941 0.944 0.947 0.949
0.083 0.089 0.044 0.100 0.106
0.167 0.178 0.189 0.200 0.211
0.164 0.174 0.182 0.192 0.201
0.493 0.492 0.490 0.487 0.483
0.19 0.19 0.19 0.19 0.19
0.30 0.30 0.30 0.30 0.29
0.145 0.143 0.142 0.140 0.139
70 71 72 73 74
Yb Lu Hf Ta W
0.951 0.953 0.955 0.957 0.958
0.112 0.120 0.128 0.137 0.147
0.222 0.234 0.246 0.258 0.270
0.210 0.220 0.231 0.243 0.255
0.478 0.472 0.465 0.457 0.447
0.19 0.19 0.18 0.18 0.17
0.29 0.28 0.28 0.28 0.28
0.138 0.136 0.135 0.134 0.133
75 76 77 78 79
Re Os Ir Pt Au
0.959 0.961 0.962 0.963 0.964
0.144 0.130 0.120 0.114 0.107
0.283 0.295 0.308 0.321 0.334
0.268 0.281 0.294 0.306 0.320
0.485 0.552 0.603 0.640 0.672
0.16 0.16 0.15 0.14 0.14
0.33 0.39 0.45 0.50 0.53
0.130 0.128 0.126 0.124 0.122
80 81 82 83 84
Hg Tl Pb Bi Po
0.965 0.966 0.967 0.968 0.968
0.107 0.107 0.112 0.117 0.122
0.347 0.360 0.373 0.387 0.401
0.333 0.347 0.360 0.373 0.386
0.690 0.696 0.696 0.694 0.689
0.13 0.13 0.12 0.11 0.11
0.56 0.57 0.58 0.58 0.58
0.120 0.118 0.116 0.113 0.111
85 86 87 88 89
At Rn Fr Ra Ac
0.969 0.969 0.970 0.970 0.971
0.128 0.134 0.139 0.146 0.153
0.415 0.429 0.443 0.456 0.468
0.399 0.411 0.424 0.437 0.450
0.685 0.682 0.677 0.672 0.664
0.10 0.10 0.10 0.09 0.09
0.59 0.58 0.58 0.58 0.58
0.111 0.110 0.109 0.108 0.108
90 91 92 93 94
Th Pa U Np Pu
0.971 0.972 0.972 0.973 0.973
0.161 0.162 0.176 0.187 0.205
0.479 0.472 0.467 0.466 0.464
0.463 0.476 0.489 0.502 0.514
0.660 0.664 0.652 0.642 0.605
0.09 0.08 0.08 0.07 0.05
0.57 0.58 0.57 0.57 0.56
0.108 0.139 0.167 0.192 0.198
Copyright © 2002 Marcel Dekker, Inc.
Table 6
Continued
Z
ok
o1
o2
o3
f1
f12
f13
f23
95 96 97 98 99
Am Cm Bk Cf Es
0.974 0.974 0.975 0.975 0.975
0.218 0.228 0.236 0.244 0.253
0.471 0.479 0.485 0.490 0.497
0.526 0.539 0.550 0.560 0.570
0.595 0.587 0.580 0.573 0.565
0.05 0.04 0.04 0.03 0.03
0.55 0.55 0.54 0.54 0.54
0.203 0.200 0.198 0.197 0.196
100 101 102 103 104
Fm Md No Lr
0.976 0.976 0.976 0.977 0.977
0.263 0.272 0.280 0.282 0.291
0.506 0.515 0.524 0.533 0.544
0.579 0.588 0.596 0.604 0.611
0.556 0.548 0.540 0.538 0.531
0.03 0.02 0.02 0.01 0.01
0.53 0.53 0.52 0.53 0.52
0.194 0.191 0.189 0.185 0.181
105 106 107 108 109
0.977 0.978 0.978 0.978 0.978
0.300 0.310 0.320 0.331 0.343
0.553 0.562 0.573 0.584 0.590
0.618 0.624 0.630 0.635 0.640
0.522 0.513 0.505 0.497 0.488
0.01
0.51 0.51 0.50 0.50 0.49
0.178 0.174 0.171 0.165 0.163
110
0.979
0.354
0.598
0.644
0.477
0.48
0.158
a Note: 1.7E-03 ¼ 1.761073; designation for L shell is omitted, read, for example, o1 ¼ oL1 ; f1 ¼ f12 þ f13 . Source: From Krause, 1979.
Copyright © 2002 Marcel Dekker, Inc.
Table 7 Z
Auger Yields for K, L1, L2 and L3 Levelsa aK
a1
a2
a3
Z
aK
a1
a2
a3
B C N O F
0.998 0.997 0.995 0.992 0.987
50 51 52 53 54
Sn Sb Te I Xe
0.138 0.130 0.123 0.116 0.109
0.524 0.513 0.504 0.495 0.488
0.778 0.775 0.771 0.767 0.763
0.936 0.931 0.926 0.921 0.915
10 11 12 13 14
Ne Na Mg Al Si
0.982 0.977 0.970 0.961 0.950
0.038 0.035 0.041
0.999 0.999 1.000
0.999 0.999 1.000
55 56 57 58 59
Cs Ba La Ce Pr
0.103 0.098 0.093 0.088 0.083
0.481 0.474 0.467 0.460 0.454
0.756 0.751 0.744 0.737 0.730
0.909 0.903 0.896 0.889 0.882
15 16 17 18 19
P S Cl Ag K
0.937 0.922 0.903 0.882 0.860
0.049 0.056 0.061 0.066 0.071
1.000 1.000 1.000 1.000 1.000
1.000 1.000 1.000 1.000 1.000
60 61 62 63 64
Nd Pm Sm Eu Gd
0.079 0.075 0.071 0.068 0.065
0.448 0.444 0.437 0.432 0.428
0.724 0.717 0.710 0.702 0.695
0.875 0.868 0.861 0.853 0.845
20 21 22 23 24
Ca Sc Ti V Cr
0.837 0.812 0.786 0.757 0.725
0.080 0.088 0.098 0.105 0.114
1.000 0.999 0.999 0.997 0.996
1.000 0.999 0.999 0.997 0.996
65 66 67 68 69
Tb Dy Ho Er Tm
0.062 0.059 0.056 0.053 0.051
0.424 0.419 0.416 0.413 0.411
0.688 0.679 0.669 0.660 0.650
0.836 0.826 0.818 0.808 0.799
25 26 27 28 29
Mn Fe Co Ni Cu
0.692 0.660 0.627 0.394 0.560
0.122 0.131 0.143 0.152 0.159
0.995 0.994 0.992 0.963 0.962
0.995 0.994 0.992 0.991 0.989
70 71 72 73 74
Yb Lu Hf Ta W
0.049 0.047 0.045 0.043 0.042
0.410 0.408 0.407 0.406 0.406
0.640 0.630 0.619 0.608 0.597
0.790 0.780 0.769 0.757 0.745
30 31 32 33 34
Zn Ga Ge As Se
0.526 0.493 0.465 0.438 0.411
0.167 0.176 0.183 0.188 0.193
0.963 0.956 0.937 0.923 0.908
0.988 0.987 0.985 0.984 0.982
75 76 77 78 79
Re Os Ir Pt Au
0.041 0.039 0.038 0.037 0.036
0.371 0.318 0.277 0.246 0.221
0.587 0.577 0.566 0.555 0.344
0.732 0.719 0.706 0.694 0.680
35 36 37 38 39
Br Kr Rb Sr Y
0.382 0.357 0.333 0.310 0.290
0.196 0.199 0.201 0.205 0.209
0.894 0.880 0.869 0.859 0.848
0.980 0.978 0.976 0.974 0.972
80 81 82 83 84
Hg Tl Pb Bi Po
0.035 0.034 0.033 0.032 0.032
0.203 0.197 0.192 0.189 0.189
0.533 0.522 0.511 0.500 0.488
0.667 0.653 0.640 0.627 0.614
40 41 42 43 44
Zr Nb Mo Tc Ru
0.270 0.253 0.235 0.220 0.206
0.214 0.278 0.278 0.278 0.279
0.840 0.832 0.825 0.819 0.812
0.969 0.966 0.963 0.960 0.957
85 86 87 88 89
At Rn Fr Ra Ac
0.031 0.031 0.030 0.030 0.029
0.187 0.184 0.184 0.182 0.183
0.474 0.461 0.448 0.436 0.424
0.601 0.589 0.576 0.563 0.590
45 46 47 48 49
Rh Pd Ag Cd In
0.192 0.180 0.169 0.157 0.147
0.282 0.286 0.290 0.294 0.299
0.807 0.802 0.796 0.789 0.782
0.954 0.951 0.948 0.944 0.940
90 91 92 93 94
Th Pa U Np Pu
0.029 0.028 0.028 0.027 0.027
0.179 0.174 0.172 0.171 0.190
0.413 0.389 0.366 0.342 0.338
0.537 0.524 0.511 0.498 0.486
5 6 7 8 9
Copyright © 2002 Marcel Dekker, Inc.
Table 7 Z
Continued aK
a1
a2
a3
Z
aK
a1
a2
a3
95 96 97 98 99
Am Cm Bk Cf Es
0.026 0.026 0.025 0.025 0.025
0.167 0.185 0.184 0.183 0.182
0.326 0.321 0.317 0.313 0.307
0.474 0.461 0.450 0.440 0.430
105 106 107 108 109
0.023 0.022 0.022 0.022 0.022
0.178 0.177 0.175 0.172 0.169
0.269 0.264 0.256 0.251 0.247
0.382 0.376 0.370 0.365 0.360
100 101 102 103 104
Fm Md No Lr
0.024 0.024 0.024 0.023 0.023
0.181 0.180 0.180 0.180 0.178
0.300 0.294 0.287 0.282 0.275
0.421 0.412 0.404 0.396 0.389
110
0.021
0.169
0.244
0.356
a Designation for L shell is omitted; read, for example, a1 aL1. Source: From Krause, 1979.
Copyright © 2002 Marcel Dekker, Inc.
APPENDIX VII: COEFFICIENTS FOR CALCULATING THE PHOTOELECTRIC ABSORPTION CROSS SECTIONS (BARNS=ATOM) VIA ln^ln REPRESENTATION Table 1
E > EK a
Atomic number
Element
Atomic weight
H He Li Be B C N O F Ne Na Mg Al Si P S Cl Ar K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc Ru Rh
1.008 4.003 6.940 9.012 10.811 12.010 14.008 16.000 19.000 20.183 22.997 24.320 26.970 28.086 30.975 32.066 35.457 39.944 39.102 40.080 44.960 47.900 50.942 51.996 54.940 55.850 58.933 58.690 63.540 65.380 69.720 72.590 74.920 78.960 79.920 83.800 85.480 87.620 88.905 91.220 92.906 95.950 99.000 101.070 102.910
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45
Copyright © 2002 Marcel Dekker, Inc.
A0 2.44964 6.06488 7.75370 9.04511 9.95057 1.06879 þ 1 1.12765 þ 1 1.17130 þ 1 1.20963 þ 1 1.24485 þ 1 1.26777 þ 1 1.28793 þ 1 1.31738 þ 1 1.32682 þ 1 1.33735 þ 1 1.37394 þ 1 1.36188 þ 1 1.39491 þ 1 1.37976 þ 1 1.42950 þ 1 1.39664 þ 1 1.43506 þ 1 1.47601 þ 1 1.48019 þ 1 1.48965 þ 1 1.43456 þ 1 1.47047 þ 1 1.42388 þ 1 1.45808 þ 1 1.44118 þ 1 1.36182 þ 1 1.39288 þ 1 1.34722 þ 1 1.30756 þ 1 1.32273 þ 1 1.35927 þ 1 1.30204 þ 1 1.35888 þ 1 1.34674 þ 1 1.27538 þ 1 1.33843 þ 1 1.39853 þ 1 1.28214 þ 1 1.26658 þ 1 1.21760 þ 1
A1
A2
A3
73.34953 73.29055 72.81801 72.83487 72.74173 72.71400 72.65400 72.57229 72.44148 72.45819 72.24521 72.12574 72.18203 71.98174 71.86342 72.04786 71.71937 71.82276 71.54015 71.88644 71.40872 71.66322 71.88867 71.82430 71.79872 71.23491 71.38933 79.67736-1 71.18375 79.33083-1 73.18459-1 74.79613-1 77.73513-2 1.83235-1 1.37130-1 73.05214-2 3.82736-1 2.20194-3 1.91023-1 6.97409-1 2.81028-1 71.17426-1 7.51993-1 8.85020-1 1.19682
74.71370-2 71.07256-1 72.41738-1 72.10021-1 72.15138-1 72.00530-1 72.00445-1 72.05893-1 72.34461-1 72.12591-1 72.74873-1 72.99392-1 72.58960-1 73.16950-1 73.39440-1 72.73259-1 73.54154-1 73.28827-1 73.94528-1 72.83647-1 74.14365-1 73.31539-1 72.71861-1 72.79116-1 72.83664-1 74.18785-1 73.86631-1 74.78070-1 74.13850-1 74.77357-1 76.11348-1 75.72897-1 76.60456-1 76.94264-1 76.83203-1 76.51340-1 77.32427-1 76.38940-1 76.86616-1 77.89307-1 76.86607-1 75.91094-1 77.87006-1 78.11144-1 78.66697-1
7.09962-3 1.44465-2 2.62542-2 2.29526-2 2.27845-2 2.07248-2 2.00765-2 1.99244-2 2.19537-2 1.96489-2 2.50270-2 2.67643-2 2.22840-2 2.73928-2 2.88858-2 2.29976-2 2.90841-2 2.74382-2 3.23561-2 2.26263-2 3.34355-2 2.62065-2 2.15792-2 2.17324-2 2.22095-2 3.21662-2 3.03286-2 3.66138-2 3.12088-2 3.62829-2 4.58138-2 4.31277-2 4.92177-2 5.02280-2 4.95424-2 4.77616-2 5.29874-2 4.60070-2 4.97356-2 5.64531-2 4.86607-2 4.17843-2 5.58668-2 5.73759-2 6.06931-2
Table 1
Continued
Atomic number
Element
Atomic weight
Pd Ag Cd In Sn Sb Te I Xe Cs Ba La Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Rn Th U Pu
106.400 107.880 112.410 114.820 118.690 121.760 127.600 126.910 131.300 132.910 137.360 138.920 140.130 140.920 144.270 147.000 150.350 152.000 157.260 158.930 162.510 164.940 167.270 168.940 173.040 174.990 178.500 180.950 183.920 186.200 190.200 192.200 195.090 197.200 200.610 204.390 207.210 209.000 222.000 232.000 238.070 239.100
46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 86 90 92 94 a
A0 1.39389 þ 1 1.33926 þ 1 1.15254 þ 1 1.18198 þ 1 1.30323 þ 1 9.06999 1.16656 þ 1 1.21075 þ 1 1.10857 þ 1 1.13757 þ 1 1.02250 þ 1 1.09780 þ 1 1.02725 þ 1 1.10156 þ 1 1.17632 þ 1 1.13864 þ 1 1.19223 þ 1 1.16168 þ 1 9.91968 1.13818 þ 1 1.14845 þ 1 8.75203 1.20195 þ 1 1.25613 þ 1 7.42791 1.26387 þ 1 7.58160 8.65271 7.57541 1.36944 1.37534 þ 1 1.25506 þ 1 1.27882 þ 1 4.96352 1.97594 þ 1 1.52879 þ 1 8.63374 9.44293 1.51782 þ 1 1.34336 þ 1 1.37951 þ 1 1.82787 þ 1
Notation abbreviated as in Appendix VI, Table 4.
Copyright © 2002 Marcel Dekker, Inc.
A1
A2
A3
1.64528-1 4.41380-1 1.07714 1.45768 7.90788-1 3.28791 1.71052 1.43635 2.08356 1.94161 2.67835 2.23814 2.74562 2.22056 1.79481 2.05593 1.79546 1.97533 3.03111 2.14447 2.10451 3.71822 1.84815 1.57523 4.28955 1.55476 4.47037 3.73117 4.28874 7.79444 1.02122 1.63090 1.63605 5.79212 71.97990 2.73664-1 3.69400 3.44965 3.49021-1 1.34805 1.23983 71.17371
76.62170-1 76.93711-1 78.31424-1 78.88529-1 77.62349-1 71.26203 79.48281-1 78.82038-1 71.01209 79.83232-1 71.12648 71.03549 71.14174 71.02216 79.36661-1 79.88180-1 79.42902-1 79.70901-1 71.17520 79.99222-1 79.89870-1 71.29273 79.39582-1 78.90467-1 71.35167 78.81094-1 71.42808 71.26359 71.34998 71.99822 77.77126-1 78.75676-1 78.98523-1 71.61842 72.76981-1 76.38890-1 71.21312 71.19886 76.37638-1 78.13282-1 78.01545-1 73.68344-1
4.76289-2 4.82085-2 5.79120-2 6.05982-2 5.27872-2 8.53470-2 6.53213-2 6.03575-2 6.90310-2 6.7l986-2 7.62669-2 7.02339-2 7.74162-2 6.90465-2 6.35332-2 6.69106-2 6.44202-2 6.58459-2 7.86751-2 6.75569-2 6.69382-2 8.55026-2 6.38106-2 6.09779-2 8.66136-2 6.02036-2 9.39044-2 8.23539-2 8.65200-2 1.26225-1 5.38811-2 5.92011-2 6.18550-2 1.02911-1 2.68856-2 4.57495-2 7.74601-2 7.83484-2 4.51377-2 5.55664-2 5.53596-2 2.98738-2
Table 2
EL1 < E < EKa
Atomic number
Element
Atomic weight
Na Mg Al Si P S Cl Ar K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te I Xe Cs Ba La Ce
22.997 24.310 26.970 28.086 30.975 32.066 35.457 39.944 39.102 40.080 44.960 47.900 50.942 51.996 54.040 55.850 58.933 58.690 63.540 65.380 69.720 72.590 74.920 78.960 79.920 83.800 85.480 87.620 88.905 91.220 92.906 95.950 99.000 101.070 102.910 106.400 107.880 112.410 114.820 118.690 121.760 127.600 126.910 131.300 132.910 137.360 138.920 140.130
11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58
Copyright © 2002 Marcel Dekker, Inc.
A0 1.02355 þ 1 1.05973 þ 1 1.08711 þ 1 1.12237 þ 1 1.15508 þ 1 1.18181 þ 1 1.20031 þ 1 1.22960 þ 1 1.24878 þ 1 1.27044 þ 1 1.28949 þ 1 1.31075 þ 1 1.32514 þ 1 1.34236 þ 1 1.35761 þ 1 1.36696 þ 1 1.38699 þ 1 1.39848 þ 1 1.42439 þ 1 1.43221 þ 1 1.44795 þ 1 1.46813 þ 1 1.46431 þ 1 1.47048 þ 1 1.48136 þ 1 1.49190 þ 1 1.49985 þ 1 1.50114 þ 1 1.51822 þ 1 1.52906 þ 1 1.52088 þ 1 1.53494 þ 1 1.55086 þ 1 1.54734 þ 1 1.55757 þ 1 1.55649 þ 1 1.56869 þ 1 1.59668 þ 1 1.62101 þ 1 1.58638 þ 1 1.57557 þ 1 1.61087 þ 1 1.64086 þ 1 1.63098 þ 1 1.65418 þ 1 1.66217 þ 1 1.63134 þ 1 1.65862 þ 1
A1
A2
72.55905 72.89818 72.77860 72.73694 72.92200 72.64618 72.41694 72.63279 72.53656 72.55011 72.40609 72.53576 72.49765 72.51532 72.49761 72.39195 72.50669 72.48080 72.58677 72.62384 72.54469 72.69285 72.48397 72.38853 72.42347 72.42418 72.39108 72.28169 72.38946 72.38703 72.20278 72.26646 72.33733 72.23080 72.24976 72.17229 72.22636 72.38363 72.51838 72.19010 72.04460 72.27876 72.48214 72.31679 72.46363 72.48972 72.20156 72.36288
71.19524-1 2.34506-1 1.75853-1 1.27557-1 2.54262-1 79.68049-2 72.40897-1 77.36600-2 71.04892-1 79.43195-2 71.77791-1 79.57177-2 7 1.06383-1 71.01999-1 71.05943-1 71.37648-1 78.69945-2 78.88115-2 76.67398-2 72.64926-2 77.57204-2 72.08355-2 77.96180-2 71.05877-1 79.14590-2 78.76447-2 79.59473-2 71.26485-1 78.81174-2 79.12292-2 71.36759-1 71.16881-1 79.87857-1 71.19454-1 71.13377-1 71.27652-1 71.12223-1 78.01104-2 75.49951-2 71.13539-1 71.40745-1 79.29405-2 75.07179-2 78.54498-2 75.42849-2 74.49623-2 79.80569-2 76.54708-2
Table 2 Atomic number 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 86 90 92 94 a
Continued Element
Atomic weight
Pr Nd Pm Sm Eu Cd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Rn Th U Pu
140.920 144.270 147.000 150.350 152.000 157.260 158.930 162.510 164.940 167.270 168.940 173.040 174.990 178.500 180.950 183.920 186.200 190.200 192.200 195.090 197.200 200.610 204.390 207.210 209.000 222.000 232.000 238.070 239.100
Notation abbreviated as in Appendix VI, Table 4.
Copyright © 2002 Marcel Dekker, Inc.
A0 1.67179 þ 1 1.65964 þ 1 1.68368 þ 1 1.68725 þ 1 1.70692 þ 1 1.71159 þ 1 1.71499 þ 1 1.73446 þ 1 1.76583 þ 1 1.77988 þ 1 1.74250 þ 1 1.69795 þ 1 1.72638 þ 1 1.64329 þ 1 1.72410 þ 1 1.72533 þ 1 1.78750 þ 1 1.73525 þ 1 1.65270 þ 1 1.73636 þ 1 1.74240 þ 1 1.71857 þ 1 1.77379 þ 1 1.77963 þ 1 1.75348 þ 1 1.75028 þ 1 1.85481 þ 1 1.75258 þ 1 1.75519 þ 1
A1
A2
72.40326 72.26073 72.38881 72.39051 72.48046 72.47838 72.45507 72.54821 72.72523 72.74671 72.51103 72.22577 72.37189 71.82851 72.30313 72.23874 72.61051 72.28550 71.76315 72.21112 72.23911 72.08470 72.37745 72.37691 72.23353 72.13876 72.61281 72.07237 72.02162
76.12619-2 78.72426-2 76.45041-2 76.01080-2 74.47055-2 74.37107-2 74.71370-2 73.17606-2 78.19409-4 72.87580-3 73.29454-2 77.32557-2 74.95994-2 71.32268-1 75.91006-2 77.27338-2 71.36093-2 75.88047-2 71.35232-1 77.30934-2 76.63720-2 78.53294-2 74.33223-2 74.55883-2 75.96161-2 77.24638-2 77.90574-3 77.23932-2 78.22940-2
Table 3
EM1 < E < EL3 a
Atomic number 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79
Element Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te I Xe Cs Ba La Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt Au
Copyright © 2002 Marcel Dekker, Inc.
Atomic weight 69.720 72.590 74.920 78.960 79.920 83.800 85.480 87.620 88.905 91.220 92.906 95.950 99.000 101.070 102.910 106.400 107.880 112.410 114.820 118.690 121.760 127.600 126.910 131.300 132.910 137.360 138.920 140.130 140.920 144.270 147.000 150.350 152.000 157.260 158.930 162.510 164.940 167.270 168.940 173.040 174.990 178.500 180.950 183.920 186.200 190.200 192.200 195.090 197.200
A0
A1
1.22646 þ 1 1.24133 þ 1 1.25392 þ 1 1.26773 þ 1 1.27612 þ 1 1.28898 þ 1 1.30286 þ 1 1.31565 þ 1 1.32775 þ 1 1.34508 þ 1 1.35434 þ 1 1.36568 þ 1 1.37498 þ 1 1.38782 þ 1 1.40312 þ 1 1.41392 þ 1 1.41673 þ 1 1.43497 þ 1 1.44115 þ 1 1.45572 þ 1 1.46268 þ 1 1.47125 þ 1 1.47496 þ 1 1.47603 þ 1 1.49713 þ 1 1.50844 þ 1 1.51863 þ 1 1.52693 þ 1 1.53379 þ 1 1.54353 þ 1 1.55131 þ 1 1.56006 þ 1 1.57063 þ 1 1.57159 þ 1 1.58415 þ 1 1.59225 þ 1 1.60140 þ 1 1.60672 þ 1 1.61269 þ 1 1.61794 þ 1 1.62289 þ 1 1.62758 þ 1 1.63068 þ 1 1.62613 þ 1 1.63564 þ 1 1.64233 þ 1 1.65144 þ 1 1.67024 þ 1 1.64734 þ 1
72.68965 72.53085 72.41380 72.39750 72.37730 72.26021 72.38693 72.36655 72.43174 72.50201 72.50135 72.48982 72.44737 72.48066 72.61303 72.57206 72.48078 72.52756 72.49401 72.56792 72.55562 72.54324 72.48179 72.45068 72.53145 72.56341 72.58287 72.58174 72.57086 72.59006 72.59623 72.61328 72.63481 72.60843 72.64040 72.65289 72.67903 72.67587 72.67886 72.67715 72.67128 72.66622 72.66148 72.60672 72.62453 72.63163 72.64832 72.71631 72.57834
Table 3
Continued
Atomic number 80 81 82 83 86 90 92 94 a
Element
Atomic weight
Hg Tl Pb Bi Rn Th U Pu
200.610 204.390 207.210 209.000 222.000 232.000 238.070 239.100
A0
A1
1.65903 þ 1 1.66564 þ 1 1.67131 þ 1 1.67078 þ 1 1.69000 þ 1 1.70483 þ 1 1.70353 þ 1 1.72953 þ 1
72.60670 72.61593 72.61538 72.58648 72.60945 72.58569 72.56903 72.62164
Notation abbreviated in Appendix VI, Table 4.
Table 4
E < EM5 a
Atomic number 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 86 90 92 94
Element Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Rn Th U Pu
a
Atomic weight 147.000 150.350 152.000 157.260 158.930 162.510 164.940 167.270 168.940 173.040 174.990 178.500 180.950 183.920 186.200 190.200 192.200 195.090 197.200 200.610 204.390 207.210 209.000 222.000 232.000 238.070 239.100
Notation abbreviated as in Appendix VI, Table 4. Source: From McMaster et al., 1969.
Copyright © 2002 Marcel Dekker, Inc.
A0 1.55131 þ 1 1.56006 þ 1 1.57063 þ 1 1.57159 þ 1 1.58415 þ 1 1.59225 þ 1 1.60140 þ 1 1.60672 þ 1 1.61269 þ 1 1.39111 þ 1 1.39813 þ 1 1.40548 þ 1 1.41313 þ 1 1.42536 þ 1 1.42392 þ 1 1.42795 þ 1 1.43422 þ 1 1.43785 þ 1 1.44398 þ 1 1.45195 þ 1 1.45473 þ 1 1.45771 þ 1 1.46832 þ 1 1.47243 þ 1 1.47730 þ 1 1.49036 þ 1 1.48535 þ 1
A1 72.59623 72.61328 72.63481 72.60843 72.64040 72.65289 72.67903 72.67587 72.67886 72.40380 72.40841 72.42829 72.47214 72.32582 72.35326 72.21971 72.40183 72.34834 72.32838 72.33016 72.26773 72.25279 72.30940 72.12905 71.91192 72.12148 71.87733
APPENDIX VIII: COEFFICIENTS FOR CALCULATING THE INCOHERENT COLLISION CROSS SECTIONS C (BARNS=ATOM) VIA THE ln^ln REPRESENTATION Atomic Number 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46
Element H He Li Be B C N O F Ne Na Mg Al Si P S Cl Ar K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc Ru Rh Pd
Copyright © 2002 Marcel Dekker, Inc.
A0
A1
A2
A3
72.l5772 72.56357 71.08740 76.90079-1 77.91177-1 79.87878-1 71.23693 71.73679 71.87570 71.75510 79.67717-1 75.71611-1 74.39322-1 74.14971-1 74.76903-1 76.56419-1 77.18627-1 76.82105-1 73.44007-1 79.82420-2 71.59831-1 72.30573-1 73.08103-1 73.87641-1 72.47059-1 73.42379-1 74.28804-1 75.04360-1 75.70210-1 74.20535-1 73.58218-1 73.34383-1 73.39189-1 74.32927-1 74.48001-1 73.91810-1 71.28039-1 7.99161-2 6.29057-2 3.66697-2 2.02289-4 75.62860-2 7.57616-2 74.24981-2 71.60399-1 72.67564-1
1.32685 2.02536 1.03368 9.46448-1 1.21611 1.46693 1.74510 2.17686 2.32016 2.24226 1.61794 1.35498 1.30867 1.34868 1.46032 1.65408 1.74294 1.74279 1.49236 1.32829 1.39055 1.45848 1.52879 1.59727 1.49722 1.57245 1.64129 1.70040 1.75042 1.63400 1.60050 1.60237 1.62535 1.72833 1.76082 1.73010 1.53044 1.38397 1.41577 1.45207 1.49347 1.55778 1.44950 1.54639 1.64861 1.73740
73.05620-1a 74.48710-1 71.90377-1 71.71142-1 72.39087-1 72.93743-1 73.54660-1 74.49050-1 74.75412-1 74.47640-1 72.87191-1 72.22491-1 72.11648-1 72.22315-1 72.51331-1 72.98623-1 73.19429-1 73.17646-1 72.54135-1 72.13747-1 72.25849-1 72.39160-1 72.52768-1 72.66240-1 72.38781-1 72.53198-1 72.66013-1 72.76443-1 72.84555-1 72.53646-1 72.44908-1 72.45555-1 72.50783-1 72.77138-1 72.85099-1 72.76824-1 2.27403-1 71.92225-1 71.99713-1 72.08122-1 72.17419-1 72.33341-1 72.04890-1 72.26470-1 72.50238-1 72.69883-1
1.85025-2 2.79691-2 7.79955-3 6.51413-3 1.17686-2 1.56005-2 1.98705-2 2.64733-2 2.80680-2 2.55801-2 1.31526-2 8.30141-3 7.54210-3 8.41959-3 1.07202-2 1.42979-2 1.58429-2 1.56467-2 1.07684-2 7.73065-3 8.51954-3 9.38528-3 1.02571-2 1.11523-2 8.93208-3 9.85822-3 1.06512-2 1.12628-2 1.16930-2 9.27233-3 8.61898-3 8.71239-3 9.09103-3 1.11735-2 1.17865-2 1.11280-2 7.39033-3 4.78611-3 5.33312-3 5.95139-3 6.62245-3 7.85506-3 5.64745-3 7.18375-3 8.93818-3 1.03248-2
Appendix VIII Atomic Number 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 86 90 92 94
Continued Element Ag Ce In Sn Sb Te I Xe Cs Ba La Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Rn Th U Pu
a Notation as in Appendix VI, Table 4. Source: From McMaster et al., 1969.
Copyright © 2002 Marcel Dekker, Inc.
A0
A1
A2
A3
71.66475-1 75.16701-2 78.17283-2 1.42151-2 1.56362-2 74.07579-2 74.04420-2 72.82407-3 1.84861-1 3.44376-1 4.09104-1 4.39881-1 4.49124-1 4.37283-1 4.05823-1 3.55383-1 2.80316-1 2.73133-1 2.57539-1 2.42685-1 2.28493-1 2.15233-1 2.02656-1 2.02248-1 1.97176-1 1.99469-1 1.96871-1 1.91015-1 1.89644-1 1.16448-1 7.19908-2 4.20186-2 1.56916-2 1.14587-1 1.47052-1 1.82167-1 1.89860-1 1.96619-1 1.70890-1 1.08277-1 3.88791-2
1.65794 1.57426 1.55865 1.55754 1.57175 1.64267 1.65596 1.64039 1.50030 1.38742 1.33075 1.30925 1.30351 1.31370 1.33837 1.37733 1.44016 1.43842 1.45064 1.46266 1.47438 1.48545 1.49625 1.48804 1.50264 1.50233 1.50623 1.51240 1.50867 1.57615 1.61204 1.63611 1.65406 1.58076 1.56695 1.54661 1.56125 1.60080 1.65561 1.74158 1.82229
72.48740-1 72.27646-1 72.24492-1 72.24736-1 72.28753-1 72.47897-1 72.51067-1 72.47642-1 72.13333-1 71.86356-1 71.70883-1 71.64548-1 71.61841-1 71.62866-1 71.67229-1 71.74941-1 71.88641-1 71.86137-1 71.87591-1 71.89102-1 71.90559-1 71.91908-1 71.93234-1 71.89143-1 71.92474-1 71.91385-1 71.91396-1 71.91922-1 71.89570-1 72.05532-1 72.13186-1 72.17964-1 72.20982-1 72.02968-1 72.00347-1 71.95793-1 72.00932-1 72.13800-1 72.29702-1 72.54104-1 72.76009-1
8.66218-3 7.05650-3 6.85776-3 6.91395-3 7.26386-3 8.80567-3 9.04874-3 8.82144-3 6.24264-3 4.24917-3 3.04111-3 2.52641-3 2.27394-3 2.29377-3 2.55570-3 3.06213-3 4.01226-3 3.75240-3 3.79932-3 3.85628-3 3.90903-3 3.95645-3 4.00233-3 3.62264-3 3.85751-3 3.74011-1 3.70889-3 3.71450-3 3.49584-3 4.66731-3 5.20497-3 5.52670-3 5.70751-3 4.35692-3 4.20901-3 3.90772-3 4.36768-3 5.51717-3 6.92516-3 8.95056-3 1.07392-2
APPENDIX IX: COEFFICIENTS FOR CALCULATING THE COHERENT SCATTERING CROSS SECTIONS R (BARNS/ATOM) VIATHE ln^ln REPRESENTATION Atomic number 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47
Element H He Li Be B C N O F Ne Na Mg Al Si P S Cl Ar K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag
Copyright © 2002 Marcel Dekker, Inc.
A0 71.19075-1 1.04768 1.34366 2.00860 2.61862 3.10861 3.47760 3.77239 4.00716 4.20151 4.26374 4.39404 4.51995 4.64678 4.76525 4.92707 5.07222 5.21079 5.25587 5.32375 5.43942 5.55039 5.65514 5.77399 5.84604 5.93292 6.01478 6.09204 6.17739 6.23402 6.28298 6.33896 6.39750 6.45637 6.51444 6.57129 6.59750 6.62203 6.67096 6.72275 6.79013 6.84600 6.87599 6.93136 6.97547 7.03216 7.06446
a
A1
A2
A3
79.37086-1 78.51805-2 1.81557-1 74.61920-2 72.07916-1 72.60580-1 72.15762-1 71.48539-1 75.60908-2 4.16247-2 1.34662-1 1.37858-1 1.40549-1 1.62780-1 1.68708-1 1.65746-1 1.49127-1 1.35618-1 1.88040-1 2.06685-1 2.00174-1 1.97697-1 1.99533-1 2.03858-1 2.13814-1 2.25048-1 2.37959-1 2.52277-1 2.73123-1 2.84312-1 2.91334-1 2.91512-1 2.88866-1 2.86737-1 2.86324-1 2.87711-1 3.02389-1 3.24559-1 3.25075-1 3.23964-1 3.11282-1 3.02797-1 3.26165-1 3.34794-1 3.46394-1 3.49838-1 3.63456-1
72.00538-1 74.03527-1 74.23981-1 73.37018-1 72.86283-1 72.71974-1 72.88874-1 73.07124-1 73.32017-1 73.56754-1 73.70080-1 73.59540-1 73.52441-1 73.58563-1 73.60383-1 73.59424-1 73.52858-1 73.47214-1 73.59623-1 73.61664-1 73.59064-1 73.57694-1 73.57487-1 73.59699-1 73.59718-1 73.61748-1 73.64056-1 73.66568-1 73.72360-1 73.72143-1 73.69391-1 73.65643-1 73.61747-1 73.58794-1 73.57027-1 73.56311-1 73.56755-1 73.61651-1 73.60613-1 73.59463-1 73.55233-1 73.51131-1 73.58969-1 73.63497-1 73.67794-1 73.70099-1 73.73597-1
1.06587-2 2.69398-2 2.66190-2 1.86939-2 1.44966-2 1.35181-2 1.51312-2 1.67303-2 1.87934-2 2.07585-2 2.14467-2 2.02380-2 1.93692-2 1.96926-2 1.97155-2 1.95505-2 1.89439-2 1.84333-2 1.93085-2 1.93328-2 1.91027-2 1.89866-2 1.89691-2 1.92225-2 1.91459-2 1.93024-2 1.94754-2 1.96586-2 2.01638-2 2.00525-2 1.97029-2 1.92895-2 1.88788-2 1.85618-2 1.83557-2 1.82470-2 1.81706-2 1.84800-2 1.83325-2 1.81890-2 1.78231-2 1.74403-2 1.80482-2 1.84429-2 1.87885-2 1.89983-2 1.92478-2
Appendix IX Atomic number 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 86 90 92 94
Continued Element
A0
A1
A2
A3
Cd In Sn Sb Te I Xe Cs Ba La Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Rn Th U Pu
7.09856 7.12708 7.16085 7.19665 7.23464 7.27415 7.31469 7.33490 7.35812 7.39532 7.44255 7.48347 7.52334 7.56222 7.60020 7.63711 7.66938 7.70798 7.74188 7.77470 7.80643 7.83711 7.86662 7.89137 7.91803 7.94534 7.97266 7.99940 8.02574 8.05150 8.08084 8.10524 8.12542 8.14399 8.15996 8.17489 8.22553 8.27843 8.33010 8.38174
3.72199-1 3.82082-1 3.85512-1 3.85543-1 3.82493-1 3.77223-1 3.70315-1 3.76825-1 3.79361-1 3.69895-1 3.71328-1 3.68431-1 3.66462-1 3.65055-1 3.64134-1 3.63957-1 3.59752-1 3.65345-1 3.67107-1 3.69722-1 3.73226-1 3.77547-1 3.82933-1 3.86034-1 3.87021-1 3.87299-1 3.87704-1 3.88739-1 3.90458-1 3.93143-1 3.95790-1 4.00576-1 4.05858-1 4.08692-1 4.18031-1 4.27916-1 4.51478-1 4.79056-1 4.78314-1 4.77085-1
73.75345-1 73.76855-1 73.76481-1 73.75054-1 73.72715-1 73.69728-1 73.66280-1 73.65713-1 73.64099-1 73.59376-1 73.59642-1 73.57689-1 73.56048-1 73.54511-1 73.53086-1 73.51909-1 73.48899-1 73.50031-1 73.49433-1 73.49132-1 73.49147-1 73.49441-1 73.50126-1 73.49756-1 73.48881-1 73.47926-1 73.47155-1 73.46726-1 73.46658-1 73.47052-1 73.48032-1 73.49340-1 73.50329-1 73.49802-1 73.52330-1 73.55068-1 73.62056-1 73.67657-1 73.67250-1 73.66556-1
1.93481-2 1.94151-2 1.93305-2 1.91608-2 1.89194-2 1.86280-2 1.83025-2 1.81843-2 1.79817-2 1.75406-2 1.75852-2 1.74099-2 1.72620-2 1.71214-2 1.69894-2 1.68783-2 1.65890-2 1.66927-2 1.66273-2 1.65862-2 1.65710-2 1.65780-2 1.66173-2 1.65480-2 1.64406-2 1.63299-2 1.62372-2 1.61751-2 1.61455-2 1.61573-2 1.62345-2 1.63264-2 1.63772-2 1.62888-2 1.64660-2 1.66601-1 1.71556-2 1.74621-2 1.74129-2 1.73422-2
a
Notation as in Appendix VI, Table 4. Source: From McMaster et al., 1969.
Copyright © 2002 Marcel Dekker, Inc.
APPENDIX X: PARAMETERS FOR CALCULATING THE TOTAL MASS ATTENUATION COEFFICIENTS IN THE ENERGY RANGE 0.1^1000 keV [VIA EQ. (78)] Table1 Parameters for Calculating the Total Mass Attenuation Coefficients for Photon Energies Above the K Absorption Edge (k1, k2, k3, k4) and for Energies Between L1 and EK Absorption Edge (l1, l2, l3, l4); Z ¼ Atomic Number, Sy ¼ Symbol of Corresponding Element EL1 < E < EK Z
Sy
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25
H He Li Be B C N O F Ne Na Mg Al Si P S Cl Ar K Ca Sc Ti V Cr Mn
EL1
0.2292 0.2702 0.3200 0.3771 0.4378 0.5004 0.5637 0.6282 0.6946 0.7690
l1
77.00266 72.38938 72.50341 72.03564 72.91417 0.15632 2.10634 3.34768 3.81364 3.05305 2.76125 3.00797 2.62786 0.68453 2.44853 2.03061 2.33976 2.38673 2.53782 2.65431
Copyright © 2002 Marcel Dekker, Inc.
l2
5.19086 3.24818 3.43696 3.43789 4.13834 2.00149 0.28244 70.83777 71.04042 70.24003 0.29160 0.19834 0.68939 3.86200 1.46227 2.27680 1.88017 1.99445 1.85115 1.83552
E > EK l3
70.34413 70.11121 70.12806 70.13520 70.22931 0.32080 0.92539 1.34549 1.42378 1.25060 1.09998 1.11670 1.01331 70.43637 0.69759 0.31534 0.55084 0.50415 0.62868 0.65972
l4
70.04820 70.11707 70.16880 70.17980 70.17486 70.16451 70.16546 70.16625 0.04226 70.12878 70.07434 70.11829 70.11481 70.14231 70.15525
EK
k1
k2
k3
k4
0.2838 0.4016 0.5320 0.6854 0.8669 1.0721 1.3050 1.5596 1.8389 2.1455 2.4720 2.8224 3.2029 3.6074 4.0381 4.4928 4.9664 5.4651 5.9892 6.5390
76.48071 73.51566 72.36523 71.38914 70.58244 0.11476 0.63927 1.08730 1.42132 1.81194 2.07992 2.39196 2.62043 2.88077 3.06672 3.28801 3.42667 3.53219 3.75863 3.93329 4.00179 4.11127 4.21463 4.35405 4.44148
3.27538 2.60589 2.88866 3.05349 3.07947 3.09245 3.08731 3.07389 3.06182 3.04034 3.04059 3.00737 2.99387 2.96887 2.95078 2.92821 2.90289 2.88277 2.85955 2.83775 2.81618 2.79692 2.77190 2.75878 2.72589
0.04385 0.26361 0.15389 0.07919 0.05871 0.03213 0.01587 0.00035 70.00946 70.02320 70.02978 70.04492 70.05681 70.06174 70.07943 70.08514 70.09061 70.10191 70.09992 70.11226 70.11678 70.11885 70.13140 70.13335 70.14641
70.00866 70.03787 70.02935 70.02484 70.02647 70.02494 70.02538 70.02438 70.02708 70.02568 70.03575 70.02734 70.03063 70.02724 70.03227 70.03094 70.02881 70.03142 70.02689 70.02972 70.02898 70.02784 70.03049 70.03142 70.03205
26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
Fe Co Ni Cu Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Te Ru Rh Pd Ag Cd In Sn Sb Te I Xe Cs Ba La Ce Pr Nd
0.8461 0.9256 1.0081 1.0961 1.1936 1.2977 1.4143 1.5265 1.6539 1.7820 1.9210 2.0651 2.2163 2.3725 2.5316 2.6977 2.8655 3.0425 3.2240 3.4119 3.6043 3.8058 3.0180 4.2375 4.4647 4.6983 4.9392 5.1881 5.4528 5.7143 5.9888 6.2663 6.5488 6.8348 7.1260
2.41689 2.60591 2.69756 2.75840 2.87225 2.93846 3.02673 3.12353 3.19106 3.30117 3.36926 3.46256 3.54696 3.63965 3.71677 3.79880 3.86387 3.93865 3.99914 4.07189 4.12607 4.19974 4.24309 4.30447 4.35252 4.40598 4.43690 4.51763 4.53737 4.61780 4.65770 4.71490 4.77277 4.83442 4.87359
Copyright © 2002 Marcel Dekker, Inc.
2.80062 2.55718 2.72893 2.72647 2.72029 2.73406 2.74409 2.74414 2.76121 2.73599 2.74007 2.74320 2.73914 2.73646 2.73020 2.72885 2.72657 2.72262 2.70967 2.71531 2.71194 2.70350 2.70139 2.69926 2.69160 2.68835 2.68323 2.67963 2.66911 2.66299 2.65946 2.65505 2.64710 2.64088 2.63355
70.00696 0.18669 0.04244 0.04391 0.04441 0.03264 0.01747 0.01165 70.01683 0.01398 0.00224 70.10161 70.00741 70.01290 70.01273 70.01869 70.02278 70.02487 70.02706 70.02854 70.03442 70.03661 70.04152 70.04319 70.04781 70.04963 70.05424 70.05663 70.05780 70.06277 70.07172 70.07121 70.07113 70.07717 70.07273
70.01904 70.06696 70.03220 70.03423 70.03527 70.03438 70.03117 70.03096 70.02249 70.03428 70.03125 70.02723 70.03104 70.03036 70.03038 70.02896 70.02881 70.03004 70.03l74 70.03173 70.02941 70.02694 70.02776 70.03320 70.02689 70.03001 70.03161 70.03296 70.02658 70.02884 70.03577 70.03442 70.03077 70.03325 70.02815
7.1120 7.7089 8.3328 8.9789 9.6586 10.3671 11.1031 11.8667 12.6578 13.4737 14.3256 15.1997 16.1046 17.0384 17.9976 18.9856 19.9995 21.0440 22.1172 23.2199 24.3503 25.5140 26.7112 27.9399 29.2001 30.4912 31.8138 33.1694 34.5614 35.9846 37.4406 38.9246 40.4430 41.9906 43.5689
4.56600 4.64390 4.77308 4.81074 4.89850 4.93990 5.00743 5.07635 5.11759 5.19474 5.23284 5.29354 5.34695 5.41166 5.45290 5.50180 6.47503 5.57355 5.60095 5.63478 5.65532 5.68392 5.68137 5.72030 5.76598 5.74957 3.74700 5.79401 5.80690 5.82191 5.84551 5.85321 5.87476 5.90293 5.90905
2.70395 2.67587 2.65503 2.62267 2.60534 2.57045 2.55814 2.53376 2.50569 2.48145 2.44844 2.41378 2.38309 2.37364 2.33638 2.30251 3.03067 2.24202 2.21314 2.17349 2.14514 2.05134 2.09840 2.04363 2.05890 1.98639 1.96282 1.93222 1.91476 1.87715 1.87466 1.82364 1.79700 1.77311 1.74652
70.15515 70.16255 70.16936 70.18383 70.18393 71.19961 70.19972 70.20666 70.21581 70.22156 70.23505 70.24899 70.25992 70.25521 70.26879 70.28165 70.04965 70.29909 70.30745 70.32262 70.32915 70.34706 70.36341 70.36014 70.34949 70.37446 70.37979 70.38924 70.39242 70.040213 70.39966 70.41682 70.42193 70.42727 70.43341
70.03352 70.03403 70.03476 70.03729 70.03594 70.03856 70.03827 70.03872 70.03969 70.04024 70.04200 70.04399 70.04534 70.04393 70.04558 70.04737 70.03050 70.04910 70.04983 70.05195 70.05230 70.05473 70.05665 70.05573 70.05406 70.05693 70.05738 70.05845 70.05865 70.05948 70.05904 70.06098 70.06126 70.06168 70.06218
Table 1
Continued EL1 < E < EK
E > EK
Z
Sy
EL1
l1
l2
l3
l4
EK
k1
k2
k3
k4
61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88
Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn Fr Ra
7.4279 7.7368 8.0520 8.3756 8.7080 9.0458 9.3942 9.7513 10.1157 10.4864 10.8704 11.2707 11.6815 12.0998 12.5267 12.9680 13.4185 13.8799 14.3528 14.8393 15.3467 15.8608 16.3875 16.9393 17.4930 18.0490 18.6390 19.2367
4.93318 4.95712 5.00931 5.03443 5.08055 5.11495 5.15537 5.19648 5.23938 5.26855 5.31163 5.34305 5.37806 5.41320 5.44913 5.47715 5.50865 5.54158 5.57564 5.60041 5.62415 5.65349 5.69400 5.73466 5.76863 5.74682 5.78319 5.80588
2.62474 2.61647 2.61226 2.60481 2.60115 2.59238 2.58071 2.57148 2.56466 2.55541 2.55490 2.54268 2.53595 2.53128 2.52042 2.52666 2.50428 2.50551 2.48980 2.47759 2.46173 2.45575 2.47086 2.46840 2.45600 2.43686 2.43220 2.41932
70.08002 70.07625 70.08680 70.08912 70.07597 70.08230 70.08885 70.09459 70.09047 70.10174 70.09208 70.10077 70.09548 70.10036 70.10577 70.09007 70.10492 70.09847 70.10793 70.11241 70.12251 70.12044 70.10349 70.09608 70.10153 70.11031 70.10993 70.11330
70.02998 70.02487 70.03205 70.03162 70.02286 70.02541 70.02644 70.02802 70.02454 70.03012 70.02484 70.02734 70.02359 70.02631 70.02724 70.02154 70.02495 70.02303 70.02561 70.02570 70.02850 70.02698 70.02277 70.01973 70.02099 70.02272 70.02259 70.02255
45.1840 46.8342 48.5190 50.2391 51.9957 53.7885 55.6177 57.4855 59.3896 61.3323 63.3138 65.3508 67.4164 69.5250 71.6764 73.8708 76.1110 78.3948 80.7249 83.1023 85.5304 88.0045 90.5259 93.1050 95.7299 98.4040 101.1370 103.9219
5.92214 5.92934 5.94975 5.92827 5.95498 5.95991 5.94804 5.96408 5.88779 6.04546 6.07395 6.00771 6.05159 6.00250 6.03030 6.11400 6.09889 6.15378 6.15303 6.19862 6.17641 6.17648 6.30427 6.21512 6.18945 6.24492 6.31471 6.26760
1.70715 1.69904 1.67827 1.64215 1.63471 1.61100 1.55826 1.54243 1.43854 1.57974 1.57891 1.48367 1.50856 1.43225 1.43738 1.50446 1.45990 1.53187 1.47494 1.50676 1.46912 1.45115 1.55736 1.43087 1.38290 1.47699 1.51865 1.45136
70.44407 70.44407 70.44795 70.45629 70.45467 70.46029 70.47717 70.47949 70.50808 70.45878 70.45631 70.48598 70.47273 70.49552 70.48952 70.46496 70.47717 70.44176 70.46441 70.45079 70.46009 70.46277 70.42520 70.46426 70.47598 70.43939 70.42408 70.44373
70.06326 70.06312 70.06335 70.06402 70.06359 70.06408 70.06595 70.06606 70.06870 70.06334 70.06301 70.06608 70.06439 70.06672 70.06580 70.06309 70.06429 70.05974 70.06253 70.06097 70.06183 70.06196 70.05792 70.06195 70.06300 70.05879 70.05721 70.05917
Copyright © 2002 Marcel Dekker, Inc.
89 90 91 92
Ac Th Pa U
19.8400 20.4721 21.1046 21.7574
5.83228 5.85156 5.87741 5.88224
2.39643 2.40038 2.35466 2.34307
70.12322 70.11549 70.14418 70.15054
70.02407 70.02204 70.02817 70.03027
106.7553 109.6509 112.6014 115.6061
6.32123 6.39823 6.28773 6.30701
1.48583 1.55418 1.41526 1.44776
70.42883 70.40447 70.44597 70.43028
70.05744 70.05479 70.05907 70.05716
Source: From Orlic et al., 1993. Reprinted with kind permission from Elsevier Science – NL, Sara Burgehartstraat 25, 1055 KV Amsterdam, The Netherlands.
Copyright © 2002 Marcel Dekker, Inc.
Table 2 Parameters for Calculating the Total Mass Attentuation Coefficients for Photon Energy Between L1, L2, and L3 Absorption Edges and for Energies Between L3 and Lower Energy Limit E0 or M Absorption Edge EL3 < E < EL2
E0<E<EL3 Z
Sy E0 =Eedge
16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43
S Cl Ar K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc
0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1 0.1085 0.1487 0.2122 0.2118 0.2370 0.2631
m1
m2
m3
m4
3.98381 72.71337 2.07772 8.53825 5.10006 7.28382 6.39592 2.83998 77.38556 2.13747 2.21508 4.76476 4.28916 2.32551 4.23694 2.60598 4.27624 5.14580 6.74588 7.26941 11.95287 15.89070 11.09104 5.18978 6.15438 5.83085 6.35098 5.57003
71.96207 2.78307 70.84611 74.62277 71.85097 73.26298 72.45282 0.44991 8.17416 1.11467 1.10224 71.05687 70.66344 1.22642 70.64439 0.84383 70.65284 71.51272 73.03410 73.69837 78.69443 712.8788 78.15432 71.99566 72.90381 72.56434 73.21302 72.03397
1.29140 0.23822 1.21097 1.97295 1.29907 1.60507 1.38399 0.61932 71.29292 0.46032 0.47160 1.09423 1.02071 0.45413 1.08840 0.68376 1.15806 1.47391 1.98313 2.28437 4.04104 5.51561 4.08776 2.09278 2.42622 2.36756 2.67445 2.13536
70.13371 70.05767 70.14523 70.19719 70.14560 70.16673 70.14698 70.08033 0.07794 70.06687 70.06697 70.12565 70.12207 70.06650 70.13570 70.10262 70.15244 70.19017 70.24674 70.28824 70.48832 70.65735 70.52451 70.32184 70.36813 70.37176 70.41914 70.33831
Copyright © 2002 Marcel Dekker, Inc.
EL3
1.0197 1.1154 1.2167 1.3231 1.4358 1.5499 1.6749 1.8044 1.9396 2.0800 2.2223 2.3705 2.5202 2.6769
l31
77.206 47.827 20.461 6.9330 0.0065 74.6062 75.6370 3.2959 2.0366 1.9840 2.2290 2.724 2.813 3.044
l32
727.639 716.433 75.2040 0.6640 3.8833 6.2403 6.9490 2.4609 3.1926 3.2996 3.2290 2.999 2.999 2.907
EL2 < E < EL1 EL2
1.0428 1.1423 1.2478 1.3586 1.4762 1.5960 1.7272 1.8639 2.0068 2.1555 2.3067 2.4647 2.6251 2.7932
l21
9.4520 6.2330 5.4560 5.1580 3.9491 3.8764 3.4100 4.0979 3.9355 3.9621 3.9270 3.981 3.996 4.051
l22
EL1
70.2320 1.1180 1.4580 1.6080 2.1968 2.2734 2.5440 2.2097 2.3328 2.3596 2.4170 2.425 2.447 2.452
0.2292 0.2702 0.3200 0.3771 0.4378 0.5004 0.5637 0.6282 0.6946 0.7690 0.8461 0.9256 1.0081 1.0961 1.1936 1.2977 1.4143 1.5265 1.6539 1.7820 1.9210 2.0651 2.2163 2.3725 2.5316 2.6977 2.8655 3.0425
44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78
Ru Rh Pd Ag Cd In Sn Sb Te I Xe Cs Ba La Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt
0.2901 0.3180 0.3425 0.3766 0.4129 0.4530 0.4946 0.5379 0.5827 0.6293 0.6775 1.2171 1.2928 1.3613 1.4346 1.5110 1.5753 1.6540 1.7228 1.8000 1.8808 1.9675 2.0468 2.1283 2.2065 2.3068 2.3981 2.4912 2.6009 2.7080 2.8196 2.9317 3.0485 3.1737 3.2960
9.99349 6.55219 6.32821 5.65378 3.12801 2.78289 2.57173 2.52338 4.98562 3.00387 3.00203 3.16785 3.19080 3.23265 3.29411 3.35891 3.41765 3.45084 3.49847 3.54762 3.58135 3.61719 3.67219 3.72181 3.77845 3.82800 3.86458 3.91793 3.94527 3.99165 4.02988 4.07043 4.10244 4.14636 4.18356
Copyright © 2002 Marcel Dekker, Inc.
78.27969 73.56318 73.29827 72.26740 1.54731 2.29146 2.70219 2.80299 71.29951 2.36374 2.56894 2.42799 2.48392 2.53710 2.53524 2.54117 2.51947 2.59000 2.55440 2.56798 2.57591 2.61430 2.57359 2.55757 2.52004 2.53261 2.50726 2.48402 2.52598 2.50855 2.51689 2.52499 2.53312 2.52893 2.53390
5.01647 2.98931 2.95809 2.49736 0.70511 0.30064 0.11208 0.07373 2.27938 0.26067 0.10775 0.12755 0.08776 0.05145 0.05465 0.04839 0.05777 0.00563 0.02671 0.01585 0.00022 70.03278 70.00002 0.010760 0.05588 0.03019 0.06374 0.08777 0.03404 0.05734 0.04574 0.03041 0.01448 0.01458 0.00520
70.76636 70.48554 70.49645 70.43066 70.16236 70.09544 70.07092 70.06562 70.44589 70.09483 70.06179 70.05218 70.04429 70.03744 70.03926 70.03842 70.04029 0.02894 70.03367 70.03181 70.02718 70.01918 70.02988 70.03537 70.04817 70.03886 70.05192 70.06146 70.04239 70.05289 70.05071 70.04512 70.03948 70.04009 70.03783
2.8379 3.0038 3.1733 3.3511 3.5375 3.7301 3.9288 4.1322 4.3414 4.5571 4.7822 5.0119 5.2470 5.4827 5.7234 5.9642 6.2079 6.4593 6.7162 6.9769 7.2428 7.5140 7.7901 8.0711 8.3579 8.6480 8.9436 9.2441 9.5607 9.8811 10.2068 10.5353 10.8709 11.2152 11.5637
3.320 3.653 4.146 4.111 3.971 4.293 4.834 4.637 4.618 4.577 4.572 4.215 4.363 4.401 4.449 4.469 4.463 4.500 4.536 4.553 4.583 4.633 4.655 4.695 4.727 4.767 4.802 4.847 4.866 4.911 4.940 4.979 5.006 5.050 5.084
2.770 2.586 2.250 2.321 2.454 2.217 1.753 1.940 1.955 2.048 2.073 2.528 2.384 2.397 2.402 2.438 2.500 2.528 2.502 2.556 2.543 2.537 2.572 2.558 2.581 2.592 2.569 2.560 2.629 2.584 2.624 2.629 2.661 2.673 2.666
2.9669 3.1469 3.3303 3.5237 3.7270 3.9380 4.1561 4.3804 4.6120 4.8521 5.1037 5.3594 5.6236 5.8906 6.1642 6.4404 6.7215 7.0128 7.3118 7.6171 7.9303 5.2516 8.5806 8.9178 9.2643 9.6169 9.9782 10.3386 10.7394 11.1361 11.5440 11.9687 12.3850 12.8241 13.2726
4.303 4.091 4.344 4.217 4.328 4.620 4.592 4.584 4.554 4.625 4.599 4.563 4.601 4.698 4.702 4.757 4.778 4.854 4.861 4.924 4.927 4.985 5.014 5.047 5.086 5.128 5.082 5.181 5.214 5.237 5.276 5.302 5.323 5.355 5.375
2.302 2.502 2.337 2.487 2.420 2.196 2.238 2.279 2.323 2.318 2.375 2.482 2.470 2.404 2.470 2.467 2.490 2.449 2.469 2.428 2.467 2.423 2.420 2.426 2.414 2.398 2.815 2.431 2.377 2.468 2.318 2.373 2.314 2.361 2.278
3.2240 3.4119 3.6043 3.8058 3.0180 4.2375 4.4647 4.6983 4.9392 5.1881 5.4528 5.7143 5.9888 6.2663 6.5488 6.8348 7.1260 7.4279 7.7368 8.0520 8.3756 8.7080 9.0458 9.3942 9.7513 10.1157 10.4864 10.8704 11.2707 11.6815 12.0998 12.5267 12.9680 13.4185 13.8799
Table 2
Continued EL3 < E < EL2
E0<E<EL3
EL2 < E < EL1
Z
Sy
E0 =Eedge
m1
m2
m3
m4
EL3
l31
l32
EL2
l21
l22
EL1
79 80 81 82 83 84 85 86 87 88 89 90 91 92
Au Hg Tl Pb Bi Po At Rn Fr Ra Ac Th Pa U
3.4249 3.5616 3.7041 3.8507 3.9991 4.1494 4.3170 4.4820 4.6520 4.8220 5.0020 5.1823 5.3669 5.5480
4.22770 4.26093 4.29302 4.32934 4.36991 4.41866 4.46126 4.45433 4.49562 4.52666 4.56932 4.59153 4.63772 4.65216
2.52409 2.51954 2.51974 2.51778 2.52184 2.52158 2.51394 2.51411 2.51108 2.50733 2.50607 2.50190 2.49776 2.49170
0.01426 0.02092 0.01369 0.00460 70.01213 70.00971 0.00640 70.00666 70.01358 0.00692 70.01644 70.00815 70.00936 70.02703
70.04262 70.04665 70.04376 70.03614 70.02654 70.03298 70.04593 70.03871 70.02823 70.05016 70.03124 70.03948 70.03490 70.01144
11.9187 12.2839 12.6575 13.0352 13.4186 13.8138 14.2135 14.6194 15.0312 15.4444 15.8710 16.3003 16.7331 17.1663
5.125 5.156 5.184 5.220 5.257 5.301 5.347 5.334 5.376 5.410 5.454 5.490 5.525 5.541
2.690 2.643 2.630 2.644 2.642 2.625 2.642 2.637 2.647 2.653 2.662 2.696 2.666 2.672
13.7336 14.2087 14.6979 15.2000 15.7111 16.2443 16.7847 17.3371 17.9065 18.4843 19.0832 19.6932 20.3137 20.9476
5.406 5.430 5.451 5.512 5.536 5.555 5.574 5.535 5.650 5.644 5.761 5.883 5.798 5.829
2.303 2.337 2.323 2.502 2.453 2.387 2.335 2.281 2.494 2.411 2.594 2.800 2.534 2.571
14.3528 14.8393 15.3467 15.8608 16.3875 16.9393 17.4930 18.0490 18.6390 19.2367 19.8400 20.4721 21.1046 21.7574
Source: From Orlic et al., 1993. Reprinted with kind permission from Elsevier Science – NL, Sara Burgehartstraat 25, 1055 KV Amsterdam, The Netherlands.
Copyright © 2002 Marcel Dekker, Inc.
Table 3 Parameters for Calculating the Total Mass Attenuation Coefficients for Photon Energies Between M1, M2, . . ., M5 Absorption Edges and for Energies Between M5 and Lower Energy Limit E0 E0 < E < Eedge Z
E0
41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68
0.1085 0.1085 0.1085 0.1085 0.1085 0.1328 0.1328 0.1487 0.1717 0.1717 0.1511 0.1511 0.1511 0.1833 0.2122 0.2770 0.1926 0.2122 0.3117 0.3117 0.2770 0.2770 0.2770 0.2770 0.1717 0.2122 0.2770 0.2122
n1
n2
n3
EM5 < E < EM4 n4
EM5
m42
m41
EM4 < E < EM3 EM4
m32
m31
EM3 < E < EM2 EM3
m22
115.013 769.4208 14.9364 71.0543 106.662 767.3126 15.2602 71.1408 3.5107 2.3889 70.4178 0.0324 89.5750 757.9740 13.8424 71.1025 80.5898 757.4238 15.0813 71.3052 76.3873 1.0408 1.9749 70.3214 29.0389 718.5282 5.5664 70.5470 11.0289 76.3556 2.6658 70.3046 174.946 7138.963 38.3880 73.5077 42.9485 731.1679 9.1280 70.8669 39.8331 730.6663 9.5340 70.9502 37.0196 728.3109 8.8955 70.8939 14.2756 79.7352 3.9179 70.4554 23.1086 718.4405 6.7995 70.7732 32.5527 727.5567 9.7373 71.0874 0.7330 2.364 31.6390 728.1233 10.4351 71.2218 0.7905 0.7961 3.768 2.088 1.0622 3.897 28.6047 724.1290 8.8027 71.0079 0.8473 0.8485 3.554 2.190 1.1234 3.870 3.9181 70.9835 1.6398 70.2735 0.8861 0.9013 3.194 2.376 1.1854 3.852 4.6993 0.0130 0.6969 70.0975 0.9349 0.9511 3.317 2.355 1.2422 5.314 20.5246 716.5997 6.5042 70.7721 0.9843 0.9999 5.484 1.419 1.2974 3.879 16.9684 713.1825 5.4255 70.6587 1.0269 96.373 735.542 1.0515 6.609 0.921 1.3569 3.845 11.8798 77.7898 3.5371 70.4381 1.0802 89.021 733.267 1.1060 6.352 1.023 1.4198 3.825 5.6437 71.4828 1.4474 70.2096 1.1309 83.564 731.660 1.1606 6.316 1.039 1.4806 3.825 6.6183 72.4260 1.7586 70.2443 1.1852 76.154 729.235 1.2172 8.597 0.048 1.5440 3.792 75.6112 10.4330 72.6765 0.2602 1.2412 73.753 728.683 1.2750 6.289 1.062 1.6113 3.863 1.4535 3.1211 70.1907 70.0138 1.2949 69.682 727.456 1.3325 6.096 1.157 1.6756 3.809 4.077 0.1743 0.9232 70.1528 1.3514 65.046 725.898 1.3915 5.446 1.471 1.7412 3.811 1.6315 2.7196 0.0767 70.0607 1.4093 60.848 724.477 1.4533 5.353 1.523 1.8118 3.847
Copyright © 2002 Marcel Dekker, Inc.
EM2 < E < EM1
m21
EM2
m12
m11
2.621 2.098 2.132 2.162 1.551 2.193 2.234 2.255 2.278 2.308 2.296 2.342 2.363 2.368
1.0650 1.1367 1.2044 1.2728 1.3774 1.4028 1.4714 1.5407 1.6139 1.6883 1.7677 1.8418 1.9228 2.0058
3.983 3.957 3.970 4.007 2.489 3.993 4.039 3.986 4.005 4.032 4.027 4.060 4.092 4.087
2.072 2.097 2.114 2.120 2.863 2.167 2.170 2.206 2.220 2.217 2.242 2.243 2.245 2.269
EM1
1.2171 1.2928 1.3613 1.4346 1.5110 1.5753 1.6540 1.7228 1.8000 1.8808 1.9675 2.0468 2.1283 2.2065
Table 3
Continued E0 < E < Eedge
EM5 < E < EM4
EM4 < E < EM3
EM3 < E < EM2
EM2 < E < EM1
Z
E0
n1
n2
n3
n4
EM5
m42
m41
EM4
m32
m31
EM3
m22
m21
EM2
m12
m11
69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92
0.1926 0.1085 0.1085 0.1085 0.1085 1.1085 0.1085 0.1085 0.1085 0.1328 0.1511 0.2122 0.1926 0.1717 0.1926 0.2122 0.3117 0.3924 0.3924 0.3924 0.3924 0.2770 0.2770 0.2770
75.0355 71.8380 70.3276 70.1868 71.9405 71.1407 70.3842 71.8806 72.9368 1.5113 3.5698 3.6427 3.6296 4.6393 5.7697 5.6131 5.1757 4.2011 4.5198 5.5170 6.3317 4.1761 4.3146 3.9954
9.4382 6.2849 4.8522 4.8402 6.5977 5.9151 5.1929 6.8129 7.9106 3.4177 1.1260 1.0962 1.1014 70.1055 71.3933 71.3279 70.6452 0.6074 0.1763 71.1906 72.2638 0.8164 0.8409 1.2801
72.1146 71.0736 70.6175 70.6304 71.1795 70.9671 70.7225 71.2684 71.6078 70.1106 0.7499 0.7581 0.7912 1.2868 1.7810 1.8314 1.5552 1.0490 1.2693 1.9034 2.3875 1.0031 0.9488 0.7662
0.1726 0.0593 0.0113 0.0127 0.0672 0.0439 0.0155 0.0733 0.1054 70.0581 70.1652 70.1655 70.1770 70.2451 70.3078 70.3245 70.2925 70.2281 70.2654 70.3623 70.4354 70.2383 70.2273 70.2038
1.4677 1.5278 1.5885 1.6517 1.7351 1.8092 1.8829 1.9601 2.0404 2.1216 2.2057 2.2949 2.3893 2.4840 2.5796 2.6839 2.7867 2.8924 2.9999 3.1049 3.2190 3.3320 3.4418 3.5517
63.572 54.494 57.769 55.928 53.589 50.662 30.913 54.605 48.393 45.233 43.502 40.458 31.751 20.254 9.171 70.200 74.893 77.059 72.615 0.627 2.581 3.793 3.639 3.662
726.240 722.411 724.531 724.196 723.567 722.588 712.446 725.802 722.958 721.695 721.199 719.895 715.013 78.092 71.075 5.161 8.501 10.172 7.198 4.924 3.529 2.621 2.790 2.788
1.5146 1.5763 1.6394 1.7164 1.7932 1.8716 1.9489 2.0308 2.1161 2.2019 2.2911 2.3849 2.4851 2.5856 2.6876 2.7980 2.9087 3.0215 3.1362 3.2484 3.3702 3.4908 3.6112 3.7276
5.841 5.900 6.893 7.170 6.974 6.894 6.854 6.900 7.249 7.209 7.102 6.847 6.055 5.339 4.475 3.586 3.141 3.222 3.649 3.982 4.255 4.355 4.369 4.409
1.257 1.222 0.681 0.527 0.627 0.671 0.666 0.581 0.343 0.344 0.393 0.540 1.064 1.574 2.212 2.895 3.276 3.251 2.961 2.716 2.525 2.455 2.484 2.460
1.8845 1.9498 2.0236 2.1076 2.1940 2.2810 2.3673 2.4572 2.5507 2.6454 2.7430 2.8471 2.9566 3.0664 3.1769 3.3019 3.4260 3.5280 3.6638 3.7918 3.9098 4.0461 4.1738 4.3034
3.889 3.916 3.876 3.916 3.957 4.019 4.063 4.032 4.126 4.166 4.199 4.187 4.191 4.259 4.277 4.327 4.372 4.382 4.405 4.391 4.460 4.495 4.526 4.527
2.370 2.373 2.422 2.420 2.420 2.408 2.405 2.445 2.410 2.408 2.414 2.445 2.467 2.443 2.459 2.459 2.457 2.442 2.458 2.501 2.475 2.462 2.477 2.492
2.0998 2.1730 2.2635 2.3654 2.4687 2.5749 2.6816 2.7922 2.9087 3.0265 3.1478 3.2785 3.4157 3.5542 3.6963 3.8541 4.0080 4.1590 4.3270 4.4895 4.6560 4.8304 5.0009 5.1822
4.114 4.125 4.129 4.158 4.146 4.178 4.214 4.228 4.307 4.323 4.317 4.382 4.314 4.453 4.383 4.460 4.494 4.500 4.523 4.511 4.599 4.576 4.603 4.614
2.278 2.287 2.309 2.311 2.343 2.344 2.346 2.356 2.327 2.338 2.370 2.344 2.419 2.332 2.420 2.395 2.402 2.386 2.398 2.442 2.387 2.437 2.457 2.458
Source: From Orlic et al., 1993. Reprinted with kind permission from Elsevier Science – NL, Sara Burgehartstraat 25, 1055 KV Amsterdam, The Netherlands.
Copyright © 2002 Marcel Dekker, Inc.
EM1 2.3068 2.3981 2.4912 2.6009 2.7080 2.8196 2.9317 3.0485 3.1737 3.2960 3.4249 3.5616 3.7041 3.8507 3.9991 4.1494 4.3170 4.4820 4.6520 4.8220 5.0020 5.1823 5.3669 5.5480
APPENDIX XI: TOTAL MASS ATTENUATION COEFFICIENTS FOR LOW-ENERGY K LINES Emitter wavelength (A˚) energy (keV)
Absorber
Atomic number 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27
Element
B 6.67 þ 1a 1.83 7 1
C 4.48 þ 1 2.77 7 1
N 3.16 þ 1 3.92 7 1
O 2.36 þ 1 5.25 7 1
F 1.83 þ 1 6.77 7 1
Ne 1.46 þ 1 8.49 7 1
Na 1.19 þ 1 1.04 þ 0
Mg 9.89 þ 0 1.25 þ 0
Al 8.34 þ 0 1.49 þ 0
H He Li Be B C N O F Ne Na Mg Al Si P S Cl Ar K Ca Sc Ti V Cr Mn Fe Co
1.81 þ 3 1.20 þ 4 2.98 þ 4 5.82 þ 4 3.93 þ 3 7.32 þ 3 1.19 þ 4 1.74 þ 4 2.45 þ 4 3.72 þ 4 4.25 þ 4 5.62 þ 4 6.44 þ 4 7.48 þ 4 7.49 þ 4 8.18 þ 4 7.44 þ 3 9.51 þ 3 1.14 þ 4 1.31 þ 4 1.43 þ 4 1.54 þ 4 1.68 þ 4 2.09 þ 4 2.23 þ 4 2.61 þ 4 2.86 þ 4
4.89 þ 2 4.06 þ 3 1.08 þ 4 2.29 þ 4 3.75 þ 4 2.75 þ 3 4.71 þ 3 6.90 þ 3 9.73 þ 3 1.36 þ 4 1.80 þ 4 2.44 þ 4 2.92 þ 4 3.47 þ 4 4.19 þ 4 4.95 þ 4 5.17 þ 4 5.24 þ 4 5.92 þ 3 7.15 þ 3 7.80 þ 3 8.45 þ 3 9.22 þ 3 1.11 þ 4 1.20 þ 4 1.40 þ 4 1.54 þ 4
1.51 þ 2 1.25 þ 3 3.67 þ 3 8.40 þ 3 1.53 þ 4 2.44 þ 4 1.83 þ 3 2.68 þ 3 3.73 þ 3 5.70 þ 3 7.26 þ 3 1.00 þ 4 1.23 þ 4 1.58 þ 4 1.93 þ 4 2.38 þ 4 2.67 þ 4 3.00 þ 4 3.51 þ 4 3.49 þ 4 3.97 þ 3 4.35 þ 3 4.77 þ 3 5.61 þ 3 6.13 þ 3 7.10 þ 3 7.85 þ 3
5.85 þ 1 5.07 þ 2 1.60 þ 3 3.78 þ 3 7.22 þ 3 1.22 þ 4 1.73 þ 4 1.27 þ 3 1.79 þ 3 2.71 þ 3 3.52 þ 3 4.89 þ 3 6.10 þ 3 7.93 þ 3 9.74 þ 3 1.21 þ 4 1.39 þ 4 1.59 þ 4 1.94 þ 4 2.21 þ 4 2.34 þ 4 2.20 þ 4 1.66 þ 4 3.13 þ 3 3.46 þ 3 4.00 þ 3 4.41 þ 3
2.62 þ 1 2.23 þ 2 7.57 þ 2 1.86 þ 3 3.63 þ 3 6.29 þ 3 9.23 þ 3 1.21 þ 4 9.12 þ 2 1.39 þ 3 1.83 þ 3 2.55 þ 3 3.22 þ 3 4.22 þ 3 5.19 þ 3 6.52 þ 3 7.50 þ 3 8.59 þ 3 1.07 þ 4 1.24 þ 4 1.34 þ 4 1.45 þ 4 1.57 þ 4 1.58 þ 4 1.68 þ 4 2.31 þ 3 2.58 þ 3
1.22 þ 1 1.06 þ 2 3.84 þ 2 9.59 þ 2 1.93 þ 3 3.43 þ 3 5.14 þ 3 6.94 þ 3 8.54 þ 3 7.54 þ 2 1.01 þ 3 1.42 þ 3 1.80 þ 3 2.37 þ 3 2.92 þ 3 3.69 þ 3 4.27 þ 3 4.89 þ 3 6.13 þ 3 7.19 þ 3 7.79 þ 3 8.59 þ 3 9.40 þ 3 1.09 þ 4 1.16 þ 4 1.31 þ 4 1.22 þ 4
6.32 þ 0 5.58 þ 1 2.09 þ 2 5.44 þ 2 1.10 þ 3 1.98 þ 3 3.02 þ 3 4.17 þ 3 5.23 þ 3 7.11 þ 3 5.93 þ 2 8.37 þ 2 1.07 þ 3 1.41 þ 3 1.74 þ 3 2.21 þ 3 2.57 þ 3 2.93 þ 3 3.71 þ 3 4.38 þ 3 4.74 þ 3 5.27 þ 3 5.81 þ 3 6.75 þ 3 7.27 þ 3 8.22 þ 3 8.84 þ 3
3.76 þ 0 3.15 þ 1 1.20 þ 2 3.26 þ 2 6.73 þ 2 1.21 þ 3 1.90 þ 3 2.67 þ 3 3.44 þ 3 4.55 þ 3 5.65 þ 3 5.06 þ 2 6.63 þ 2 8.93 þ 2 1.09 þ 3 1.39 þ 3 1.63 þ 3 1.85 þ 3 2.36 þ 3 2.75 þ 3 2.99 þ 3 3.34 þ 3 3.70 þ 3 4.30 þ 3 4.67 þ 3 5.31 þ 3 5.73 þ 3
2.23 þ 0 1.77 þ 1 6.82 þ 1 1.89 þ 2 3.94 þ 2 7.22 þ 2 1.14 þ 3 1.63 þ 3 2.11 þ 3 2.84 þ 3 3.53 þ 3 4.39 þ 3 4.07 þ 2 5.52 þ 2 6.78 þ 2 8.64 þ 2 1.01 þ 3 1.15 þ 3 1.47 þ 3 1.74 þ 3 1.90 þ 3 2.13 þ 3 2.36 þ 3 2.74 þ 3 3.00 þ 3 3.42 þ 3 3.71 þ 3
Copyright © 2002 Marcel Dekker, Inc.
Appendix XI
Continued Emitter wavelength (A˚) energy (keV)
Absorber
Atomic number 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55
Element Ni Cu Zn Ga Ge As Se Br Kr Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te I Xe Cs
B 6.67 þ 1a 1.83 7 1
C 4.48 þ 1 2.77 7 1
N 3.16 þ 1 3.92 7 1
O 2.36 þ 1 5.25 7 1
F 1.83 þ 1 6.77 7 1
Ne 1.46 þ 1 8.49 7 1
Na 1.19 þ 1 1.04 þ 0
Mg 9.89 þ 0 1.25 þ 0
Al 8.34 þ 0 1.49 þ 0
3.34 þ 4 3.65 þ 4 4.03 þ 4 4.18 þ 4 4.41 þ 4 4.41 þ 4 4.49 þ 4 4.19 þ 4 3.94 þ 4 3.52 þ 4 2.83 þ 4 2.04 þ 4 6.41 þ 3 4.28 þ 3 4.66 þ 3 4.99 þ 3 5.49 þ 3 5.87 þ 3 6.09 þ 3 6.34 þ 3 6.45 þ 3 6.61 þ 3 6.64 þ 3 6.71 þ 3 6.65 þ 3 5.77 þ 3 5.79 þ 3 5.91 þ 3
1.81 þ 4 1.96 þ 4 2.26 þ 4 2.42 þ 4 2.64 þ 4 2.86 þ 4 3.03 þ 4 3.22 þ 4 3.28 þ 4 3.56 þ 4 3.47 þ 4 3.10 þ 4 3.06 þ 4 2.96 þ 4 2.23 þ 4 1.41 þ 4 4.18 þ 3 4.64 þ 3 5.22 þ 3 5.48 þ 3 5.75 þ 3 6.06 þ 3 6.27 þ 3 6.50 þ 3 6.58 þ 3 6.99 þ 3 7.27 þ 3 7.52 þ 3
9.19 þ 3 9.96 þ 3 1.17 þ 4 1.26 þ 4 1.40 þ 4 1.55 þ 4 1.69 þ 4 1.86 þ 4 2.05 þ 4 2.17 þ 4 2.29 þ 4 2.34 þ 4 2.46 þ 4 2.67 þ 4 2.29 þ 4 2.29 þ 4 2.47 þ 4 2.48 þ 4 2.21 þ 4 1.10 þ 4 4.10 þ 3 4.42 þ 3 4.70 þ 3 4.94 þ 3 5.09 þ 3 5.52 þ 3 5.88 þ 3 6.13 þ 3
5.25 þ 3 5.94 þ 3 6.56 þ 3 7.10 þ 3 7.89 þ 3 8.80 þ 3 9.64 þ 3 1.07 þ 4 1.19 þ 4 1.28 þ 4 1.37 þ 4 1.51 þ 4 1.61 þ 4 1.78 þ 4 1.85 þ 4 1.78 þ 4 1.95 þ 4 1.94 þ 4 1.80 þ 4 1.91 þ 4 2.01 þ 4 2.23 þ 4 9.12 þ 3 3.36 þ 3 3.53 þ 3 3.88 þ 3 4.17 þ 3 4.40 þ 3
3.01 þ 3 3.20 þ 3 3.73 þ 3 4.08 þ 3 4.56 þ 3 5.10 þ 3 5.61 þ 3 6.28 þ 3 6.98 þ 3 7.57 þ 3 8.21 þ 3 9.06 þ 3 9.77 þ 3 1.08 þ 4 1.14 þ 4 1.19 þ 4 1.30 þ 4 1.38 þ 4 1.46 þ 4 1.47 þ 4 1.47 þ 4 1.36 þ 4 1.38 þ 4 1.48 þ 4 1.75 þ 4 1.18 þ 4 2.84 þ 3 3.02 þ 3
1.80 þ 3 2.01 þ 3 2.25 þ 3 2.47 þ 3 2.75 þ 3 3.08 þ 3 3.38 þ 3 3.97 þ 3 4.23 þ 3 4.63 þ 3 5.04 þ 3 5.58 þ 3 6.03 þ 3 6.69 þ 3 7.09 þ 3 7.45 þ 3 8.16 þ 3 8.79 þ 3 9.42 þ 3 1.00 þ 4 1.04 þ 4 1.10 þ 4 1.08 þ 4 1.13 þ 4 1.13 þ 4 1.02 þ 4 1.08 þ 4 1.15 þ 4
1.02 þ 4 9.47 þ 3 5.26 þ 3 1.52 þ 3 1.70 þ 3 1.95 þ 3 2.17 þ 3 2.44 þ 3 2.66 þ 3 2.92 þ 3 3.20 þ 3 3.55 þ 3 3.86 þ 3 4.27 þ 3 4.55 þ 3 4.90 þ 3 5.37 þ 3 5.80 þ 3 6.23 þ 3 6.66 þ 3 7.09 þ 3 7.51 þ 3 7.88 þ 3 8.26 þ 3 8.35 þ 3 8.24 þ 3 8.72 þ 3 9.79 þ 3
6.58 þ 3 7.04 þ 3 7.51 þ 3 6.92 þ 3 4.72 þ 3 1.30 þ 3 1.53 þ 3 1.68 þ 3 1.72 þ 3 1.89 þ 3 2.08 þ 3 2.32 þ 3 2.53 þ 3 2.80 þ 3 2.99 þ 3 3.57 þ 3 3.92 þ 3 4.23 þ 3 4.54 þ 3 4.86 þ 3 5.84 þ 3 6.20 þ 3 6.50 þ 3 6.82 þ 3 6.90 þ 3 6.93 þ 3 6.75 þ 3 6.67 þ 3
4.26 þ 3 4.52 þ 3 4.96 þ 3 5.23 þ 3 5.63 þ 3 5.32 þ 3 4.86 þ 3 1.01 þ 3 1.14 þ 3 1.25 þ 3 1.38 þ 3 1.52 þ 3 1.66 þ 3 1.84 þ 3 1.97 þ 3 2.12 þ 3 2.32 þ 3 2.51 þ 3 2.69 þ 3 2.90 þ 3 4.48 þ 3 4.75 þ 3 4.99 þ 3 5.25 þ 3 5.32 þ 3 5.39 þ 3 5.30 þ 3 5.29 þ 3
Copyright © 2002 Marcel Dekker, Inc.
56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90
Ba La Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn Fr Ra Ac Th
4.09 þ 3 3.76 þ 3 6.79 þ 3 8.74 þ 3 1.05 þ 4 1.26 þ 4 1.48 þ 4 1.58 þ 4 1.49 þ 4 1.87 þ 4 2.04 þ 4 2.16 þ 4 2.32 þ 4 2.43 þ 4 2.28 þ 4 2.15 þ 4 2.13 þ 4 2.04 þ 4 1.93 þ 4 1.76 þ 4 1.58 þ 4 1.38 þ 4 1.13 þ 4 9.09 þ 3 6.62 þ 3 5.42 þ 3 4.19 þ 3 3.17 þ 3 2.54 þ 3 2.28 þ 3 2.52 þ 3 2.69 þ 3 2.66 þ 3 2.49 þ 3 1.99 þ 3
Copyright © 2002 Marcel Dekker, Inc.
7.64 þ 3 7.74 þ 3 7.98 þ 3 8.77 þ 3 9.43 þ 3 1.04 þ 4 1.13 þ 4 1.20 þ 4 1.12 þ 4 1.24 þ 4 1.37 þ 4 1.47 þ 4 1.60 þ 4 1.72 þ 4 1.79 þ 4 1.75 þ 4 1.80 þ 4 1.82 þ 4 1.85 þ 4 1.84 þ 4 1.67 þ 4 1.61 þ 4 1.57 þ 4 1.49 þ 4 1.47 þ 4 1.29 þ 4 1.14 þ 4 1.01 þ 4 8.62 þ 3 6.67 þ 3 4.13 þ 3 3.02 þ 3 2.42 þ 3 2.59 þ 3 2.12 þ 3
6.21 þ 3 6.32 þ 3 7.14 þ 3 7.60 þ 3 7.92 þ 3 8.40 þ 3 8.95 þ 3 9.42 þ 3 9.39 þ 3 9.93 þ 3 1.06 þ 4 1.13 þ 4 1.21 þ 4 1.29 þ 4 1.35 þ 4 1.36 þ 4 1.27 þ 4 1.32 þ 4 1.37 þ 4 1.39 þ 4 1.39 þ 4 1.44 þ 4 1.50 þ 4 1.51 þ 4 1.50 þ 4 1.40 þ 4 1.38 þ 4 1.35 þ 4 1.34 þ 4 1.31 þ 4 1.21 þ 4 1.10 þ 4 9.31 þ 3 8.17 þ 3 6.44 þ 3
4.52 þ 3 4.66 þ 3 5.16 þ 3 5.45 þ 3 5.64 þ 3 5.92 þ 3 6.26 þ 3 6.58 þ 3 6.60 þ 3 7.38 þ 3 7.83 þ 3 8.29 þ 3 8.86 þ 3 9.49 þ 3 9.94 þ 3 1.01 þ 4 9.96 þ 3 1.04 þ 4 1.09 þ 4 1.12 þ 4 1.12 þ 4 1.15 þ 4 1.12 þ 4 1.16 þ 4 1.17 þ 4 1.21 þ 4 1.23 þ 4 1.26 þ 4 1.33 þ 4 1.23 þ 4 1.16 þ 4 1.17 þ 4 1.17 þ 4 1.15 þ 4 1.11 þ 4
3.13 þ 3 3.26 þ 3 3.57 þ 3 3.77 þ 3 3.89 þ 3 4.07 þ 3 4.27 þ 3 4.49 þ 3 4.53 þ 3 4.99 þ 3 5.26 þ 3 5.56 þ 3 5.94 þ 3 6.36 þ 3 6.67 þ 3 6.86 þ 3 7.20 þ 3 7.54 þ 3 7.90 þ 3 8.18 þ 3 8.36 þ 3 8.36 þ 3 8.83 þ 3 9.23 þ 3 9.30 þ 3 9.55 þ 3 8.93 þ 3 9.23 þ 3 9.68 þ 3 1.02 þ 4 1.02 þ 4 1.05 þ 4 1.04 þ 4 1.01 þ 4 8.83 þ 3
1.16 þ 4 4.21 þ 3 2.45 þ 3 2.59 þ 3 2.68 þ 3 2.79 þ 3 2.93 þ 3 3.08 þ 3 3.12 þ 3 3.39 þ 3 3.56 þ 3 3.76 þ 3 4.00 þ 3 4.27 þ 3 4.47 þ 3 4.62 þ 3 4.86 þ 3 5.11 þ 3 5.36 þ 3 5.57 þ 3 5.74 þ 3 6.09 þ 3 6.43 þ 3 6.74 þ 3 6.95 þ 3 7.24 þ 3 7.10 þ 3 7.35 þ 3 7.72 þ 3 8.14 þ 3 8.26 þ 3 7.52 þ 3 7.75 þ 3 7.88 þ 3 7.85 þ 3
7.51 þ 3 7.89 þ 3 8.81 þ 3 9.49 þ 3 8.29 þ 3 2.51 þ 3 2.01 þ 3 2.12 þ 3 2.17 þ 3 2.34 þ 3 2.47 þ 3 2.61 þ 3 2.77 þ 3 2.96 þ 3 3.10 þ 3 3.22 þ 3 3.38 þ 3 3.53 þ 3 3.70 þ 3 3.87 þ 3 4.02 þ 3 4.27 þ 3 4.51 þ 3 4.68 þ 3 4.77 þ 3 5.13 þ 3 5.34 þ 3 5.55 þ 3 5.82 þ 3 5.70 þ 3 5.72 þ 3 5.99 þ 3 6.22 þ 3 6.35 þ 3 6.22 þ 3
6.36 þ 3 6.56 þ 3 7.12 þ 3 6.10 þ 3 6.19 þ 3 6.59 þ 3 6.94 þ 3 7.32 þ 3 4.57 þ 3 1.61 þ 3 1.73 þ 3 1.87 þ 3 2.01 þ 3 2.14 þ 3 2.32 þ 3 2.42 þ 3 2.54 þ 3 2.45 þ 3 2.57 þ 3 2.69 þ 3 2.92 þ 3 3.11 þ 3 3.28 þ 3 3.28 þ 3 3.31 þ 3 3.88 þ 3 4.04 þ 3 4.20 þ 3 4.42 þ 3 4.57 þ 3 4.46 þ 3 4.57 þ 3 4.68 þ 3 4.49 þ 3 4.64 þ 3
5.21 þ 3 5.22 þ 3 5.43 þ 3 5.26 þ 3 5.47 þ 3 5.73 þ 3 5.84 þ 3 4.88 þ 3 5.01 þ 3 5.49 þ 3 5.68 þ 3 5.70 þ 3 2.84 þ 3 1.56 þ 3 1.47 þ 3 1.54 þ 3 1.62 þ 3 1.77 þ 3 1.86 þ 3 1.95 þ 3 1.99 þ 3 2.12 þ 3 2.24 þ 3 2.29 þ 3 2.36 þ 3 2.50 þ 3 2.61 þ 3 2.72 þ 3 2.88 þ 3 3.04 þ 3 3.04 þ 3 3.18 þ 3 3.30 þ 3 3.40 þ 3 3.48 þ 3
Continued
Appendix XI
Emitter wavelength (A˚) energy (keV)
Absorber
Atomic number 91 92 93 94
Element
B 6.67 þ 1a 1.83 7 1
C 4.48 þ 1 2.77 7 1
N 3.16 þ 1 3.92 7 1
O 2.36 þ 1 5.25 7 1
F 1.83 þ 1 6.77 7 1
Ne 1.46 þ 1 8.49 7 1
Na 1.19 þ 1 1.04 þ 0
Mg 9.89 þ 0 1.25 þ 0
Al 8.34 þ 0 1.49 þ 0
Pa U Np Pu
2.34 þ 3 2.57 þ 3 2.93 þ 3 3.73 þ 3
2.38 þ 3 2.30 þ 3 2.41 þ 3 2.39 þ 3
3.44 þ 3 2.70 þ 3 2.62 þ 3 2.72 þ 3
1.13 þ 4 1.09 þ 4 9.65 þ 3 6.59 þ 3
9.45 þ 3 9.39 þ 3 9.74 þ 3 1.01 þ 4
8.35 þ 3 8.36 þ 3 8.59 þ 3 8.24 þ 3
5.96 þ 3 6.03 þ 3 6.34 þ 3 6.66 þ 3
4.97 þ 3 4.96 þ 3 5.21 þ 3 5.40 þ 3
3.71 þ 3 3.76 þ 3 3.74 þ 3 3.89 þ 3
a Notation as in Appendix VI, Table 4. Source: From Veigele, 1974.
Copyright © 2002 Marcel Dekker, Inc.
APPENDIX XII: CORRESPONDENCE BETWEEN OLD SIEGBAHN AND NEW IUPAC NOTATION X-RAY DIAGRAM LINES Siegbahn
IUPAC
Siegbahn
IUPAC
Ka1 Ka2 Kb1 KbI2
K-L3 K-L2 K-M3 K-N3
La1 La2 Lb1 Lb2
L3-M5 L3-M4 L2-M4 L3-N5
Lg1 Lg2 Lg3 Lg4
L2-N4 L1-N2 L1-N3 L1-O3
KbII2 Kb3 KbI4 KbII4 Kb4x KbI5 KbII5
K-N2 K-M2 K-N5 K-N4 K-N4 K-M5 K-M4
Lb3 Lb4 Lb5 Lb6 Lb7 Lb07 Lb9 Lb10 Lb15 Lb17
L1-M3 L1-M2 L3-O4,5 L3-N1 L3-O1 L3-N6,7 L1-M5 L1-M4 L3-N4 L2-M3
Lg04 Lg5 Lg6 Lg8 Lg08 LZ Ll Ls Lt Lu Lv Ma1 Ma2 Mb Mg Mz
L1-O2 L2-N1 L2-O4 L2-O1 L2-N6(7) L2-M1 L3-M1 L3-M3 L3-M2 L3-N6,7 L2-N6(7) M5-N7 M5-N6 M4-N6 M3-N5 M4,5-N2,3
Source: From Jenkins et al., 1991.
Copyright © 2002 Marcel Dekker, Inc.
Siegbahn
IUPAC
REFERENCES Arndt UW, Brown PJ, Colliex C, Cowley JM, Creagh DC, Dolling G, Fink M, Freund AK, Hilderbrandt R, Howie A, Hubbell JH, Gjønnes, Lander G, Lynch DF, McAuley WJ, Pynn R, Ross AW, Rowe JM, Sears VF, Spence JCH, Steeds JW, Valvoda V, Willis BTM, Wilson AJC, Zvyagin BB. In: Wilson AJC, ed. International Tables for Crystallography. Vol. C. Mathematical, Physical and Chemical Tables. Dordrecht: Kluwer Academic Publishers, 1992, p 162. Auger P. J Phys 6:205, 1925. Bambynek W, Crasemann B, Fink RW, Freund HU, Mark H, Swift CD, Price RE, Venugopala Rao P. Rev Mod Phys 44:716, 1972; erratum: 46:853, 1974. Behrens P. Trends Anal Chem 11:218, 1992a. Behrens P. Trends Anal Chem 11:237, 1992b. Bethe HA. In: Annalen der Physik, Band 5. Leipzig: Verlag von Johann Ambrosius Barth, 1930, p 325. Bethe HA, Ashkin J. Experimental Nuclear Physics, Vol. 1. New York: Wiley, 1953, p 252. Birks LS. Electron Probe Microanalysis. New York: Wiley–Interscience, 1971a, p 51. Birks LS. Electron Probe Microanalysis. New York: Wiley–Interscience, 1971b, p 172. Burhop EHS. The Auger Effect and Other Radiationless Transitions, New York: Cambridge University Press, 1952. Burr A. In: Robinson JW, ed. Handbook of Spectroscopy, Vol. I. Cleveland, OH: CRC Press, 1974, p 24. Campbell JI, Cookson JA. Nucl Instrum Methods Phys Res B3:185, 1984. Clark GL. Applied X-Rays, New York: McGraw-Hill, 1955, p 144. Clark GL. In: Clark GL, ed. Encyclopedia of X-Rays and Gamma Rays. New York: Reinhold Publishing, 1963, p 1124. Compton AH. Phys Rev 21:15, 1923a. Compton AH. Phys Rev 22:409, 1923b. Compton AH, Allison SK. X-Rays in Theory and Experiment. New York: Van Nostrand, 1935, p 537. Condon EU. In: Condon EU, Odishaw H, eds. Handbook of Physics, Part 7. New York: McGrawHill, 1958, p 7. Davis B, Mitchell DP. Phys Rev 32:331, 1928. Deslattes RD. Acta Crystallogr A25:89, 1969. Dirac PAM. The Principles of Quantum Mechanics. 4th ed. Oxford: Clarendon Press=Oxford University Press, 1958, p 136. Dyson NA. X-rays in Atomic and Nuclear Physics. London: Longman Group, 1973, p 42. Evans RD. In: Flu¨gge S, ed. Encyclopedia of Physics, Vol. XXXIV. Berlin: Springer-Verlag, 1958, p 218. Evans RD. In: Gray DE, ed. American Institute of Physics Handbook. New York: McGraw-Hill, 1963, p 8. Fink RW. In: Robinson, JW, ed. Handbook of Spectroscopy, Vol. I. Cleveland, OH: CRC Press, 1974, p 219. Goldstein JI, Newbury DE, Echlin P, Joy DC, Fiori C, Lifshin E. Scanning Electron Microscopy and X-ray Microanalysis. New York: Plenum Press, 1981, p 105. Green M, Cosslett VE. Proc Phys Soc 78:1206, 1961. Hanke W, Wernisch J, Po¨hn C. X-ray Spectrom 14(1):43, 1985. Hasnain SS, Ed. X-ray Absorption Fine Structure, Proceedings of the Conference Held in York, August 1990. Chichester: Ellis Horwood, 1991. Heinrich KFJ. Electron Beam X-Ray Microanalysis. New York: Van Nostrand–Reinhold, l98l, p 234. Heitler W. The Quantum Theory of Radiation. 3rd ed. London: Oxford University Press, 1954, p 207.
Copyright © 2002 Marcel Dekker, Inc.
Hirsch PB. In: Thewlis J, ed. Encyclopaedic Dictionary of Physics. Oxford: Pergamon Press, 1962, p 800. Hubbell JH. Photon Cross Sections, Attenuation Coefficients, and Energy Absorption Coefficients from 10 keV to 100 GeV. Washington, DC: National Bureau of Standards, l969. Hubbell JH. Int J Appl Rad Isot 33:1269, 1982. Hubbell JH. Proceedings of the Workshop on New Directions in Soft X-Ray Photoabsorption, Asilomar Conference Center, Pacific Grove, CA, 1984. Hubbell JH. Report NISTIR-89-4144, U.S. Department of Commerce, National Institute of Standards and Technology, August 1989. Hubbell JH, McMaster WH, Kerr Del Grande N, Mallett JH. In: Ibers JA, Hamilton WC, eds. International Tables for X-Ray Crystallography, Vol. 4. Birmingham, UK: Kynoch Press, 1974, p 47. Hubbell JH, Veigele VJ, Briggs EA, Brown RT, Cromer DT, Howerton RJ. J Phys Chem Ref Data 4(3):471, 1975. Hubbell JH, Trehan PN, Singh N, Chand B, Mehta D, Garg ML, Garg RR, Singh S, Puri S. J Phys Chem Ref Data 23(2):339, 1994. Jenkins R, Gould RW, Gedcke D. Quantitative X-Ray Spectrometry. New York: Marcel Dekker, 1981. Jenkins R, Manne R, Robin J, Senemaud C. Pure Appl Chem 63(5):735, 1991. Kallithrakas-Kontos N. Spectrochim Acta Part B51:1655, 1996. Kawai J, Gohshi Y. Spectrochim Acta 41B:265, 1986. Kawai J, Nakajima K, Gohshi Y. Spectrochim Acta 48B:1281, 1993. Kenney RW. In: Besanc¸on RM, ed. Encyclopedia of Physics. New York: Reinhold Publishing, 1966, p 86. Klein O, Nishina Y. Z Phys 52:853. 1929. Koningsberger DC, Prins R, eds. X-ray Absorption. Chemical Analysis Vol. 92, New York: Wiley, 1988. Kramers HA. Philos Mag 46:836, 1923. Krause MO. J Phys Chem Ref Data 8(2):307, 1979. Krause MO, Oliver JH. J Phys Chem Ref Data 8(2):329, 1979. Krishnan KS. Nature 122:961, 1928. Kulenkampff H. Ann Phys 69:548, 1922. Kulenkampff H. Ann Phys 87:597, 1928. Kulenkampff H. Handbuch der Physik, Bd. 23. Berlin: Geiger-Scheel 1933, p 433. Lagarde P. Nucl Instrum Methods 208:621, 1983. Lee PA, Pendry JB. Phys Rev B11:2795, 1975. Lifshin L. Adv X-ray Anal 19:113, 1976. Love G, Scott VD, Scanning 4:111, 1981. Love G, Cox MG, Scott VD. J Phys D 11:7, 1978. Markowicz AA, Van Grieken RE. Anal Chem 56:2049, 1984. Markowicz AA, Storms HM, Van Grieken RE. Anal Chem 57:2885, 1985. Markowicz AA, Storms H, Van Grieken R. X-ray Spectrom 15:131, 1986. McGuire EJ. Phys Rev A9: 1840, 1974. McMaster WH, Kerr Del Grande N, Mallett JH, Hubbell JH. Compilation of X-Ray Cross Sections. Lawrence Radiation Laboratory (Livermore) Report UCRL-50174, Sec. II, Rev. 1, University of California, 1969. Merzbacher E, Lewis HW. In: Flu¨gge S, ed. Encyclopedia of Physics, Vol. 34. Berlin: SpringerVerlag, 1958, p 166. Mizuno Y, Ohmura Y. J Phys Soc Japan 22:445, 1967. Moseley HGJ. Philos Mag 27:703, 1914. Norrish K, Tao GY. X-ray Spectrom 22:410, 1993. Orlic I, Loh KK, Sow CH, Tang SM, Thong P. Nucl Instrum Methods Phys Res B74:352, 1993.
Copyright © 2002 Marcel Dekker, Inc.
Pirenne MN. The Diffraction of X-Rays and Electrons by Free Molecules. New York: Cambridge University Press, 1946. Poehn C, Wernisch J, Hanke W. X-ray Spectrom 14(3):120, 1985. Rao-Sahib TS, Wittry DB. Proceedings of the 6th International Conference on X-Ray Optics and Microanalysis. Tokyo: University of Tokyo Press, 1972, p 131. Rebohle L, Lehnert U, Zschornack G. X-ray Spectrom 25:295, 1996. Sandstro¨m AE. In: Flu¨gge S, ed. Encyclopedia of Physics, Vol. 30. Berlin: Springer-Verlag, 1957, p 78. Sauter F. In: Annalen der Physik, Band 9. Leipzig: Verlag von Johann Ambrosius Barth, 1931, p 217. Sayers DE, Lytle FW, Stern EA. Adv X-ray Anal 13:248, 1970. Scherzer O. Ann Phys 13:137. 1932. Schreiber TP, Wims AM. X-ray Spectrom 11:42, 1982. Sherry WM, Vander Sande JB. X-ray Spectrom 6:154, 1977. Smith DGW, Reed SJB. X-ray Spectrom 10:198, 1981. Sparks CJ Jr. Phys Rev Lett 33:262, 1974. Sparks CJ Jr. In: Ramaseshan S, Abrahams SC, Hodgkin D, eds. Proc. Inter-Congress Conference on Anomalous Scattering held in Madrid, Spain, 22–26 April l974. Copenhagen: Munksgaard, 1975, p 175. Statham PJ. X-ray Spectrom 5:154, 1976. Statham PJ. Proc Annu Conf Microbeam Anal Soc 14:247, 1979. Stephenson ST. In: Flu¨gge S, ed. Encyclopedia of Physics, Vol. 30. Berlin, Springer-Verlag, 1957, p 337. Stobbe M. In: Annalen der Physik, Band 7. Leipzig: Verlag von Johann Ambrosius Barth, 1930, p 661. Stoev KN, Dlouhy JF. X-ray Anal 37:697, 1994. Storm E, Israel HI. Nucl Data Tables A7:565, 1970. Suzuki T. J Phys Soc Japan 21:2087, 1966. Suzuki T, Nagasawa H. J Phys Soc Japan 39:1579, 1975. Suzuki T, Kishimoto T, Kaji T, Suzuki T. J Phys Soc Japan 29:730, 1970. Szalo´ki I. X-ray Spectrom 25:21, 1996. Tertian R, Claisse F. Principles of Quantitative X-Ray Fluorescence Analysis. London: Heyden, 1982. Tertian R, Broll N. X-ray Spectrom 13:134, 1984. Veigele WJ. In: Robinson JW, ed. Handbook of Spectroscopy, Vol. I. Cleveland, OH: CRC Press, 1974, p 155. Ware NG, Reed SJB. J Phys E 6:286, 1973. Wentzel G. Z Phys 27:257, 1924. Wernisch J. X-ray Spectrom 14:109, 1985. Wernisch J, Po¨hn C, Hanke W, Ebel H. X-ray Spectrom 13:180, 1984. West RC, ed. Handbook of Chemistry and Physics. 63rd ed. Cleveland, OH: Chemical Rubber Co, 1982–83, p E-l88.
Copyright © 2002 Marcel Dekker, Inc.
2 Wavelength-Dispersive X-ray Fluorescence Jozef A. Helsen Catholic University of Leuven, Leuven, Belgium
Andrzej Kuczumow Lublin Catholic University, Lublin, Poland
I.
INTRODUCTION
Moseley’s law, formulated soon after the discovery of x-rays by Ro¨ntgen (Ro¨ntgen, 1896), represented the direct and definite onset of the use of x-ray spectrometry in chemistry. The first successful period was completed in the beginning of the 1920s by creating order in the periodic table of the elements and by the discovery of the missing elements (Weebs and Leicester, 1967). X-ray fluorescence (XRF) spectrometry has some unique features so that it became an irreplaceable tool for the analyst. Not may techniques had such a brilliant start to their analytical careers! gThe first commercial instrument became available around 1940 and was derived from the x-ray goniometer of a diffraction instrument. A reasonable estimation of the number of wavelength-dispersive (WDXRF) instruments working today is 15000, of which some 20% are multichannel spectrometers. In addition, about 3000 instruments working in this mode are attached to the electron microprobes, mainly in Japan. Some 3000 energy-dispersive (EDXRF) instruments are working independently, and some 1500–2000 others are attached to electron and other [e.g., proton (Uda et al., 1987, Folkmann and Frederiksen, 1990)] microprobes; a few hundred instruments are connected to synchrotrons as excitation sources. About 300 instruments are working now in total reflection (TRXRF) and several in grazing emission modes (Klockenka¨mper, 1997). The average price of a new WDXPF instrument in 1998 is of the order of magnitude of $250,000, $100,000 for EDXRF, and $200,000 for TRXRF (Claes et al., 1997) and this is not expected to rise much in the coming years. The calculation of the total money involved in the x-ray instrument market made hereafter was based on the 50% amortization estimate per instrument. We consider in our account only these parts of Particle-induced x-ray emission (PIXE), electron microscope, and synchrotron arrangements which can be treated as x-ray attachments. An auxiliary market was created also around the main equipment business, covering the spare parts, recent supplements, and service connected with the maintenance. Then, the commercial value of the instruments running today may be estimated at some US$4 billion (Markowicz and Van Grieken, 1986; Van Grieken et al., 1986). An important body of scientific literature relates to x-ray spectrometry Copyright © 2002 Marcel Dekker, Inc.
(Markowicz and Van Grieken, 1986; van Grieken et al., 1986; Jenkins, 1984; de Vries, 1986; Klockenka¨mper, 1987; Kikkert, 1983; Potts and Webb, 1992; To¨ro¨k et al., 1996, To¨ro¨k et al., 1998; Szalo´ki et al., 2000; Brill, 1996; Ellis et al., 1997 and Quisefit, 1996). Literature about essentially diffraction and scattering problems can also be very useful for the researchers working in the field of XRF (Fewster, 1996). About 10% of the WDXRF instruments are applied in geological research and prospecting (sequential and simultaneous spectrometers). Most of the simultaneous instruments are installed in industry, and the sequential instruments are distributed over industry, analytical laboratories, and research institutes at universities and public services. X-ray fluorescence spectrometry is unique in element analysis of inanimate matter: In qualitative analysis, it has unrivaled selectivity for all elements between beryllium and uranium and also for transuranium elements and an extremely wide dynamic range in quantitative analysis (ppm to 100%). Although the plasma emission spectrometry [Direct Current Plasma (DCP) or Inductively Coupled Plasma (ICP)], atomic absorption spectrometry (AAS), neutron activation analysis (NAA), and different kinds of mass spectrometry (MS) have lower or much lower detection limits, the wide dynamic range remains a unique feature of XRF (over five decades). Moreover, it possesses the scarce virtue among spectral methods—the upper useful limit reaches the concentration of several tens of percent of the total concentration for the main component. The detection limits at the ppm level without preconcentration and the limits of precision and accuracy that can be obtained (connected, in turn, to sample homogeneity and counting statistics) determine the boundaries of its application. Earlier books describing the techniques are mentioned in the reference list (Compton and Allison, 1935; Jenkins and De Vries, 1967; Birks, 1969; Dyson, 1973; Jenkins, 1974; Bertin, 1975; Tertian and Claisse, 1982; Lachance and Claisse, 1995, Jenkins et al., 1995; Buhrke et al., 1998—the latter is a kind of a guide for sample preparation in XRF and XRD, but is cited here, because the problem is very important and nontrivial). The purpose here is to describe the basic principles that allow wavelength dispersion of x-rays and to describe all the components of a spectrometer. The phenomenon of x-ray fluorescence is only a small part of a much wider problem of the interaction of charged particles or photons with matter (Feldman and Mayer, 1986). Different kinds of secondary particles are then emitted (Fig. 1a). Various combinations of exciting=emitted particles are possible, giving birth to different analytical techniques, with the combination ‘‘photons as exciting particles=emitted photons’’ as a base for the x-ray fluorescence. The angles the trajectories of the exciting or emitted particles create with the surface of the sample define normal XRF and PIXE methods and total reflection versions: TXRF (Chapter 9) and TRPIXE (Vis and van Langevelde, 1991; van Kan and Vis, 1996a, 1996b, 1997). In addition, for the x-ray fluorescence, a grazingemission method can be mentioned (GEXRF) (Sasaki, 1990; Becker et al., 1983; de Bokx and Urbach, 1995; Urbach and de Bokx, 1996; Tsuji et al., 1995; Pe´rez and Sa´nchez, 1997) (Fig. 1b). Insofar as the fundamental principles are concerned, we may recall the law of Moseley (Moseley, 1913, 1914), which is the basis of the qualitative application of XRF: pffiffiffi n¼
rffiffiffi c ¼ kðZ sÞ l
ð1Þ
where n is the frequency (in Hz), c is the speed of light (in m=s), l is the wavelength (in m), k is the constant for a given series, and s is the screening constant. The relationship Copyright © 2002 Marcel Dekker, Inc.
Figure 1 Phenomena accompanying the interaction of x-rays with matter: (a) different applications for analytical purposes (abbreviations of methods coupled with the use of particular phenomena are mentioned; (b) different versions according to the incidence and emergence angles (I—total reflection XRF or PIXE; II—grazing emission XRF; III—grazing incidence–grazing emission XRF).
Copyright © 2002 Marcel Dekker, Inc.
between the wavelength of the lines belonging to the same spectral series and the atomic number is illustrated by Figure 2. For high-Z elements, the XRF spectra become more complex and full exploitation for qualitative analysis of mixtures is not always simple. Practical unraveling of spectra is dealt with in Chapter 4. The intensity of any spectral line is proportional to the number of atoms emitting photons of energies attributed to this line. However, a simple linear proportionality is not the rule. From the concepts of mass attenuation coefficients introduced in Chapter 1, it should be clear that the intensity of an analyte line of one species is attenuated by atoms of the same species and by any other atoms present in the matrix. Matrix refers to all elements present in a sample except the analyte element. Attenuation by absorption represents a first complication compromising the proportionality. If the matrix contains elements with absorption edges of slightly lower energy than the energy of the characteristic line of the analyte, strong absorption can occur. On the one hand, this results in further attenuation of the analyte line and, on the other hand, in enhancement of the spectral line related to the said absorption edge of the other matrix element. This phenomenon is known as secondary fluorescence. The process may be repeated with respect to all matrix elements with an absorption edge with a lower energy than the energy of a fluorescence line emitted by another matrix element. Secondary and higher-order fluorescence is the major complication deteriorating the simple proportionality between intensity and concentration. In Figure 3, three typical kinds of relationship between the relative intensity (measured intensity divided by the intensity obtained for the pure element) and concentration (expressed as weight fraction) are represented. Curve I is obtained when the analyte line undergoes absorption only. If enhancement occurs, the analyte line intensity is higher than expected from primary excitation only and the curve is situated above the diagonal. A similar curve, however, can be obtained when the mass attenuation coefficient of the matrix is lower than the related coefficient of the element for its own radiation, as shown in the case for Fe in FeO (curve II). In a special case, the mutual relation between attenuation
Figure 2
Representation of Moseley’s law of K and L spectral lines.
Copyright © 2002 Marcel Dekker, Inc.
Figure 3 Absorption phenomena in a heavy matrix (curve I), in a light matrix (curve II), and in a neural matrix (curve III).
and enhancement coefficients may be balanced such that both effects cancel precisely and then the diagonal is approximated (curve III). The effects just discussed are all energy (wavelength) dependent. This means that any calculations converting intensities into concentration should include the attenuation and higher-order fluorescence effects and also integrate over all energies present in the exciting primary beam above the value of the absorption edge and over all fluorescence lines. This is an ab initio approach for conversion of intensities into concentration implemented by several authors. We mention here the articles by Gillam and Heal (1952), Sherman (1954, 1955), Shiraiwa and Fujino (1966, 1967), who in the 1950s and the 1960s proposed suitable equations. Sparks (1976) and Li-Xing (1984) implemented small corrections and refinements concerning the details of these expressions, and de Boer (1990) and de Boer and Brouwer (1990) gave solutions to some difficult exponential integrals related to the enhancement. Important contributions to the numerical side of quantitative XRF are due to Fernandez (1989), Fernandez and Molinari (1990), Rousseau et al. (1996), and Mantler (1984). The first two authors prepared commercial, customer-oriented versions of the numerical programs XRFPc and CiROU, CiLAC, CiLT, and CiREG, respectively, and Mantler paved the way for the analyses of multilayers. It should be emphasized that Fernandez solved the transport equations for x-ray beams, the very general and powerful but difficult tool for the description of particle beams. Likewise, Gardner and Hawthorne (1975) obtained similar results by Monte Carlo simulation of x-ray excitation. A specific approach was presented by Dane et al. (1996). These authors solved the problem of quantification in situations where either an incomplete data set is available or where a reasonable assumption about composition and=or depth profile cannot be done. Then, the idea of processing whole sets of solutions (populations or strings or generations) has been introduced, instead of processing the particular concentration values in consecutive iterations. In the fundamental parameter methods, the incidence and takeoff angles are the important parameters. As a rule, they are kept constant during the traditional run of the analysis. However, there have been significant efforts to use the angle-resolved version Copyright © 2002 Marcel Dekker, Inc.
of x-ray XRF spectrometers for depth profiling of the samples (Gries, 1992; Schmitt et al., 1994). For a complete treatment of equations deriving the line intensities in x-ray fluorescence spectrum from the elemental sample composition, the reader referred is to Chapter 5.
II.
FUNDAMENTALS OF WAVELENGTH DISPERSION
The crystal monochromator is the heart of a wavelength-dispersive spectrometer. Wavelength dispersion of electromagnetic radiation in the x-ray region cannot be performed, as a rule, by normal gratings but only by diffraction on crystals or, for the long-wavelength regions, on multilayers. We briefly explain the principles because the construction features of the monochromator are directly derived from them. Consider a monochromatic beam of x-rays of wavelength l with their electrical vectors of equal amplitude in phase along any point of the direction of propagation. Assume further that the beam is parallel and is incident on a crystal at an angle W between a given crystal plane (and all the planes parallel to this first) and the incident beam direction. The beam is scattered and diffracted rays of equal l result but interfering constructively only in those directions for which the phase relationship is conserved. This happens at an angle W for scattered rays 1 and 2 (Fig. 4), for which the path difference ABC of ray 2 with ray 1 is equal to an integral number n of wavelengths. From Figure 4, it is clear that AB þ BC ¼ d sin W þ d sin W ¼ nl or, according to Bragg and Bragg (1913), who first formulated this relation, written as nl ¼ 2d sin W
ð2Þ
where n denotes the number of wavelength differences between the rays scattered by the adjacent planes. Following the recommendation of the IUPAC (Freiser and Nancollas, 1987), n is the order of diffraction. If n ¼ 1, the difference is one wavelength and the diffraction is said to be of first order. If n ¼ 2, the difference is two wavelengths and the diffraction is second order, and so on. All x-rays emitted at angles different than W cancel because they are out of phase and destructive interference occurs. From Figure 4, one can note that in the Bragg scattering,
Figure 4
Derivation of the Braggs’ law. ABC is the path difference.
Copyright © 2002 Marcel Dekker, Inc.
the incident and exit angles are equal and, in this sense, there is some analogy to mirror reflection in the classical optics. However, diffraction is by no means equal to reflection in classical optics, because the diffraction is a volume not a surface process. It is less efficient (with a great loss of intensity) and is performed only under particular angular conditions; that is, according to Bragg’s law. Full analogy between the reflection of x-rays and the reflection of optical rays happens only at grazing incident angles (i.e., smaller than the critical Snell’s angle). This is intrinsically a consequence of the refraction index, which is slightly smaller than 1 for x-rays. There are continuous efforts to construct a real mirror on the basis of multilayers, which would efficiently reflect x-rays under large incident angles, at least for soft x-rays (Kearney et al., 1991). Thus, the static condition for obtaining diffraction of a monochromatic x-ray beam in some direction in the volume surrounding the analyzing crystal is given. What happens in the case of a polychromatic beam of x-rays? For a crystal, one set of planes is selected (for different reasons) and d is constant. If only first-order diffraction is considered and constructive interference must be realized for all l present in the incident beam, then W is the only variable: l ¼ ðconstantÞðsin WÞ
ð2aÞ
The signal arriving from the diffraction angle W is detected by a detector placed on a goniometer arm. The detector rotates around an axis through the macroscopic plane of the analyzing crystal. For a source at a fixed position, the detector rotates over an angle 2W. The wavelength is calculated from constant ¼ 2d and sin W. Note that this holds only for the first-order diffraction. If the second-order diffraction is used, the wavelength is equal to half of that value. The maximum wavelength lmax that can be diffracted in the first-order diffraction by a crystal is equal to 2d because sin W 1. Because wavelength and energy are related, one more equations must be given, allowing us to convert wavelengths into energy units: E¼
12398:5 l
ð3Þ
where the energy E is in electron volts (eV) and l is in angstroms (A˚). It is a common practice that the units used by x-ray spectroscopists are still electron volts or kiloelectron volts for energy and angstroms for wavelength. When joules and nanometers are used, as required by the international rules, the numerical value of the conversion constant becomes 1.9864561016 (nm=J). The presence of different wavelengths of different order on the same goniometer position has a particular consequence for wavelength-dispersive spectrometry. For a given position of the goniometer (and detector), one may have a first-order wavelength, say 0.6 A˚, second-order diffraction of l ¼ 0.3 A˚, third-order diffraction of l ¼ 0.2 A˚, or values of energy 20.66, 41.32, or 61.98 keV, respectively. Figures 5 and 6, reproduced from the work of Arai (1987) demonstrate a practical situation and Table 1 reprints a fragment from x-ray tables by Cauchois and Senemaud (1978). Similar compendia of spectral lines and attenuation coefficients are easily available (Birks, 1974), as a rule covering the range of elements between lithium (Z ¼ 3) to plutonium (Z ¼ 94). Much more controversial are the values of the mass attenuation coefficients for low energies and for low-Z elements at the same time (Henke et al. 1982, 1988). If necessary, one can have access to the calculation for even heavier elements (Soff et al., 1977). It is easy to understand the practical value of such calculations for the analysis. However, it is much more difficult to estimate the accuracy of numerical calculations for transuranium elements. To sort out x-rays with Copyright © 2002 Marcel Dekker, Inc.
Figure 5 Higher-order overlapping of L series x-rays of Cr, Fe, and Ni to Ka of carbon. (From Arai, 1987.)
Figure 6
Overlapping of CKa by higher-order x-rays of Ni, Fe, and Cr. (From Arai, 1987.)
wavelengths belonging to different spectral orders, a pulse-amplitude selector or detector is needed that can make the distinction between energy levels. A detector is a device of which the principle can be most easily explained by assuming the particle character of electromagnetic radiation. This is one reason that the characteristic of the impacting photons is often expressed in terms of energy, not wavelength. Copyright © 2002 Marcel Dekker, Inc.
Table 1
Fragment of Detailed Tables of X-ray Lines
l (uX) 1251.56 1251.64(a) 1252.192 1252.58 1252.6 1252.82 1255.17 1255.43 1255.50 1256.60 1256.71 1257.1(a) 1257.12 1257.66 1258.74 1259.210 1259.8 1260.30 1260.6 1260.620(a) 1261.15(a) 1261.23 1261.956 1262.68 1263.086 1263.428(a) 1264.6 1264.99(a) 1265.0 1265.24(a) 1268.520(s) 1268.66 1268.68(a) 1270.2 1270.35(a)
l (mA˚) 1254.19 1254.27 1254.824 1255.21 1255.2 1255.45 1257.81 1258.07 1258.14 1259.24 1259.35 1259.7 1259.76 1260.30 1261.39 1261.856 1262.4 1262.95 1263.2 1263.269 1263.80 1263.88 1264.608 1265.33 1265.741 1266.083 1267.3 1267.65 1267.7 1267.90 1271.186 1271.33 1271.35 1272.9 1273.02
Element 42 58 62 73 90 73 73 32 69 75 73 71 90 73 54 68 90 74 90 58 64 73 42 90 42 58 74 70 91 69 92 90 68 74 64
Mo Ce Sm Ta Th Ta Ta Ge Tm Re Ta Lu Th Ta Xe Er Th W Th Ce Gd Ta Mo Th Mo Ce W Yb Pa Tm U Th Er W Gd
Transition
Notation usuelle
Ordre
KMIV KMIV KLII abs. LIII LINIV LIIIOIV,V LIIINVI,VII KLII KLI LIMII LIIIOIII LIINI abs. LII LIIIOII KLII KLIII LIIPII,III LIMIII LIIPI KMIII KMIV,V LIIIOI KMIII LIIOIV KMII KMII LIIINIII LIINIV LIINIV LINIII LIINI LINIII LIOIV,V LIIMV KMIII
KbII5 KbII5 Ka2 — — Lb5 Lb7 Ka2 — Lb4 — Lg4 — — Ka2 Ka1 — Lb3 — Kb1 Kb5 Lb7 Kb1 Lg6 Kb3 Kb3 — Lg1 Lg1 Lg3 Lg5 Lg3 — — Kb1
2 4 4 1 2 1 1 1 5 1 1 1 2 1 3 5 2 1 2 4 5 1 2 2 2 4 1 1 2 1 2 2 1 1 5
E (keV)
n=R
ðn=RÞ1=2
9.8776
725.99
26.9441
9.8758 9.8573 9.8552
725.85 724.49 724.34
26.9416 26.9163 26.9135
9.8460 9.8452 9.842
723.66 723.60 723.4
26.9010 26.8998 26.8956
9.8377
723.05
26.8897
9.8171
721.54
26.8615
9.8099
721.01
26.8516
9.784 9.7807
719.1 718.86
26.8158 26.8116
9.7788
718.72
26.8090
9.7523 9.741
716.77 715.9
26.7726 26.7566
Source: From Cauchois and Senemaud, 1978.
The differences between wavelength-dispersive (WD) and energy-dispersive (ED) XRF are due not only to the different detectors but also to other factors: 1. The brightness of a WD spectrometer is very low, the attenuation in a crystal being responsible for an important part of losses. This problem may be overcome by the use of radiation sources of significant intensity, the synchrotron being the best known but bound to the availability of this large facility. 2. The crystal is only the dispersive device, not the detecting device. The situation is different in EDXRF, for which detectors play a double role: as the dispersive device and as the detector at the same time. Copyright © 2002 Marcel Dekker, Inc.
3.
4.
Because Bragg’s law is geometric in character, the angular conditions for the collimation of primary and secondary beams are very severe for WDXRF (8– 25 msr), not the same as for EDXRF (150 msr) (Wollman et al., 1997). This restriction on the geometrical efficiency still makes the overall photon collection efficiency of a wavelength-dispersive spectrometer worse. An attractive aspect of EDXRF is the simultaneous collection of the whole spectrum, whereas a typical WDXRF device is exclusively sequential. However, the maximum count rate for an EDXRF instrument is 30 kcps for the whole spectrum, which severely limits the total number of accumulated counts and, consequently, limits the precision (counting statistics). Similarly, the limitation on the total count rate leaves a small margin for trace element analysis. From the assumption, the trace element, if present in a sample, can emit only a very small part of the total radiation emitted by a whole sample (and there is also the background, participating in the total amount of counts).
Simultaneous WD instruments are composed of a series of individual crystal spectrometer (channels) operating simultaneously, but the number of channels is limited. A comparison of wavelength- and energy-dispersive versions of x-ray spectrometers is performed in a recent publication by Brill (1996).
III.
LAYOUT OF A SPECTROMETER
The main parts of a spectrometer can be represented in the simplest form by Figure 7. The analyzing crystal is the central point of the wavelength-dispersive instrument. On the lefthand side of the crystal, we find (1) the source of excitation, (2) the filters and devices for shaping the exciting radiation (collimators and masks), and (3) the sample. On the righthand side, we find (4) devices for shaping the diffracted beam (collimators) and (5) the detector. Signals from the detector are fed into the electronic circuitry where they are shaped to be processed by the computer software for data analysis. This arrangement can be reduced or made more complicated according to the demand. However, all possible reductions lead to less flexible devices, and complication does not necessarily enhance the quality of the instrument.
Figure 7 A wavelength-dispersive spectrometer: FC ¼ proportional flow counter; SD ¼ scintillation detector. (Reprinted with permission of Siemens AG.)
Copyright © 2002 Marcel Dekker, Inc.
A.
Sources
A variety of radiation sources, emitting either charged particles or g- or x-rays of sufficient energy, are used for excitation of some or all elements of the periodic table and some or most of the spectral lines of analytical interest. Other chapters deal with excitation by protons [proton-induced or sometimes particle-induced x-ray emission (PIXE) (Chapter 12)], by electrons [electron microprobe (Chapter 13)], or by x-rays emitted from secondary targets or from a synchrotron (Chapter 8). Excitation by x-rays or soft g-rays from radioisotopes and x-rays from low-power tubes is mainly restricted to energy-dispersive spectrometers (Chapter 3). The ideal excitation source would be a tunable x-ray laser (monochromatic and intense, allowing the best choice of exciting wavelength and often selective excitation), but this cannot be expected in the near future. Nevertheless, it is worth reading some treatises about recent and future progress in the field (Nagel, 1982; Jamelot, 1995; Fill, 1995; Crasemann, 1994; London, 1993). There are serious reasons for the slow progress in x-ray laser construction. The primary reason is that the population inversion of electrons means a much larger deviation from the energetical equilibrium if done between the levels, allowing the emission of x-rays. Consequently, it demands an excessive pumping power, turning the pumped matter into a plasma. One must be aware that the x-ray laser would be even much more useful in the fields of x-ray microscopy, holography, litography, or for the research of time-resolved phenomena than as a source for XRF analysis on a routine basis. The synchrotron is another modern source of x-ray radiation, but it cannot be considered as a tabletop instrument and a device for easy, inexpensive, and routine-based applications. From a practical point of view, vacuum x-ray tubes are the overwhelming choice among other potential excitation sources. High-power tubes are the only ones dealt with in detail in this chapter; low-power tubes are discussed in Chapter 3. All modern tubes owe their existence to Coolidge’s hot-cathode x-ray tube as presented in Physical Review some 85 years ago (Coolidge, 1913). They essentially consist of a sealed glass tube containing a hot tungsten filament for the production of electrons, a cooled anode, and a beryllium window. From a variety of modifications proposed over more than three-quarters of the century, two geometries have emerged as the most suitable for all practical purposes [the end-window tube (EWT) and the side-window tube (SWT)], but now the preponderance of the EWT pushes the SWT out of the market. Perhaps, a future deeper orientation of the WDXRF toward the microprobe applications will renew the interest in SWT, which is much better in deriving parallel x-ray beams. The general requirements of the x-ray tube as a source are as follows: 1. Sufficient photon flux over a wide spectral range, with increasing emphasis on the intensity of the long-wavelength tail of the white spectrum. The actual vivid interest in low-Z element analysis will certainly activate research in this direction. 2. Good stability of the photon flux (< 0.1% at least). Long-term stability reduces the frequency of recalibration in routine analyses; short-term stability is an absolute requirement for obtaining an acceptable precision. 3. Switchable tube potential (10–100 kV), allowing the creation of the most effective excitation conditions for each element. Still, in more high-power constructions available on the market, putting a filter in the beam path is an easier solution than switching the voltage (Shimadzu XRF-1700). For sure, the selection of the spectral region for the targeted excitation by the use of a single x-ray tube will never be as good as in dedicated synchrotron beam lines with Copyright © 2002 Marcel Dekker, Inc.
4.
monochromators of the adequate quality, but the low cost of tube operation is an obvious advantage. The intensity of the analyte lines varies considerably with excitation conditions. An extreme example is given by Vrebos and Helsen (1985a, 1985b) for simulated Al–Mo alloys. Freedom from too many interfering lines from the characteristic spectrum of the tube anode.
Freedom from interfering lines is important. The scattered characteristic lines of the anode may spectroscopically interfere with analyte lines, disturbing the qualitative recognition, the peak intensity estimation for the line of interest, and the accuracy of subsequent conversion of intensity to concentration. This is found in the results obtained after correction by some algorithms as well as by fundamental parameter calculations. Although there is some kind of a remedy by the use of more efficient spectral deconvolution methods (Remond et al., 1993; 1996), it occurs at the expense of a greatly increased computational effort and depends heavily on the assumptions made a priori. An x-ray tube is characterized by its anode element (a single element or two elements as in dual-anode tubes), its input power [expressed in watts (W) or kilowatts (kW), typically between 0.2 and 5 kW for high-power tubes], the voltage range between anode and cathode (10–100 kV for a SWT; limited to 60 kV for an EWT for the majority of commercial instruments), tube current [milliampere (mA), typically up to 60 mA or sometimes to 150 mA for high-power tubes], and an open (SWT) or closed (EWT) anode cooling circuit. The photon output of the tube or, more importantly, the photon flux hitting the sample (expressed in counts per second per watt of input power) is determined by a, the incidence angle of an electron beam on the anode, the takeoff angle b (for SWT), the distance t to the beryllium window, the thickness d of the beryllium window, and the distance between window and sample, t0 (Fig. 8). The radiation output power is rather poor with respect to the input power and is of the order of 1%, making the device a very inefficient transformer of electron current to electromagnetic radiation. For an energy balance, see Bertin (1975) and some remarks by Wollman et al. (1997). The impact of the electrons creates an excitation volume from which the white and the characteristic radiation of the target escape. This phenomenon represents the first distinct difference between a SWT and an EWT. As shown in Figure 9a, the ‘‘escape’’ path of the photons in SWT geometry is, on average, less than in the EWT geometry. An immediate consequence of this characteristic is that dual anodes are used only in a SWT, where a light element (e.g., scandium) covers a heavier element (e.g., molybdenum). By switching the excitation voltage, x-rays are produced either in the upper (lighter) element layer or in the substrate (higher-Z element), resulting in two distinct tube spectra with different yields in the low- and high-wavelength regions (Fig. 9b). The popular dual-target tube arrangements are now Rh=Cr and Rh=W. The smaller the distance t, the higher the output. A decrease of t to half of the original value increases the intensity roughly by a factor of 4. However, the reduction of t in the classical setup is limited. The smaller the value of t, the more intense the bombardment by the electrons and subsequent heating of the window by the scattered electrons (Fig. 8). In a SWT, this bombardment is rather intensive because both the anode and the window are in this geometry at ground potential. In a EWT, to the contrary, the filament and the window are both at ground potential, and heating of the window is negligible. For an EWT, however, t can be reduced to a certain extent only because it faces the anode at high potential (Fig. 10). Copyright © 2002 Marcel Dekker, Inc.
Figure 8
Details of the geometry of a side-window x-ray tube and sample.
1. Alternative Con¢gurations To overcome the absorption of the low-wavelength tail of the continuum, a windowless configuration was considered. This is an obvious solution, but it requires that the whole spectrometer with sample, collimators, crystals, and flow counter be evacuated to the low pressure suitable for securing an acceptably long life for the tube filament and conducting analyses in the required range of the soft x-rays. Nordfors (1956) advocated a dual-anode tube with two anodes physically separated. They were excited by deviating the electron beam to either of the two anodes (Fig. 11). Lack of stability prevented this solution from gaining commercial interest. Probably, a modified version of this idea was introduced in the series 2000 x-ray spectrometers of Diano (dual target þ dual filament; see Section VI). Tubes with exchangeable anodes are another possibility for solving the problem of comfortable switching between different anodes. To the best of our knowledge, this solution is not commercially available because of the tedious exchange operation and difficulties in rigorous repeating of previous conditions of the tube action. Using the sample as an anode is another solution, but then the application is bound to conducting materials. In a modified form, this solution is implemented in scanning electron microprobes. Because this ‘‘anode’’ cannot be adequately cooled, the input power must be restricted to very low values, entailing low count rates and, consequently, reducing the precision that can be obtained within reasonable counting times. Another possibility is to apply an existing 18-kW rotating anode x-ray tube generator. Sufficient intensity is obtained. Part of this gain can be sacrificed and Boehme (1987) and Nichols et al. (1987) introduced the microdiffraction beam collimator into the beam Copyright © 2002 Marcel Dekker, Inc.
Figure 9 (a) Principle of excitation in a side-window tube. (b) Principle of excitation in a dualanode x-ray tube (side window); Mo, W, or Au (for A) and Sc, Rh, or Cr (for B) are commercially available anode components. The spectral feature of the tube spectrum depends on the energy of the incident electrons.
derived from such a high-power tube. This collimates the beam to about 30 mm and allows microfluorescence. Squeezing the beam to more or less the same diameter is possible when applying the polycapillary lens, even more so with the single capillary. These solutions are very restrictive for the beam intensity; thus, a reasonable device configuration includes an energy-dispersive detector instead.
2. Optimization Because the takeoff angle is a very important parameter, the surface of the anode can be formed in steps in such a way that the yield of low-wavelength radiation is optimized. As is clear from Figure 12, the step-shaped surface decreases the escape depths. In a conventional EWT, the reduction of the anode to window distance t is limited (Fig. 10). According to an old idea of Thordarson (1939), Botden et al. (1952) developed an x-ray tube with a gold anode evaporated directly on the beryllium window. The tube was used for surface radiotherapy. After many years, Phillips returned to Botden’s construction and again reduced t to zero by sputtering the anode target element directly on the beryllium window (Fig. 13). In a SWT, the anode is at ground potential, which Copyright © 2002 Marcel Dekker, Inc.
Figure 10
Principle of end-window x-ray tubes.
Figure 11 X-ray tube with two separate anodes: (1) anodes: (2) filaments; (3) cooling; (4) separating brass wall; (5) insulator. (After Norfords, 1956.)
Figure 12
X-ray tube with stepwise configuration of the anode.
considerably facilitates and simplifies the safety requirements. This geometry enhances the intensity by an order of magnitude, allowing a much lower input power: 200 W for an intensity equivalent to the earlier 3-kW EWT. As a result of the low input power, the heat produced in the target by the electron beam is easily dissipated and no supplementary water cooling is required. The tube construction is reduced and only the most basic components are left. The commercial name for this development is T3 or TTT (target transmission tube; Fig. 14c). It is used exclusively in simultaneous WDXRF instruments, especially in the new version of the PW1660 (Philips Technical Materials Simultaneous X-Ray Spectrometer System PW 1660) (see Section VI). Copyright © 2002 Marcel Dekker, Inc.
Figure 13
Anode elements deposited on the beryllium window.
3. End-WindowTube The essential parts of an EWT are displayed in Figure 14a. The ring-shaped cathode and the cooling circuit of the anode are typical features of EWT construction. The latter consists of a closed circuit filled with deionized water. Deionized water is necessary to reduce the conductivity and enhance the safety because the circuit is at anode potential! Electrons are diverted to the anode surface by electron-optics elements. Because the filament and window are at equal potential, no electrons hit the window. Heating is mainly by radiation from the anode, and an external cooling circuit is provided, which may be connected to normal water supplies because it is not in contact with a voltage source. Although the construction just described is in a common use now, it cannot be considered as an ideal source of x-rays. Recently, van Sprang and Bekkers (1998) and Kuczumow et al. (2000) have discussed the deviations from the homogeneity in the endwindow x-ray tube output. The geometry of the excitation system, as described by an angle supported on the tube axis – the sample point – the goniometer entrance axis, deviates from the formal geometry (about 83 ). The ‘‘real,’’ not formal, geometry of the device can be read out from the shift in the Compton-scatter energy. It creates the special demands for all the constructors of microprobes and milliprobes. They must avoid using the conventional end-window tube. There are also the consequences for the normal analysis: The uniformity of the samples and spinning operation are essential for getting reproducible results. Otherwise, particular spots on a surface of the sample can be irradiated in a different manner. 4. Side-WindowTube An expanded view of a SWT is shown in Figure 14b. The essential differences from an EWT are the distance between cathode and anode, an earthed anode, and a single cooling circuit for the anode only. This circuit may be directly connected to the normal water supply because it is at ground potential. The geometric arrangement of the cathode supports higher potentials of the cathode with respect to the anode. This is the basic reason that tube potentials of 100 kV are allowed. Side-window tubes with fine-focus anodes are used in many microprobe constructions. They are appreciated for their photon output, easy to transform into a parallel, collimated beam. A consequence of the regulated anode potentials is the possibility of use of dual anodes. The available dual-anode tubes are summarized later in Table 10. 5. Tubes with Rotating Anode The electrical input in an x-ray generator is high; however, a final output of about 1% of primary power is not very impressive. This must be due to the low efficiency of conversion of the electrical energy into the energy of x-ray photons. One should remember that photon losses are even much greater in further parts of the spectrometer (especially Copyright © 2002 Marcel Dekker, Inc.
Figure 14 Different types of construction of commercial x-ray tubes: (a) end-window tube (EWT) (courtesy of Siemens AG); (b) side-window tube (SWT) (courtesy of Philips Analytical); (c) target transmission tube (TTT) (courtesy of Philips Analytical); (d) tube with a rotating anode (Vekemans 2000 courtesy of B. Vekemans).
Copyright © 2002 Marcel Dekker, Inc.
Figure 14
Continued
in collimators and crystals). The desparate need for more efficient excitation sources emerges with many more subtle applications of x-rays: the microprobe constructions, (micro)diffraction experiments with organic or biochemical materials, research on the fine structure of the absorption edge, focusing of x-rays, or the application of the total reflection, for example. If one does not have access to the synchrotron, no patience to wait for the beam time or no money for paying great sums for simple analyses, then the tube with a rotating anode seems to be the best solution. Copyright © 2002 Marcel Dekker, Inc.
Figure 14
Continued
According to our knowledge, three companies, Rigaku, Siemens, and Enraf-Nonius, constructed high-power tubes with a rotating anode. The anode, in the shape of a wide cylinder placed on a narrower cylindrical rotary shaft, can rotate with a speed of 100 rps. Mo, Cu, and Ag are the materials most frequently applied for the construction of the anodes. The sides of the anode are under intensive electron bombardment, but the high heat load is dissipated by the intensive rotation coupled with cooling by water, flowing inside the anode with a rate of 8–15 L=min. The system is evacuated by a prevacuum oilrotary pump followed by a turbo-molecular pump. It enables one to work with a voltage of 60 kV and a current of 300 mA, sometimes even greater. It gives a power of the order of 20–30 kW. Recently, the most powerful generator available is the Ru-1500 by Rigaku with a 90-kW load. The application of power of the order of 20–30 kW means roughly a 10-fold increase in the intensity as compared to the conventional x-ray tubes. When coupled with a Kumakhov lens, rotating anode tubes can be used as a powerful microfocusing source. 6. Compact Flash X-Ray Sources An x-ray tube is a very conventional x-ray source, dating back to the times of Ro¨ntgen and Coolidge. The efforts described earlier aimed at a better and more efficient construction of the x-ray tube. Different versions of the tube were by no means the only solution for constructing a better and more convenient source. The most pronounced progress of the last decades is the synchrotron, but it is still far from an ideal device in terms of price and size. Much more promising for most XRF laboratories seems to be a compact flash x-ray source (Germer, 1979). Such sources are now being developed intensively, giving, for example, a flux of photons with energies between a few kiloelectron volts to 200 keV, in pulses lasting between 10 and 50 ns, with a repetition frequency up to 50 Hz. Doses are very intensive, up to several roentgens per shot (Pouvesle et al., 1996). Primary x-rays from Copyright © 2002 Marcel Dekker, Inc.
such sources can be used for the excitation of x-ray fluorescence, being especially useful for real-time monitoring of kinetics of fast processes in ballistics and in the simulation of accidents. It seems that the polycapillary x-ray focusing waveguide would be an ideal supplement to this kind of x-ray source, giving the impetus for the development of a new generation of pulse x-ray spectrometer. This device would combine the time resolution of the source with the spatial resolution of the focusing polycapillary semilens. 7. High-Voltage Generator A high-voltage generator supplying up to 60 kV for an EWT or up to 100 kV for a SWT and an output power of 3–4 kW (or much lower, 0.2 kW for TTT tubes) is required. A special generator is needed if one wants the high-power rotating anode tube. The conventional power supplies were very cumbersome because of the size of the transformer. The new generation belongs to the so-called switched power supplies. Such a power supply is basically a dc=dc converter in which a dc voltage is electronically switched at high frequencies (several kilohertz) and fed into an inductance–capacitance network. The output power is not continuously regulated but is controlled through pulse-width modulation. One possible scheme of a switched power supply is given in Figure 15. The mains are rectified (a), high-frequency switched (b), transformed (c), and rectified and smoothed (d). The output voltage is sensed (e) and compared to a reference voltage (f), and this signal monitors the pulse-width modulator (g), which, in turn, commands the switching circuit (b). In the high-frequency transformer, a ferrite core is used; this is a lightweight component. The whole setup allows the size of the generator to be reduced to a fraction of the volume of classical generators. The whole system is generally monitored by microprocessor. Power switching of the tube is done preferentially along the isowatt curve (constant input power), all under microprocessor control. The system has a number of safety switches on flow and temperature controllers of cooling water with microswitches on all panels, which when
Figure 15
Principle of switched power supply: dc=dc convertor.
Copyright © 2002 Marcel Dekker, Inc.
removed could expose the customer either to the electrical circuitry (high voltage of the tube supply or the mains voltage) or to unacceptable radiation levels. B.
X-ray Tube Spectrum
The first attempts to propose a quantitative description of the tube intensity as a function of wavelength were made by Kulenkampff (1922) and Kramers (1923). The quantitative treatment of intensities of fluorescent x-rays (obtained by polychromatic excitation) by fundamental parameter programs requires knowledge of the intensity function of the exciting beam. This explains the continuous interest in the rigorous functional or numerical description of the tube spectrum. The description of the tube spectrum is divided into two parts, the first concerning the continuum spectrum [bremsstrahlung; see Seltzer and Berger (1985)] and the second concerning the characteristic lines of the anode material superimposed on the continuum. The relative importance of these two contributions depends on the target element: For a tungsten target, the characteristic L lines constitute the minority of the emitted radiation, about 24%, and the K lines from the copper tube represent about 60% of the emitted radiation at 45 kV. The most popular Rh x-ray tube derives about 4% of the emitted x-rays as RhK and 20% as RhL photons; the rest is emitted as bremsstrahlung (45 kV). The positive side of the latter tube is its relatively good output of the low-energy photons, with a possibility of the excitation of the lines of heavier elements by RhK lines. In the continuum research, the main effort was focused on the accuracy of Kramers’ law. Some deviations of this law from experimental results were corrected by the introduction of a nonintegral exponent m in the wavelength-dependent term in Kramers’ law (Brunetto and Riveros, 1984): Zr l Nl / 2 1 Kramers law l l0 m Zr l 1 Brunetto’s result Nl / 2 l l0
ð4aÞ ð4bÞ
where l0 is the wavelength at the short-wavelength (high energy) end of the continuum and r and m are constants related to the power-dependent terms: atomic number Z and energy function, respectively. The constants were determined by fitting the theoretical curves to the experimental results. Brunetto and Riveros’ result was confirmed later by Tertian and Broll (Tertian and Broll, 1984). By no means does it exhaust the solutions; Many other descriptions of bremsstrahlung spectra have also been published; to mention only the best known, there are those by Rao-Sahib and Wittry (1972), Reed and Ware (1973), Fiori et al. (1976), and Statham (1976a, 1976b). Finally, the reader can find the interesting articles by Small et al. (1987) and Trincavelli et al. (1998) summarizing the state of the art in this area. Other corrections must be added to Eqs. (4): the absorption correction [often that of Philibert; see Pella et al. (1985) and Markowicz and Van Grieken (1984)] and the exponential correction term including the absorption of the beryllium window. Furthermore, in the article by Pella et al. (1986), an interesting algorithm was proposed for taking the characteristic line intensities into account. For this purpose, these authors made use of the equation proposed by Green and Cosslett (1961). The ratio of peak to background intensity was found to be a function of overvoltage U0 and atomic number Z: Copyright © 2002 Marcel Dekker, Inc.
" 2 # Nchar U0 1 a U0 ln U0 1 ¼ exp 0:5 þ d Ncon U0 1 1:17U0 þ 3:2 b þ Z4
ð5Þ
Overvoltage U0 is the ratio of the initial electron energy E0 to the excitation energy of a given shell Eex . The symbols a, b, and d are experimentally determined constants. The intrinsic merit of this formula is its dependence on the physical parameter U0 and on Z [see also the work by Murata and Shibahara (1981) with Monte Carlo estimations of the penetration depth of electrons]. Recently, a series of papers on the numerical description of the tube output has been published by Ebel et al. (1989) and Schoßmann et al. (1995, 1997). The solutions were derived from the different depth distribution functions approximating the electron penetration in anode materials. Apart from all efforts concentrating on the numerical estimation of x-ray tube emission, experimentally recorded spectra may always be used [collected in Gilfrich and Birks (1968), Gilfrich et al. (1971), Gilfrich (1974), Brown et al. (1975), and Loomis and Keith (1976)]. The same kind of calculation as for the spectral x-ray tube output is very helpful for the description of the background in wavelength-dispersive XRF analysis. The background is mainly determined by the scattered bremsstrahlung radiation from the tube. The theoretical estimation of the continuous radiation should always be helpful for solving background problems (Arai and Omote, 1988; Omote and Arai, 1989; Arai and Shoji, 1990; Arai, 1991). Sometimes, the knowledge of background can be applied for analytical aims (Kuczumow et al., 1992). In this case, a selected bremsstrahlung channel provides quantitative (but not qualitative!) information about the sample contents. A similar approach has been applied in an article by Kuczumow et al. (1995) for the extraction of the information included in the coherently scattered signal in the energydispersive mode on the capillary version of x-ray spectrometry. A similar extraction of the information from the Rayleigh signal in wavelength-dispersive XRF and from the selected bremsstrahlung channel in the electron microprobe is possible as well (Kuczumow et al., 1999). It concerns special samples (e.g., the biological ones) and the information obtained is connected with the description of the matrix. It brings additional knowledge about the density of matrix, which is sometimes of essential significance in data evaluation. The articles just mentioned described the analysis of tree rings and other similar periodic structures, where the variability of the density of the matrix was, by far, more important than the variability in the chemical composition.
C.
Collimators and Masks
The wavelength-dispersive mode of operation depends strongly on rigorous geometric constraints on goniometer construction and analyzing crystal. Ideally, only a parallel beam of fluorescence radiation should be diffracted and stray radiation should be absent. Stray radiation enhances the background. Diverging beams result in worsening of the spectral resolution. Converging beams act in the same direction, but the latter are often necessary for making microanalysis. If applied in such conditions, they demand a special numerical treatment to obtain analytical results with a given level of confidence. Useful versions of fundamental parameter calculations for converging beams are now available (Chang and Wittry, 1994). Copyright © 2002 Marcel Dekker, Inc.
Masks, which reduce the size of the fluorescent beam in front of the collimator, cut off x-rays emitted or scattered by an area larger than the sample area. Three positions for collimators in the optical path of the spectrometer are possible: (1) between the sample and the crystal, (2) between the crystal and the detector, and sometimes (3) an auxiliary position between two detectors working in tandem. Collimators are not the ideal devices because they deliver a divergent photon beam with the angle of divergence a given by a a ¼ arctan ð6Þ l where a is the spacing between the blades and l is the length of the blades. In addition to divergence, there is also a substantial loss of counts while passing the collimator; for example, for a radiation of about 10 keV passing by a typical 10060.25 mm Soller collimator, the transmission is close to 105. On the bonus side of the use of collimators, it must be said that the resolution is improved very substantially (see Fig. 16). Thus, one needs to find a compromise between better resolution and higher brightness. In the determination of light elements, for which the intensity is relatively low (because poor excitation and efficient attenuation) but relatively well-spaced spectral lines exist, the demand for intensity prevails over resolution and, whenever possible, a coarse collimator (or none at all) is usually sufficient. Further considerations on this topic are in Section III.D.2. related to multilayers. In all focusing spectrometers (with curved crystals), the use of collimators is superfluous and the role of collimators is taken over by pinholes or slits. D.
Dispersive Elements
Both the primary photon beam derived from the x-ray tube (or another source of radiation) and the secondary or fluorescence beam derived from the sample can be monochromatized. The monochromatization of the primary beam is not essential but can be made for more efficient excitation of the selected elements by photons of chosen wavelengths or for radical simplification of calculation procedures (the fundamental parameter calculation is then transformed to simple Lachance–Traill–Tertian correction). Monochromatization may be compared to the action of a narrow bandpass filter by which only a small band of the whole spectrum is transmitted. The width of the bandpass is related to the spectral resolution. A number of parameters characterizes this process.
Figure 16 Two spectra from the same sample demonstrate how an auxilliary collimator reduces line overlap when measuring heavy elements (a) without and (b) with auxillary collimator. (Courtesy of Phillips Analytical. From Phillips Materials.)
Copyright © 2002 Marcel Dekker, Inc.
The ratio DE=E, where DE denotes the energy width of transmitted radiation band and E the energy of photons to be transmitted (the mean energy in case of a wide band), is called the relative spectral resolution of the monochromator. DE is generally expressed as full width at half-maximum (FWHM) of the transmitted band (a peak in intensity–energy representation). The reciprocal of spectral resolution, E=DE, is called the resolving power. The same can be expressed in wavelength units as Dl=l by transformation through Eq. (3). Both expressions are numerically equivalent but of opposite sign. For the wavelengthdispersive spectrometer, the angular dispersion (Bragg’s dispersion) is important and is obtained by differentiation of Bragg’s law: dW n ¼ dl 2d cos W
ð7Þ
The monochromatization of the primary beam, if required, is made by the use of filters or secondary targets. The absorption edges of the elements present in the filter should be at lower wavelengths than the wavelength domain to be transmitted. Secondary targets emit their own characteristic radiation without a white background (except the scattered part of the primary beam) and with a population density of peaks depending on the atomic number of the secondary target. In both of these cases, the monochromatization is far from being perfect. A relatively simple application of double total reflection from the quartz plates was proposed with success for monochromatization in TXRF (Schwenke and Knoth, 1982). The criteria for the monochromatization of the primary beam differ substantially, depending on the application. They become very sharp in some methods of research [photoelectron spectroscopy, extended x-ray absorption fine structure (EXAFS)]. Then, the methods of monochromatization of the beam are different also, depending very much on new discoveries in x-ray and related optics. In general, the strong monochromatization of the beam is associated with a great intensity loss and can be made for the very efficient sources of x-rays as synchrotrons. The efforts for the virtual numerical monochromatization of the real exciting beam by the introduction of the so-called ‘‘equivalent wavelength’’ concept failed. A new equivalent wavelength is needed for each analytical situation (e.g., change in composition). The gain from the simplification of the spectrum is doubtful as compared with the great increase in the numerical effort. The monochromatization of the secondary beam is the key problem. Dispersion of radiation is needed for the specific detection of characteristic lines. There are, however, a number of analytical problems in which the requisite dispersion is low. In limited cases, it may be sufficient to subdivide the spectrum into two spectral windows, from which one is to be detected. In other cases, the cutting off of a relatively wide spectral window may be sufficient, as is done in the Philips MiniMate device, where only a sealed gas proportional counter is applied for the photon energy discrimination. In practice, this selection of wide spectral domains is easily performed by energy-dispersive systems and pulse-height selection, by filters or by a set of balanced filters. A spectrometer for general use over the widest range of elements, however, should be equipped with monochromators of sufficient overall resolution, as is obtained by crystal diffraction. 1. Crystals According to the orientation to the possible crystallographic planes of the crystal, different values for d are preferentially diffracting and, coupled with it, a different resolution results Copyright © 2002 Marcel Dekker, Inc.
for a given wavelength. A few values has been calculated for LiF in different orientations and these are listed in Table 2. From the dispersion, it can be expected that the Ka line of manganese can be separated from the chromium Kb by LiF (220), not by LiF (200). The values in Table 2 are calculated for ideal crystals and a parallel beam. For real crystals, the dispersion and resolving power are less favorable; some real values for a complete spectrometer are provided later. In Table 3, a list is presented for crystals currently available for spectrometers: lithium fluoride (LiF), silicon (Si), germanium (Ge), pyrolytic graphite (PG), indium antimonide (InSb), pentaerythritol (PE or PET), ethylenediamine-d-tartrate (EDDT), ammonium dihydrogen phosphate (ADP), thallium hydrogen phosphate (TlAP), and multilayers with their commercial designations. Important characteristics other than the interplanar distances codetermine the ultimate usefulness of an analyzing crystal: spectral resolution, mosaicity, reflectivity, stability, the thermal expansion coefficient, and the spectral range. a. Resolution The spectral resolution DE=E of the crystals oscillates about the value 102, with few exceptions. The data for the resolution of different crystals are included later in Figure 26a, a summarizing figure on resolution [see also Kuczumow and Helsen (1989)]. The spectral resolution for the crystals is, generally speaking, better than that of other dispersive devices, such as detectors or filters, but the difference with respect to the energy-dispersive spectrometers becomes less favorable for the short wavelengths. The progress in superconducting tunnel junctions detectors (Finkbeiner et al., 1991; De Korte, 1992; Le Grand et al., 1993) and microcalorimeter detectors (Lessyna et al., 1993; McCammon et al., 1993; Silver et al., 1996) during the last decade shows that these devices would have even better spectral resolution in the medium energy range (i.e., above 4–5 keV) than the crystals (see Fig. 26b). However, it is a question of compromise among price, necessity to work at liquid-helium temperatures, the very value of spectral resolution necessary for a given analysis, and the loss of intensity in crystals on one hand and the poor count rates allowed in the newest detectors on the other hand, which decides what kind of resolving device will be the future choice for commercial applications. b. Mosaicity and Reflectivity Real crystals have all kinds of imperfections, and mosaicity is one of them. Mosaicity refers to the existence of ‘‘blocks’’ within the crystal with sizes of the order of magnitude of 100 A˚, which have slightly different orientations and lead to widening the diffracted peaks. This happens to widely differing degress: topaz, EDDT, ADP, and gypsum exhibit low mosaicity, quartz and LiF a little higher mosaicity, and PET a significantly higher
Table 2 LiF: d Spacings for Different Orientations and Angular Dispersion Crystal
2d (A˚)
Dispersion
LiF LiF LiF LiF
1.652 1.800 2.848 4.027
— — 0.5154 0.2902
(422) (420) (220) (200)
Note: Calculated for W angles corresponding to MnKa (2.102 A˚) and n ¼ 1.
Copyright © 2002 Marcel Dekker, Inc.
Table 3
Currently Available Crystals
Crystal
2d (nm)
Element range
LiF(420)
0.180
Ni–U
LiF(220)a LiF(200)a
0.2848 0.4027
V–U K–U
Si(111) Ge(111)a,b
0.6271 0.6532
P, S, Cl P, S, Cl
PG(002)a InSb(111)a,b PET(002)b
0.6708 0.784 0.8742
P, S, Cl Si Al–Cl
EDDT(020)
0.8808
Al–Cl
ADP(101) TIAP(100) PbSt LTC
1.064 2.575 10.0 15.6
Mg O–Mg F, C Up to Be
PX-1a PX-3a PX-4a OVO 55 OVO 100 OVO 160 OVO H300 PX6
4.93 19.5 12.2 5.5 10.0 16.0 30.0
O–Mg B C Mg, Na, F C, O B, C Be, B
a
Remarks High resolution, special applications for heavy-element K lines High resolution, good intensity General purpose, wide range, best diffracted intensity Supresses even orders Supresses even orders, good for intermediate- and low-Z elements Good intensity, poor resolution Good intensity, low spectral contamination, soft Lower intensity than PET, problems with stability Still lower intensity Especially F, Na, Mg, poisonous Good resolution, low intensity Longest wavelengths known for semicrystals’ structures Low resolution, good intensity Low resolution, good intensity Low resolution, good intensity Low resolution, good intensity Low resolution, good intensity Low resolution, good intensity One of the largest 2d available up to now
Also available as curved in simultaneous instruments. Also available as transversely curved for sequential instruments.
b
mosaicity. A result is a slight angular widening of the diffracted peak on one hand, but on the other hand, the intensity of the peak is generally enhanced. Widening is smaller than widening due to other geometric factors of the whole spectrometer and the loss can be tolerated while compensated by the increase in intensity. Some mechanical treatments enhance the mosaicity, and LiF crystals are easily improved in that way. The angular intensity distribution of a diffracted peak, the rocking curve, can be determined by doublecrystal spectrometer. In Figure 17, the peak height of the represented curve obtained by a high-resolution spectrometer is determined by atomic scatter factors and the space group of the analyzing crystal, whereas the width at half-maximum is determined by the mosaicity of the analyzing crystal and the divergence of the incident and diffracted beams. The surface under the curve is the integral reflection or, when the curve is normalized to the intensity of the incident beam, the integral reflection coefficient. This coefficient is very much dependent on mosaicity. Abraded LiF has a 10-fold increase in the reflection coefficient compared to freshly cleaved LiF (36105 to 46104 rad), measured by using CuKa radiation (Birks, 1969). LiF crystals have also the great advantage of low absorption by their constituting atoms. Topaz and quartz, otherwise good crystals, have Copyright © 2002 Marcel Dekker, Inc.
Figure 17 Rocking curve that would be obtained from a real crystal-diffracting parallel monochromatic radiation (rocking curves are actually measured on double-crystal spectrometers): P ¼ the peak diffraction coefficient; R ¼ the integral reflection coefficient. (From Birks, 1969.)
poor reflectivity properties, whereas PET is very good in this respect. Reflection constants as a function of wavelength are given in Figure 18. c.
Stability and Temperature
The mechanical stability of most crystals is satisfactory, but there are exceptions. Gypsum can effloresce (especially in a high vacuum); PET has a tendency to change phases on aging
Figure18 Single-crystal integral reflection coefficients of graphite, LiF (200), LiF (220), and KAP. [From Gilfrich et al., 1971). Reprinted by permission of Analytical Chemistry. Copyright # American Chemical Society.]
Copyright © 2002 Marcel Dekker, Inc.
and it is soft so that it is easily damaged when manipulated. The temperature inside the measuring chamber of a spectrometer is kept as constant as necessary (< 1 or < 0.01 C if chemical peak shifts have to be measured). In these circumstances, damage of the organic crystals by high temperature is not likely to occur. When very precise measurements of peak position are to be made for speciation purposes, the thermal-expansion coefficient is important. PET is very disadvantageous in this respect; topaz is in the opposite situation. It should be noted that the characteristic spectral lines of the crystal can be superimposed on the spectrum, making it obscure or even unfit to use, especially when it suggests the presence of some elements in a false way. Crystals of excellent resolving features, Be and Ge (111), have their own lines in the soft x-ray region. d. Spectral Range The useful spectral range covered by one crystal is limited. The maximum wavelength is of trigonometric origin and is imposed by Bragg’s law, namely by the condition that sin W < 1 and lmax < 2d. In practice, the goniometer is scanned for values of 2W varying from a few degrees to about 150 . However, for higher values of 2W, an angular dispersion widens the peak profiles. On the contrary, in the low-2W range, only a small fraction of x-rays emitted by the sample is intercepted unless the analyzing crystal is very long. The surface of the crystal projected on a plane perpendicular to the x-ray beam may be smaller than the width of the beam emerging from the collimator. Thus, 2W values as well as the macroscopic dimension determine the useful spectral range of a crystal. For x-rays of low energies, the interplanar distances of real crystals become too small. Some substances, namely salts of heavy metals with organic acids with long chains (otherwise the soaps of heavy metals), can take over the role of dispersive structures if their organic chains are regularly arranged. There are special techniques for making such quasicrystals. The individual layers are called Langmuir–Blodgett films. The ends of hydrophobic chains of adjacent layers join each other, and the heavy metal ends are connected to the next heavy-atom ends from the next molecules. Many features of such films depend on the length of the chain. A great number of similar structures with a wide variety of interplanar distances 2d (up to 156 A˚ in the case of lead melissate) have been synthesized. Langmuir–Blodgett structures have different negative properties: for example, they are soft and not very stable, strongly hydrophobic, and of low reflectivity. Some of them, like thallium adipate, are highly poisonous substances. e. Curved Crystals and Other Focusing Systems Not only are flat crystals used in the goniometer. Some focusing systems, such as the Johann, Johansson, and Cauchois arrangements, may be used (Fig. 19). Then, the crystal must be curved, with the curvature radius equal to the diameter of the Rowland circle and, at the same time, to the Bragg curvature, being half that of the Rowland circle. Threedimensional crystals can sometimes meet both of these conditions. The focusing arrangements are extremely useful in all cases in which the sample area is small, losses of intensity are prohibitive, and scanning over large angles is not necessary. In these arrangements, x-rays are collected in one point by focusing on the detector. An obvious virtue of such an arrangement is found in the analysis of small sample areas and by a using focused exciting beam, as in electron or proton microprobes. In such situations, the curved crystal geometry allows the collection of x-rays from a relatively large solid angle of diverging x-rays to a small spot (Fig. 20). A consequence is that the collimators normally placed between the sample and the crystal, and the crystal and the detector must be replaced by slits and pinholes. The focusing geometry is a very economical arrangement for saving intensity (the brightness of such a device is estimated to be up to two orders of magnitude greater than Copyright © 2002 Marcel Dekker, Inc.
Figure19 Curved crystal x-ray arrangements in (a) Johann, (b) Cauchois (in transmission), and (c) Johansson version. S ¼ entrance slit of radiation; D ¼ exit slit of diffracted radiation; d ¼ crystal lattice spacing.
Figure 20
Focusing of x-rays by a logarithmically curved crystal. (Courtesy of Siemens AG.)
that of conventional flat geometry). The curved crystal geometry is applied in simultaneous instruments in which scanning is not used and an optimized arrangement is chosen for each wavelength (element) implemented. Then, logarithmically curved crystals are used. In a new version of the sequential instrument (Venus), Philips used the reverted channels taken from the simultaneous instrument. Crystals applied in that construction are curved as in a simultaneous device and their position is additionally adjusted by a screw to trap as much intensity as possible. In general, during the last two decades, great progress took place in the planning and construction of the optical elements based on curved reflecting or diffracting surfaces of crystals and multilayers; most of the developments were in the synchrotron version of x-ray spectroscopy (Ice and Sparks, 1984). The curved crystal version of the x-ray microprobe is not the only known solution. The progress in x-ray optics during last decades enabled the easy concentration of x-rays while being transmitted through a single capillary (Rindby, 1986; Carpenter 1989; Thiel et al., 1989) or even through a semilens based on the bundle of capillaries (Kumakhov and Komarov, 1990; Yiming et al., 1994; Dagabov et al., 1995; Xunliang et al., 1995; X-Ray Optical Systems Inc., 1997, 1998). This first possibility, due to the great loss of intensity resulting from a poor interception of the primary beam by the narrow inlet of a single capillary, is limited to the energy-dispersive version of x-ray spectrometry (Attaelmanan et al., 1994). The second version is more interesting from the point of view of WDXRF. A properly curved capillary bundle can serve as a kind of semifocusing lens with a well-defined Copyright © 2002 Marcel Dekker, Inc.
Figure 21 Cross section of a typical multicapillary ‘‘lens’’: (a) focusing configuration, from wide spot (divergent source) to another spot (semifocus); (b) configuration transforming the divergent into the parallel beam. (Courtesy of X-Ray Optical Systems Inc., Albany, NY. and personally from Dr. J. P. Bly.)
curvature and moderately-defined focus (Fig. 21). The intensity of the beam leaving the ‘‘lens’’ is, of course, lower than the intensity derived from the x-ray tube, but still sufficient for many applications. Moreover, the primary collimator is removed in this system and the gain from this removal is partially balancing the loss resulting from the bundle construction and operation. One can estimate the focal distance f of such a lens as df ¼ d0 þ 2 f ycr
ð8Þ
where d is the spot size, d0 is the diameter of the capillary end, and ycr is the critical angle of the total reflection. As a rule, the polycapillary array is introduced to concentrate the primary beam, but it can also be used for the collection of secondary x-rays outgoing from the sample, before reaching the curved crystals in nonconventional arrangements. 2. Multilayers The limitation of XRF analysis in the region of soft x-rays is one of the most serious limitations of crystal monochromators, although none of the ‘‘first principles’’ constituting the background of XRF method puts this limitation explicitly. Moseley’s law determines the rules for qualitative analysis of any element except hydrogen and helium, and the Shiraiwa and Fujino equations govern quantitative analysis. Irresolvable problems occurred until the late eighties because of troublesome absorption, lack of an adequate dispersive device, and bad detection while passing to longer wavelengths. The concept of multilayers or, strictly speaking, layered synthetic microstructures (LSMs) or multilayer interference mirrors (MIMs) solved at least part of the problems. Two approaches to the idea of a suitable analyzing device in the soft x-ray domain have been applied from the very beginning (Underwood and Barbee, 1981). The first resulted from consideration of Bragg’s equation [Eq. (2)]. This indicates that if any crystal of a longer interplanar distance is found, it will be possible to disperse longer wavelengths in the soft x-ray range. The potential candidates for long-period crystals or quasicrystals, such as clay minerals of the chlorite (14 A˚) or illite–montmorillonite (25–30 A˚) type, are not sufficiently ordered in this respect (Weaver and Pollard, 1973). Unfortunately, natural crystals, irrespective of their different crystallographic planes, offer only limited possibilities in this spectral range; specifically, their reflectivity is very low. The class of structures called Langmuir–Blodgett films was more promising, but the reflectivity and thermal, mechanical, and chemical stabilities were rather poor. Copyright © 2002 Marcel Dekker, Inc.
A correct solution resulted from generalization of Bragg’s law, namely that diffraction also occurs in media consisting of layers of different refraction coefficients—in other words, in sites of different periodically changing electron densities. Such structures may be formed in an artificial way. They may be treated as crystallike substances ordered along the c axis (i.e., spacing d or 2d), but with uncontrolled arrangement within the ab plane. The structure in the ab plane can be amorphous, provided that the optical and mechanical properties are uniform in these directions. The first effort of synthesis was made in 1940, but the Au–Cu multilayer of Du Mond and Youtz (1940) did not survive more than a few days because of deterioration by interdiffusion. The problems of thermal, radiative, diffusional, and chemical instabilities of multilayers have recently been solved. Significant progress in this field has taken place in the last three decades, hence some years before the application of multilayers in WDXRF. The second approach was the result of an increasing interest in x-ray optics. As a branch of physics, x-ray optics is nearly as old as the discovery by Ro¨ntgen, but because of the specific properties of x-rays, its progress has been slow. All existing substances exhibit complex refractive indices: m ¼ 1 d ib
ð9Þ
where m is the refraction index, d is the so-called unit decreament, a real number, and the imaginary term b is related to absorption. The refraction indices in the x-ray region are all below but very close to 1. This makes all efforts of x-ray focusing by lenses in a traditional sense an impractical task. Franks (1977) gives an example of a lens with a curvature radius of 10 mm built of a material with d ¼ 5 105 . The focal length of this lens is 100 m. In such a situation, only glancing, grazing, and diffracting features of materials have been of practical interest for a long time. Otherwise, the grazing optics conditions for x-rays are extremely severe. However, the need for mirrors, monochromators, and focusing devices for the x-ray region is increasing rapidly in the fields of synchrotron radiation, plasma research and diagnostics, x-ray microscopy, x-ray fluorescence analysis, diffraction, thermonuclear fusion, x-ray laser, x-ray astronomy, and even x-ray waveguides. The invention of multilayers defined this time as a complex set of particular layers consisting of materials of periodically different refractive indices gave great impetus to progress in this branch of optics. Some important properties of the materials used to produce the multilayers are discussed in the next section. a. Nature of Materials In principle, three kinds of material are necessary for the synthesis of multilayers. First, there must be a support (sometimes called a substrate), such as a piece of flat or curved silica layer, with an ideally polished surface. This condition is very important because the roughness of the surface is translated to the deposited layers, correctable almost exclusively when one of the layers consists of amorphous materials such as carbon. After years of progress in the field, when the roughness of the interfaces have probably reached the limit of about 2 A˚, the most important contribution in it is from the deviations of the substrate from flatness (Kortright, 1996). Two sources of relatively inexpensive and precisely polished supports were found: silicon wafers as used in microelectronics (curved) and glass or fused-silica optical elements for low-scatter mirrors (flat or curved). The additional polishing is the last step in the preparation of the substrate for further manufacturing of the multilayer. An even more complicated role for the substrate is described by Nicolosi et al. (1986) in whole coal analysis. The authors determined the contents of Copyright © 2002 Marcel Dekker, Inc.
C, N, O, and F using multilayered microstructure PX2 and the contents of P, S, K, and Ca by taking advantage of diffracting properties of the supporting wafer of silicon (111) on which the multilayer was settled. Heavy metals (Hf, Pb, W, Mo, Pt, Ir, Ru, Rh, Os, Ni, Co, Fe, and Cr) and light elements or compounds [Al, B, B4C (Ovonic Synthetic Materials Co., 1987), BN, Be, C, Si, SiO2, SiC, Ti and Sc—the last with the high reflection] may be used as materials for alternating high and low optical densities, at the same time being strong and weak scatterers, respectively. Sometimes, materials from the first group are called absorbers, and those belonging to the second group are called spacers. The reactive materials and substances with a low melting point are excluded from the list. Both components of the multilayer should not share a common crystalographic structure, to avoid the epitaxial growth. The selection rule for the materials can easily be understood from Figure 22 for boron. The mass attenuation coefficients for the boron K line, as presented versus the atomic number of the absorber, have several minima. One falls in the region of boron (weak self-absorption) and carbon, and the next very deep minimum is in the region of niobium=molybdenum. It puts the constraints on the materials, which can be used for the determination of boron. The first, a low-density material, can be carbon or even better B4C; the second, a high-density material, should be niobium or molybdenum. Indeed, Philips made the structure PX3, composed of Mo=B4C, with a period of 195 A˚, adviced for the determination of boron. However, when a new multilayer should be synthesized
Figure 22 Mass attenuation coefficient for the BKa line drawn against the possible absorbers in the periodic table. Arrows show the deep minima in coefficient value. (Courtesy of Bruno Vrebos’ inspiration.)
Copyright © 2002 Marcel Dekker, Inc.
with high reflectivity in mind, then the attention is put, instead, on a pair of materials with a small value of the scattering Debye–Wallner factor [see Eq. (10)]. Here, the W=B4C tandem is a very good one. As a rule, the low-electron-density compounds are selected from the elements with K absorption edges above the energy of spectral lines to be measured. In that sense, boron and carbon are selected for multilayers of spacing above 50 A˚ and beryllium and carbon for spacings above 120 A˚. b. Smoothness This parameter is of great importance on each level, starting from the supporting substrate to the last upper layer. All rough places, uneven layers, disturbances in interfaces, interdiffusion, and chemical interaction regions or defects, act in a manner analogous to the thermal motions in classical diffraction analysis. During synthesis, the extent of roughness can vary in a different manner: In some cases, it tends to smoothness, in others to further growth. The use of amorphous materials to form some layers favors smoothing. The greater interplanar distances are, the more pronounced the roughness is. On the other hand, the smoothness probably has its limits, even in the case of multilayers with relatively small interplanar distances, with the Debye– Wallner parameter s tending to reach a value of about 2 A˚. It probably results from the smoothness level obtained during the substrate production. Finally, it puts some domination of the Debye–Wallner parameter over other possible factors influencing the quality of multilayers in the region where the interplanar distance is not very long (say, a few tens of angstroms). c. Reflectivity and Thickness of a Multilayer Classical dispersive crystals exhibit low reflectivity in the soft x-ray range. The situation was greatly improved in multilayer reflectors in which a single layer at normal incidence angles exhibits a reflection coefficient of the order of 104 rad. This coefficient is, however, proportional to the square of the electrical field amplitude (i.e., to a value of the order of 102 for the amplitude reflection coefficient). Thus, a multilayer consisting of 100 single layers may provide almost the maximum reflectivity if the absorption can be neglected (Underwood and Atwood, 1984; Underwood, 1986, http:==www-cxro.lbl.gov==multilayer=survey.html, 1998). For long wavelengths, the results are not so impressive; for example, in the very important region of the so-called water window (284–530 eV), the reflection coefficient was lower than 3% (Kozhevnikov et al., 1994). The region is of essential significance for the construction of an x-ray microscope based on zone plates for biological research (Kirz et al., 1995). These results are in accordance with the calculations by Rosenbluth concerning numbers of layers necessary for ‘‘saturation’’ of reflectivity [data cited by Barbee (1986)]. Peak reflectivities of some commercially available multilayers [e.g., Philips materials PX1, PX2 and PX3 (Nicolosi et al., 1986; Van Eenbergen and Volbert, 1987) and the OVONYX line of multilayers [(Ovonic Synthetic Materials Co., 1987)] are often compared to classical pseudocrystal dispersive structures, like lead octodecanate and thallium acid phthalate. The predominance of the multilayer structures over classical structures is striking, and spectral resolution is the only parameter in which the classic crystals are better (Van Eenbergen and Volbert, 1987; Henke et al., 1986). Still, when using multilayers, one must remember the weak points inherently coupled with those devices: great refraction effects (Martins and Urch, 1992; Luck et al., 1992), deviating the effective 2d value of the structure from the real value by as much as 5%, and poorly shaped background with ‘‘ghost’’ lines in some cases. Copyright © 2002 Marcel Dekker, Inc.
d.
Quality of Layers and Boundaries
The structure of the multilayer ought to exhibit geometric features. This condition is very strict. Thus, flat layers or figured structures must be uniform, the latter with controlled curvature, without defects but with clearly defined optical parameters. All deviations from an ideal model, easy to estimate by the reflectivity characteristics, are, as a rule, coupled by the use of the Debye–Waller factor into the effective roughness parameter seff, including both thickness errors and imperfect boundaries (Spiller and Rosenbluth, 1986). The Debye–Waller factor DW is the Gaussian-type term describing the reduction of reflected amplitude at a boundary: " # 2ps cos a 2 DW ¼ exp 2 ð10Þ l where s is the width of a smooth transition layer and a is the propagation angle of electromagnetic radiation, measured from the normal to the boundary between a medium with one refractive index to another. The question arises, however, whether the errors result from a rapid or from a mild change of the optical features on the boundaries. Any roughness of a given layer decreases its reflectivity (Payne and Clemens, 1993), but a random distribution of roughness may bear higher reflectivities of the higher-order peaks. Moreover, many other physical factors can influence the uniformity and expected layer characteristics: Both the support and the particular monolayers can have different slopes, densities are subject to fluctuations, the elements can diffuse, or some chemical elements can mutually react (e.g., W with C or Mo with Si) (Rosen et al., 1993). It should be emphasized that despite of so many obstacles, the magnitude of the roughness in such a totally artificial structure can be kept in the range of an atom radius (Barbee, 1986). The numerical description of multilayers is based on the matrix method (a matrix expresses the features of a layer or interlayer region, and for the characteristics of a multilayer, the product of matrices is written) or on the recursive use of layer–reflection– transmission characteristics. Deviations of empirical results (mainly reflectivity measurements) from the model predictions are associated with roughness. In this case, seff, the roughness parameter, results from the best fit of empirical data to the model. The selection of an adequate method of synthesis and subsequent control of production and quality is a key question. Many methods were checked, and the aim of all these processes was the deposition in which one atom follows another. The physical vapor condensation, chemical vapor deposition, electrochemical covering, and sequential adsorption of films are the most promising methods. However, for the time being, the first method prevails. Two basic versions of physical vapor deposition are used—with thermal and sputter sources. In the thermal source technique, the covering materials are evaporated by electron beam or laser evaporation. The vapor source and samples are in a fixed position. The time of evaporation is a parameter controlled by shutters or by pulsed laser irradiation. An in situ monitoring system was used as a feedback system for controlling the shutter. Spiller (1981) invented a system for thermal deposition. He used a soft x-ray reflectometer (a source plus a detector), placed in an evaporation chamber to register the reflectivity of successive layers. The open loop is set according to the indications of maximum reflectivity for a given layer. It is possible by this method to obtain a multilayer with an area up to 25 cm2. The rates of deposition vary in the range of 2–100 A˚=s and must be carefully controlled because the successive layers must be commensurate. Sometimes, especially Copyright © 2002 Marcel Dekker, Inc.
when multilayers of large surface are produced, a special motion of the substrate in the evaporation chamber may be necessary to guarantee the same evaporation rate for all points of the surface. Other difficulties arise from this type of deposition: the high vacuum required in the evaporation chamber (< 102 Pa), radiation heating of the deposited film, and the problems with thermal sources (alloys may segregate during melting). The sputter source systems (Barbee, 1981) belong to another processing method. The glow discharge, magnetron, electron cyclotron, ion beam, and triode devices may serve as sputter sources. The plasma discharge is always the first step of the process. The plasma flux is directed by the electrical potential difference to the cathode surface, where the ions and atoms dislodge the maternal atoms. The secondary atoms and ions establish the proper flux (Sigmund, 1981), which passes to the substrate to deposit a layer on it. Samples are placed on a special rotatory table, appearing at fixed intervals under the secondary flux courses. The rotation speed, precisely controlled, is the regulating factor for a designated thickness of the layer. The purity of secondary fluxes and their energies also influence the quality of the multilayer. The main errors in this method result from the ion and atom inclusion into layer (there are high-energy tails in the fluxes) and from heating caused by flux energy deposition in thin layers. However, substrate rotation may overcome many of these defects. Sputter rates range from 1 to 20 A˚=s; the covered surfaces may be quite large. Today, the magnetron sputtering is the most common method, with the best results obtained as far as the reflectivity of the products is concerned. After presenting two basic methods of layer formation, some attention must be paid to methods of control. As mentioned earlier, reflectivity estimations are of fundamental importance in multilayer quality determination, both experimentally and in theory (Spiller, 1981). A comparison of reflectivity predicted and obtained in an experiment is instructive, allowing, at least, understanding the source of the roughness (Henke et al., 1986; Luck et al., 1992). Different peak, absolute, and integrated reflectivities were measured with the use of both laboratory x-ray sources and synchrotron facilities. These results confirmed the layered structures of the synthesized materials; as a rule, reflectivity was smaller than predicted, but the results show that the optical parameters estimated until now (Henke et al., 1982, 1988) are reasonably correct, with the shorter the wavelength concerned, the better. The methods just discussed only allow the deduction of an indirect image of the multilayer structure. Other methods give direct insight into the structure. Thus, the diffraction data may be very useful, but they require large sample volumes. Progress in capillary version of the diffractometer (Bilderback et al., 1994), allowing the observation of the samples as small as 56103 mm3, may bring a solution to the structural problems. Very small samples are needed for obtaining the structural information by electron microscopy, especially in combination with a special technique of sample preparation, socalled microcleavage (Lepeˆtre et al., 1986). Wedge-shaped strips of materials are placed in the electron beam in such a manner that the planes of particular layers are parallel to the beam direction. Both the transmission image and the diffraction pattern can be obtained. The in-depth and lateral variations, the roughness, and the visual image of the structure are in the field of observation. High-resolution electron microscopy can give even better results, and electron energy loss spectroscopy (EELS) may provide the complementary chemical information on the subsequent layers (Lepeˆtre et al., 1986). Requirements for the application of multilayers in XRF may be summarized as follows: They should satisfy Bragg’s law for the assumed wavelength domain within permissible goniometer angles, have a high reflectivity and a good spectral resolution, Copyright © 2002 Marcel Dekker, Inc.
suppress diffraction peaks of higher order, and do not absorb soft radiation too intensively. Flat or curved multilayers can be used. Dedicated designs are described in the literature for the application of curved multilayer optics (Van Eenbergen and Volbert, 1987; Gilfrich et al., 1982). The geometric requirements imposed on the curvature of multilayers for XRF are modest in comparison with their applications in other fields of x-ray optics. The increasing spectral distance between K lines of the neighboring elements when going to lower-Z elements may require a dedicated multilayer for each element. Sometimes, great differences in the mass attenuation coefficients for the lines of neighbouring elements of interest exist for a given material—attenuator. Then, one can construct a multilayer efficiently diffracting the radiation of one element while deeply suppressing it for other elements. The dedicated multilayers are the preferable choice for the simultaneous WDXRF instrument, with an optimized channel for each element. Multilayers can be also applied in a special version of WDXRF analysis, called grazing emission x-ray analysis (de Bokx and Urbach, 1995). Curved multilayers are also used in electron microprobes (Ovonic Synthetic Materials Co., 1987). From what has just been described, it may be concluded that in order to analyze a useful range of soft x-rays, at least three different multilayers have to be selected. Such combinations are currently available commercially (e.g., for example from Osmic, Philips, Siemens, Rigaku, and Shimadzu) (Ovonic Synthetic Materials Co., 1987; Van Eenbergen and Volbert, 1987; D. K. G. de Boer, personal communication, 1998): PX1 PX3 PX4 PX5 PX6
2d ¼ 49:3 A˚ 2d ¼ 195 A˚ 2d ¼ 122 A˚ 2d ¼ 112 A˚ 2d ¼ 300 A˚
oxygen to magnesium boron and possibly for beryllium (Nicolosi et al., 1987) carbon nitrogen beryllium
and from the Ovonic Synthetic Materials Company (1987): OV-040A
2d ¼ 40 A˚
fluorine to silicon
with multilayers changing 2d spacing every 20 A˚ up to OV-140B
2d ¼ 140 A˚
carbon to oxygen
and also the MoB4C structure: OV-H series
2d ¼ 244 A˚
beryllium to boron
up to OV-300H
2d ¼ 300 A˚
Probably, materials with even greater interplanar distances can also be supplied. However, recent scientific (except the XRF field) interest in multilayers is centered, instead, more on multilayers with periods of 10–20 A˚ for their excellent reflectivity or for their polarizing abilities (Scha¨fers et al., 1998). These materials are described in the current literature as completely thermally stable; thus, they do not need temperature stabilization when used in x-ray spectrometers (Nicolosi et al., 1986; Barbee, 1986; Van Eenbergen and Volbert, 1987). Some sensitivity to the damaging influence of intense radiation [above 0.15 J=cm2 (Kohler et al., 1985)], confirmed later in an article by Kortright et al. (1991) and by a comprehensive report by MacGowan et al. (1993), may limit the application of the multilayer for monochromatization of dense flux of x-rays, which usually happens in synchrotron, plasma, Copyright © 2002 Marcel Dekker, Inc.
or laser research. Perhaps, the more important conclusion from the latter study is that, after the impact of high power on multilayers, the amorphous carbon layer transforms partially into a graphite form. It improves the reflectivity but greatly deteriorates the stability and changes the interplanar distances. The phenomenon of ultrafast graphitization ( 40 ps) can, in the future, be applied for the construction of x-ray switches. These data can be compared with the results cited in a report by Barbee (1986). He claims that during the sputtering process of multilayer synthesis, very high temporary doses of energy can be deposited in the layers (power density ffi 105 W=cm3=s1), which involves a significant increase in local temperatures (50–180 C). The relative spectral resolution of dispersive devices is often expressed in the form of the important parameter DE=E or Dl=l, which is the basis for their choice in XRF applications. Unfortunately, this is the weakest point of the multilayer. In a commercially available specimen, the relative spectral resolution changes from about 0.025 to over 0.1 while passing from structures with 2d ¼ 40 A˚ to those with 2d ¼ 244 A˚, respectively, for different measured lines (Ovonic Synthetic Materials Co., 1987), and for the structure with 2d ¼ 75 A˚, it changes from 0.025 to 0.04 in the wavelength range 7–75 A˚ (Henke et al., 1988), which is about three times worse (Ovonic Synthetic Materials Co., 1987) or even more (Henke et al., 1986) than for a good classical pseudocrystal. The comparison of spectra of the same sample by using multilayers with different spacings, as in Figure 23, is very instructive (Van Eerbergen and Volbert, 1987). Relatively broad internal bandpasses of multilayers diminish the resolution of the whole spectrographic device, seriously increasing the detection limits. This is of special importance to the analysis of the L or M series, and avoids the efforts to search the chemical shifts in x-ray spectra [although see Habulibaz et al. (1996), where chemical shift is investigated for the SiL spectrum, diffracted on a multilayer with an interplanar distance 300 A˚]. The relatively wide bandpass of the multilayer can be used in a constructive way. As proven by Vrebos (personal communication, 1997; see Fig. 24), the imminent width of the BKa peak is so great that the contribution from the secondary beam collimation to the total width of the signal is negligible. In that situation, the analyst can remove the fine or medium collimation or
Figure 23 Comparison of sensitivity and resolution for layered synthetic microstructures with 2d values of 5, 12, and 16 nm from bottom to top, respectively. Increased sensitivity together with decreased resolution on the drop in 2d is evident. (From Van Eenbergen and Volbert, 1987.)
Copyright © 2002 Marcel Dekker, Inc.
Figure 24 Comparison of the spectra of the boron K line, obtained by a multilayer PX-3 with a fine, medium, and coarse collimator. (Courtesy of Bruno Vrebos and Philips Analytical.)
even the collimation completely. The gain is double. It results from the greater intensity given by the multilayer as such, as well as from the lack of the collimator, which is always one of the most intensity-restrictive parts of the WD spectrometer. This kind of work with a limited collimation is possible only in the region of soft x-rays, where the population of lines is rare and in the absence of the very soft L and M lines of heavier elements. The suppression of higher-order diffraction peaks gives better spectral purity and may at least partially compensate for the poorer resolution of multilayers. The isolation of the weak MgKa line from the dominant CaKa line is possible by the use of PX1 structure for the analysis of cement (Fig. 25). This is an extreme example of the superiority of multilayers over a classical crystal in particular cases (Nicolosi et al., 1986). Another great advantage of multilayers is their great reflectivity. It should be added that the peak reflectivity is 3–10 times greater than that of corresponding classical crystals. Multilayers were frequently used in the routine determination of light elements. Many examples can be cited, including carbon (Nicolosi et al., 1986, 1987; Van Eenbergen and Volbert, 1987; Philips Application Note No. 745) and boron determination (Van Eenbergen and Volbert, 1987; Nicolosi et al., 1987; Adamson et al., 1991; Philips Application Note No. 737 and No. 803; Uhlig and Mu¨ller, 1991; van Sprang and Bekkers, 1998), and beryllium detection (Nicolosi et al., 1987; Siemens Lab Report X-ray Analysis, 1994; Anonymous article, The Rigaku J., 1997). In the analysis, advocated by Rigaku, the application of the high-power 4-kW generator was the additional factor, enhancing the detection power for beryllium determination. The detection limit for that element was found to be in the region of a few tenths of a percent. Determinations of light elements were made using both flat optics sequential spectrometers and curved optics in simultaneous instruments (Van Eenbergen and Volbert, 1987). After overcoming the technical difficulties, the results obtained from such analyses may seem quite trivial (Nicolosi et al., 1987; Bonvin et al., 1995). However, not always are things going in that usual way. Van Sprang and Bekkers, (1998) emphasized the number of problems, coupled with light-eleCopyright © 2002 Marcel Dekker, Inc.
Figure 25
2W scans for Mg in cement using TlAP and PX1. (From Nicolosi et al., 1986.)
ment determination. For the analysis of boron in glass, only the combination of methods gave good results in a whole range of the detectable concentrations: The fundamental parameter method using the fluorescent signals for the low B concentrations and the use of the incoherently scattered radiation as a measure of the light-element fraction above 4%. One should note some special features of x-ray spectrometry in the soft x-ray range: The type of analysis changes from bulk to surface measurements; the influence of the chemical state and the subsequent wavelength shifts on the analysis can be significant; the wavelength of soft x-rays is comparable to the surface roughness; and the problem of the surface quality becomes extremely important. Considering the so-called total information depth of x-rays [see Sec.IV.G.1, Eq. (19)], it can be calculated that in the case of iron determination in a steel sample, the effective energy of exciting photons being 20 keV, the secondary x-rays arriving at the detector emerge from depths up to 62 mm. A carbon signal, emitted by a coal sample excited in the same manner, emerges from depths of only 8.3 mm. The problem looks even more dramatic if carbon is determined in a steel sample: In this case, signals emerge from depths as small as 0.45 mm! This means that the analytical volume for iron is about 140 times greater than that for carbon and the information on both constituents really refers to totally differing volumes of the steel sample. Moreover, the method of excitation is also critical in this case: The excitation of light elements is most efficient if made with relatively soft x-rays, the energy of which does not substantially exceed the value of the absorption edge. Excitation by use of hard radiation is not profitable in this case, as photons are mainly stopped inside the sample. The most restricting stage is the very short escape path of the characteristic photons for low-Z elements. The effect of the chemical state can be observed during the analysis of some elements using soft x-rays. The wavelengths of La1,2 lines of chromium and manganese are 0.573 and 0.637 A˚, respectively, and should be analyzed in the region otherwise reserved for the Copyright © 2002 Marcel Dekker, Inc.
K series of oxygen and fluorine, hence by the use of the PX-1, PX-4 or OV-120B pseudocrystals. The question is whether the limited spectral resolution of multilayers is sufficient to show such subtle effects as chemical shifts and to separate the mentioned lines of metals from the disturbing lines of oxygen or fluorine if present in the sample. The wavelengths of characteristic x-rays in the K series are longer than atomic radii starting from Ga (GaKa ¼ 1.337 A˚; atomic radius of Ga ¼ 1.26 A˚) and passing to the lighter elements. Radiation with a wavelength comparable to the dimensions of an obstacle always result in surface problems, with scattering and diffraction at the head. One should take into account that even the most perfectly polished sample has a surface roughness of the order of at least one atomic radius and that the best layers in multilayer structures have a roughness, expressed as the effective roughness seff, on the order of one to several atomic radii (Barbee, 1986; Kortright, 1996). These latter remarks are not intended to weaken interest in multilayer optics and its application in XRF, but only to evoke a special awareness of potential problems. 3. Spectral Resolution Figure 26a shows the features of commonly used dispersive devise (Heinrich, 1981; Caciuffo et al., 1987; Plotnikow and Pszenicznyj, 1973; Fitzgerald and Gantzel, 1971; Burkhalter and Campbell, 1967; Gohshi et al., 1982a, 1982b; Potts et al., 1985; Sparks, 1980; Salem and Lee, 1976; Bent, 1970; Bandas et al., 1978; Gilfrich, 1987). The width of the bandpass DE (expressed as FWHM whenever possible) is plotted versus the energy of radiation E; the parameter is relative to the spectral resolution DE=E. A line of given value of DE=E is called the isoresolving line. These lines divide the whole figure into two zones of different resolutions (or bandpasses). Note that both the typical dispersing devices and typical detectors are collected in Figure 26a. Only a few detecting devices with no resolving power (the reciprocal of the relative spectral resolution) at all exist (e.g., Geiger-Mu¨ller counter), and no detector exists with perfect, infinite resolving power. Real dispersive devices exhibit very contrasting features, with the parameter DE=E varying from almost 1 (scintillator counter or semiconductor photocathode) to systems with DE=E as small as 104 [Ge (111) crystals and two- or three-crystal spectrometers]. Even better dispersive devices are needed, for example, for x-ray inelastic scattering spectroscopy in which relative energy resolution near 107 is wanted. Indeed, we can cite here some exemplary articles announcing the achievement of similar values: 1:14 107 at a level of 14:4 keV (Chumakov et al., 1996) or 2 108 for 25:7 keV (Verbeni et al., 1996). XRF spectrometry has rather modest demands for resolution. We see that for introducing the characteristics of the devices with spectral resolution on a level of 108 in our diagram, the scale of Figure 26a should be prolonged by approximately four orders of magnitude at the bottom! Another interesting conclusion can be drawn from Figure 26a. Nearly all dispersive structures exhibit variable relative spectral resolution, but only KAP, topaz, and Ge (111) crystals and, to some extent, also multilayer structures can be considered as isoresolving systems. The microcalorimeters show rather unusual behavior, compared with other detectors. They have more or less the same absolute spectral resolution in the whole useful range (2–20 keV). As we can see from the shape of the DK and DL curves, nature does not demand isoresolving behavior from the dispersive structures. Rather, these structures, with resolution curves parallel to the curves mentioned, would be optimal for spectrometric aims: The Si–Li detector seems to have the better features. Generally speaking, a bandpass range of 101–103 is considered analytically useful. Copyright © 2002 Marcel Dekker, Inc.
Figure 26 (a) Spectral resolution of different devices: DZ ¼ spectral difference between analogous lines of adjacent elements; DW ¼ width of spectral line; Ka1 7 Ka2 ¼ difference between Ka1 and Ka2 lines; (b) the imminent features of x-ray spectra extracted from (a) for better characterization of the analytical demands; (c) characteristics of new microcalorimeter detectors as compared with typical Si–Li and LiF (200) dispersive structures.
Systems with bandpasses below 103 are used only in spectrophysical work. For making a proper choice of the dispersive structure for study of chemical shifts, one has to take into account the curves DW and Ka1 7 Ka2 (or even La1 7 La2) from Figure 26b. Figure 26b shows the spectral characteristics, extracted from Figure 26a, as a better demonstration of the immanent features of x-ray spectra. Depending on what matters in analysis, the adequate choice of the resolving system should be made. The regions occupied in the diagram by the most common [SiLi and crystal (e.g., LiF)] or most promising (microcalorimeter) detectors are shown in Figure 26c for easier comparison with data from Figure 26b. Figure 27 represents some example of the parallel analysis of the same sample, based on spectra from the electron microprobe acquired by the WDXRF system (TLAP), EDXRF system with Si(Li), and EDXRF system with the Copyright © 2002 Marcel Dekker, Inc.
Figure 26
Continued
microcalorimeter. For the time being, some preponderance of the LiF detector is observed for light element determinations (below 4 keV) (Wollman et al., 1997), but even this was endangered when Wollman announced the energy resolution of 2.0þ=0.1 eV at 1.5 keV level (Wollman et al., 2000). Spectral resolution is only one parameter determining the selection of a spectrometer. Other parameters, however, such as speed of measurement, count rate capability, geometrical collection efficiency, and ease of operation may affect the choice of the most suitable instrument. E. The Goniometer The goniometer is basically a very simple construction and is the analog of a diffractometer as used in x-ray diffraction (the crystallographic research preceded the x-ray fluorescence method). The essential features are shown in Figure 7. The sample is positioned in front of the tube window. To account for possible heterogeneities, the sample is Copyright © 2002 Marcel Dekker, Inc.
Figure 26
Continued
spun about its axis at 30 rotations per minute. The analyzer crystal is positioned in the center of the diffractometer circle, rotates around an axis perpendicular to the plane of the drawing, and passes through the macroscopic plane of the crystal. The detector moves along the diffractometer circle supported by the goniometer arm. As imposed by the Bragg condition, the crystal rotates over an angle 2W, realizing equal angles of incidence and diffraction with respect to the crystal plane. The crystal is diffracting the incident fluorescent beam of the opening angle determined by the slit width and length of the collimator. Rotation of crystal and detector are either mechanically coupled through gears with a ratio of 1 : 2 or mechanically decoupled but moved by separate computer-controlled stepping motors, but both arrangements are still implemented (cf. Tables 8–10 in Sec. VI). The idea of mechanical decoupling and optical position control was first introduced by Applied Research Laboratories (see Sec. VI). This technique offers a number of advantages: Alignment of crystals is no longer necessary. They can be positioned on a known spectral line and the offsets are memorized by the computer. Copyright © 2002 Marcel Dekker, Inc.
Figure 27 (a) NIST microcalorimeter spectrum of AlGaAs (solid line) compared to parallelly determined EDS spectrum [with commercial Si(Li) detector]; (b) WDS spectrum of the same sample (with TLAP applied as a crystal), spectrum transformed from wavelength into energy representation. Observe still a little better performance of the WDS spectrometer for the considered energy region 1–1.6 keV. (From Wollman et al., 1997.)
The detectors can be juxtaposed on the goniometer arm and brought into position as required by the program. Tandem arrangement of flow and scintillation is no longer necessary. Slewing and scanning speed range is virtually unlimited. Optical position control (Moire´ fringes or optical encoder) allows very high angular precision and accuracy. Low-energy radiation is strongly attenuated by air. The Ka radiation of a pure copper sample (sample 2’’, measured on a Philips 1410 sequential spectrometer, SWT with chromium anode, 45 kV and 60 mA) produces 67800, 33200, and 2660 cps, respectively, in vacuum and 48000, 16800, and 890 cps, respectively, in air. The results mentioned are not easily repeated on recent computer-monitored instruments (e.g., Philips 2400) which do not routinely allow working under air at normal pressure. To decrease the absorption of photons, all wide-range spectrometers operate in a controlled atmosphere (vacuum or helium). The vacuum cabinet contains the entire spectrometer, except for some instruments such as the scintillator detector. Attenuation of the high-energy radiation is low, and to reduce the size of the vacuum chamber, the scintillation detector was and still is placed by some producers outside the chamber. A long rectangular window of a mechanically strong low-density polymer (Mylar) along the diffractometer circle allows the high-energy photons to reach the detector. High angular precision and accuracy is required for positioning the goniometer arm; therefore, severe mechanical constraints are imposed on its construction. Typical is an Copyright © 2002 Marcel Dekker, Inc.
angular reproducibility of 0.001 or less and a mechanical resolution of a few thousands of a degree or less (cf. Sec. VI). In most modern designs, 2W and W axes are decoupled. Scanning (step size can be software selected) or measuring at preprogrammed positions (e.g., element line, background left and right of the line for many elements) is possible. Typical angular scan speeds (2W) are in the range from 0.1 to 180 =min and slewing speeds can be up to 4800 =min. The slewing speed is the maximum rotation speed of the goniometer arm for passing from one position to another without recording. The 2W range covered is from 4 to 152 , with slight differences between manufacturers. Because of the high slewing speeds and full automation, the analysis time is severely reduced compared to the older intsruments and, when not too many elements have to measured, is often not prohibitively long. Jenkins et al. (1984) and Croke and Nicolosi (1987) announced a dual-channel sequential spectrometer. A twin primary collimator was installed consisting of an upper, fine collimator and a lower, coarse collimator. The beam is ‘‘divided’’ into two parts, each striking its own crystal. Two crystals are mounted close together but with a 15 inclination difference. The diffracted beams are detected by two detectors 30 apart. This fixed angle between both detectors, however, limits the benefits of a dual-channel instrument. Two alternative constructions should be discussed here. Simultaneous or multichannel instruments are equipped with monochromators set by the manufacturer (or by the customer) at a fixed position for the detection of a single element. Usually, curved crystal optics are used, and the whole setup is optimized for that specific element with an appropriate curved crystal, entrance and exit slits, and detector. The crystals are logarithmically or logarithmically and cylindrically curved as explained in Sec. III.D.1.e, see also Figure 20. To give some more flexibility to multichannel instruments, one or two channels can be equipped with a scanner. The optics are either focusing (e.g., Bruker) or nonfocusing (e.g., Philips). The flat crystal arrangement is principally a reduced size of the normal monochromator. The focusing arrangement is complicated by the fact that crystal and detector must be moved to focus the diffracted beam on the detector at any position. The scan range of 2W for the Bruker scanner is 30 –120 and 10 –100 for the Philips scanner. Full details of typical construction differences are given in Sec. VI on instrumentation. Two hybrid, difficult-to-classify spectrometers are discussed in this section: MDX1000 (Oxford) and the Venus 100 (Philips). The so-called ‘‘flexi-channel’’ of MDX1000 is an energy-dispersive spectrometer. The Venus 100 has only fixed channels, but they are sequentially presented to the fluorescent beam. Moreover, manufacturers are focusing more and more on dedicated hyphenated instruments combining x-ray fluorescence with diffraction measurements, such is an ARL 8600S and Philips CubiX. F.
Detectors
The objective of a detector is the transformation of photon energy into an electrical pulse. Pulses are counted over a period of time, and the count rate, expressed in counts per second or any other unit of time, is a measure of the intensity of the detected x-ray beam. From the theoretical point of view, the interaction of photons with matter is explained by the particle character of electromagnetic radiation, not by its wave nature. Four main classes of detectors may be distinguished for use in x-ray spectrometry according to the medium responsible for the energy transformation: gas detectors, scintillation detectors, semiconductor detectors, and recently introduced detectors supported on superconductive elements (tunnel junctions and microcalorimeters). They differ in the efficiency of detecting photons of a given energy, determining the range of energy or wavelength for which they Copyright © 2002 Marcel Dekker, Inc.
are suited. The width of distribution of pulse amplitudes as a result of impact of photons of the same energy determines the spectral resolution. Until recently, the best performance in energy discrimination has been obtained with semiconductor detectors and the full width at half-maximum (FWHM) is for the typical XRF domain in the range of 130 eV. This allows electronic pulse-height selection discriminating the energy of the impacting photons (at the expense of count rate, a maximum of 30 kcps). In WD spectrometers, pulse-height discrimination is required for removal of noise of low amplitude and of higher-order peaks of other elements interfering with the analytical line. It is a disadvantage typical for diffraction spectrometers. High-resolution detectors are not required in WD spectrometers because the interfering photons differ so substantially in energy that moderate discrimination is sufficient (a few hundred electron volts). High-resolution detectors constitute the heart of energy-dispersive devices and are discussed extensively in Chapter 3. Now, it is obvious that new types of device, superconductive junction detectors and microcalorimeters, exhibit energy discrimination an order of magnitude better. Of course, in the future they might be the obvious choice in applications demanding better spectral resolution, but the availability of the superconductive detectors is still a problem, as well as the maximum count rates they can manage and their routine maintenance and price. Only gas-filled and scintillation detectors will be discussed here as the devices supplementing crystal dispersive operation. Some attention will be paid also to the microcalorimeter counters, because the stormy period in their development and because of the prospects for the construction of the future generation of XRF (see Sec. VII). 1. Gas-Filled Detectors In this type of detector, the energy-exchange process occurs between photons and gas atoms or molecules in a strong electric field. When a photon strikes an atom, there is a given probability that the quantum of energy hn of the photon is imparted completely to an orbital electron of the atom. As a result of this gain in energy, the electron emerges from the atom with a kinetic energy of Ekin ¼ hn W (Einstein’s photoelectric equation). W is the work function of the electron for leaving its orbital and is a specific constant of the atom. In the case of argon, hn
Ar ! Arþ þ e The electron imparts its kinetic energy to other atoms, creating a series of electron–cation pairs on its path through the gas in a number equal to Ekin divided by the energy needed to expel an electron. Values for different gaseous and solid detector media are included in
Table 4
Effective Ionization Potentials (eV)
He Ne Ar Kr Xe NaI (Tl) Si (Li) Ge
27.8 27.4 26.4 22.8 20.8 50 3.8 3
Copyright © 2002 Marcel Dekker, Inc.
Table 4. Subsequently, the cations and electrons are accelerated toward the cathode and anode of the detector by the gradient of the electrical field between both electrodes, forming the avalanches of new ion pairs on their way to the electrodes. The discharge at the electrodes gives rise to a current pulse in the external electrical circuit. For a range of electrode potentials, the current pulse is proportional to the energy of the photon, and in spectrometers, the gas-filled detectors are exclusively used in this range. This is why these detectors are called proportional counters. Above this potential, in the Geiger–Mu¨ller plateau (Fig. 28), any photon entering the detector gives rise to a pulse of maximum amplitude without relation to the initial photon energy. The ratio of the number of ion pairs discharged at the electrodes to the ion pairs originally formed is called the amplification factor, generally denoted A. It is a complex exponential function of ionization potential of the gas, anode potential, radii of detector chamber and anode wire, and the mean free path of the electron. It implies that stability of applied potential and gas pressure is very important for reducing statistical errors on pulse amplitude. The basic parts of a gas-filled detector are a metal rectangular housing as cathode, a thin wire passing through the central axis as an anode isolated from the cathode, a filling gas, and a window either on one side or on two opposite sides of the housing. This gas flows through or is enclosed in the housing (Fig. 29). The diameter of the anode wire should be small. Realistic diameters are in the range of 40–80 mm. a. Flow-Proportional Counters These detectors cover the widest wavelength range and are used in sequential spectrometers exclusively for long wavelength ( > 2 A˚) and also in simultaneous instruments. The windows are made of thin foils of Mylar or polyester of 0.6, 1.5, 2 or 6 mm or polypropylene 1, 2, or 6 mm thick. On the inner side, windows are coated with gold or aluminum to create a homogeneous electrical field inside the detector. Because of these very thin windows, there is a leak of the filling gas that is compensated by a
Figure 28
Region of proportionality for gas detectors.
Copyright © 2002 Marcel Dekker, Inc.
Figure 29
Construction of proportional counters: (a) flow counter; (b) sealed counter.
constant flow of gas through the counter (Fig. 29a). This is called a flow-proportional counter. The gas is Ar with CH4 at a concentration of 10%, and the flow is about 0.5 dm3=h. For detectors in simultaneous instruments, helium is also used, or 88% He þ 12% CO2. For very specific applications, other combinations of gases may be useful. The gas is stabilized for pressure, flow, and temperature. This is of significance because a 1% change in gas pressure or temperature introduces a much more serious 6–7% change in pulse amplitude, which can discredit the results of the analysis (Short, 1991). The amplification factor in the proportional range is between 102 and 106. The current pulse is conducted over a resistor by which it is converted into a voltage pulse that is amplified, shaped, possibly discriminated in height, and counted. The count rate is a maximum of 26106 cps. The operating voltages range from 1000 to 3000 V. b. Sealed Proportional Detector Basically, the construction is the same as in the flow-proportional detector but the windows are thicker and no constant flow compensation for a leak is needed (Fig. 29b). The gases used are Ne, Kr, and Xe with Al, Be, mica, Mylar, or polypropylene windows. According to the element for which the radiation is to be detected, the most suitable combination of crystal, gas, and windows is chosen; for example, a PET crystal with a Ne detector with Al (for the determination of magnesium!) or a LiF (100) crystal with an Ar or Kr detector and a Be window for the detection many of medium-Z elements. If the filament is 80 mm thick, it is stiff and easy to install, has a long lifetime, and is easily exchanged. Because the proportional counter is often used in tandem with a scintillation counter, two windows are present on both sides of the main body. The absorption Copyright © 2002 Marcel Dekker, Inc.
probability of photons decreases with photon energy; the high-energy photons pass to the scintillation counter through the proportional counter without being essentially absorbed. The sealed detectors are applied in simultaneous instruments on their own and are not used in tandem with a scintillation counter. There is no need for a tandem arrangement because each channel is optimized for one element with the best-suited combination of detector and crystal. c. Pulse-Height Distribution In the detector, a photon generates a number of ion pairs proportional to the pulse amplitude. The formation of ion pairs is subject to statistical fluctuations, which means that the number of pairs formed oscillates around the most probable value. A record of pulses and their amplitudes as they emerge from the detector as a function of time is represented in Figure 30. When these pulses are collected according to amplitude, a distribution emerges as shown on the right side of Figure 30. The width of the distribution is measured at half-maximum and is designated FWHM. This determines the energy resolution of the detector. It is small for semiconductor detectors, such as Si(Li) (about 130–150 eV for the 5.9-keV MnKa lines from a standard 55Fe source; see Chapter 3), about 900 eV for gas proportional counters, and 3500 eV for a scintillation counter, all measured for the same photon energy. Note that this distribution differs from a distribution in intensity for a normal spectral line (intensity versus wavelength) because of natural and instrumental line broadening. d. Escape Peak As mentioned in the beginning of this section, a very probable method of energy exchange between a photon and an atom of the counter gas is the removal of an outer-shell electron. The probability of removal of inner-shell electrons is not zero, however, and this phenomenon happens in a number of interactions, in which case the specific K or L radiation is emitted. Because the wavelength of this radiation is on the long-wavelength side of the absorption edge, the mass attenuation coefficient is low and this characteristic photon has a good chance of escaping from the detector. The remainder of the photon energy is transferred to the electron of which the kinetic energy is generating a number of ion pairs but necessarily in a smaller number as an electron having received the whole photon energy. Thus, a number of pulses is created, distributed around an amplitude proportional to the energy of this electron. The absorption edge for ArK is 3203 eV; the energy of the FeKa line is 6403 eV. Pulses centered around 6403 7 3203 ¼ 3200 eV are generated, and
Figure 30 Amplitude–time record of impulses from the detector (left) and transformation of it into a pulse-height distribution (right).
Copyright © 2002 Marcel Dekker, Inc.
about 3200=26 ¼ 123 ion pairs are formed, 26 eV being the energy needed per creation of electron–ion pair. These pulses are present in the pulse-amplitude distribution as a second maximum on the low-energy side of the iron peak position. For Ne, Kr, or Xe, photon energies greater than 0.87, 14.32, or 35.58 keV, respectively, are needed to excite the K radiation. For lower energies, only L radiation is excited, resulting in an escape peak closer to the main pulse distribution. Due to the constant distance of escape peaks in relation to their ‘‘parent’’ fluorescence peaks and high intensity in some cases, they can be used sometimes instead of the originals. e.
Pulse-Height Discrimination
The energy resolution of a proportional counter, a few hundred electron volts at least, is not sufficient for use in energy-dispersive spectrometers (but see the simplified MiniMate construction by Philips). The spectral line separation in wavelength-dispersive spectrometers is done by the crystal monochromator. As is obvious from Bragg’s law, however, higher-order lines of other elements and the continuum radiation of the tube may be present in the same spectral window (angular position) of the crystal monochromator. In spectrometry in the ultraviolet or visible region, the higher-order diffractions of the gratings are removed by filters. This is not easily done with x-rays. For the same angle 2W, a second-order line has a substantially different wavelength or energy and is far enough apart to give two distinct pulse-amplitude distributions at the exit of the counter. For x-rays, electronic pulse-height selection or discrimination fulfills the role of filters in the optical region. The pulse-height analysis constitutes a necessary attribute to the wavelength-dispersive systems. Within an interval of 0.1 around 2W ¼ 37 for LiF (200) as analyzer crystal, we find AuLa1 in first order, E ¼ 1:2764 keV, ThLg8 in second order E ¼ 0:6390 keV, SnKb2 in third order, E ¼ 0:4259 keV, and LaKb2 in fourth order, E ¼ 0:3201 keV. The energy difference with the second-order line is 0.6374 keV and the energy resolution of a proportional counter is about 0.3 keV. The electronic circuits of all commercial instruments are able to discriminate between the pulses of both lines and, consequently, also the other higher-order lines. 2. Scintillation Detectors The energy exchange occurs in this type of detector in a medium of higher density and with high-Z elements in the matrix, namely a thallium-doped sodium iodide [NaI(Tl)], and as can be expected, this detector is only efficient for high-energy photons (< 2 A˚ or > 6 keV). The outer orbital electron for an iodide ion requires about 30 eV to be knocked out. Hence, the originally ejected electron is imparted with almost all the initial energy of the photon. The ejected electron dissipates its energy by promoting valence-band electrons to an excited state 3 eV above ground level, an energy emitted on deexcitation as a photon of 3 eV or l ¼ 410 nm. The intensity of the emitted light pulse is proportional to the number of electrons excited by the x-ray photon (ffi hv=3 eV). The pulses are detected by a photomultiplier in which the light pulse produces a few photoelectrons from the cathode material (e.g., indium antimonide). These electrons are accelerated inside the vacuum tube between the cathode and the first intermediary anode, called a dynode (Fig. 31). They gain kinetic energy and generate, in turn, a higher number of electrons from the dynode. This process is most often repeated 10 times, resulting in a substantial multiplication of the original number of electrons. The multiplication depends on the potential differences between the successive dynodes (100–150 V), and the order of magnitude of this multiplication is 106, which is called the amplification factor A. In general, the photomultiplier is a front-end Copyright © 2002 Marcel Dekker, Inc.
Figure 31
Layout of a scintillation counter.
tube, and the iodide crystal is in close contact with the tube window. The crystal itself has on its outer faces a reflective coating (but transparent for the photons above 6 keV) and a coating against moisture. In the external electronic circuit, the electrons provoke a current pulse of which the amplitude is ultimately proportional to the energy of the photon captured in the crystal. The conversion of the x-ray photons into a current pulse occurs over different steps: the initial formation of a photoelectron (I ! I0 þ e ), the quantified dissipation of the kinetic energy of this electron in 3-eV steps, emission of 410-nm light photons, the production of photoelectrons from the photocathode of the photomultiplier, and the multiplication of electrons. Although the production of the number of light photons in the crystal (hn=3 eV) is higher than the number of ion pairs created in a flow counter (hv=30 eV), the subsequent conversion into electrons is inefficient and roughly 1 electron is produced for every 10 light photons! The statistical errors of each process, particularly in the photomultiplier tube, propagate into the error of the output current pulse, thus substantially widening the distribution of pulse amplitudes and, consequently, reducing the spectral resolution. The DE is approximately two to three times worse than for a flow-proportional counter (see Fig. 26a), making a scintillation counter unsuitable for ED spectrometry. However, the resolution is still sufficient for a discrimination of higher-order spectra. Hardware for pulse amplifying, shaping, and counting are much the same as for the proportional counters. Also, an escape peak is present, with its energy equal to the initial photon energy minus the energy at the iodine K absorption edge. As already discussed, in many sequential spectrometers, flow-proportional and scintillation detectors are used in tandem, coupled in different ways. In another arrangement, both detectors are shifted over 30 . A secondary collimator is placed in front of the scintillation detector but inside the vacuum chamber; the scintillation counter is outside the chamber, but close to the exit window. According to the manufacturer (Rigaku), this allows a gain in count rate of 15%. 3. Alternative Con¢gurations Some interesting modifications of the detector have been proposed in the article by Ebel et al. (1983). These authors considered the possibility of reducing count losses between Copyright © 2002 Marcel Dekker, Inc.
analyzer crystal and detector (see Fig. 32) by removing the secondary collimator and allowing the detector to collect a divergent beam and, at the same time, by using a positionsensitive wire detector (with an energy resolution of 100 eV for detected photons with energies of about 10 keV). Position-sensitive constructions are popular in many physical applications, unfortunately not in the XRF field. Potentially, it may lead to new solutions for WDXRF. Another interesting effort is the use of a scintillation gas instead of the scintillation crystal (Kikkert, 1983; Palicarpo, 1978). This is the so-called gas-scintillation proportional counter (GPSC). As is clear from Figure 26a, the resolution is substantially better than that of its ‘‘parents.’’ The production of avalanches of ion pairs by the expelled electrons is the same as described previously. The discharge currents at the electrodes are not measured, however, but, rather, the scintillation pulses in the ultraviolet part of the optical spectrum are detected by a photomultiplier as in other scintillation counters. An essential difference with conventional gas proportional counters is that the space charge during the passage of a photon is avoided. The freedom-of-charge-related phenomena enable using higher counting rates (up to 2.56104 cps; Dos Santos et al., 1993). The lifetime of a GPSC can be quite long, up to several years, if there is no gas leakage in the meantime. The ability to use counters with large windows positively differentiates this kind of construction from room-temperature solid-state detectors, such as HgI2. Their pulse amplitude is also proportional to the energy of the incident photon. The counter can be supplied with a digital rise-time discrimination analyzer to suppress the pulses
Figure 32 Geometry with position-sensitive detector: T ¼ x-ray tube; specimen with point P irradiated by primary x-rays; crystal; mp ¼ mirror plane; M ¼ mirror point of P; PSD ¼ positionsensitive detector; l ¼ distance between mirror point and detector wire. (From Ebel et al., 1983.)
Copyright © 2002 Marcel Dekker, Inc.
forming the background (Simo˜es et al., 1997). This results in a much better performance of the counter. 4. Calorimetric X-ray Detectors Calorimetric detectors operate in extreme conditions—at temperature of liquid helium. The essential part of the device is a low-temperature, high-Z superconducting material with a minimal heat capacity. X-rays, even single photons, are absorbed and their energy is transferred to phonons. The detector is thermally coupled to a cold bath. The increase in temperature, which results from the absorption, is measured and is proportional to the energy of the photon. The temperature is measured by a doped semiconductor thermistor, as small as possible to minimize its heat capacity. It converts the experienced temperature change into a change in resistance. The thermalization time of the detector is in the range of 1–50 ms, and the time necessary to equalize the temperature of the detector to that of the cooling device is 100–500 ms. There is an analogy to the dead time of gas-filled and gasscintillation detectors. Still, there is a possibility for that type of device to allow count rates as high as 104 cps. Now, in more modern constructions (Irvin et al., 1996), the semiconductor thermistor is substituted by a superconducting transition-edge sensor (TES) thermometer. It is a strip of superconductor material biased within the transition range from the normal to superconductive state. The greater thermal sensitivity of the device allows the use of normal metal absorbers as a first stage of the detector. Unlike semiconductors and superconductors, metals can dissipate the absorbed energy into freeelectron excitations, not phonon ones. The free-electron deexcitation process is quick and efficient. The use of TES leads to a change in the amplifier: A less noisy superconducting quantum-interference-device current amplifier is introduced. As a whole, this new combinations exhibits a surprising feature: The energy resolution of the detector, as measured for the MnKa line, is better than the thermodynamic energy fluctuations in the microcalorimeter (7.2 eV in comparison with 10.4 eV). The total spectrometer efficiency obtained up to now is reaching 80% in some low-energy regions below 5 keV and can still be improved. Among all the detectors mentioned in the present chapter, this one does not make use of the ionization phenomenon. Strictly speaking, this type of detectors can measure energy transfers even in these processes, in which no ionization takes place. It has been recently demonstrated for, e.g., particle detection and impact energy measurements in time-of-flight mass spectrometry of single biomolecules (Hilton et al., 1998). The quantum efficiency of such an x-ray detector can be very good ( 95%) and the absolute spectral resolution is 7 eV in the range 0.2–20 keV; see Figure 26b (Silver et al., 1997). The expected absolute spectral resolution can potentially be even better than 1 eV in the mentioned range (Moseley, 1984). Of course, we are fully aware that this type of detector is not intended to compete with wavelength-dispersive systems, but we must emphasize here that the time of unbeaten behavior of crystals over all other spectral-resolving devices has gone. It seems to be a good place to present the comparison between wavelength and most modern energy-dispersive detectors at their present status. 5. Dead Time and Shift of Maximum Pulse Amplitude At high count rates (from a few thousand counts per second on!), the recorded count rate deviates proportionally from the real count rate. The reason for this discrepancy is the slow decay of the anode charge. Cation sheaths screen the cathode. This reduces the Copyright © 2002 Marcel Dekker, Inc.
effective voltage between the electrodes, resulting in a lower acceleration of the electrons and, consequently, in a smaller number of ion pairs formed. Fouling of the anode wire by impurities in the counter gas has the same effect. This phenomenon shifts the maximum of the pulse-amplitude distribution to pulses from apparently lower-energy photons. Moreover, when a photon is detected, the detector is, for a short time, unable to detect the next incident photon. This period of time is called the dead time. The value of the dead time is about 1–2 ms for scintillation counters, about 0.2 ms for proportional counters, and 200 ms for a Geiger–Mu¨ller detector. To overcome this problem, one can either choose another detector (but the choice is very limited), lower the intensity of the exciting beam by masks or absorbers, change the geometry and the optical arrangement (as is done in the case of synchrotron radiation), or use electronic anticoincidence circuits or appropriate mathematical corrections. Now, after the introduction of the high-power 4-kW option to the modern spectrometers, masks can be inserted for the analysis of the major elements (e.g., in Shimadzu instruments). The mask avoids problems with too great a count rate and its introduction is probably easier than tube power change during the analysis. The relation between the real and recorded count rate is Nr ¼
Nm 1 N m td
ð11Þ
where Nr is the real count rate (count=s or cps), Nm is the measured count rate (cps), and td is the dead time (s). If we allow a given level of discrepancy, say a%, then the following equation is valid: a ¼ N r td 100 a
ð12Þ
The real count rate and dead time are thus connected by a hyperbolic relation. If the allowed value of discrepancy must be kept constant, the left-hand side of the hyperbolic equation is constant. However, this is only part of the problem. As already stated, a photon, arriving just at the end of the dead time, is registered with reduced amplitude. It is only after the socalled recovery time that a new photon is registered with normal amplitude. An improved method for calculation of the dead time was proposed by Bonetto and Riveros (1984). These authors used the well-known fact that the second-order peaks are significantly less intensive than their first-order analogs, but the ratio of the intensities should be constant. If it is not constant, the deviation results fully from dead-time losses in the first-order signal. They derived the relation N m1 ¼ B AN m1 N m2
ð13Þ
where B ¼ Nr1 =Nr2 and A ¼ Btd . The indices 1 and 2 refer to the order of the spectrum. A summary of the constituting parts of spectrometers and the wavelength ranges in which they are used is given in Figure 33. Note that different excitation sources are used in different branches of XRF, namely radioisotopes for EDXRF and tubes for WDXRF; there are also examples of synchrotron radiation. The x-ray laser is mentioned, but its appropriate version for XRF applications does not yet exist. Copyright © 2002 Marcel Dekker, Inc.
Figure 33 Review of different elements serving the construction of a spectrometer: optimal range (crossed area), feasible range (hatched area), and possible range (open area). (Adapted from Bertin, 1975.)
Copyright © 2002 Marcel Dekker, Inc.
IV.
QUALITATIVE AND QUANTITATIVE ANALYSIS
In this section, we review a number of items important to both qualitative and quantitative analyses. The qualitative application of XRF spectrometry is obvious. In simple cases, quite unequivocal assignments of spectral lines can be made based on a library of Ka or La lines or, in case of doubt, when an overlap exists, Kb or other L or M lines are checked to confirm or reject a line assignment. A few more complex situations are discussed. Quantitative analysis may also be simple. In the low-concentration range, quasilinear relationships are found between intensity and concentration. The same holds for determinations of higher concentrations but in a very narrow range for samples with quite comparable matrices. Interpolation on standard curves, obtained on suitable standard samples, allows a quick and sufficiently accurate quantitative determination. Another intensity–concentration relation, valid for binary samples, is based on the relative intensities of Ka and Kb lines in the case that the absorption edge of another element is located in between these two lines. In these boundary cases, matrix effects are negligible or interbalanced but always exist. In normal practice, they compromise the conversion of intensity into concentration in such a way that more or less complicated mathematical methods must be used for obtaining any accurate quantitative determination. The approach by fundamental parameter programs or correction algorithms is dealt with in Chapter 5. Between simple qualitative work and rigorous quantitative work, however, an area of applications is situated in which knowledge of the exact concentration or the complete qualitative composition is less important than the overall aspect of a spectrum for recognizing classes of samples (i.e., pattern recognition). An introductory discussion of this aspect is given in this section. Another point of interest in qualitative as well as to quantitative work is the measurement of the background. A good estimation of the background is necessary for obtaining net count rates, but decreasing the background allows lowering the detection limits. The origin and magnitude of the background is bound to the physical processes involved in the method of excitation and, thus, indirectly to the type of spectrometer. This aspect is also treated in this section. Finally, the said disturbing phenomena associated with the excitation of fluorescence radiation, as bremsstrahlung and the scattered radiation, can be used to increase the total analytical information and they also have some interesting applications. A.
Background
The background in WDXRF has four main sources: (1) the coherent and incoherent scattering of source radiation, (2) the presence of characteristic radiation of other than sample origin (materials of sample cup masks, sample cups, collimators, spectrometer housing, for example), (3) the detector, and (4) environmental noise (i.e., resulting from the threshold natural radiation of materials and even from a cosmic origin). The background from characteristic radiation from construction materials of the spectrometer can be minimized by careful construction and geometric arrangement of the device. Scattering from sample cup masks can efficiently be reduced by masks in front of the primary collimator. This contribution is especially great when small samples are analyzed. The beam mask reduces the irradiated area and may totally eliminate scatter from sample cup masks. Higher-order scatter increases with increasing wavelength. Pulse-height selection, as explained earlier, supresses counts from higher-order diffraction, but this technique is most efficient in the long-wavelength region and is decreasingly efficient toward the short-wavelength end. The short-wavelengths contribution to the background can be reduced by Copyright © 2002 Marcel Dekker, Inc.
installing adequate filters. A 100-mm titanium foil removes the characteristic lines of chromium from a chromium anode tube, significantly improving the conditions for the detection of Cr, Mn, Fe, Co, and Cu (Tertian and Claisse, 1982). For Cr tube radiation and a LiF (200) analyzer crystal, removal of scattered chromium radiation may be necessary when detecting the neighboring first-order lines. As Table 5 shows, not very many common elements are interfering. For steel analysis, only manganese really represents a problem. Because chromium is present in prevailing concentrations in the environment (tube anode!), manganese determination is also difficult with an aluminum filter. The simplest solution in such a case is the use of a LiF (220) crystal with better resolution in this region (but also with lower intensity). If tubes with other targets are used, tables are similar to Table 5 but with even more complex interference if targets like Mo and Rh are applied. The other elements listed in Table 5 refer mostly to less common elements and less intense lines. Other higher-order lines may interfere as well if the concentration of elements, from which the lines are derived, is high. For special cases, a convenient combination of tube targets, analyzer crystals, and filters can almost always be selected. Let us look at an example for the rhodium target tube. In the determination of palladium, PdKa1 overlaps with the Compton peak of RhKb (Fig. 34). A molybdenum filter in the primary beam reduces this peak to a very small level, allowing a perfect estimation of the location of the maximum and the determination of the intensity of the palladium line. Another efficient solution for background suppression is the use of linearly polarized x-rays. An excellent example was given by Kno¨chel et al. (1983) and it is reproduced in Figure 35. The analysis of nitrogen with a small concentration of xenon was made by a synchrotron excited XRF. Some special geometries allowing the use of polarized x-rays have been proposed by Wobrauschek and Aiginger (1980, 1983, 1985, 1986) for the more conventional XRF spectrometers. For EDXRF systems, a special commercial polarizing design is available (SPECTRO A.I. Gmbh). The subject is of great significance to synchrotron-based XRF and discussed as such in Chapter 8. A careful estimation of the background is apparently the simplest way to take the background into account! A first approach, but with limited application, is measurement on blanks—simple at first glance, but the adequate blank is often a rarity. Moreover, the blank ought to have the same scattering properties. Intensive efforts for mathematical modeling of background were performed but frequently for other purposes. The most common method of background evaluation is the linear (or even nonlinear) interpolation from background countings at a suitable angle above and below the peak position (suitable means not too far away for an easy linear interpolation, outside the tailings of the peak, and not at the position of an adjacent peak). Consequently, a good
Table 5 CrKa Pm La2 La Lb2 Ba Lg5 La Lb7 Ce Lb6 Pm La1 V Kb
Interfering Lines for CrKa and CrKb with LiF 200 crystal DE (eV)
D ( 2W)
CrKb
10 30 40 30 20 20 10
0.05 0.31 0.60 0.23 0.31 0.31 0.56
Ho Pr Mn Pm Ba
Copyright © 2002 Marcel Dekker, Inc.
Ll Lb7 Ka Lb1 Lg4
DE (eV)
D ( 2W)
< 10 20 50 10 30
0.04 0.24 0.61 0.17 0.30
Figure 34 Spectra of Pd excited by a Rh tube and recorded with and without a primary beam molybdenum filter (0.1 mm Mo). The peaks are labeled PdKa, PdKb, RhKa, and RhKb with their respective order indices. (Courtesy of ARL.)
spectral resolution of the system is essential because only in this case are peaks very narrow and do not overlap, and the background level is easily reached in the intervals between peaks. Finally, we refer to a method based on the relationship of fluorescent and scattered radiation intensities versus the value of attenuation coefficients. No measurements of background for the estimation of net line intensities is required in this case (Bougault et al., 1977). If we cannot overcome the background problem, it would be advantageous to use the background or scattered tube lines for increasing the information about the sample. There are very interesting examples of such work and they will be shown in Sec. IV.C. Copyright © 2002 Marcel Dekker, Inc.
Figure 35 Synchrotron-excited spectra of gaseous nitrogen with a small amount of xenon taken in the direction of polarization and perpendicular to the orbit plane. Counts normalized to stored electron current are integrated over the lifetime of the detection system. DORIS was operated at 3.3 GeV and about 55 mA. (From Kno¨chel et al., 1983. Reprinted by permission of the author and Elsevier Science Publishers.)
B.
Qualitative Spectrometry
Although the number of spectral lines is very limited compared to atomic emission spectra in the ultraviolet or visible region, overlapping of lines in x-ray spectra does exist. A classic example can be found in the works of Gentry et al. (1976), Sparks et al. (1978), and Sparks (1980), and an example is reproduced in Figure 36. At energies at which K lines of elements between palladium and cesium (about 22 and up to 30 keV) can be present, additionally the tails of L lines of uranium and thorium are found, and also the theoretically calculated M lines of the superheavy elements of Z ¼ 110 [in the time which passed since 1976 this element has been discovered (Hofmann et al., 1994)] and 126 should be expected. The erroneous attribution of some lines to these hypothetical elements was the subject of extended research and long discussions (Sparks et al., 1978; Sparks, 1980). In the software of modern spectrometers, special searching blocks exist for the assignment of lines: The detected line is compared with data compiled in the program. After the initial assignment, other lines of the element are identified according to data from the library. The initial assignment is eventually rejected if not confirmed by other lines (Fig. 37). Multichannel wavelength-dispersive instruments are not suited for this type of work, except for elements implemented on the spectrometer (maximum 28 elements). Sequential instruments can cover all elements (from beryllium on). The spectra can be displayed on the monitor screen of all modern spectrometers, documented, memorized as files, and printed. Particular parts of the spectrum can be scanned occasionally with higher resolution and sensitivity (step scanning) according to the time available for the analysis and the information required. Parts of particular interest can be displayed separately with a linear intensity axis or rescaled to square root or logarithmic scales. The peaks can be identified on the display by the angle 2W, by element name, either by wavelength or energy. Copyright © 2002 Marcel Dekker, Inc.
Figure 36 Fluorescence spectrum from a monazite particle excited with 37-keV synchrotron radiation (ORNL DATA) shows an improved signal-to-background ratio over that excited with 5.7-MeV protons (FSU DATA). Data points every 20 eV in ORNL spectrum and every 61.8 eV in FSU spectrum. (From Sparks, 1980.)
Figure 37 Example of qualitative analysis by rejection. The suspected MoLa peak in the lower spectrum was rejected after comparison with the standard containing molybdenum (upper spectrum). (Courtesy of Phillips Analytical.)
Copyright © 2002 Marcel Dekker, Inc.
The spectrum can be scanned linearly with respect to 2W or in a more optimal way. If Bragg’s and Moseley’s laws are combined, the following relationship among Z, the atomic number, and W is found: rffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi m þs ð14Þ Z¼ k2d sin W Z is proportional to the inverse square root of sin W. Jenkins et al. (1979) proposed computer monitoring of the scan speed according to this equation. It optimizes the scan speed, which represents a high gain in time especially in the low-Z range where the angular distances between the lines are long. The newest applications of WDXRF softwares try to avoid the old-fashioned treatment of data. One such traditional manual estimation was that concerning the possible spectral interferences. Normally, the analyst has to decide which place in a spectrum is relatively empty of disturbing peaks and compromising background contributions. This job was rather time-consuming, with the wide use of spectral tables and other datasets concerning, for example, intensity ratios. The results were not always encouraging. Now, virtual synthesis of wavelength-dispersive spectra is in progress (Reed and Buckley, 1996), in the Internet demonstration version also (Buckley, 1998). The expected pure-element spectra are loaded from the memory and installed in a region of interest of the anticipated spectrum. Those partial spectra are apparently corrected for the expected concentrations of the sample constituents. From the result shown on a screen of the computer, one can easily estimate whether the region of interest is suitable for a given analysis or should be excluded. Such a procedure makes the selection of lines for the qualitative analysis much easier, as well as the estimation of the background and also the assessment of the detection limits. C.
Special Quantitative Applications
Although the main quantitative treatment of XRF has been left to Chapter 5, still there are some special and nonconventional applications which proved to have potential practical meaning. Their significance becomes obvious when, considering the excitation process, one wonders how great a part of the radiation is transformed into the scattered form and how it can supplement the essential quantification based on characteristic signals. As stated earlier, monochromatic excitation simplifies the quantitative treatment of spectral data considerably, but this case is only seldom met in WDXRF. The most common excitation source is the x-ray tube, in which a set of characteristic lines superimposed on a white spectrum is generated. For this reason, only the fundamental coefficient method can be considered as rigorous and somewhat simplified in comparison with a complete fundamental parameter method. How the correction terms can be derived in a rigorous manner for the case of monochromatic excitation was shown in an article by Kuczumow (1982). In its original formulation, the Lachance–Traill equation is (Lachance and Traill, 1966) ! X Wi ¼ R i 1 þ aij Wj ð15Þ j6¼i
with aij defined as a constant written as aij ¼
mjC1 ðEÞ þ mjC2 ðEi Þ miC1 ðEÞ miC2 ðEi Þ miC1 ðEÞ þ miC2 ðEi Þ
Copyright © 2002 Marcel Dekker, Inc.
ð15aÞ
where m denotes the mass absorption coefficients for primary or secondary photons of energy E or Ei, respectively, corrected for the incident C1 and takeoff C2 angles. It has been proved (Claisse and Quintin, 1967) that this expression is valid only for boundary conditions: monochromatic excitation and absence of enhancement. In all other conditions, a’s are not constant, as experimentally shown (Tertian and Claisse, 1982; Lachance, 1981). It is impossible to derive a more flexible algorithm with constant coefficients directly from Shiraiwa and Fujino’s equation (Kuczumow, 1982; Rousseau and Claisse, 1974). Other algorithms similar to that of Tertian (Rousseau, 1974) and Lachance COLA (Lachance, 1981) are not recognized as fundamental. An algorithm that can be considered partially fundamental is that of Claisse and Quintin (1967), but it suffers from inaccuracies in the approximations introduced. Later, Rousseau (1984, 1987) and Kuczumow and Holland (1989) gave examples of algorithms derived from fundamental assumptions. Kuczumow and Holland describe the change in a when enhancement is included (monochromatic excitation) as follows: ! X aij Wij þ Daik Wk Wi ¼ R i 1 þ ð16Þ j6¼i;k
where all a’s are defined as previously but Da is defined as Daik ¼
m0kC2 ðEi Þ mkC2 ðEi Þ miC1 ðEÞ miC2 ðEi Þ
ð16aÞ
The new term m0kC2 ðEi Þ was introduced in Eq. (16). Although dependent on the excitation condition E and composition, this new term has the great advantage of being dependent on the latter in a simple way. The possibility of deriving the Lachance–Traill equation directly from the Shiraiwa and Fujino equation under the conditions applicable to Eq. (16) is also well established (Kuczumow and Holland, 1989). The next interesting case concerns scattered radiation as a source of analytical algorithms. Similar to the derivation made by Kuczumow (1982), an expression on the basis of coherently scattered radiation is obtained (Kuczumow, 1988): X coh coh coh þ R a b ð17Þ Wi ¼ Rcoh Wj i i ij ij j6¼i
where acoh ij ¼ bcoh ij ¼
mjC1 ðE0 Þ þ mjC2 ðE0 Þ miC1 ðE0 Þ miC2 ðE0 Þ miC1 ðE0 Þ þ miC2 ðE0 Þ coh sj ðE0 Þ scoh i ðE0 Þ
ð17aÞ ð17bÞ
and Ricoh results from the comparison of a coherently scattered line from the sample and from the pure element i and scoh is the mass coherent scatter coefficient (for the set of incident and takeoff angles, as always in the conditions of the analysis). For Comptonscattered radiation, analogous expressions are obtained: X Wi ¼ Rcom þ Rcom acom bcom ð18Þ Wj ij i i ij j6¼i
Copyright © 2002 Marcel Dekker, Inc.
with mjC1 ðE0 Þ þ mjC2 ðEcom Þ miC1 ðE0 Þ miC2 ðEcom Þ miC1 ðE0 Þ þ miC2 ðEcom Þ scom ðE Þ 0 j ¼ com si ðE0 Þ
acom ¼ ij
ð18aÞ
bcom ij
ð18bÞ
is the ratio of the Compton peak intensity of the sample to that of the pure eleRcom i ment. Equations (17) and (18) may be called Lachance–Traill equations for scattered radiation by their general appearance as well as by the way they have been derived. These equations are less sensitive to the type of components and changes in compositions than their analogs for fluorescent radiation, which otherwise is an advantage in analyses of samples widely varying in composition. From assumption, they include only the terms for strictly defined energy values, even if the analysis is carried out with a polychromatic source! In Table 6, the coefficients a and b from Eqs. (15)–(18) are collected, calculated for the K lines of La and Sm, the main components of samples consisting of La2O3 þ Sm2O3 excited by 60-keV photons at the incidence and emergence angles equal to 80 and 70 , respectively (the example is taken from EDXRF, but it is also instructive here). When applied to nonhomogeneous samples, also the a’s are proved to be dependent not only on the composition and the energy but also on particle size for dispersions or layer thickness for piles of layers of different composition (Helsen and Vrebos, 1984a, 1984b). Kuczumow et al. (1992) have proved that parallely with the conventional analysis of binary Au–Cu alloys by fluorescent lines (Fig. 38), it is possible to get similar information using coherently or incoherently scattered tube radiation. Sometimes, the results are even better than for the direct fluorescence analysis. Still other possibilities of using the scattered lines exist. Kuczumow et al. (1995) have proved, by a proper mathematical transformation of the scattered Rayleigh signal, that for light, chemically homogeneous matrices (as biological materials, e.g., wood), this new function of scattered radiation, called the corrected scattered intensity, strictly follows the density changes in the analyzed material (Fig. 39). The use of Rayleigh signals eliminates the necessity of transmission measurements and brings important auxiliary information (e.g., about the density). Although the method just described has been applied up to now for x-ray capillary-microprobe-type measurements only, it may be used generally (Luggar and Gilboy, 1994).
Table 6
a and b Coefficients for Fluorescent and Scattered Signals La2O3
af Daf acoh bcoh acom bcom
(La2O3 Sm2O3)
0.292 0.895 0.285 1.153 0.281 1.000
Copyright © 2002 Marcel Dekker, Inc.
Sm2O3 (Sm2O3 La2O3)
1.009 — 0.222 0.867 0.219 1.000
Figure 38 Analytical curves in the system R=W-1 versus w w0 for the CuKa line in Au–Cu alloys. The solid line with open circles denotes the authors’ fluorescence measurements; the solid line with crosses denotes Rayleigh measurements, and the solid line with solid square denotes Compton measurements. R is the intensity relative to the intensity measured for the pure element and W is the composition expressed as weight ratio; w represents the ratio of the mass attenuation coefficient of the sample to the sine of the takeoff angle C2 . The subscript O refers to the values for the pure element o. (Modified from Kuczumow et al., 1992.)
D.
Pattern Recognition
The spectrum of a sample of complex composition is characterized by a set of lines, documented by their position and intensity. If we imagine an n-dimensional space with the intensities on the coordinate axes, the set of intensities determines a point in this space. Such an intensity space is convertible into an n-dimensional space of concentrations, and the point in the intensity representation has its equivalent in the concentration space (Klimasara and Berry, 1987). The aim of pattern recognition is to attribute a sample to a given class. A class is in a discrete way a set of points of similar features in the intensity or concentration space or, in a continuous way, a subspace inside a complete n-dimensional space. This subspace is restrained by a closed hyperplane, which can be approximated by a number of hyperplanes of smaller dimensionality. Inherent to all experimental determinations, the determination of intensity is also bound to a certain degree of uncertainty. This implies that a point, as such, cannot be determined, only a small subspace of limited size, limited by the assumed confidence levels. Equally ill-defined for the same reasons is the boundary hyperplane surrounding the subspace occupied by a class of samples. The ‘‘thickness’’ of this ‘‘space ring’’ surrounding the subspace is determined in relation to each Copyright © 2002 Marcel Dekker, Inc.
Figure 39 Line scans from x-ray microprobe measurements that show the correspondence between transmission (dotted) and corrected scattered (solid) data, and the sequence of tree rings of Silesian Pinus silvestris. The linear scan was made using a 89-mm capillary and a step size of 89 mm. (Modified from Kuczumow et al., 1995.)
coordinate axis by the appropriate multiplicity of the projection of standard deviation on this axis. 1. Delimitation of Classes If we consider classes as separate subspaces in n-dimensional configuration space, it is obvious that each of these classes has its own boundary hyperplane or closed space ring. If these hyperplanes can be determined in an analytical way, we are able to know the boundaries of classes. In simple cases, a class can be delimited by construction of a lowdimensional hyperplane that closely approximates the real boundary. The simplest example is the trivial division of the whole configuration space on two categories; occupied space–empty space (zero–one decision). It can be done by assuming the limiting surface (‘‘wall’’) inbetween the mentioned subspaces. If classes are separated by empty spaces, the differentiation of classes is even easier. 2. Method of Nearest Neighbours Here, the classes are considered as small subsets of points. The classification of each new point relies on a comparison with its nearest neighbors. Knowledge of the distance between the sample and the nearest neighbors is essential. The accuracy of the method depends on the way the distance is determined and the degree of complexity of the boundaries between classes, although it can be simplified by imposing proper statistical weights on learning samples. The method of the nearest neighbors has further implications. Copyright © 2002 Marcel Dekker, Inc.
Once the sample is placed in relation to the nearest neighbors, which are nothing but the standards, a separate program can draw the average from the composition of standards, taking the distances from the neighbors as the statistical weights for the estimation of the influence of each standard on the final result (Klimasara and Berry, 1987). However, this way of using pattern-recognition methods is the same as very conventional method of comparison with standards! Of course, standards are no longer considered in a conventional sense as some numerical values in relation to some elements, but rather as the full patterns [compare also the genetic algorithm by Dane et al. (1996)]. 3. Reduction in Number of Dimensions All points and classes are included in the multidimensional space generated by the number of elements present in the sample; however, some variables may be less dependent on the related coordinates, and then the number of variables necessary to differentiate the sample may be reduced. The number of dimensions may also be reduced if some variables are correlated; the most obvious way of coupling the elements is by joining them in a chemical compound. 4. Pattern Grammar Approach Another version of ‘‘pattern recognition’’ is also possible. The full spectra for different standards and=or samples can be collected and treated as some analytical claims (sentences), with peaks considered to be analogs of single words, and the whole set of spectra as a constituent of a ‘‘spectral language.’’ At the same time, we have the complete chemical knowledge of the same standards, with standard composition treated as the sentences, their particular components as the single words, and such second system is the part of another, ‘‘chemical language.’’ The task of the chemist is to translate sentences from spectral to chemical language while working with the real samples. Even more, the translation from one language to another can be treated as mapping (rather linear than nonlinear!). 5. Examples Samples may be classified into groups according to their origin. A primitive version of such x-ray methods was demonstrated by Lagarde et al. (1984). They observed the characteristic features of spectra of ceramics samples from western Africa and then put the samples in order ‘‘by the naked eye’’ according to their origin. A classic reference here is the work of Kowalski et al. (1972), which was followed by many other articles on archeology. Kowalski recognized the origin of obsidian artifacts by studying a set of intensities of 10 trace elements and projected the image of this set from 10-dimensional configuration space into a 2-dimensional space (nonlinear mapping). Similar applications in geology are obvious. Lo I Yin et al. (1989a, 1989b) proposed the application of the pattern-recognition-supported XRF to the initial selection and analysis of Mars Rover geological samples. They discovered on this occasion that it is not necessary to use sophisticated detection systems. On the contrary, they applied probably the most primitive version of an x-ray spectrometer with a proportional counter as detector. The resolving power of such a detecting system belongs to the worst among the existing ones (compare Fig. 26a), but even in these conditions, the results of the sample-selection procedure are good. In such classification procedures, putting emphasis on advanced data analysis can be a substitute for using a sophisticated high-resolution spectrometer for special analyses in difficult conditions. Copyright © 2002 Marcel Dekker, Inc.
If only a limited amount of data is available, samples can only be subdivided in groups. Discrimination of the alloys 663, B-1900, and 1455 can be done on the basis of their zirconium content (Kevex Material Analyzer Mode 6500 brochure). Nickel–cobalt alloys were subdivided on the basis of a logical analysis of successive spectral responses: Is the ratio ‘‘contents of elements A=contents of element B’’ greater than some assumed level X: A=B > X (Fig. 40)? Complete analysis of the alloy was not necessary. The same is often done when samples can be compared to standards [e.g., the ‘‘framing action’’ by software (description of Siemens software)]. An important application of pattern recognition may be found in ‘‘two-stage analysis.’’ For complex samples with nonlinear dependence of intensity on concentration, the analysis can be broken down in steps with a smaller number of dimensions (lower degree of complexity) or the samples may first be subdivided in classes before applying correction algorithms. The appropriate sets of differential equations exist: they establish the relation between the value of relative intensities [Claisse–Thinh systems (Claisse and Thinh, 1979)] or between increments of intensities [Kloyber et al. (1980) and Kuczumow (1984) systems] and increments of concentrations. These increments are calculated with respect to a central standard that represents the origin of a local system of coordinates. In
Figure 40 Selective procedure of the KEVEX Material Analyzer Model 6500 based on successive analysis of the element ratios. (Courtesy of KEVEX Corp.)
Copyright © 2002 Marcel Dekker, Inc.
this local system, a certain set of differential equations is valid and is of linear character (if not, it is not profitable to use it in such a restricted space). After attributing the samples to given classes, the right set of differential equations may be used for accurate determination of composition. The origin of the local system of coordinates may, at the same time, be the central point of a given class. When the delimitation of classes becomes problematical, a point in the configuration space may be designated and a set of differential equations, valid about this point, may be determined. From this, the boundaries of classes can be found and determined by some leading vectors introduced in the central points of the classes. Yap (1987) and Yap and Tang (1984, 1985) posed the reciprocal question in relation to ‘‘pattern recognition’’: how to extract some number of differentiating features from the known classes, differing them by time and place of origin. This was used to analyze pottery samples. Another name for this method is the ‘‘training phase’’ of the analysis. E.
Analysis on the Basis of Neural Networks
The pattern-recognition method leads us to quite different levels of reasoning. The most popular recent way of working with a spectrum is to transform it numerically and show it in two-dimensional intensity–wavelength or equivalent intensity–2W or intensity–energy diagrams. Some logical artifacts result from the numerical transformation of the original sets of signals. The interesting comparison between functioning of the electronic chip in the binary system of information and the biological cell in the complex system of information was presented by Cholewa during 15 ICXOM, Antwerp (Cholewa, 1998) (see Fig. 41). At the same time, without such a numerical presentation, a common naked-eye observation allows one to compare a fresh spectrum to other similar pictures compiled in our memory. Our way of reasoning is not numerical at all, but resembles some procedures of pattern recognition. It is simply a process of attributing a sample to some class. In a happy case, one can recall the same pattern in memory and then classify the sample as a part of existing and well-described objects. One of the possible ways of reasoning similar to the ways of functioning of the human brain is by applying so-called ‘‘fuzzy logic.’’ In that type of logic, the object can belong to the proper set with some probability, not necessary equal to 1 (belongs) or 0 (does not belong). The spectrum can be recognized in this manner, with some artificial probabilistic rules set in the very beginning of the process. An instructive example is taken from an article by Otto et al. (1992). For the qualitative estimation of an x-ray spectrum, the following rule is very useful: If the Ka line is supposed to be in a spectrum, then the Kb line should be associated with it, being ‘‘about 5’’-fold weaker. Neural networks open another approach to the problem. They imitate the action of neurons joined by synapses in the human brain. The artificial network has a layer of input cells and a layer of output cells. The cells from layers just mentioned are connected, some of them directly and the majority indirectly, by the additional cells, grouped in one or more additional, hidden layers. Each middle cell collects one or more signals from previous cells, gives those signals the proper statistical weight (e.g., according to previously imposed rules), and sums them up. Next, the modified signal is sent from the middle cell into one or many cells in the next hidden layer or, finally, to the output layer. The initial information (e.g., spectrum) given to the input layer is transformed into a useful set of information (e.g., about the chemical composition) in the output layer (Otto et al., 1992; Larsen et al., 1992; Bishop, 1992; Bos and Weber, 1991; Luo et al., 1997). Otherwise, Copyright © 2002 Marcel Dekker, Inc.
Figure 41 Drawing showing the different degrees of complicacy involved in the action of (left) the electronic chip and (right) the single biological cell, also from the informatic point of view. The electronic chip includes only one bit of the binary information; the cell is an incomparably more complicated system. It is a basis for the differences in computing and brain thinking. (Courtesy of Copyright © 2002Cholewa.) Marcel Dekker, Inc. Dr. Marian
it resembles the ‘‘pattern grammar’’ method previously described. The method can be applied for the classification of very complex spectra, for clustering the datasets, for coupling the sets of spectra to the groups of compounds, and for nonlinear data transformation between sets with different dimensionalities.
F.
Detection Limits
The importance of the correct determination of detection limits has been acknowledged by many chemists, especially those involved in environmental regulation. The accepted definition is as follows: ‘‘The limit of detection is the lowest concentration level that can be determined to be statistically significant from an analytical blank’’ (Nomenclature, 1978). The limit of detection is expressed as a concentration cL or an amount qL and is derived from the smallest measure IL, I being the instrument reading. How the data are treated in the statistics is described in Chapter 5. For a thorough discussion we may refer to the article by Long and Winefordner (1983). This article also contains interesting references to articles of Kaiser, Boumans, and others, which are therefore not repeated here. For the derivation of some simple statistical dependencies, the small, almost forgotten book of Beers (1958) is instructive. Finally, we can recommend the excellent discussion of the physical origin of background sources and detection limits in PIXE and synchrotron-excited XRF given by Sparks (1980), which is also valid for WDXRF. Here, we only invoke that according to the International Union of Pure and Applied Chemistry (IUPAC) recommendation, the characteristic signal is statistically attributed to the element when it is elevated not less than 3s(s ¼ standard deviation) above the background level. It makes the confidence level equal to 99.86%. The detection limit is then calculated for the standard deviation multiplicity k ¼ 3 and is expressed as a function of intensity in a unit of time (R ¼ N=t) with t as the counting time and with the sensitivity m expressed in cps=% or cps=ppm. In Table 7 a series of values for cL are given for randomly chosen elements collected from manufactures’ documents. No general list can be included because detection limits as well as sensitivities depend heavily on matrix composition and on the experimental conditions. Instead, this is intended as an indication of order of magnitude. Each discussion of detection limits provides methods for deciding whether a signal value IU for a unknown was greater than background reading IB with a given level of confidence. Another important approach is advocated and applied by Clayton et al. (1987). They state that ‘‘traditional techniques for determining detection limits have been concerned only with providing protection against type I errors, or false positive conclusions (i.e., reporting an analyte as present when it is not). They have not considered type II errors, or false negative assertions (i.e., reporting an analyte as not present when it is).’’ In their article, they discuss a test of the null hypothesis c ¼ 0 versus alternatives that c > 0, where c denotes the true but unknown concentration of an analyte in a medium and for two models of calibration curves with known or with unknown parameters. The test is applied to chromatographic data. Their theory can certainly be applied to XRF data, but the attempt has not yet been reported.
G.
Limits of XRF
Compared to ultraviolet (UV)–visible emission spectra, x-ray spectra are ‘‘oligoline’’ ones and very convenient with respect to the resolution of the spectrometers. The UV-visible Copyright © 2002 Marcel Dekker, Inc.
Table 7
Detection Limitsa
Matrix
Element
Terephthalic acid Low-alloy steel
Al–Mg alloy
Aluminum Glass powder Phosphosilicate glass Cement
Copper alloys
Fe Co C Al Si Cr Mg Si Ti Mn Cu Mg P B B2O3 Na2O MgO Al2O3 SiO2 SO3 P2O5 Be
cL 0.15 0.18 80, 240 4, 4.5 2.1, 4.1 2, 1.7 7 5 3 2 1 10.5 1.3 1% 0.4% 36 27 22 50 24 32 0.2%
Data source Philips Philips, Siemens
Philips
Siemens Broll, 1985 Philips Siemensb
Philips
Source: Data from Philips Analytical ‘‘X-Ray Spectrometry Application Notes nos. 504, 622, 623, 737, 808, 913, 916; Siemens SRS Awendung 84=8, 86=1, 86=5, 86=6; and Broll, 1985. a ppm (unless otherwise specified) for 100 s counting time and k ¼ 3. b With use of multilayer OVO 55.
emission spectrum of iron contains some 2000 lines between 200 and 800 nm; its complete x-ray counterpart has a dozen important lines between 1.757 A˚ (Kb1) and 0.615 A˚ (Lg)! For everyday work, however, only two lines, 1.936 and 1.940 A˚ (Ka1,2) and 1.757 A˚ (Kb), play a role. The densely populated L series spectrum such as that of gold has a set of about 20 lines between 0.9 and 1.5 A˚, a spectrum that can be managed by a spectrometer with a relative spectral resolution of 102. The prospects for the x-ray spectrometric range should be bright, but what are the limits to this method? 1. Application Limits All elements from beryllium on can be detected and=or quantitatively determined. On the high-Z element or low-l side, the spectral resolution may be a problem for pure spectroscopic studies, but seldom or never for qualitative or quantitative applications. Because the emitted energy of photons is high and the attenuation coefficients are low, the critical depth is great and, consequently, the analyzed volume is very representative of the composition of the bulk. The critical depth is the thickness dcrit of a layer parallel to the surface, from which 99% of the intensity sensed by the detector is recruited. Copyright © 2002 Marcel Dekker, Inc.
It is a function of the takeoff angle C2 , the linear absorption coefficient mr of the matrix, and the density: ln 1 II1t ln102 4:61 dcrit ¼ lim ! ð19Þ r=sinC2 mr=sinC2 mr=sinC2 m where It is the intensity emitted by a thin foil and I 1 is the intensity emitted by an infinitely thick sample. For long wavelengths, mr becomes high while dcrit and analyzed volume become low. If the sampled volume is low, the analysis may be called a near-surface determination, not representative for the bulk (see discussion in Sec. III.D.2.d). In that sense, the results of the determination of heavy and light elements in the sample including different kinds of elements can be nonequivalent. The heavier elements are then determined in the bulk, whereas the light ones are determined on the surface only. This can lead to surprising results if the process of surface segregation is occurring. Moreover, during recent density determinations (Tsuji and Wagatsuma, 1996) of coating layers by a similar method, Grazing emission x-ray fluorescence, striking differences have been revealed between surface and bulk densities; for example, the density of a 27-nm-thick coating lead layer was equal to hardly 35% of the normal, the bulk value of the lead density (Tsuji et al., 1997). It is a true limit by the force of nature, although it may be exploited for near-surface determination of light elements. From the concept of critical depth, it is obvious that coatings can be analyzed up to certain thicknesses depending on the absorption (and enhancement!) characteristics of the coating and substrate; the overall thickness range is from 0.01 mm to a few hundred micrometers for the normal angle versions of XRF. Apart from the analysis of macrosamples, WD spectrometers may also be fit to electron microprobes, where they supplement the ED spectrometer for better spectral resolution of light elements. Curved crystal optics is generally used because of optimum efficiency (compare the quantum efficiency of a classical WD spectrometer ffi108, curved crystal WD ffi 106, and ED spectrometer ffi104). WD spectrometers are also needed for heavy-particle-induced emission; because of the presence of many satellite lines, good resolution is required (Watson et al., 1977; Chapter 12). The dynamic range of analysis covers about five orders of magnitude, from 103 to 100%. Below 103%, preconcentration procedures must be used. The intensity and uniformity of exciting beams delivered by conventional end-window tubes were not sufficient to allow spot analysis. However, Boehme (1987) and Nichols et al. (1987) published the results of two-dimensional mapping executed with small beams from high-power sources (rotating anode tubes), allowing lateral resolution of about 30 mm. This offsets the boundaries to a new domain: high-spatial-resolution x-ray spectrometry. Still, the possibilities of the tube-excited, slit-based WDXRF microprobe are relatively poor in comparison with an x-ray capillary-based EDXRF system or especially proton and electron microprobes (Chapters 11–13). However, for many applications, the lateral resolution mentioned earlier is quite sufficient and this technique is time-saving. Moreover, the tube can be coupled with the polycapillary device. Note, that, for example, the diameter of the spot obtained from the polycapillary lens is in a range 50–500 mm, but the careful work with such a beam can give a somewhat better spatial resolution. In Sec. VI.B.3, we will discuss the milliprobes, a new solution which seems to be the most suitable for the wavelength-dispersive mode of operation. Copyright © 2002 Marcel Dekker, Inc.
2. Limits of Precision and Accuracy An analyst is always threatened by errors in the analysis outliers, random, and systematic errors. When a given operation or measurement provokes an offset between the true and measured values, a systematic error is said to be introduced. This offset may be due to a wrong calibration of any kind or to a wrong standard. Such errors can only be detected by analyzing a sample from the same batch with other techniques. Random errors and outliers can be dealt with by statistical methods. One example of such a method is described by Plesch (1981) and implemented in the software package Spectra 310 (Siemens). A calibration with n standards is assumed to give a standard deviation. The standard deviation ss is calculated according to sffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi X ðC0 Ci Þ2 i ð20Þ ss ¼ np where Ci is the true weight fraction and C0i the result obtained; p is the number of parameters used in the fit: p ¼ 2 for a linear and p ¼ 3 for a parabolic fit. If the concentration Ca of an element in one specimen differs considerably from C0a , the presence of an outlier is suspected. The suspected value is eliminated, and a new standard deviation is calculated. Ca is an outlier with a probability of P% according to the inequality (F-test) (S1=S2)2 > Fp. The threshold values for Fp are listed in the current statistical literature. Application of the F-test in fact checks whether a statistically significant difference exists between the standard deviation calculated with and without the suspected result. Possible causes of outliers are numerous, including all kinds of human mistake, temporary instrument failure, and heterogeneity of specimens. Recently, the numerical programs helping to segregate the data on the ‘‘true’’ ones and outliers are available. They are based on the Grubbs test (Grubbs, 1969; Rousseeuw and Leroy, 1987). A measurement of fluorescent radiation is, as in any other experimental determination, subject to minimal error. The standard deviation of a radiation measurement is equal to the square rootpof ffiffiffiffi the number of accumulated counts N. The real standard deviation is greater than N because of errors of instrumental origin. Another source of error is the conversion of intensity into concentration and, last but not least, inherent to the sample, microheterogeneity, which is difficult to foresee and to quantify and is present even when all normal precautions for careful sample preparation are taken. Heterogeneities caused by sample preparation are dealt with elsewhere. These artifacts should be absent here. Segregation, however, which is the origin of what we call microheterogeneity, is a bulk (or sometimes surface) phenomenon of crystallographic origin: solid solutions of compounds or elements that are not perfectly soluble in each other, segregating into multiphase components. The ‘‘degree’’ of heterogeneity in the composition of the segregated phase as well as in the size of the phases depends on the history of the solution (thermal, chemical, or mechanical treatment). It is not easily mastered and is difficult to quantify. Its importance for XRF analysis has been acknowledged by several authors; Examples are tin and lead in solders (Glade and Post, 1970), silicates in fused beads (Novosel-Radovic et al., 1984), and silicon in Si–Al alloys (Michaelis and Kilday, 1962). The first report we are aware of is an article by Claisse (1957) on the determination of FeS in a sulfur matrix. A systematic study has been devoted to this phenomenon by Helsen and Vrebos (1984a, 1984b, 1986a, 1986b) and Vrebos and Helsen (1983, 1985a, 1985b, 1986), both by Monte Carlo simulation on hypothetical mixtures and by measurements on samples of known segregation. It was found by simulation on a (hypothetical) dispersion of spherical particles of iron in a chromium matrix Copyright © 2002 Marcel Dekker, Inc.
that even for a particle diameter of 1 mm, an error is introduced in the relative intensity of FeKa of 6% (on the level of cFe ¼ 0.5) with respect to the perfectly homogeneous solution. The effect was proved experimentally on alloys of Al–Si, Al–Mo, and Al–Zn. The extent of the effect changes with time, as in fused beads. This was demonstrated by the determinations of Novosel-Radovic et al. (1984) on fluxed silicates and also by Monte Carlo simulations by Vrebos and Helsen (1985a, 1985b). Microsegregation imposes a true physical limit to the accuracy obtainable by XRF spectrometry on quite a substantial number of samples, including minerals and alloys. 3. High-Z Limit We already insisted on the low-Z limit in Secs. III.D.1 and III.D.2, but what about the high-energy end of the spectrum? Commercially available generators and side-window tubes allow working at high voltages up to 100 kV, bringing even the K line of bismuth into the potential analytical spectrometric range. If a tungsten target is used, however, characteristic lines represent only a minority of the integrated emitted intensity, thus, excitation is bound mostly to continuous radiation, the maximum of which lies at 50 keV. Thus, only elements with K absorption edges below this value can be efficiently excited (Eu, Z ¼ 63, and all elements below). The above remark is even more valid for the spectrometers with end-window tubes, most commonly used in present-day instruments, of which the high voltage is limited to 60 kV. Possible excitation of K lines for all the elements from the periodic table by the tubes operating in voltages up to 160 kV is bound to metal–ceramic low-power tubes, cooperating with EDXRF systems and is outside the scope of this chapter. Thus, the available voltage limitation is the first one. If we return to Figure 26b and compare the spectral resolution of the LiF spectrometer to spectral distances between analogous lines in the K series, it is obvious that spectral resolution is worse above 35 keV (for K lines of Pr, Z ¼ 59). This is a second limitation. Most of goniometers have their optimal 2W range between 10 –15 and 70 : At the higher 2W end, the angular dispersion becomes insufficient; at the lower 2W end, the intensity becomes very low. If we choose LiF (420) with the shortest interplanar distance, 2d ¼ 1.802 A˚, useful measurements can be made at 10.0 corresponding to photons of 39.6 keV (Sm, Z ¼ 62 in the first-order spectrum). This is a third limitation. Studying Figure 26 further, we note that the spectral resolution of the LiF spectrometer is below the spectral resolution of the semiconductor Si–Li detector for photons of energy exceeding 17 keV, which corresponds to MoKa, Z ¼ 42. This is the fourth and most serious limitation in the K series. For analysis in the L series, there is no such limitation because the whole energy range lies within the capabilities of wavelength-dispersive devices. However, using the new microcalorimetric detectors with extraordinary and relatively constant spectral resolution, one can anticipate the preponderance of the spectral resolution factor for wavelength-dispersive crystals only in a region below 4–5 keV, which means the Ka line of V, Z ¼ 23 (see Fig. 27). Then, the ability to make the efficient trace analysis (because of higher count rates) and the availability of the device are still the advantageous sides of application of the crystal dispersion in the future. The wide voltage range of generators should not be used without some awareness of inherent drawbacks. For example, the K absorption edge of Mo is 20 keV. Thus, each x-ray tube with characteristic lines above 20 keV and working above twice this 20 kV for the maximum of bremsstrahlung generation is quite good for the excitation of Mo. The potential, however, should not exceed this value substantially much because the maximum Copyright © 2002 Marcel Dekker, Inc.
of the white radiation would shift more and more away from the molybdenum absorption edge and even the spectral characteristics do not improve. If we divide the mass coefficient of photoelectron excitation by the sum of mass coefficients for Compton and coherent scatters (this is a kind of measure of peak-to-background ratio), we obtain the values 76 and 18 for tube-excitation potentials of 40 and 100 kV, respectively! This is a substantial loss in peak-to-background ratio while passing to higher voltages. The discussion in this section shows that the advantage of high-energy photons (generated by very high voltages in SWT) does not compensate for the drawbacks and this probably explains why most manufacturers fit EWT tubes to the instruments. During the last years, manufacturers rather resigned from high-voltage generators and tubes, even in high-power 4-kW constructions. The use of a higher current is now a preferred option. Thus, we expect that only spectrometers with a high voltage up to 60 kV will dominate the market in the future. The effective spectral range above 30–40 keV should probably be left for typical energy-dispersive detectors. In the preceding paragraphs, we tried to show that although the spectrometers are almost perfect, there are limits to precision and accuracy due to the nature of the sample and to the mathematical conversion of intensity to concentration. These uncertainties are generally greater than those introduced by instrumental parameters.
V.
CHEMICAL SHIFT AND SPECIATION
The photoelectric excitation of atoms and subsequent specific deexcitation by emission of x-rays is the basis of all qualitative and quantitative uses of XRF. The deexcitation process was considered to take place between unique excited and ground quantum states. If these states were always stable, x-ray emission lines were to be found at strictly determined wavelengths or energies. However, electron density and quantum states are influenced by the environment of the atom with consequent wavelength shifts, or so-called chemical shifts, of the emitted lines. Such techniques as nuclear magnetic resonance or Mo¨ssbauer spectroscopy collect their spectroscopic information essentially from measurement of chemical shifts. As a matter of fact, all kinds of spectral lines of electromagnetic radiation are subject to chemical shifts. All these shifts are small and detectable only if the spectral linewidths are sufficiently small and the instrument has the required spectral resolution. For x-rays, the line shift is at most a few electron volts. An angular error DW ¼ 0.02 on the determination of a peak position at l ¼ 1 A˚ corresponds to 5 eV. Moreover, small temperature changes of the analyzing crystal already shifts the position over similar values. The sensitivity of peak position to the chemical environment of the atom was observed at the very beginning of x-ray spectrometry by Lindh and Lundquist (1924). The following characteristics of lines are sensitive to changes: line position (chemical shift) (Sumbaev et al., 1968; Sumbaev, 1970; 1976), line shape (Urch and Wood, 1978), and mutual ratios of line intensities (Asada et al., 1975). A detailed summary of the effect can be found in the book by Agarwal (1979), for example. In spectrometers with limited resolution, such as different energy-dispersive systems, most spectral shifts are hardly detectable. Chemical shift research has been done on high-resolution spectrometers. Gohshi et al. (1982a, 1982b) constructed an excellent two-flat-crystal instrument for studying chemical shifts. Asada et al. (1975) and Haycock and Urch (1978) improved the use of commercially available instruments for this purpose: by Fourier transform and iterative procedures for peak position detection or by using finer collimators. When analyzing Copyright © 2002 Marcel Dekker, Inc.
crystals with large thermal expansion coefficients, such as PET, are used, a temperature variation of 1 C gives rise to peak shifts greater than the chemical shifts. Helsen and Wijnhoven (1972) solved this problem by inserting standards at regular intervals and by using statistical techniques for the determination of the peak position. Sumbaev and Coworkers (1968, 1970, 1976) made experiments with an instrument in a Cauchois arrangement. In an article by Kataria et al. (1986), it was proved that even using a common solid-state Si–Li detector with moderate energy resolution, it was possible to detect chemical effects. Similarly, Habulibaz et al. (1996) performed experiments with a multilayer with an exact 2d spacing of 300 A˚, which means with poor spectral resolution. They managed to register SiL spectra for different compounds, clearly showing the influence of the chemical environment. Theoretically, chemical shifts can be calculated by the self-consistent field approach for the four quantum states involved (Manne, 1981). Unfortunately, the calculations are largely dependent on initial assumptions, the choice of which is difficult and somewhat random and the results are far from unequivocal. The experimental results of Sumbaev seem to be more spectacular. He connected the chemical shift DEKL to Pauling ionicity, i, valency m, and the parameter p, which was determined by lifetime ratios of electrons on the levels K or L: DEKL ¼ im½ pK DEK;ðZ:Z1Þ pL DEL;ðZ:Z1Þ
ð21Þ
He achieved an even more surprising result by presenting chemical shifts per unit ionicity versus valence (Fig. 42). This dependence is linear and may serve as proof of both the importance of chemical shift and the accuracy of Pauling’s scale. The shift can be related to the character of the bond in the molecule or the crystal. During the formation of a chemical bond, a rearrangement of electrons occurs, thus the formation of ions or bonds of varying degree of ionicity, the influence of which is reflected on all electron energy levels, sometimes more on the K level than on others levels. An example may be the shift between the Ka position of metallic aluminum and aluminum in different coordination states (IV and VI) (Helsen and Wijnhoven, 1972). Asada and Co-workers (1973, 1975, 1976) also made measurements in which the resulting chemical shift was attributed to the ionic bond character. They investigated the
Figure 42 Chemical shift as a function of valence. The energy shift divided by the ionicity is plotted (solid line). The dotted curve illustrates the uncorrected values of DE. (From Sumbaev, 1970. Reprinted with permission of J. B. Adashko.)
Copyright © 2002 Marcel Dekker, Inc.
changes in the ratio of some spectral lines, such as those for the Kb5 and Kb1 satellite line, and they connected them to the oxidation number of the analyte element. The results are again surprisingly elegant (Fig. 43). Clear proof for the dependence between the shape of lines and the oxidation number was given for manganese (Urch and Wood, 1978) and also between the chemical shift and the coordination number for different elements (Gohshi, 1981). The shift of AlKa for coordination states 0, IV, and VI was given by Helsen and Wijnhoven (1972). The application of the determination of chemical shift to chemical speciation is obvious. In geological samples, sulfur can be distinguished according to its different oxidation states (sulfur, sulfide, sulfite, sulfate, and hyposulfate). Birks and Gilfrich (1978) made measurements for sulfur using a portable low-power WDXRF instrument. In environmental and biological samples, the oxidation state of chromium is important: Cr(III) does not pass the cell wall, but Cr(VI) does and is suspected of being carcinogenic. Iwatsuki and Fukasawa (1987) reported the determination of chemical states of arsenic, selenium, and bromine. The determination of the valence of iron in welding fumes by measuring the energy shift and the intensity ratio of Kb0 to Kb1,3 has yielded very fine results (Fig. 44) (Tanninen et al., 1985). Kataria et al. (1986) managed to measure the varying ratio between the total line intensities of Kb and Ka as due to
Figure 43 1975.)
Satellite peak intensity ratio Kb5 =Kb1 versus oxidation number. (From Asada et al.,
Copyright © 2002 Marcel Dekker, Inc.
Figure 44 Relative chemical shift DE and the average valence of iron in standard samples and four welding fumes. The squares indicate the statistical uncertainties: SS ¼ stainless steel; MS ¼ mild steel; MMA ¼ manual metal arc technique; MIG ¼ metal inert gas technique. (From Tanninen et al., 1985.)
different oxidation states of manganese using an instrument of very moderate resolving power. This discussion of other than elemental or quantitative information from x-ray spectra has been intentionally restricted to x-ray spectrometry. A counterpart exists in the spectrometry of the ejected electrons, but that discussion belongs to the domain of spectrometry of Auger electrons or primary ejected electrons, for which higher resolutions are available or better speciation is possible. Another and sometimes severe drawback of these techniques is the required high vacuum and surface charging for nonconducting specimens. It should also be mentioned that Raman spectrometry is possible in the x-ray domain, but this field of research is only really accessible when synchrotron radiation is available (Sparks, 1974; Eisenberger et al., 1976a, 1976b), although performance on conventional instruments is feasible (Suzuki et al., 1970). In analytical determinations by synchrotron radiation, the Raman signals are often viewed as an additional disturbance (Jaklevic et al., 1988). Not many researchers are working in the field of speciation by WDXRF; nevertheless, it is a field of increasing importance, with theoretical as well as practical interest. It gives more than simple elemental information. Moreover, the shift of lines affects the intensity measurements for quantitative analysis, and stresses the use of the right standards for specific cases. In extreme cases, it may even be compulsory for quantitative determinations to scan over the peak maximum for each determination and locate in each case the intensity at the maximum for good precision and accuracy. If ultimate accuracy should be required, correction for peak heights may be involved. The few examples given point out that for studying the chemical shift, shape, and intensity measurements of spectral lines, the use of dedicated instruments is indicated, although a certain amount of useful work can be performed on commercially available instruments. A very interesting discussion of the problems immanently connected with the analytical application of soft x-rays can be found in the work by van Sprang and Bekkers (1998). These remarks have the obvious implications for the speciation analysis. Copyright © 2002 Marcel Dekker, Inc.
VI.
INSTRUMENTATION
The physical principles of a WD spectrometer do not allow dramatic differences in the general layout of the spectrometer. A practical instrument unites a number of construction compromises for speed, versatility, resolution, intensity, cost, user friendliness, and other factors, and all combinations have their own merits and their own drawbacks. Buying an instrument is a decision that in itself is a compromise because all benefits wanted by a customer are seldom or never united within a single instrument. In this section, the main features of the most important commercially available instruments are juxtaposed without formulating a global or final judgment.
A.
Electrical and Electronic Features
In modern instruments, all or almost all functions are controlled by a computer. As the electronic configuration is subject to continuous change, no details will be given here. Mentioned here are a few aspects which were difficult to automate or, when feasible, were hardly automated in older instruments: High voltage generators: The costly electromechanical control of conventional generators is replaced by computer-controlled solid-state generators, switched or high frequency. An example of exploitation of this flexible control is the switching of the tube along the isowatt curve, modifying the high voltage but keeping the power constant, an interesting feature if different intensities of the primary x-ray beam are required for different elements in the analysis of a complex sample (major, minor elements) by a sequential spectrometer. A consequence of the mechanical decoupling of W and 2W axes lifts the necessity of manual alignment of the crystals by giving (and memorizing) a correcting offset angle to the W axis for each crystal. A tandem disposition for gas and scintillation counters is no longer a necessity. The one or the other is brought into the right position under software control. The sin W potentiometer for continuous adjustment of the manual setting of the lower level and window of the pulse-height discriminator is replaced by the multichannel analyzer and electronic and electro-optical detection of the position of the goniometer. Automatic digital gain control corrects for pulse shrinking. Dead time is automatically corrected. This feature ensures linear response in most instruments to 2 Mcps for a gas-flow-proportional counter and to 1 or more Mcps for a scintillation counter. The main remaining electromechanical parts are the motor(s) of the goniometer in the sequential spectrometer (dc or stepper) or alignment motors for the crystals in the simultaneous spectrometers, the vacuum pumps, the electrovalves for gas flow (He, N2, counter gas), cooling water and compressed air, temperature control and fans, safety switches, sample lock and changer (internal and external), crystal turret, collimators, diaphragms, attenuator, and primary beam filters. An aspect totally absent in the older instruments was remote servicing. The full computer control allows remote servicing over a modem which substantially reduced the mean service time, especially in situations, where the service engineer is not close to the user’s facility. Copyright © 2002 Marcel Dekker, Inc.
B.
Spectrometer Layout
Commercially available spectrometers are classified according to the x-ray tube used, the position of the sample, sequential or simultaneous acquisition of the fluorescent radiation, hybrid forms, and some special dedicated designs. They will be discussed in the following order: Sequential instruments with EWT or SWT, sample down or sample up Simultaneous instruments with sample surface up or down Hybrid or dedicated designs 1. Sequential Spectrometers with EWTor SWT A typical example of geometry with sample surface down (i.e., the x-ray tube positioned under the sample) is represented in Figure 45 (layout of the Siemens-Bruker SRS-3400). The merit of the sample-surface-down geometry is in the analysis of solutions, not easily feasible with a sample-surface-up geometry. The positions of primary beam filters, sample=spectrometer seal, collimator masks or diaphragms, and spectrometer optics are clearly shown. The front end of the x-ray tube and all other parts except the scintillation counter are enclosed in a vacuum chamber. The sample=spectrometer seal is inserted in order to isolate the sample lock from the goniometer during sample loading. The sample lock and goniometer are evacuated by their own vacuum pump. The aim of this particularity is the maintenance of a monochromator atmosphere of constant composition,
Figure 45 A spectrometer layout with EWT and sample-surface-down geometry. (Siemens SRS3400. Courtesy of Bruker.)
Copyright © 2002 Marcel Dekker, Inc.
which may contribute to the reproducibility of the determination of light elements and prevent the contamination of the monochromator, particularly the collimator, from contamination by powdery samples. A drawback is the x-ray absorption when light elements are to be determined. Philips recently introduced a spectrometer with the same geometry (PW 2400 or 2404; Fig. 46) with an end-window x-ray tube, called the Super Sharp Tube (SST). The suffix ‘‘super sharp’’ points to the angle of conical nose of the tube, allowing a decrease of the sample distance to the window. The duplex detector (replacing the classical tandem fpc=sc construction) is a combination of a flow counter and a sealed xenon detector (improved performance for Ka for Ti to Cu and for La to W). The scintillation counter is giving an offset angle with respect to the gas counters. By this, the scintillation counter can be enclosed in the vacuum chamber without increasing the volume of the chamber. The selection of the detector is software controlled and became possible by the decoupling of the W and 2W axes. Shown in Figure 47 is the method by which one or two additional fixed channels can be added. Fixed channel(s) reduce(s) the analysis time for routine analysis of samples with one or two elements constantly present or for samples with an internal standard. ARL
Figure 46 Spectrometer with EWT and sample-surface-down position but with the addition of a fixed channel. This channel is the similar to the ones used in simultaneous spectrometers. (Philips PW2404) (Philips Technical Materials. Sequential X-Ray Spectrometer System PW 2404. Courtesy of Philips.)
Copyright © 2002 Marcel Dekker, Inc.
Figure 47 Heart of the ARL 9400 spectrometer (EWT, sample surface down). Two such monochromators, operating simultaneously, are integrated in one spectrometer, model, ARL 8420S. (Courtesy of ARL.)
realizes the simultaneous determination of more than one element by enclosing within the vacuum chamber a complete second monochromator of their unique extremely compact design as represented in Figure 47. It is characterized by the complete absence of gears and a positioning system by counting Moire´fringes resulting from interference of two grating systems. A development of the last few years is the substitution of diaphragms by a smalldiameter aperture allowing analysis of an area as small as approximately 1 mm2 (see Sec. VI.A.3 about milliprobes). An appropriate mechanism is moving the sample, which allows element mapping. This feature is realized in the XRF-1700 or 1500 series of Shimadzu (Fig. 48). It is an instrument with sample-surface-up geometry and fitted with an attenuator, reducing the intensity by a factor of 10 (for a major element determination); the merit of this is that no tube power switching is required for major and minor element determination. Copyright © 2002 Marcel Dekker, Inc.
Figure 48 Sample-surface-up spectrometer, fitted with small-aperture collimator diaphragms and an attenuator; all parts are enclosed in the vacuum chamber (Shimadzu XRF-1700). (Courtesy of Shimadzu.)
Only one manufacturer still produces an instrument with SWT. A number of instruments of a former generation are still in use and the tubes are still available. Diano Corporation continues the production of a spectrometer with sample-surface-up geometry and SWT (cf. Table 10). The merits and disadvantages of EWT or SWT were previously discussed. However, if a good homogeneity of the beam is required, a fine-focus SWT is by far a better solution. Sample holders of slightly different design for solids and liquids, for sample surface up or down, are offered by the manufacturers. The holders for instruments with samplesurface-up geometry are fitted with a spring mechanism to press the sample against the sample mask, both for sequential and simultaneous instruments. The holders adapting insert for liquids have a shape that can be recognized automatically and will not be accepted if the spectrometer is not under a He atmosphere (or eventually switches automatically from vacuum to He). 2. Simultaneous Instruments All simultaneous instruments are equipped with end-window tubes with very similar construction features. Up to 30 monochromators may be arranged radially around the tube head, with alternating high and low takeoff angles. One or more monochromators may be replaced by programmable goniometers of the flat (a normal flat crystal goniometer of reduced size) or curved crystal type. The latter is small but limited in angular range; the former has a large angular range but is large and occupies at least two positions otherwise taken by fixed monochromators. As already mentioned, Philips offers a small, low-power target transmission tube (TTT tube) used in the spectrometer PW 1660 (Philips Technical Copyright © 2002 Marcel Dekker, Inc.
Materials. Simultaneous X-ray Spectrometer System PW 1660) or X0 CEM capable of simultaneous determination of only 14 fixed elements or 12 elements when equipped with a programmable goniometer. In most instruments the samples are placed in the surface-up position. Some instruments (e.g., Oxford and PW2600 from Philips) irradiate in the sampledown position. The monochromators can be fitted to the outside of the vacuum cabinet, in which case they need not be vacuum tight. Figure 49 shows a generally used geometry. In the ARL sequential instruments, fixed channels (up to eight) or a complete second monochromator can be added. In Philips’ sequential spectrometer (PW 2400), an optional two fixed channels may be installed, making this instrument partly simultaneous. This is made feasible by the use of the special x-ray front end tube, the model Super Sharp Tube. 3. Milliprobes In an obvious effort toward the new fields of application, two companies introduced options, which seem to be restricted up to now to the very special, laboratory-made instruments. Rigaku and Shimadzu independently constructed the analytical systems with special masks: nonmovable (Rigaku) and scanning type (Shimadzu). In both versions, the diameter of the opening is 1 mm. Coupled with a 4-kW generator and high-power tube, it can carry out the analysis of the small pieces of a sample. In the Shimadzu version,
Figure 49 Layout of a simultaneous spectrometer with sample-surface-up geometry, typical for Bruker MRS4000 or Philips 1606 and with the arrangement at two levels with different takeoff angles. (Courtesy of Bruker.)
Copyright © 2002 Marcel Dekker, Inc.
the opening is scanned before a front of the sample, thus enabling the elemental mapping of the sample. The suggested application would be for the spot composition and density tracing of the sample, as it has been checked elsewhere (Kuczumow et al., 1995; Luggar and Gilboy, 1994). The construction of the milliprobe requires special attention concerning the x-ray tube used. The conventional end-window tube, due to its output inhomogeneouity, might be used only in the Rigaku arrangement, with a constant mask and the sample moving in front of the tube. Otherwise, the relatively small output of such tubes prevents their use as a radiation source for the milliprobes. C.
Intercomparison of Instruments
Only a limited number of manufacturers produce x-ray spectrometers. Hereafter, we will discuss the instruments of seven manufacturers subdivided in three classes: sequential and simultaneous instruments and a third class that we called hybrid spectrometers, because they have characteristics belonging to both former classes. The series numbers together with the names of the manufacturer are given in Table 8. In Tables 9–11, the numbers in the top row of the third, fifth, and last column refer to the manufacturer and series number of Table 8. The reader will find the actual addresses of the manufacturer’s headquarters at the end of this subsection. 1. Sequential Instruments In Table 9, a summary is given of the main characteristics of the current instruments as a guide to the market for the interested reader and for those who are considering buying a new instrument or replacing an older instrument. a.
General Remarks and Items Common to All Instruments A question mark in the table means that the data were not available in the brochures. When nothing is marked, it means that the given item is not applicable.
Table 8
Available Wavelength-Dispersive Spectrometers Sequential
Manufacturer ARL Diano Oxford Philips
Rigaku Siemens Shimadzu
Series
Simultaneous
Table 9
Series
Hybrid
Table 10
9400 2000
1 2
8600S
1
PW2400 PW2404
3 30
PW1606 PW2600 X-CEM
2 3 4
RIX2100 RIX3100 SRS3400 XRF1700 XRF1500
4 40 5 6 60
MRS 4000
5
Copyright © 2002 Marcel Dekker, Inc.
Series
Table 11
MDX1000 Venus 100 PW4100=20 PW4120=20 PW4110=10
1 2 20 3
Table 9
Presentation of Sequential Instruments Manufacturer 1
Generator Power (kW) Voltage range=step (kV=step) Current range=step (mA) Switch time (s) Switch mode Stability D%/% mains Source X-ray tube: geometry, load (kW) Anode, type, Be window thickness (mm) Option other anodes Primary beam filters number Elements Sample Sample surface Sample turret: number of positions Sample=spectrometer seal (yes, no) Optics Diaphragms: number of positions Collimators: selectable (yes, no), number
Copyright © 2002 Marcel Dekker, Inc.
3, 30
2
4, 40
6, 60
5
3 3 ?–60=? (opt.: ?–70=?) 15–60=1
3, 4 20–60=1
3 or 4 60=?
3 or 4 10–60=1
3 or 4 60=?
?–100 ? ? 0.0001
0–80=0.1 ? ? 0.0005
10–125=1 < 2,5 Isowatt curve 0.0005
100 or 150=? ? ? 0.0005
5–150=1 Fast — 0.0001
100 or 140=? — — 0.0002
EWT, 3
SWT
EWT, 3
EWT, 3 or 4
EWT, 3 or 4
Rh, 3GN, 75
Cr, EA-75, 250
Rh, SS, 75 or 125 Rh
W, Mo, 125 3
W,Pt,W=Cr, Pt=Cr,Mo=Cr None
Others on request Rh=Cr,Rh,=C, Rh=W 5 1 or 5
Cu, Al, Fe
—
Al, Cu
Down 2
Up 4
y
Rh, AGRh66G, Rh 75 or 125 Mo, W, Au, Cr Rh=Cr, Rh=W 10
4
Zr, Ti, Al, Cu
Cu, Al
Al, Ti, Ni, Cr
Down 2
Up 6
Down 2
Up 8 or 2
n
n
n
y
n
3
—
4
6
4
5
y, 4
1
y, 3
y, 2 or 3
y, 4
3
Attenuator (yes, no) Crystal changer: number of positions Detectors: extra (yes, no) Goniometer angular range ( 2y) fpc angular range ( 2y) sc and 2 axes (coupled=decoupled) Positioning technology Continuous scanning ( 2y=min) Step scanning ( 2y=min) Slewing speed ( 2y=min) Angular accuracy ( 2y, y) Angular reproducibility ( 2y, y) Temperature stabilization ( C) Additional fixed channels (yes, no, or number) Parts included in vaccum Counting electronics Count rate (kcps): fpc sc Sealed Duplex Pulse-shift correction Dead-time correction Computer instrument monitoring peripherals Remote servicing (yes, no) Sample changer: type (x, y), positions
Copyright © 2002 Marcel Dekker, Inc.
n
n
n
y
n
y
9 n
4 or 6 Sealed Xe
10 n
8 n
10 n
17–152 0–115 dec
0–146 c
8 Sealed Xe, Ne, Kr 13–148 8–104 dec
8–148 5–118 ?
4–150 4–112 dec
7–148 0–118 dec
Moire´ fringe 0.25–128
? 0.1–10
Optical sector 0.006–120
? 0.1–240
Stepper motor 1–200
Stepper motor 0.01–180
0.001–1.00 4800 0.001 0.0002
0.01–? 600 0.001 ?
0.0001–2 2400 0.0025 0.0001
0.005–0.1 1000 ? 0.0005
— 1000 0.00001 0.001
0.002–1 1200 ? 0.0003
0.1
None
< 0.05
?
y
None
2
None
None
None
All
All
All
All except sc
All except sc
All
2000 1500
? ?
0.01
0.3
2000 1000
2000 2000 4000
2000 1000
Pulse-height discrim. Automatic Full
Automatic
Automatic
Automatic Full
2000 1500 1500 3500 Graphical display Automatic, dynamic Graphical display Automatic Partly Full
Automatic Full
Automatic Full
y xy, 12–98
n x, 4 or 10
? x(?), 50, 100
y xy, 8, 58, 100, 110, 138
n ?
Automatic
y xy, 6, 24–168
Table 9
Continued Manufacturer 1
Installation Instrument dimensions: H*W*D (mm) Weight (kg) Electricity supply (V, kVA) Cooling water: bar, max. temp. ( C), flow (L=min) Compressed air: bar, flow (L=min)
Copyright © 2002 Marcel Dekker, Inc.
3, 30
2
4, 40
6, 60
5
920*1080*760
1400*860*1346
1092*1410*920
1400*1420*1054 1070*1890*880
450 220, 6 > 2, 20, 0.5–5
590 208–480, 10 > 1, 29, 19
580 750 200–240, 8, 5 200–220, 10 > 3.5–8, 20, 6–11 ?, < 30, 3.5–11
700 115–400, 8 4–6, —, 1.5
780 200–220, 8 or 9 > 1.5–3, 30, 3–10
—
—
4–5, ?
6–10, 20
—
None
1350*1770*1080
The analyzing range of all instruments goes from Be to U and the transuranians, except for Diano 2000, for which the range is limited to B–U. The mains power requires a stability from 10% to þ10%, exceptionally from 15% to þ10%. A single phase is required, except for Rigaku, which requires three phases. Most commonly offered by all manufacturers are rhodium anodes for x-ray tubes. Primary beam filters are used to remove interfering lines from the tube target element and=or to improve signal-to-background ratio: Al: improves peak-to-background ratio for Pb and As in light matrices; removes RhKa line and the CrK lines Cu: determination of Ru, Rh, Pd, Ag, and Cd; removal of interfering lines from Rh Fe: removes the Ka line of Rh Ti: determination of Cr, Co, Fe, Zn; improves peak-to-background ratio Zr: measurement of CdKa, removes the Ka line of Rh Most manufacturers supply other filters on request. Sample dimensions: Most instrument allow a sample size about 52 mm in diameter (Diano: 63 mm) and about 30 mm in height (Bruker: 57 mm). To cope with small heterogeneities, the samples are spun at 30 rpm (or 60 rpm for Shimadzu) on most instruments (for Diano, the spinner is only available on the sample changer with 10 positions). The collimators are quoted in most brochures as extra-coarse, coarse, medium, fine (high intensity to high resolution). Only Bruker mentions the divergence angles: 0.077 , 0.15 , 0.46 , 0.93 , 1.0 , 1.90 , and 1.54 . Crystals: All sequential spectrometers make use of flat crystals. Counters: A gas-flow-proportional counter (fpc) and a scintillation counter (sc) are standard on all instruments, except Diano, in which the sealed Xe detector is standard and a scintillation counter is an option. The gas flow (Ar þ 10% CH4) is 1–2 L=min. The electronic counters are all based on a multichannel analyzer (Shimadzu and Rigaku probably still use a pulse-height discriminator) with an automatic pulse shift and dead-time correction. Available on all spectrometers is the choice of vacuum or He as the spectrometer atmosphere, with nitrogen as option for many. All companies supply the control unit by personal computer in more or less sophisticated versions with software for monitoring the instrument, for qualitative analysis, and for a number of possibilities for conversion of intensity to concentration. All of these items are subject to change and not discussed in detail here. The conversion algorithms are discussed in detail in Chapter 5. The software systems supplied by Bruker, ARL, Rigaku, and Shimadzu are all very user friendly. Installation data are added to give the potential buyer an idea of the required space and infrastructure. The dimensions and weight of the instruments are given without an external sample changed fit. b.
Remarks About the Individual Instruments (in Alphabetical Order as in Table 8) ARL 9400: As far as we are aware, ARL was the first manufacturer (and still the only one) to use counting Moire´ fringes for its goniometer positioning. The goniometer is a complete gearless construction and it has the highest slew rates
Copyright © 2002 Marcel Dekker, Inc.
Table 10
Presentation of Simultaneous Instruments Manufacturer 1
Generator Power (kW) Voltage (kV), step (kV) Current (mA), step (mA) Stability D%/% mains Source X-ray tube: Load (kW) Anodes Be windows (mm) Primary beam filters positions Elements Primary beam attenuators (Yes, No) Samples Sample surface Size: diameter, height (mm) Sample changer (positions) Monochromator (fixed channels) Takeoff angle ( ) Maximum number Crystals Detectors Element range Goniometer
Copyright © 2002 Marcel Dekker, Inc.
2
3
4
5
3 ?–60=? ?–100=? 0.0001
3 ?–60=? ?–75=? 0.0005
3 5–60=1 (a) 5–125=1 0.0005
0.2 50 4 0.0005
3 20–60=1 5–150=1 0.0001
3 Rh, W
3 Rh, Cr, Mo
TTT* Cr
125, 75 4 (a)
125, 75 10
3 Rh, other on request 125, 75 3
Rh 66 G Rh, Mo, W, Au, Cr 125, 75
Cu, Al, Fe y, 14
Up 75, 50 11, 32, or 8 (b)
Up 51, 40 6, 72, 300
Down (b) 51, 40
? 30, 20, 11 (c) Curved Gas counter, sc
29, 44 28 (a) x
29, 44 28 Log. curved
Be–U
Be–U
y
40, ?
52, 57 8, 58, 100, 138
25, 39, 5 14
26, 24 (a) 28
Gas (Ar flow, Ne, Kr sealed)
fc, sc Be–U
Maximum number Collimators Crystals number
1 (d) 3 9
Type Scan speed ( 2y=min) Step scan ( 2y=step) Slewing speed ( y=min) Angular accuracy ( 2y) (LiF) Angular resolution ( 2y) Angular reproducibility ( ) Angular range gas counter ( 2y) sc ( 2y) Detector Element range Spectrometer environment
Flat 0.25–128 0.001–1 4800 0.015 0.001 0.0002 17–155
0.06–76 0.01 1000 ? ? 0.003 5–120
0–115
5–120
B–U Vac
B–U Vac, He, air
XRD channel (yes, no) Installation Dimensions: H*W*D (mm) Weight (kg) Water supply(bar, max. C, L=min) Compressed air (bar, L=min)
Copyright © 2002 Marcel Dekker, Inc.
(a)
2 (c) 1 1 (LiF 200, 220, Ge) Flat 0.06–76 0.01 480 ? ? < 0.01 10–100
2 (d) Slits 4 (PE, InSb, Ge, LiF 200, 220) Cyl. curved 0.06–76 0.01 480 ? ? < 0.005 85–47
sc, Kr-sealed
fpc
Vac, other on request
Vac, other on request
8
Up to 4
F–U Vac, air
y
n
1350*1400*950
1804*1232*877
1000 > 3, < 18, 5
900
?
2 (a)
305–8, < 20, 6–10 4–5, ?
1190*815*1230
1800*1810*950
580
900 3–8, 20, 3 6–10, 20
Table 11
Hybrid Instruments Manufacturer 2, 20
1 Generator Voltage (kV)=current (mA)=power (W) Stability D%=%mains Max. mains variation (V) X-ray source Tube Anode External Cooling Sample Sample surface up, down Spinning (rpm) Dimension (diameter, height in mm) Sample changer Spectrometer chamber Monochromator (fixed) Element covered
Crystals Additional spectrometers Dimensions (H*W*D in mm) Weight (kg)
3
40–50=4=200
50=4=200
id.
0.007 210–250
0.0005 180–264
id. id.
Similar (a) Rh None
Target transmission Sc None
id. Cr id.
Down
Down
48, 40
26–47, 40
id. 30 id.
? Vac, He optional
Single, 10 Vac, He
Single Vac
F–U
S, S-bg., Zr, Ni, Ni-bg., Mn, Mn-bg., V, V-bg Curved None
Al, Si, Na, Mg, Ti, Mn, K, Ca, P, S, Fe id. id.
id.
1200*750*960
id.
180
Flat, curved optional Flexi-channel (EDXRF) 1090*800*800 160
of all spectrometers. In the series 9420S, two goniometers or additional fixed channels can be fitted, adaptable to a great variety of analytical situations, excellent for rapid nonroutine industrial control, reducing the analysis time by a factor of 2 with respect to a conventional sequential spectrometer. Separate vacuum pumps support the vacuum for the sample lock and the monochromator. Diano 2000: This offers dual-target tubes. The targets are push-button selected and no moving parts are present (dual filament). Two collimators are offered: a 0.032-in. spacing ‘‘beam tunnel’’ or a 0.010-in.-spacing ‘‘diffracted beam soller slit.’’ The instrument is not temperature stabilized because the 45-kg bronze bull gear goniometer has a sufficiently high thermal capacity, avoiding rapid temperature changes. It is, however, not sufficient when one wants to measure peak shifts. The crystal in the turret must be individually tuned. The W and 2W axes are coupled but with a bisector arrangement and moved by a dc motor. The standard detectors are a flow-proportional counter and sealed Xe, the scintillation counter is an option. The counting is also by a pulse-height selector. Diano offers a fixed-sample and fixed-element mode for analysis. Copyright © 2002 Marcel Dekker, Inc.
Philips PW 2400, 2404: Both types differ only in the power output of the generator. The tube power is switched along the isowatt curve (i.e., by keeping the power constant when either the voltage or current is changed). The fixed channels that can be added have as the detector either a flow-proportional counter (3000 kcps), Xe-sealed (1000 kcps), Kr-sealed (2000 kcps), Ne-sealed (2000 kcps) proportional counters, or a scintillation counter (1500 kcps). The positioning of the goniometer is done by optical disk, a sophisticated highperformance system, giving the highest angular accuracy and reproducibility. The goniometer is driven by dc motors. The external sample changer accepts samples of nonuniform size. Rigaku 2100, 3100: Both types differ only in power output of 3 and 4 kW. The collimators are either standard=high resolution or standard high resolution=ultralight elements. The attenuator in front of the monochromator crystal reduces the intensity by a factor 10. The advantage should be that it avoids power switching of the x-ray tube. The sample-surface-up geometry makes it difficult, if not impossible, to analyze solutions. A diaphragm of 1 mm diameter allows small-spot analysis. A built-in cleaning system removes the impurities of the wire in seconds in this instrument because the flow-counter efficiency deteriorates by contamination of the wire electrode. Siemens (Bruker) SRS 3400: The instrument is rather conventional and easy to service. The positioning of the decoupled arms of the goniometer is done by stepper motors. The positioning of the sample holders is very reproducible. The external sample changer accepts nonuniform samples also. An excellent software SPECTRA PLUS is supplied for qualitative, semiquantitative, and quantitative analyses. The instrument was formerly sold under the name of Siemens but is taken over by Bruker Analytical X-Ray Systems (Bruker AXS). Shimadzu XRF 1700, 1500: The difference between both instruments is that the selection of filters, diaphragms, attenuator, and crystals is automated in the 1700 model. A special feature of this instrument is the presence of diaphragms of diameter 1, 3, 10, 20, and 30 mm. The 1